Library › Devices and mechanical support › LVAD complications
LVAD complications: pump thrombosis, bleeding, infection, and right heart failure
Complications of durable left ventricular assist device support in adults, from bleeding, pump thrombosis, stroke, and infection to right heart failure, aortic insufficiency, arrhythmias, outflow graft obstruction, and device malfunction.
Complications have replaced pump failure as the main limit on durable left ventricular assist device (LVAD) therapy. Bleeding is the most frequent: in MOMENTUM 3, gastrointestinal bleeding occurred in about one-quarter of HeartMate 3 recipients within 2 years. Infection, stroke, right heart failure, aortic insufficiency, arrhythmias, and outflow graft obstruction follow. Pump thrombosis, the defining failure of axial-flow pumps, occurred in 1.4% of HeartMate 3 recipients at 2 years, compared with 13.9% with the HeartMate II. The 2023 International Society for Heart and Lung Transplantation (ISHLT) guidelines grade the evidence for the individual complications. Most recommendations rest on level C evidence, so care depends on a systematic reading of device parameters, hemolysis markers, imaging, and log files.
Candidacy, implantation, continuous-flow physiology, blood pressure targets, the baseline warfarin regimen, echocardiographic surveillance, and ramp studies are described on the durable left ventricular assist devices page.
Adverse events during LVAD support: definitions and overall burden
Adverse events after LVAD implantation are reported with standardized definitions so that trials and registries can be compared. The Mechanical Circulatory Support Academic Research Consortium (MCS-ARC) updated the Intermacs definitions in 2020 (Kormos 2020), and the 2023 ISHLT guidelines direct that these definitions be used for neurologic events, right heart failure, and device malfunction. The definitions classify events by type, timing, severity, and attribution. A stroke, for example, is classified as ischemic or hemorrhagic, as disabling or nondisabling by the modified Rankin scale, and as patient-related, management-related, or device-related.
Hemocompatibility-related adverse events form a distinct group because they share a mechanism in the interaction between blood and the pump. They comprise stroke, pump thrombosis, major bleeding, and arterial peripheral thromboembolism. This composite was the primary end point of the ARIES-HM3 trial (Mehra 2023).
| Measure | HeartMate 3 | HeartMate II | Source |
|---|---|---|---|
| Survival free of disabling stroke or reoperation to replace or remove a malfunctioning pump, 2 years | 76.9% | 64.8% | MOMENTUM 3 final report (Mehra 2019) |
| Pump replacement, 2 years | 2.3% | 11.3% | MOMENTUM 3 (Mehra 2019) |
| Suspected or confirmed pump thrombosis, 2 years | 1.4% | 13.9% | MOMENTUM 3 (Mehra 2019) |
| Gastrointestinal bleeding, 2 years | 24.5% | 30.9% | MOMENTUM 3 (Mehra 2019) |
| Major bleeding, 2 years | 43.7% | 55.0% | MOMENTUM 3 (Mehra 2019) |
| Major infection, 2 years | 58% | 56% | MOMENTUM 3 infection analysis (Patel 2020) |
| Survival, 5 years (patients on support at 2 years) | 58.4% | 43.7% | MOMENTUM 3 extended phase (Mehra 2022) |
| Freedom from gastrointestinal bleeding, device malfunction, and stroke, 5 years | 73.5%, 81.8%, and 87.1% | — | Society of Thoracic Surgeons (STS) Intermacs 2025 report, commercial HeartMate 3 (Nayak 2026) |
Event rates with the HeartMate 3 compared with the axial-flow HeartMate II. The HVAD centrifugal pump was withdrawn from the market in 2021, and its remaining recipients carry higher stroke and thrombosis risks than HeartMate 3 recipients.
Events cluster by time after implantation. Bleeding from the operation, tamponade, early right heart failure, acute kidney injury, vasoplegia, and early ventricular arrhythmias cluster in the first weeks. Gastrointestinal bleeding and driveline infection accumulate through the first years. Aortic insufficiency, late right heart failure, extrinsic outflow graft obstruction, and driveline damage become more frequent with longer support.
Hemocompatibility and the shear environment of a rotary pump
A continuous-flow pump exposes blood to supraphysiologic shear stress, to artificial surfaces, and to zones of stasis. The same environment predisposes to both bleeding and thrombosis. Shear unfolds von Willebrand factor and exposes it to cleavage by the protease ADAMTS13, which depletes the largest and most hemostatically active multimers. Shear also activates platelets and can lyse red cells. Contact with the titanium and polyester surfaces of the pump and graft activates coagulation, and regions of low flow in the inflow cannula, the pump, the outflow graft, and a closed aortic root allow thrombus to form.
The HeartMate 3 was engineered to reduce this exposure. Its rotor is fully magnetically levitated, with wide blood-flow gaps and no mechanical bearings, and an artificial pulse changes rotor speed at 2-second intervals to wash the pump. These features reduced pump thrombosis and stroke compared with the HeartMate II but did not eliminate bleeding, because the acquired von Willebrand defect and low pulsatility persist with any continuous-flow pump. Antithrombotic therapy is the second determinant: anticoagulation intensity, the presence or absence of aspirin, and drug interactions shift the patient's position on the balance between bleeding and thrombosis. In ARIES-HM3, omitting aspirin in HeartMate 3 recipients taking a vitamin K antagonist reduced nonsurgical bleeding (relative risk 0.66) without increasing thromboembolic events (Mehra 2023). The warfarin target and the aspirin-free regimen are described on the durable LVAD page.
Early postoperative bleeding and pericardial tamponade
Bleeding in the first days after implantation results from the operation itself, the need for cardiopulmonary bypass, hepatic congestion, preoperative anticoagulants and antiplatelet agents, and the acquired von Willebrand defect, which develops soon after continuous-flow support begins. Re-exploration for bleeding is common, and transfusion adds volume load to the right ventricle and can cause sensitization to human leukocyte antigens, which complicates later transplantation. Unfractionated heparin is usually started on postoperative day 1 or 2 when surgical hemostasis is secure, and warfarin is begun around postoperative days 2 to 3 once the patient is stable and no further procedures are planned (Saeed 2023).
Pericardial tamponade after LVAD implantation often presents without the classic findings. Clot may be localized behind the right atrium or right ventricle and compress one chamber, the narrow pulse pressure removes pulsus paradoxus as a sign, and transthoracic windows are limited by the device and dressings. Tamponade appears as falling pump flow, a low pulsatility index, rising central venous pressure, falling blood pressure, oliguria, and an increasing vasoactive requirement. It can be indistinguishable from right heart failure until imaging is performed. Transesophageal echocardiography identifies a regional collection when transthoracic imaging is inconclusive, and CT is used in stable patients. The 2023 ISHLT guidelines recommend evaluation for tamponade from bleeding or sternal compression as a cause of early right ventricular failure (Class I, level C). Delayed sternal closure is a reasonable consideration when bleeding, thoracic edema, or right ventricular distension threatens compression of the heart.
Gastrointestinal bleeding: incidence and sources
Gastrointestinal bleeding is the most common nonsurgical adverse event during continuous-flow support. The 2023 ISHLT guidelines cite an incidence of 18% to 26%, and recurrent bleeding follows a first episode in 30% to 40% of patients (Saeed 2023). In a meta-analysis of 1,839 patients, the pooled prevalence was 23% (Draper 2014). Bleeding arose from the upper gastrointestinal tract in 48% of events, the lower tract in 22%, and the small bowel in 15%. Angiodysplasia in the proximal gastrointestinal tract was the most common identified cause, and a definitive source is not found in 20% to 30% of episodes (Saeed 2023). The HeartMate 3 lowered but did not remove the risk (burden table above).
The sources resemble those of any adult with anemia on anticoagulation: peptic ulcer, gastritis, diverticular bleeding, colonic polyps and cancer, and hemorrhoids. Arteriovenous malformations and angiodysplasias are disproportionately frequent, multiple, and recurrent, and they explain most recurrent bleeding. Bleeding typically presents as melena, iron-deficiency anemia, or occult blood loss, and less often as hematochezia or hematemesis. A falling hemoglobin can also lower pump flow and pulsatility by reducing left ventricular filling. Severe bleeding causes low-flow alarms and suction events.
Mechanisms of mucosal bleeding during continuous flow
Three mechanisms converge. First, all continuous-flow pumps produce an acquired von Willebrand syndrome. In 31 HeartMate II recipients tested, high-molecular-weight von Willebrand multimers were reduced in all, and 58% had bleeding (Uriel 2010). The defect resembles type 2A von Willebrand disease. It impairs platelet adhesion at high-shear mucosal sites and resolves after the pump is removed at transplantation. The same mechanism links severe aortic stenosis to bleeding from angiodysplasia in Heyde syndrome.
Second, angiodysplasia forms under conditions of low pulsatility. Diminished pulse pressure is thought to cause relative mucosal hypoxia and to activate hypoxia-inducible factor 1α, and thrombin generated by the pump surface increases endothelial expression of angiopoietin-2. In a translational study, angiopoietin-2 levels were higher in LVAD recipients than in controls, and patients with levels above the mean had more nonsurgical bleeding (Tabit 2016). Increased angiotensin II signaling through transforming growth factor-β is a further proposed pathway.
Third, anticoagulation, antiplatelet therapy, and mucosal congestion convert fragile vessels into bleeding sources. Right heart failure with elevated venous pressure congests the gut mucosa, and the guidelines list this vasodilated, congested state among the causes of bleeding (Saeed 2023). The foreign surface also consumes platelets, and kidney dysfunction adds a platelet defect.
Evaluation of the first gastrointestinal bleed
Evaluation begins with hemodynamic stability, the hemoglobin trend, the international normalized ratio (INR), and pump parameters. The 2023 ISHLT guidelines recommend examination of the nasopharynx and assessment of stool for visible or occult blood in patients with unexplained anemia or suspected mucocutaneous bleeding (Class I, level C). Nasopharyngeal bleeding can be swallowed and present as melena. Spontaneous hematomas, especially in patients treated with low-molecular-weight heparin, are sought on examination of the torso, and CT identifies retroperitoneal bleeding when the source is not apparent.
Patients with a suspected gastrointestinal source are managed with a gastroenterology team, and colonoscopy, upper endoscopy, or both are performed after a first gastrointestinal bleeding event (Class I, level C). Endoscopy is performed with the pump at its usual speed and with attention to volume status, because hypovolemia during sedation and bowel preparation causes suction events. When upper and lower endoscopy are negative, evaluation of the small bowel by enteroscopy should be considered (Class IIa, level C). Persistent bleeding with a negative evaluation makes a tagged red blood cell scan, capsule endoscopy, or angiography reasonable (Class IIa, level C).
| Test | Role in LVAD recipients | Limitations |
|---|---|---|
| Upper endoscopy | First test for melena or hematemesis; treats angiodysplasia, ulcers, and Dieulafoy lesions with argon plasma coagulation, clips, or injection | Misses jejunal lesions beyond reach of a standard endoscope |
| Colonoscopy | First test for hematochezia; with upper endoscopy for occult loss after a first bleed | Bowel preparation requires attention to volume status, because hypovolemia causes suction |
| Push or device-assisted enteroscopy | Diagnosis and treatment of proximal and mid small bowel angiodysplasia | Longer sedation; experience varies between centers |
| Capsule endoscopy | Survey of the small bowel when bidirectional endoscopy is negative | Diagnostic only; a lesion found requires enteroscopy for treatment |
| Tagged red blood cell scan or CT angiography | Localization of active bleeding at a sufficient rate before angiography | Requires active bleeding at the time of imaging |
| Mesenteric angiography | Localization and embolization of active bleeding | Contrast exposure and access-site bleeding during anticoagulation |
Endoscopic and radiologic evaluation of gastrointestinal bleeding during LVAD support. The sequence follows the 2023 ISHLT recommendations.
Antithrombotic therapy during and after bleeding
Antithrombotic changes after a bleed follow the severity of bleeding, the INR, the pump type, and the patient's thrombotic history. The table lists the 2023 ISHLT recommendations.
| Situation | Recommendation | Class, level |
|---|---|---|
| Clinically significant bleeding | Anticoagulation and antiplatelet therapy are held | I, C |
| Life-threatening bleeding with an elevated INR | Anticoagulation is reversed | I, C |
| While therapy is held | Therapy remains held until clinically significant bleeding resolves, provided there is no evidence of pump dysfunction; the patient, pump parameters, and markers of hemolysis are monitored | I, C |
| Acute anemia, clinically stable, no deficit | Monitoring with cessation of anticoagulation, without urgent INR reversal | IIa, C |
| After a first bleed resolves | Anticoagulation and antiplatelet therapy are reintroduced with monitoring of hemoglobin and stool | I, C |
| Recurrent bleeding without a treatable source | A lower warfarin goal or an amended antiplatelet regimen may be reasonable, individualized to pump type and history | IIb, C |
| Recurrent bleeding with reduced or stopped therapy | The patient, the device, and markers of hemolysis are monitored | I, C |
| Pump speed reduction to restore pulsatility | Effectiveness not established; may be harmful | III, C |
Recommendations of the 2023 ISHLT guidelines for antithrombotic management of mucocutaneous bleeding. Most recommendations were written when aspirin was routine; after ARIES-HM3, most HeartMate 3 recipients take warfarin alone.
Warfarin is reversed with four-factor prothrombin complex concentrate and intravenous vitamin K when bleeding is life-threatening, and plasma is a slower alternative. Reversal is reserved for bleeding that threatens life or a critical organ, because the interval without anticoagulation carries a risk of pump and aortic root thrombosis. Agents and doses are described on the anticoagulant reversal page. Desmopressin transiently raises von Willebrand factor and factor VIII levels, and von Willebrand factor concentrate has been used for refractory mucosal bleeding. Both have limited evidence in LVAD recipients and add thrombotic risk. Platelet transfusion is used for thrombocytopenia with active bleeding.
Anticoagulation is resumed when hemostasis is secure, usually without a heparin bridge in stable patients with a HeartMate 3. The INR goal of patients with older axial-flow or HVAD pumps is lowered with greater caution because their thrombosis risk is higher. Direct oral anticoagulants are not an established alternative to warfarin.
Preventing recurrent bleeding: pharmacologic options
Recurrent bleeding from angiodysplasia that persists despite endoscopic therapy and antithrombotic adjustment is managed with drugs that act on angiogenesis, splanchnic blood flow, or the renin-angiotensin system. No randomized trial supports any of them in LVAD recipients, and the evidence consists of retrospective cohorts and before-and-after comparisons.
Angiotensin-converting enzyme inhibitors and angiotensin receptor blockers. Angiotensin II promotes angiogenesis through transforming growth factor-β and angiopoietin-2. In 111 HeartMate II recipients, an angiotensin-converting enzyme inhibitor or angiotensin receptor blocker started within 30 days of implantation was associated with a 57% lower risk of major gastrointestinal bleeding and a 63% lower risk of bleeding from arteriovenous malformations (adjusted hazard ratios 0.43 and 0.37). The effect appeared above a daily dose equivalent to 5 mg of lisinopril (Converse 2019). The 2023 ISHLT guidelines state that these drugs may be beneficial in reducing the incidence or recurrence of gastrointestinal bleeding (Class IIb, level B).
Octreotide. The somatostatin analogue reduces splanchnic blood flow and inhibits angiogenesis. In a multicenter cohort of 51 patients given secondary prophylaxis after a first gastrointestinal bleed, 24% rebled, compared with 43% of matched historical controls from the HeartMate II trials (Shah 2017). In 30 patients with recurrent bleeding treated with depot octreotide 20 mg intramuscularly at 4-week intervals, bleeding fell from 3.4 to 0.7 events per year, with fewer transfusions, hospital days, and endoscopies (Juricek 2018). The 2023 ISHLT guidelines describe subcutaneous dosing of 100 to 200 µg two or three times daily and monthly depot dosing of 20 mg, and state that octreotide may be considered for recurrent bleeding (Class IIb, level B). Diarrhea, abdominal pain, flatulence, injection-site pain, gallstones, and cost limit its use.
Digoxin. Digoxin inhibits synthesis of hypoxia-inducible factor 1α. In a single-center cohort of 199 continuous-flow recipients, digoxin use was associated with less bleeding from angiodysplasia (5% compared with 25%; adjusted hazard ratio 0.18), without a difference in other bleeding (Vukelic 2018). A 2026 meta-analysis of 4 studies and 14,917 patients found a nonsignificant reduction in all gastrointestinal bleeding (hazard ratio 0.70; 95% confidence interval 0.49 to 1.01) and a significant reduction in bleeding from angiodysplasia (hazard ratio 0.33) (Marchetti 2026). The guidelines judge its usefulness for this purpose not established (Class IIb, level C). Digoxin toxicity is more likely with kidney dysfunction, hypokalemia, and amiodarone, all common in this population.
Thalidomide. Thalidomide inhibits vascular endothelial growth factor. In 17 patients with recurrent bleeding from arteriovenous malformations, the hazard of rebleeding on thalidomide was 0.23 relative to periods off the drug, and bleeding fell from a median of 4.6 to 0.4 episodes per year. Adverse effects occurred in 59% and usually resolved with dose reduction (Namdaran 2020). Thalidomide is teratogenic, increases venous thromboembolism in other populations, causes neuropathy, fatigue, and constipation, and is dispensed in the United States only through a restricted distribution program. Its usefulness is not established in the guidelines (Class IIb, level C), and it is reserved for transfusion-dependent bleeding after other measures fail.
| Therapy | Proposed mechanism | Best available evidence in LVAD recipients | 2023 ISHLT |
|---|---|---|---|
| ACE inhibitor or ARB | Blocks angiotensin II–driven angiogenesis; lowers blood pressure | Retrospective cohorts; adjusted HR 0.43 for major GIB in HeartMate II recipients | IIb, B |
| Octreotide | Reduces splanchnic flow; antiangiogenic | Historical-control cohort (24% vs 43% rebleeding); before-and-after series | IIb, B |
| Digoxin | Inhibits HIF-1α synthesis | Cohorts and meta-analysis; HR 0.33 for angiodysplasia bleeding | IIb, C (not established) |
| Thalidomide | Inhibits VEGF | Small before-and-after series; frequent adverse effects | IIb, C (not established) |
| Omega-3 fatty acids | Anti-inflammatory and antiangiogenic effects proposed | Single-center cohort | IIb, C (not established) |
| Pump speed reduction | Restores pulse pressure | No consistent benefit; risks inadequate unloading | III, C |
Pharmacologic and device strategies for recurrent gastrointestinal bleeding during LVAD support. ACE, angiotensin-converting enzyme; ARB, angiotensin receptor blocker; GIB, gastrointestinal bleeding; HIF-1α, hypoxia-inducible factor 1α; HR, hazard ratio; VEGF, vascular endothelial growth factor.
Repeated endoscopy may be reasonable for recurrent bleeding, particularly when earlier studies found treatable lesions, but its yield is low after repeated negative examinations (Class IIb, level C). Hormonal therapy, bevacizumab, and other antiangiogenic agents have been reported in individual patients only. For patients eligible for transplantation, recurrent transfusion-dependent bleeding is a reason to reassess listing status, and transplantation removes the device and with it the acquired von Willebrand defect.
Epistaxis and other mucocutaneous bleeding
The mucosal bleeding diathesis of continuous-flow support is not confined to the gut. Epistaxis, gingival bleeding, hematuria, and menorrhagia or other vaginal bleeding occur at rates higher than in other patients with similar anticoagulation (Saeed 2023). In a pilot study using nasal endoscopy, mucosal hypervascularity was present in 63% of LVAD recipients, and gastrointestinal bleeding occurred in 32% of those with hypervascularity and none of those with normal mucosa (Patel 2016).
Epistaxis is treated with pressure, topical vasoconstrictors, cautery, and nasal packing, with otolaryngology involvement for posterior or recurrent bleeding. Humidification and avoidance of nasal trauma reduce recurrences. Anticoagulation is usually continued for minor bleeding and held only when bleeding is significant. Heavy menstrual bleeding is managed in consultation with gynecology, often with a levonorgestrel intrauterine system or endometrial ablation, because estrogen-containing contraception increases thrombotic risk. Hematuria prompts urologic evaluation for structural causes, and dark urine is distinguished from hematuria by microscopy, because hemoglobinuria indicates hemolysis.
Pump thrombosis: definition and incidence
Pump thrombosis is the ingestion or formation of clot within the flow path of the device: the inflow cannula, the rotor and housing, or the outflow graft. The MCS-ARC definitions separate suspected thrombosis, based on clinical findings, laboratory evidence of hemolysis, and pump parameters, from confirmed thrombosis, documented at device explant or by imaging. Thrombus at the inflow cannula is termed prepump thrombosis; thrombus within the rotor or housing, intrapump thrombosis; and thrombus or obstruction in the graft, postpump thrombosis.
The incidence depends on the pump and the era. In a three-center series, confirmed HeartMate II thrombosis at 3 months after implantation rose from 2.2% before March 2011 to 8.4% by January 2013 (Starling 2014). In MOMENTUM 3, suspected or confirmed thrombosis at 2 years occurred in 13.9% of HeartMate II recipients and 1.4% of HeartMate 3 recipients (Mehra 2019). In the same series, patients with thrombosis who were managed medically had an actuarial mortality of 48.2% over the next 6 months. Mortality after pump exchange or transplantation was similar to that of patients without thrombosis (Starling 2014). Pump thrombosis is now rare with the HeartMate 3 (1.4% at 2 years, burden table above). A thrombosis-like presentation in a HeartMate 3 recipient therefore more often has another cause, most often outflow graft obstruction.
Risk factors include subtherapeutic anticoagulation and nonadherence, infection and sepsis, atrial fibrillation, hypertension and low pump flow, low pump speed, hypercoagulable states, malposition of the inflow cannula, and heparin-induced thrombocytopenia. Thrombus can also form in the left ventricle or in a closed aortic root and be ingested by the pump.
Hemolysis markers: lactate dehydrogenase and plasma free hemoglobin
Intravascular hemolysis is the laboratory signature of intrapump thrombosis. Red cells are sheared against clot on the rotor, releasing hemoglobin and lactate dehydrogenase into plasma. Lactate dehydrogenase is the more sensitive and earlier marker. Plasma free hemoglobin is more specific but less stable, because traumatic phlebotomy and delayed processing raise it artifactually. Haptoglobin falls, indirect bilirubin rises, and urinalysis shows hemoglobinuria without red cells. Dark, tea-colored urine and acute kidney injury follow when hemolysis is brisk.
Thresholds are device specific. Baseline lactate dehydrogenase during normal function is higher with the HeartMate II than with centrifugal pumps. A level of 2.5 times the upper limit of normal is commonly regarded as significant hemolysis in HeartMate II recipients, with lower thresholds for centrifugal pumps (Saeed 2023). In 182 HeartMate II recipients, hemolysis by Intermacs criteria (plasma free hemoglobin above 40 mg/dL with clinical signs) occurred in 18%. Hemolysis defined by lactate dehydrogenase of 600 IU/L or more, 2.5 times the laboratory upper limit, occurred in 37%. One-year event-free survival after Intermacs-defined hemolysis was 16%, compared with 85% without hemolysis (Cowger 2014). Lactate dehydrogenase is therefore measured at all clinic visits and trended against the patient's own baseline.
Early treatment of moderate elevations is effective. In the PREVENT study of HeartMate II recipients, 14% developed lactate dehydrogenase of 2.5 times the upper limit of normal or more on two consecutive measurements. Intensified medical therapy at moderate elevations (2.5 to 3.2 times the upper limit) gave sustained resolution in 91% at 6 months, without later pump thrombosis or surgery. When treatment began at higher levels, the figure was 26% (Thenappan 2018). Some rises in lactate dehydrogenase do not reflect pump thrombosis. Hepatic congestion, hemolysis from suction events, hemolysis from a prosthetic valve, malignancy, myocardial or skeletal muscle injury, and pulmonary infarction raise it. Isoenzyme analysis, plasma free hemoglobin, haptoglobin, and pump power separate these causes.
Pump power, flow, and log files in suspected thrombosis
Pump power is a direct measurement of the motor's electrical consumption, while displayed flow is an estimate calculated from speed, power, and the entered hematocrit. Thrombus on the rotor adds friction and raises the power needed to maintain speed. Because the controller interprets power as flow, intrapump thrombosis produces an apparently high flow at the moment that true forward flow falls. An abrupt power spike suggests acute thrombus, and a gradual rise over days to weeks suggests accumulating thrombus. In contrast, a gradual or abrupt fall in power suggests obstruction before the pump (inflow occlusion) or after it (outflow graft kink, twist, or compression) (Saeed 2023).
Interrogation of the controller and analysis of the manufacturer's log files are an obligatory part of the evaluation of suspected thrombosis (Class I, level B). Log files display trends in power, speed, flow, pulsatility index, and alarm history over weeks, and the manufacturer provides graphical analysis. Aortic insufficiency also raises power and flow, because blood recirculates through the pump, and must be excluded by echocardiography before a power rise is attributed to thrombus.
Imaging and ramp testing in suspected pump thrombosis
Echocardiography is the first imaging test. Findings that suggest impeded pump flow include a dilating left ventricle, worsening mitral regurgitation, more frequent aortic valve opening, and lower inflow and outflow cannula velocities than the patient's baseline (Saeed 2023; Estep 2024). Thrombus within the pump cannot be seen, so echocardiography excludes alternative explanations: malposition of the inflow cannula, left ventricular thrombus, right ventricular failure, and aortic regurgitation.
The ramp test records left ventricular end-diastolic dimension, septal position, aortic valve opening, and valvular regurgitation as speed is increased in steps. With a patent circuit the left ventricle becomes smaller as speed rises. In the HeartMate II validation study, all patients with confirmed thrombosis, and no patient without it, had a slope of left ventricular end-diastolic dimension against speed flatter than −0.16 (Uriel 2012). A flat ramp indicates obstruction anywhere in the circuit and does not localize it. Ramp criteria for the HeartMate 3 have been studied less, and in HeartMate 3 recipients a flat ramp more often reflects outflow graft obstruction (Estep 2024). The protocols are described on the durable LVAD page.
The 2023 ISHLT guidelines recommend hospital admission for expedited assessment of suspected thrombosis and serial examinations for hypoperfusion (Class I, level C). The initial evaluation includes log file analysis, laboratory tests for hemolysis, hemoglobinuria, and end-organ function, and an echocardiogram with speed adjustment (Class I, level B). CT angiography of the outflow graft is recommended when suspicion is high (Class I, level C). Right heart catheterization with or without speed changes should be considered in patients with heart failure (Class IIa, level C), and left heart catheterization with graft angiography can be considered to assess graft patency (Class IIb, level C). Echocardiography cannot exclude pump thrombosis, so invasive hemodynamics are added when suspicion persists after a normal or inconclusive study.
Treatment of pump thrombosis
Treatment depends on the device, the hemodynamic state, and the patient's candidacy for surgery or transplantation (Goldstein 2013; Saeed 2023). Patients with suspected or confirmed thrombosis are admitted for intravenous anticoagulation with unfractionated heparin (Class I, level C). A direct thrombin inhibitor such as bivalirudin or argatroban is a reasonable alternative and is required in heparin-induced thrombocytopenia; choosing bivalirudin over heparin as the initial agent might be reasonable (Class IIb, level C). Direct thrombin inhibitors act on clot-bound thrombin and have antiplatelet effects. When hemolysis resolves, warfarin is resumed, often with a higher INR target. Many programs also add or intensify antiplatelet therapy, although evidence for this is insufficient. Glycoprotein IIb/IIIa inhibitors have largely failed to resolve thrombus and cause frequent bleeding, and their safety and efficacy are not established (Class IIb, level B).
Pump exchange is the definitive treatment of confirmed thrombosis in patients who are surgical candidates (Class I, level B). The HeartMate II rarely responds to escalated medical therapy, and persistent hemolysis on medical therapy is associated with stroke and death. In hemodynamically stable patients with a short expected wait, deferring exchange for urgent transplantation may be reasonable (Class IIa, level C). Thrombolysis, systemic or delivered into the left ventricle, can be considered as initial therapy in carefully selected patients with a hydrodynamically levitated centrifugal pump such as the HVAD (Class IIa, level C). In an HVAD registry, initial medical therapy, most often tissue plasminogen activator, succeeded in 48% of episodes and was complicated by hemorrhagic stroke in 21%, whereas all initial pump exchanges succeeded (Stulak 2015). Thrombolysis carries a risk of intracranial hemorrhage and can delay surgery if it fails. Recurrent thrombosis is more common after exchange, and exchange from an older pump to a HeartMate 3 may be reasonable in recurrent cases (Class IIb, level C). Evidence for thrombolysis in the rare HeartMate 3 thrombosis is insufficient to guide practice.
| Feature | Intrapump thrombosis | Inflow obstruction | Outflow graft obstruction | Aortic insufficiency |
|---|---|---|---|---|
| Power | Rises, abruptly or gradually | Falls | Falls, usually gradually | Rises |
| Displayed flow | Falsely high | Low, often with suction events | Low | High, ineffective |
| Hemolysis | Marked | Mild or absent | Absent unless secondary thrombosis | Absent |
| Left ventricle | Dilating; aortic valve opens more often | Variable; cannula contact or thrombus | Dilating; aortic valve opens with all beats | Dilated; valve usually closed |
| Ramp study | Flat | Flat or suction at low speeds | Flat | Dimension falls little; regurgitation increases with speed |
| Confirmatory test | Log files, hemolysis markers, exclusion of other causes; explant | Echocardiography, CT | CT angiography; graft angiography with pressure gradient | Echocardiography |
Distinguishing intrapump thrombosis from other causes of impaired pump output. Patterns are typical, with exceptions, and differ between pump models.
Outflow graft obstruction: twist, kink, and extrinsic compression
Obstruction of the outflow graft has become a leading cause of impaired pump output in HeartMate 3 recipients. It can occur early or late after implantation, and patients present with a gradual or abrupt fall in pump flow, low-flow alarms, and right heart failure (Saeed 2023). The mechanisms are a twist or kink of the graft, thrombosis within it, stenosis at the aortic anastomosis, and extrinsic compression.
Twist. The original HeartMate 3 design allowed the graft to rotate freely at its attachment to the pump, and in a small number of MOMENTUM 3 patients, movement of the pump after implantation twisted the graft, producing a "candy wrapper" obstruction on CT (Mehra 2018). A design modification now locks the graft in its orientation, and twisting is no longer being reported (Saeed 2023).
Extrinsic obstruction at the bend relief. Acellular biodebris and proteinaceous material can accumulate between the outflow graft and the bend relief that covers its origin from the pump, compressing the graft from outside (Goldstein 2024). The process is insidious and late. In a multicenter study of 2,108 HeartMate 3 recipients, extrinsic obstruction, defined as narrowing of more than 25% of the cross-sectional area, was diagnosed in 62 patients (3.0%) at a median of 953 days of support. The cumulative incidence reached 9.1% at 5 years. Of the 62 patients, 27 underwent surgical revision and 15 percutaneous stenting; mortality among those who received an intervention was 17.0% (Wert 2024). Wrapping the graft at implantation, for example with a protective sleeve, can promote accumulation of blood and debris under the wrap (Saeed 2023).
Recognition. The parameter signature is a gradual fall in power and flow at constant speed, without hemolysis (Saeed 2023). Lactate dehydrogenase is not markedly elevated unless stasis has caused secondary thrombosis of the pump. Echocardiography shows a dilating left ventricle, an aortic valve that opens with all beats, more mitral regurgitation, and a flat ramp response. The outflow graft, imaged from a high right parasternal window, may show a focal, turbulent, sustained rise in velocity compared with the patient's own baseline (Estep 2024). The baseline peak velocity is usually below 2 m/s, and the artificial pulse produces intermittent peaks above 2 m/s, so a single value is not diagnostic.
Confirmation. CT angiography timed for the aorta and graft defines twist, kink, external compression, anastomotic stenosis, and graft thrombus (Class I, level B for CT to identify device complications). CT can miss an anastomotic stenosis and may not separate intraluminal thrombus from external compression. Graft angiography with measurement of the pressure gradient across the obstruction can be useful to visualize and guide treatment such as endovascular stenting (Class IIa, level B).
Treatment. Surgical intervention is indicated for documented hemodynamically significant obstruction (Class I, level C). Options include release of the bend relief, graft revision, and pump exchange, and reoperation carries substantial risk in older patients and those with destination therapy. Percutaneous treatment with balloon-expandable or covered stents may be considered in selected patients (Class IIb, level B) and is increasingly used for extrinsic compression, because it avoids redo sternotomy. The stent is deployed within the graft and does not remove the external material, and later progression of compression beyond the stented segment has been reported.
Stroke during LVAD support: incidence, timing, and risk factors
Stroke is cumulatively the leading cause of death after implantation (Saeed 2023). Strokes are ischemic or hemorrhagic, and neurologic events also include transient ischemic attacks, subdural and subarachnoid hemorrhage, and covert infarcts found only on imaging. Acute ischemic stroke can transform into hemorrhage.
In a cohort of 402 HeartMate II and HVAD recipients, 17% had a stroke, a rate of 0.14 events per patient-year, divided almost evenly between ischemic and hemorrhagic events. The risk was bimodal: highest immediately after implantation and rising again after 9 to 12 months (Frontera 2017). In a single-institution analysis cited by the ISHLT guidelines, 96% of early strokes occurred in the first 30 days. The HVAD carried a higher stroke risk: in ENDURANCE, ischemic stroke occurred in 17.6% and hemorrhagic stroke in 14.9% of HVAD recipients at 2 years. MOMENTUM 3 showed no difference in strokes during the first 6 months, but after 180 days the incidence was 3.3 times lower with the HeartMate 3 than with the HeartMate II (Saeed 2023).
| Risk factor | Evidence and mechanism |
|---|---|
| Hypertension | Raises the pressure head across the pump and lowers flow; in HVAD recipients a mean pressure above 85 to 90 mm Hg was associated with stroke and thrombosis, and a blood pressure protocol in the ENDURANCE Supplemental trial reduced neurologic events (Milano 2018) |
| Infection | Among 16,597 Intermacs patients, those with infection had more stroke (18% vs 11%); hemorrhagic stroke was more frequent than ischemic stroke in all infection categories (Shah 2019); in another cohort, stroke followed documented infection by a median of 4 days (Saeed 2023) |
| Female sex | Hazard ratio 1.52 in Intermacs; independent predictor of ischemic and hemorrhagic stroke |
| Anticoagulation outside the target range | A subtherapeutic INR predisposes to embolic stroke and a supratherapeutic INR to hemorrhage |
| Pump thrombosis and hemolysis | Thrombus in the pump embolizes; ongoing hemolysis with watchful waiting was associated with stroke or death |
| Atrial fibrillation, aortic root thrombus, prior stroke | Additional embolic sources; a valve that never opens allows stasis in the aortic root |
| Device type | Higher with the HVAD than with axial-flow and fully levitated pumps |
Risk factors for stroke during LVAD support, from the 2023 ISHLT guidelines and the cited studies. Hypertension and infection are the major modifiable factors.
Blood pressure control is the main preventive measure. The 2023 ISHLT guidelines recommend arterial line monitoring early after implantation (Class I, level C) and consider a mean arterial pressure of 75 to 90 mm Hg reasonable to reduce stroke in hospitalized patients (Class IIa, level B) and to prevent recurrent stroke (Class IIa, level C). Excessive pharmacologic hypotension below a mean of 75 mm Hg is avoided (Class III, level C), because low pressure was associated with higher mortality in Intermacs. Measurement methods and drug choices are described on the durable LVAD page.
Acute management of ischemic and hemorrhagic stroke
LVAD recipients were excluded from the trials behind general stroke guidelines, so centers adapt them through protocols built with neurology. A new deficit or sudden severe headache prompts rapid evaluation that includes a baseline examination, CT and CT angiography of the head and neck, coagulation parameters, platelet count, and neurologic consultation (Class I, level C). The HeartMate 3 is not compatible with MRI. Pump log files are interrogated for evidence of thrombosis, malfunction, or a power interruption (Class I, level C), and the National Institutes of Health Stroke Scale documents severity.
Ischemic stroke. Large-vessel occlusion is common: in one series, nearly one-quarter of LVAD recipients with acute ischemic stroke had an occlusion amenable to clot retrieval (Saeed 2023). Mechanical thrombectomy has been performed safely in LVAD recipients (Al-Mufti 2016). The guidelines state that cerebral angiography with intervention should be considered for thrombotic cerebrovascular occlusion (Class I, level B). They also state that routine interventional treatment of thrombotic strokes, without such selection, is not recommended (Class III, level B). Intravenous thrombolysis is used less often, because most patients are anticoagulated and the risk of intracranial hemorrhage is high. Selective use may be considered when imaging shows no hemorrhage (Class IIb, level B), and routine use without imaging and neurologic discussion is not recommended (Class III, level C). Permissive hypertension, used to support collateral flow, conflicts with the usual LVAD blood pressure target, and the timing of a return to a mean pressure of 75 to 90 mm Hg is decided with neurology. The search for an embolic source includes the INR history, missed doses, carotid ultrasound, transesophageal echocardiography, and defibrillator interrogation for atrial fibrillation (Class IIa, level C).
Hemorrhagic stroke. Anticoagulation is stopped or reversed and antiplatelet therapy is stopped unless neurology advises otherwise (Class I, level C). Reversal uses four-factor prothrombin complex concentrate and vitamin K, and blood pressure is lowered promptly. An arterial line is reasonable for continuous monitoring (Class IIa, level C). Resumption of anticoagulation is decided with neurology or neurosurgery (Class I, level C). The timing balances hematoma expansion against pump thrombosis. Rehabilitation with neurologic follow-up is recommended (Class I, level B), and anticoagulation is monitored closely after any embolic or hemorrhagic event. Escalation of antiplatelet therapy may be considered when an embolic event occurs despite documented adherence to anticoagulation (Class IIb, level C).
Classification and epidemiology of infection
Infection is the most common adverse event of contemporary LVAD support and the only major adverse event that the HeartMate 3 did not reduce. In MOMENTUM 3, major infection occurred in 58% of HeartMate 3 and 56% of HeartMate II recipients over 2 years. Local infections unrelated to pump components were most common (56% of events), followed by driveline infection (27%) and sepsis (16%). Women, patients with a preimplant intra-aortic balloon pump or prior cardiac surgery, and those with a body mass index of 30 kg/m² or more were at higher risk (Patel 2020). Infection is the leading cause of readmission at 30 days and is associated with stroke and death (Saeed 2023).
The ISHLT classification divides infections by their relation to the device (Kusne 2017), and the 2023 guidelines recommend it (Class I).
| Category | Includes | Outcome in Intermacs |
|---|---|---|
| VAD-specific | Infections of the pump or cannula, the pump pocket, or the driveline; occur only in device recipients | 1-year survival 87%; stroke 0.11 events per patient-year |
| VAD-related | Infections that occur in other patients but have special considerations in device recipients: infective endocarditis, bloodstream infection, mediastinitis | 1-year survival 71%; stroke 0.17 events per patient-year |
| Non-VAD | Infections unrelated to the device: pneumonia, urinary tract infection, cholecystitis, Clostridioides difficile colitis | 1-year survival 72%; stroke 0.15 events per patient-year |
ISHLT categories of infection in device recipients. Outcomes from an Intermacs analysis of 16,597 continuous-flow recipients, 42% of whom developed an infection at a median of 69 days (Shah 2019). Bloodstream infection accounted for 92% of VAD-related infections, and 59% of bloodstream infections had no other source.
The most common organisms are Staphylococcus aureus, coagulase-negative staphylococci, and Pseudomonas aeruginosa, with Enterococcus, other gram-negative bacilli, and Candida species in patients with long antibiotic exposure. Biofilm on the driveline velour and other device surfaces protects organisms from antibiotics and host defenses, which explains why deep device infection is rarely cured without removal of the infected component.
Driveline infection: presentation, pathogens, and treatment
The driveline is a permanent breach of the skin. Trauma to the exit site from pulling, twisting, or movement of an unsecured cable disrupts the seal between skin and velour and lets organisms track along the tunnel. Driveline infection is the most common VAD-specific infection, its cumulative incidence rises with duration of support, and untreated infection can progress along the tunnel to the pump (Saeed 2023). Younger patients, obese patients, and patients with a history of substance use are at higher risk.
Infection ranges from superficial exit-site infection, with erythema, tenderness, and drainage, to deep infection of the subcutaneous tunnel and fascia, to involvement of the pump and bloodstream. Fever and leukocytosis are often absent in superficial infection.
Evaluation. The 2023 ISHLT guidelines recommend a complete blood count with differential, creatinine, C-reactive protein, lactate dehydrogenase, INR, and urinalysis (Class I, level C), and at least 3 sets of blood cultures over 24 hours, even without fever (Class I, level C). Exit-site drainage is sent for Gram stain and bacterial culture, with fungal cultures for persistent infection and negative bacterial cultures (Class I, level B). Ultrasound of the exit site characterizes the depth and loculation of an infection confined to the driveline (Class IIa, level C). CT of the chest, abdomen, and pelvis defines the extent of a suspected VAD-specific infection (Class I, level C), and transesophageal echocardiography is recommended to examine the valves and other intracardiac devices (Class I, level C). For driveline cellulitis or dermatitis without evidence of deeper infection, routine CT and transesophageal echocardiography are not recommended (Class III, level C).
Antimicrobial treatment. Empiric therapy covers staphylococci and P aeruginosa and starts promptly once infection is suspected (Class I, level B). An infection limited to the superficial fascia, without systemic signs, is treated with a 2-week course of pathogen-specific antibiotics (Class I, level C). Deep driveline infection and infection of the pocket, pump, or cannula require admission, systemic antibiotics for 6 to 8 weeks, and consideration of surgical drainage with or without vacuum-assisted closure, followed by long-term suppression (Class I, level C).
Imaging device infection: ultrasound, CT, and FDG PET/CT
Imaging defines how far infection extends along the driveline and whether it reaches the pump, the outflow graft, or the bloodstream. Ultrasound shows fluid collections at the exit site and along the superficial tunnel. CT shows fluid and gas along the driveline, a pocket collection, mediastinal involvement, and fluid around the outflow graft, although postoperative changes and metal artifact limit its specificity early after implantation.
Fluorine-18 fluorodeoxyglucose (FDG) PET/CT shows metabolic activity along the whole device and separates isolated driveline infection from infection that extends to the pump or graft. In a meta-analysis of 4 studies and 119 scans, pooled sensitivity was 92% and specificity 83%, with a wide confidence interval for specificity (24% to 99%) (Tam 2020). The 2023 ISHLT guidelines state that FDG PET/CT may be considered to distinguish infection of the pump and pocket from infection of the driveline or the outflow cannula and to guide treatment and prognosis (Class IIb, level B). Radiolabeled leukocyte scintigraphy is an alternative with poorer count statistics and lower sensitivity (Class IIb, level B). Mild, uniform uptake around the outflow graft and along the cannula can reflect a sterile foreign-body reaction, so the pattern, intensity, and clinical context are read together. Patient preparation and interpretation are described on the cardiac FDG PET page.
Pump, pocket, and bloodstream infection and LVAD endocarditis
Infection of the pump, cannulas, or outflow graft is uncommon but carries high mortality; in Intermacs, infection of internal pump components occurred at 0.003 events per patient-year (Shah 2019). Pump pocket infection affects pumps placed in a preperitoneal or abdominal-wall pocket. The HeartMate 3 sits within the pericardium, so pocket infection does not apply. Infection can still involve the pericardial space, the mediastinum, and the outflow graft.
Bloodstream infection is the most frequent VAD-related infection. Persistent or relapsing bacteremia in a device recipient is treated as possible infection of the device. LVAD endocarditis involves the native or prosthetic valves, the inflow cannula, or the left ventricle. Transesophageal echocardiography examines the valves, the cannula, and any intracardiac leads, while PET/CT and CT examine the extracardiac components. A defibrillator system in a patient with bacteremia may harbor lead infection, and management follows the cardiac device infection and infective endocarditis pages, coordinated with the LVAD team.
Suppression, surgical debridement, driveline relocation, and device exchange
After treatment of a deep or device infection, long-term oral suppressive antibiotics are usually continued while the device remains, often indefinitely. Suppression is chosen from culture results and adjusted for drug interactions, especially with warfarin.
Surgery addresses infection that antibiotics cannot sterilize. Debridement of the exit site and tunnel removes infected tissue and velour, and negative pressure wound therapy promotes closure. Relocation of the driveline creates a new tunnel and exit site away from the infected tract, sometimes with the cable embedded beneath the rectus muscle and fascia, and muscle or omental flaps cover exposed components. Pump exchange with removal of all infected hardware, including defibrillators and pacemakers, is considered for recalcitrant infection, as is listing for transplantation (Class I, level C). Device infection does not preclude transplantation, which removes the infected hardware.
Prevention of device infection
Prevention begins at implantation with perioperative antibiotic prophylaxis and an exit site placed away from skin folds. The 2023 ISHLT guidelines recommend assessment of the exit site, the cable, and the connections at all clinic visits, along with assessment of position and immobilization with an anchoring device or binder (Class I, level C). They also recommend training of the patient and caregiver in a standardized dressing technique before discharge, with reinforcement throughout support (Class I, level C). The dressing is not opened routinely in clinic. Showering technique is taught once the site has healed, and immersion of the exit site in baths, pools, or natural water is avoided.
The guidelines consider antibiotic prophylaxis reasonable for dental procedures that manipulate the gingiva or periapical region or perforate the oral mucosa, and for incision and drainage of an abscess (Class IIa, level C). Prophylaxis is not recommended for respiratory, gastrointestinal, or genitourinary procedures (Class III, level C).
Right heart failure after LVAD implantation: definitions and incidence
The LVAD unloads the left ventricle, but the pump can deliver only what the right ventricle presents to it. Right heart failure is a major cause of early morbidity and mortality after implantation. MCS-ARC classifies it by timing (Kormos 2020; Saeed 2023):
| Category | Definition |
|---|---|
| Early acute | Need for a right ventricular assist device (RVAD) implanted at the time of LVAD implantation |
| Early postimplant | Within 30 days: need for a temporary or durable RVAD or extracorporeal membrane oxygenation (ECMO), failure to wean from inotropes or inhaled pulmonary vasodilators within 14 days, or death from right ventricular failure |
| Late | Right heart failure that begins more than 30 days after implantation |
| Clinical signs required | Elevated jugular venous pressure, ascites, or edema; low pump flow or low pulsatility; or laboratory evidence of cardiorenal syndrome or hepatic congestion |
MCS-ARC categories of right heart failure after LVAD implantation, as adopted by the 2023 ISHLT guidelines (Class I, level C for their use).
Incidence depends on the definition used. In 484 HeartMate II bridge-to-transplant patients, about 20% developed right ventricular failure, with 6% requiring an RVAD. Survival to transplantation, recovery, or ongoing support at 180 days was 71% with right ventricular failure and 89% without (Kormos 2010). Independent predictors were a central venous to wedge pressure ratio above 0.63, preoperative ventilator support, and blood urea nitrogen above 39 mg/dL. Preoperative predictors of right ventricular failure are described on the durable LVAD page.
Early right heart failure: mechanisms and management
Several mechanisms converge in the first days. The LVAD increases venous return to a right ventricle that may already be diseased. Unloading of the left ventricle shifts the septum leftward and removes the septal contribution to right ventricular contraction, the largest contributor to right ventricular function. Cardiopulmonary bypass, transfusion, protamine, hypoxemia, and acidosis raise pulmonary vascular resistance and injure the myocardium. Loss of pericardial restraint dilates the right ventricle, and tamponade or sternal compression impairs filling. Excessive pump speed worsens septal shift (Saeed 2023).
Echocardiography shows a small left ventricle with leftward septal bowing, a dilated and hypokinetic right ventricle, severe tricuspid regurgitation, and a dilated inferior vena cava with little respiratory variation, the pattern that distinguishes right ventricular failure from hypovolemia (Estep 2024). Hemodynamic criteria are not standardized. The 2023 ISHLT guidelines cite one combination that has been used: cardiac index below 2.0 L/min/m², mean arterial pressure of 55 mm Hg or less, central venous pressure above 15 mm Hg, mixed venous saturation of 55% or less, and high inotrope requirements. The pulmonary artery pulsatility index (PAPi), calculated as pulmonary artery pulse pressure divided by right atrial pressure, has mean values of 1.3 to 1.8 in severe right heart failure after implantation. It depends on venous, ventricular, and pulmonary factors, so no single threshold applies across populations.
Treatment has five targets:
- Preload. Central venous pressure is kept high enough to fill a stiff ventricle and low enough to avoid distension; a central venous to wedge pressure ratio below 1 after cardiac surgery is associated with a better cardiac index. Diuretics or ultrafiltration remove excess volume.
- Contractility. Inotropes (dobutamine, low-dose dopamine, or milrinone) are started early and liberally, with vasopressors added for systemic hypotension. Inotropes may need to continue beyond 2 weeks when signs of right heart dysfunction persist (Class I, level C).
- Afterload. Inhaled nitric oxide, inhaled or intravenous prostacyclins, and inhaled milrinone lower pulmonary vascular resistance; ventilation, oxygenation, and acid-base balance are optimized. Inhaled nitric oxide improves hemodynamics acutely, but a randomized trial did not show clinical benefit (Saeed 2023).
- Pump speed. Speed is set to keep the septum near the midline without overdecompression of the left ventricle.
- Rhythm. Atrioventricular synchrony and an adequate heart rate are maintained, with pacing through epicardial wires when needed, and atrial and ventricular arrhythmias are treated promptly.
Inotropes and vasopressors for right ventricular failure are described on the vasopressors and inotropes page.
Temporary right ventricular support
An RVAD is indicated when hemodynamics do not improve with medical therapy, before end-organ malperfusion develops or progresses. Early implantation of a surgical or percutaneous RVAD should be considered for refractory right ventricular failure (Class I, level C), and planned RVAD implantation may be considered in patients with high-risk preoperative features (Class IIb, level C). Delayed, unplanned RVAD support has been associated with poor outcomes, including 50% in-hospital mortality in one series, and outcomes appear comparable when support begins before severe end-organ dysfunction (Saeed 2023).
| Device | Configuration | Features |
|---|---|---|
| Surgical centrifugal RVAD (for example, CentriMag) | Right atrium to pulmonary artery, placed at sternotomy; cannulation techniques allow removal without redo sternotomy | Full right-sided support for weeks; oxygenator can be added |
| Impella RP | Percutaneous microaxial pump from the inferior vena cava to the pulmonary artery via the femoral vein | RECOVER RIGHT: 18 of 30 patients had right ventricular failure after LVAD; cardiac index rose from 1.8 to 3.3 L/min/m², 30-day survival 73.3% (Anderson 2015) |
| ProtekDuo dual-lumen cannula | Percutaneous right atrium to pulmonary artery cannula via the internal jugular vein, with an extracorporeal centrifugal pump | Allows ambulation; oxygenator can be added |
| Venoarterial ECMO | Femoral or central cannulation | Supports oxygenation and both ventricles; raises left ventricular afterload |
| Durable biventricular support or total artificial heart | Surgical | Persistent biventricular failure; total artificial heart for device failure, thrombosis, or malignant arrhythmias |
Mechanical right ventricular support after LVAD implantation. Percutaneous devices provide unloading comparable to surgical devices with less morbidity in selected patients; long-term data are limited. Device principles are covered on the temporary mechanical circulatory support and cardiogenic shock pages.
Late right heart failure: causes and management
Late right heart failure appears more than 30 days after implantation, often after a period of apparent stability. It presents with edema, ascites, hepatic congestion, diuretic resistance, cardiorenal syndrome, malnutrition, low pump flow, and readmission, and it impairs quality of life and drug metabolism (Saeed 2023). In a single-center cohort of 183 patients, late right heart failure by MCS-ARC criteria occurred in 17% at a median of 8.8 months; early right heart failure and a larger preoperative right ventricular basal diameter were independent predictors (Wang 2026).
Late right heart failure is not always intrinsic right ventricular disease. Causes that are specific to the device are excluded first: excessive speed with septal shift, inadequate speed with residual left-sided congestion and pulmonary hypertension, aortic insufficiency, outflow graft obstruction, and pump thrombosis. Tricuspid regurgitation, atrial fibrillation, ventricular arrhythmias, volume overload, and pulmonary disease are the other common contributors. Echocardiography and an invasive ramp study with right heart catheterization separate these mechanisms, and speed is adjusted to keep the septum near the midline. Treatment adds loop diuretics with mineralocorticoid receptor antagonists, pulmonary vasodilators such as phosphodiesterase-5 inhibitors in selected patients, and rhythm control. Refractory cases are treated with chronic inotropes, listing for transplantation, or palliation.
De novo aortic insufficiency
Aortic insufficiency develops during support in a native valve that was competent at implantation. The pump holds aortic pressure above left ventricular pressure for most of the cardiac cycle, so the valve opens rarely or never. A valve that stays closed develops commissural fusion, leaflet thickening, and remodeling of the root, and continuous high transvalvular pressure drives regurgitation (Saeed 2023). The regurgitant blood re-enters the pump, creating a closed loop in which pump flow and power rise while effective systemic flow falls and the left ventricle remains dilated.
In an Intermacs analysis of 10,603 continuous-flow recipients, 1,399 developed moderate or severe aortic insufficiency, and its prevalence rose steadily with time on support. Older age, female sex, smaller body mass index, mild aortic insufficiency before implantation, and destination therapy predicted it. Among patients who survived to 1 year, significant aortic insufficiency was associated with more rehospitalization and lower survival (Truby 2018). Lower speeds that allow intermittent opening may reduce the risk of fusion and late insufficiency, provided organ perfusion is preserved (Class IIb, level B).
Grading requires LVAD-specific interpretation. When the valve remains closed, regurgitation is often continuous through systole and diastole. Pressure half-time and descending aortic flow reversal are distorted by continuous pump flow. The American Society of Echocardiography (ASE) recommends integrating vena contracta width, jet duration, left ventricular size, and the outflow graft Doppler signal. An outflow graft systolic-to-diastolic peak velocity ratio below 5.0 supports at least moderate regurgitation (Estep 2024). Rising pump flow and power, a left ventricle that dilates as speed rises, and invasive hemodynamics showing a raised wedge pressure support significance. The 2023 ISHLT guidelines recommend echocardiography at regular intervals to evaluate aortic insufficiency (Class I, level B), at a frequency set by the program but no less than annually (Class IIa, level C).
Treatment of aortic insufficiency, including transcatheter valves
Prevention starts at implantation: more than mild aortic regurgitation is corrected at the time of the operation, with a bioprosthetic valve when replacement is needed (Class I, level C), and closure techniques may be considered in selected patients (Class IIb, level C). Closure leaves the patient completely dependent on the pump and is avoided when myocardial recovery is possible.
Medical management of de novo insufficiency includes control of blood pressure, which lowers the gradient across the closed valve, and diuretics. Increasing speed overcomes mild to moderate regurgitation for a time but increases the regurgitant volume and the closed-loop flow. Speed is adjusted with echocardiography and right heart catheterization to find the setting that best balances unloading against regurgitation.
Hemodynamically significant regurgitation with heart failure is treated by valve intervention or transplantation. Surgical aortic valve replacement or closure through redo sternotomy carries high risk in this population. Transcatheter aortic valve replacement has become the main alternative. Without leaflet calcium, a conventional transcatheter valve has little to anchor against in a dilated annulus, so migration, embolization, and paravalvular regurgitation are risks. In 27 LVAD recipients treated at one center, 37.5% of those who received a standard device needed a second valve, whereas none of those treated with a valve designed for regurgitation or with prestenting of the landing zone did. In-hospital mortality was 3.7% (Hinkov 2024). In a multicenter registry of 30 recipients treated with the JenaValve Trilogy, the first transfemoral system with regulatory approval for aortic regurgitation, device success by Valve Academic Research Consortium 3 (VARC-3) criteria at 30 days was 93.3%. Mortality at 30 days was 3.3% (Hinkov 2026). Because the new valve opens rarely at usual pump speeds, leaflet thrombosis is a concern, and the evidence consists of registries and case series.
Suction events: mechanism, causes, and management
A suction event occurs when the pump draws more blood than the left ventricle receives. The ventricular wall or septum collapses onto the inflow cannula, briefly obstructing inflow. The controller detects a sudden drop in pulsatility as a pulsatility index event, reduces speed, and returns to the set speed; frequent events appear in the log (Estep 2024). Patients may notice palpitations, presyncope, or nothing. Suction triggers ventricular ectopy and ventricular tachycardia through mechanical contact, can cause hemolysis, and in sustained form lowers flow and blood pressure.
The common causes are hypovolemia from bleeding, diarrhea, vomiting, or overdiuresis; right ventricular failure; tamponade; arrhythmia; malposition of the inflow cannula toward the septum or free wall; and a speed set too high for the ventricle's size. Echocardiography shows a small, collapsing left ventricle with septal bowing toward the cannula and a small inferior vena cava when the cause is hypovolemia. Treatment is volume replacement and correction of the cause, with a temporary reduction in speed. Recurrent suction with a normal volume status suggests a cannula or speed problem and prompts echocardiographic speed optimization and CT of cannula position.
Ventricular arrhythmias during LVAD support
Ventricular arrhythmias occur in 20% to 50% of LVAD recipients. The risk is highest early after implantation and decreases with time, and arrhythmias before implantation strongly predict arrhythmias afterward (Saeed 2023). Early arrhythmias arise from the preexisting scar substrate, the postoperative adrenergic state, inotropes, electrolyte shifts, injury at the cannulation site, and QT prolongation. Late arrhythmias reflect the underlying cardiomyopathy, suction, and cannula contact. Most ventricular tachycardia is reentry around preexisting scar, and a minority arises around the inflow cannula.
Hemodynamic tolerance. Because the pump maintains systemic output, ventricular tachycardia and even ventricular fibrillation can be tolerated for minutes to hours. Tolerance depends on the right ventricle: in sustained arrhythmia the right ventricle fails to deliver preload, pump flow falls, and right heart failure, syncope, and end-organ injury follow. The 2023 ISHLT guidelines note that urgent cardioversion may be deferred in favor of less aggressive measures when three conditions hold: cardiac index or indexed pump flow remains above 2 L/min/m², mean arterial pressure exceeds 65 to 70 mm Hg, and end-organ function is preserved. Prolonged arrhythmia of more than 30 minutes on isolated LVAD support is generally avoided regardless of stability (Class IIa, level C), and repeated shocks can acutely depress right ventricular function.
Acute management. New ventricular arrhythmias prompt evaluation of hemodynamic stability (Class I, level C) and LVAD interrogation with echocardiography to detect suction events (Class I, level C). Urgent cardioversion, defibrillation, or antitachycardia pacing is recommended when the arrhythmia causes hemodynamic instability, defined by acute kidney injury, low flows, new or worsening right ventricular failure, symptomatic hypotension, or hypoxia (Class I, level C). Stable patients receive electrolyte correction and an intravenous antiarrhythmic, usually amiodarone, before cardioversion, and a 12-lead electrocardiogram records the morphology (Class IIa, level C). Defibrillation is performed without disconnecting the pump. Reversible causes, including suction, electrolyte disturbance, drug toxicity, and inotropes, are sought (Class I, level C).
For recurrent arrhythmia, amiodarone is a reasonable chronic outpatient treatment (Class IIa, level C), and beta blockers may be useful (Class IIa, level C). Because the pump supports the circulation, LVAD recipients tolerate high doses of beta blockers and other negatively inotropic drugs. Thyroid function is monitored in patients taking amiodarone. Electrical storm, defined in the guidelines as more than 3 episodes of sustained ventricular tachycardia (30 seconds or longer) in 24 hours, may require sedation and intubation to reduce adrenergic drive, catheter ablation, and listing at higher priority. Stellate ganglion block and thoracic epidural anesthesia have limited supporting data. Storm management is described on the electrical storm page.
Catheter ablation of ventricular tachycardia in LVAD recipients
Catheter ablation should be considered for recurrent symptomatic ventricular tachycardia, especially with hemodynamic compromise (Class IIa, level B), and may be needed for refractory unstable tachycardia, performed by electrophysiologists experienced with device recipients (Class IIa, level C). Ablation can also be performed surgically at the time of implantation in selected patients with recurrent arrhythmia.
A systematic review of 18 studies and 110 patients found scar-related reentry as the mechanism in 90.3% of cases and arrhythmia related to the inflow cannula in 19.3% (Anderson 2019). The clinical arrhythmia was rendered noninducible in 77.9%, complications occurred in 9.4%, and ablation terminated electrical storm in 90%. Over a mean follow-up of about 9 months, ventricular tachycardia recurred in 43.6%, 23.4% underwent transplantation, and 48.1% died, without procedure-related deaths. Defibrillator therapies fell from 57.1% to 23.8%.
The apex is inaccessible to mapping where the cannula sits, so circuits around the cannula are approached from its margins. Retrograde aortic access is limited by a closed aortic valve and the outflow graft, so transseptal access is used. Electromagnetic noise from the pump degrades signals, the pump interferes with magnetic mapping systems, and an arterial line monitors blood pressure in the absence of a pulse. Ventricular tachycardia is often hemodynamically tolerated, which allows activation and entrainment mapping that would be impossible without support. Ablation strategy for scar-related tachycardia is covered on the ventricular tachycardia page.
Defibrillator shocks and programming during LVAD support
Defibrillator management during support is summarized on the durable LVAD page. Frequent shocks for arrhythmias that the patient tolerates are a common and distressing problem. Longer detection times, higher detection rates, and additional antitachycardia pacing reduce shocks, and the guidelines state that inactivation of the defibrillator should be considered in recipients with frequent shocks for sustained ventricular tachycardia that causes few or no symptoms (Class IIa, level C). A subcutaneous defibrillator is reprogrammed after implantation because electromagnetic interference from the pump can cause oversensing or undersensing. Biventricular pacing has no proven benefit during support, and turning off the left ventricular lead to save battery life may be considered (Class IIb, level C).
Atrial fibrillation and other atrial arrhythmias
Atrial fibrillation is common: 42% of MOMENTUM 3 patients had it before implantation, and new atrial arrhythmias after implantation are frequent (Saeed 2023). Atrial fibrillation is better tolerated than ventricular arrhythmia, but rapid rates and loss of atrial contraction reduce right ventricular output and can lower pump flow. Atrial arrhythmias have not been associated with increased mortality, thromboembolism, or stroke during support. Rhythm control has not improved outcomes, and exercise capacity at 6 months is lower in patients who have atrial fibrillation at implantation.
Synchronized cardioversion is performed for sudden instability (Class I, level C) and for rapid rates that compromise device performance (Class I, level C). Stable patients undergo volume optimization and medication adjustment first (Class IIa, level C). Rate control uses beta blockers and digoxin, and amiodarone can provide rate control and chemical cardioversion; when amiodarone is used for atrial fibrillation alone, it can often be stopped after 6 to 12 weeks of remission. Otherwise, management follows the American College of Cardiology/American Heart Association (ACC/AHA) atrial fibrillation guidelines (Class IIa, level C). Warfarin for the pump also protects against atrial thromboembolism, so additional anticoagulation is not needed. Data on catheter ablation and left atrial appendage occlusion in LVAD recipients are lacking. General management is described on the atrial fibrillation page.
Device malfunction: driveline damage, controller failure, and power loss
Device malfunction is present when any component fails to operate as specified by the manufacturer. Major malfunction causes inadequate support and leads to death, urgent transplantation, pump exchange, or driveline repair; minor malfunction involves external components that can be repaired or replaced (Saeed 2023). Malfunction is broader than pump failure. In a single-center analysis of 213 HeartMate II and HVAD recipients, controller failure accounted for 30% of malfunctions, battery failure for 19%, and patient cable failure for 14%, whereas only 13% involved the pump itself (Kormos 2017). At least one malfunction occurred in 51% of HeartMate II and 36% of HVAD recipients in that cohort (Saeed 2023).
The external driveline is the most vulnerable component. Repeated bending at the exit site or near the controller fractures the conductors or damages the insulation. Partial damage causes intermittent driveline fault alarms, short power interruptions recorded in the log, and erratic speed or power readings. Complete fracture of both conductor sets stops the pump. Power loss also follows disconnection of both power sources, depleted batteries, and controller failure. Patients carry a backup controller and charged batteries and are trained in controller exchange.
The 2023 ISHLT guidelines describe a structured inspection after the patient is stabilized. The controller is interrogated for alarm type, frequency, and duration; connections are inspected for corrosion, debris, and damage; the driveline is followed from exit site to controller to inspect its sheath; and batteries are checked for charge and secure contacts. A suspected driveline fault is imaged with radiography of the whole driveline, and the manufacturer is notified. External driveline fractures can be repaired by the manufacturer, while other causes of major malfunction require pump exchange or transplantation.
Pump stoppage is a medical emergency: the patient is assessed for circulatory stability and transferred rapidly to the implanting center or another expert center (Class I, level C). With the pump stopped, the valveless outflow graft allows blood to flow backward from the aorta into the left ventricle, adding a regurgitant load to a failing ventricle. Patients with a functioning pump and alarms that suggest impending electrical or mechanical failure are also transferred urgently (Class I, level C). Definitive therapy for major malfunction that cannot be resolved by external repair is pump exchange or transplantation (Class I, level C). Inotropes or extracorporeal membrane oxygenation are reasonable bridges in patients with malperfusion (Class IIa, level C). For patients who cannot undergo surgery, percutaneous occlusion of the outflow graft to stop backflow might be considered (Class IIb, level C), and inotropes might be considered for palliation (Class IIb, level C).
Troubleshooting alarms and parameter changes by complication
The table organizes the controller values (speed, power, estimated flow, and pulsatility index) by complication, with the first tests that confirm the patterns. Values are compared with the patient's own baseline, because normal ranges differ among patients and pump models. Alarm categories and the meaning of the parameters are described on the durable LVAD page.
| Complication | Flow | Power | Pulsatility | Central venous pressure | First tests |
|---|---|---|---|---|---|
| Hypovolemia or bleeding | Low | Low | Low; PI events | Low | Hemoglobin, stool examination, echocardiography (small LV, small IVC) |
| Right heart failure | Low | Low | Low | High | Echocardiography (small LV, dilated RV, plethoric IVC), right heart catheterization |
| Tamponade | Low | Low | Low | High | Echocardiography, often transesophageal; CT |
| Hypertension | Low | Low | Variable | Normal | Arterial line or Doppler pressure |
| Vasodilation or sepsis | High | High | Low | Low or normal | Blood cultures, lactate, source search |
| Intrapump thrombosis | Falsely high | High | Low | Variable | LDH, plasma free hemoglobin, urinalysis, log files, ramp study |
| Outflow graft obstruction | Low | Low, gradual fall | Variable | Often high | Outflow graft Doppler, CT angiography, graft angiography |
| Aortic insufficiency | High | High | Low | Normal or high | Echocardiography of the aortic valve and outflow graft signal |
| Ventricular arrhythmia | Low | Low | Low | High with RV failure | ECG, defibrillator interrogation, electrolytes, echocardiography for suction |
| Driveline or controller fault | Intermittent loss | Intermittent loss | Erratic | Variable | Controller log, driveline inspection, radiography |
Parameter patterns by complication, adapted from the 2023 ISHLT guidelines (Table 2) and the ASE recommendations (Estep 2024). Flow is an estimate derived from power, speed, and hematocrit, so any process that raises power also raises displayed flow. IVC, inferior vena cava; LDH, lactate dehydrogenase; LV, left ventricle; PI, pulsatility index; RV, right ventricle.
Hypotension and the low-flow state
Hypotension in an LVAD recipient is classified by pump flow. Low flow with low pressure points to a preload problem: bleeding, hypovolemia, right ventricular failure, tamponade, sustained arrhythmia, or inflow obstruction. It can also reflect an outflow problem, such as graft obstruction. High flow with low pressure points to vasodilation, most often sepsis, and also to aortic insufficiency with ineffective forward flow, or to a falsely high estimate from intrapump thrombosis. The echocardiographic loading patterns described on the durable LVAD page separate suction, right ventricular failure, and inadequate unloading.
Pressure is measured by arterial line or Doppler, because the narrow pulse pressure defeats automated cuffs and many recipients have no palpable pulse. A mean arterial pressure of 75 mm Hg or more usually indicates adequate perfusion (Class I, level C). Mental status, skin perfusion, capillary refill, lactate, and urine output complete the assessment. Volume is given in small boluses with repeated assessment of flow and the left ventricle. Vasopressors treat vasodilation, and inotropes treat right ventricular failure. Pump speed is not increased to treat low flow until suction and right ventricular failure are excluded.
Cardiac arrest and chest compressions in LVAD recipients
In-hospital cardiac arrest occurred in 14% of 111 LVAD recipients in one center, and delays in starting cardiopulmonary resuscitation were common and associated with higher mortality (Saeed 2023). Assessment relies on perfusion, the pump hum, the controller, and the power connections. In an unresponsive patient, a Doppler opening pressure below 50 mm Hg or an end-tidal carbon dioxide below 20 mm Hg indicates markedly reduced perfusion. Chest compressions are generally advised in an unconscious patient on isolated LVAD support without adequate perfusion, unless advance directives state otherwise (Class IIa, level B). Defibrillation can be performed safely (Class I, level B), and attempts to restore pump function proceed in parallel. Reported cannula dislodgement from compressions is rare (Peberdy 2017). The algorithm is described on the resuscitation in special circumstances page.
Acute kidney injury after LVAD implantation
Acute kidney injury after LVAD implantation results from preoperative cardiorenal syndrome, cardiopulmonary bypass, hypotension and vasoplegia, nephrotoxic contrast and antibiotics, right ventricular failure with renal venous congestion, and pigment injury from hemolysis. In a meta-analysis of 56 cohort studies and 63,663 patients, the pooled incidence was 24.9%, rising to 36.9% when standard criteria were applied (Risk, Injury, Failure, Loss, End-stage kidney disease [RIFLE]; Acute Kidney Injury Network [AKIN]; or Kidney Disease: Improving Global Outcomes [KDIGO]). Kidney replacement therapy was needed in 12.6%, a proportion that has fallen over time (Thongprayoon 2020). Acute kidney injury was associated with higher mortality at 30 days (odds ratio 3.66) and 1 year (odds ratio 2.22), and kidney replacement therapy with still higher mortality.
Sources of error in the management of LVAD complications
- Repeating negative endoscopies indefinitely in recurrent bleeding, when small-bowel evaluation, antiangiogenic therapy, and antithrombotic adjustment are the remaining options.
- Treating a raised plasma free hemoglobin from a hemolyzed sample as device hemolysis. Lactate dehydrogenase, haptoglobin, and urinalysis confirm the finding.
- Treating driveline cellulitis with a short course when the infection extends deeper. Ultrasound or CT defines the depth, and deep infection needs 6 to 8 weeks of antibiotics and surgical assessment.
- Ordering a 2-week antibiotic course for bacteremia in a device recipient without excluding device infection by transesophageal echocardiography, CT, or PET/CT.
- Escalating antiarrhythmic therapy for ventricular tachycardia caused by suction from hypovolemia or cannula contact.
References
- Saeed D, Feldman D, El Banayosy A, et al. The 2023 International Society for Heart and Lung Transplantation Guidelines for Mechanical Circulatory Support: a 10-year update. J Heart Lung Transplant. 2023;42:e1–e222. doi:10.1016/j.healun.2022.12.004
- Kormos RL, Antonides CFJ, Goldstein DJ, et al. Updated definitions of adverse events for trials and registries of mechanical circulatory support: a consensus statement of the Mechanical Circulatory Support Academic Research Consortium. J Heart Lung Transplant. 2020;39:735–750. doi:10.1016/j.healun.2020.03.010
- Mehra MR, Uriel N, Naka Y, et al. A Fully Magnetically Levitated Left Ventricular Assist Device – Final Report. N Engl J Med. 2019;380:1618–1627. doi:10.1056/nejmoa1900486
- Mehra MR, Goldstein DJ, Cleveland JC, et al. Five-Year Outcomes in Patients With Fully Magnetically Levitated vs Axial-Flow Left Ventricular Assist Devices in the MOMENTUM 3 Randomized Trial. JAMA. 2022;328:1233–1242. doi:10.1001/jama.2022.16197
- Nayak A, Cascino TM, DeFilippis EM, et al. The Society of Thoracic Surgeons Intermacs 2025 Annual Report: Focus on Outcomes in Older Adults. Ann Thorac Surg. 2026;121:15–38. doi:10.1016/j.athoracsur.2025.10.005
- Mehra MR, Netuka I, Uriel N, et al. Aspirin and Hemocompatibility Events With a Left Ventricular Assist Device in Advanced Heart Failure: the ARIES-HM3 randomized clinical trial. JAMA. 2023;330:2171–2181. doi:10.1001/jama.2023.23204
- Uriel N, Pak SW, Jorde UP, et al. Acquired von Willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation. J Am Coll Cardiol. 2010;56:1207–1213. doi:10.1016/j.jacc.2010.05.016
- Tabit CE, Chen P, Kim GH, et al. Elevated angiopoietin-2 level in patients with continuous-flow left ventricular assist devices leads to altered angiogenesis and is associated with higher nonsurgical bleeding. Circulation. 2016;134:141–152. doi:10.1161/CIRCULATIONAHA.115.019692
- Draper KV, Huang RJ, Gerson LB. GI bleeding in patients with continuous-flow left ventricular assist devices: a systematic review and meta-analysis. Gastrointest Endosc. 2014;80:435–446.e1. doi:10.1016/j.gie.2014.03.040
- Converse MP, Sobhanian M, Taber DJ, et al. Effect of angiotensin II inhibitors on gastrointestinal bleeding in patients with left ventricular assist devices. J Am Coll Cardiol. 2019;73:1769–1778. doi:10.1016/j.jacc.2019.01.051
- Shah KB, Gunda S, Emani S, et al. Multicenter evaluation of octreotide as secondary prophylaxis in patients with left ventricular assist devices and gastrointestinal bleeding. Circ Heart Fail. 2017;10:e004500. doi:10.1161/CIRCHEARTFAILURE.117.004500
- Juricek C, Imamura T, Nguyen A, et al. Long-acting octreotide reduces the recurrence of gastrointestinal bleeding in patients with a continuous-flow left ventricular assist device. J Card Fail. 2018;24:249–254. doi:10.1016/j.cardfail.2018.01.011
- Vukelic S, Vlismas PP, Patel SR, et al. Digoxin is associated with a decreased incidence of angiodysplasia-related gastrointestinal bleeding in patients with continuous-flow left ventricular assist devices. Circ Heart Fail. 2018;11:e004899. doi:10.1161/CIRCHEARTFAILURE.118.004899
- El Rafei A, Trachtenberg BH, Schultz J, et al. Association between digoxin use and gastrointestinal bleeding in contemporary continuous flow left ventricular assist device support. J Heart Lung Transplant. 2021;40:671–676. doi:10.1016/j.healun.2021.03.005
- Marchetti M, Pitta Gros B, Muller O, et al. Impact of digoxin utilization on gastrointestinal bleeding in patients with continuous-flow left ventricular assist devices: a systematic review and meta-analysis. ASAIO J. 2026;72:137–144. doi:10.1097/MAT.0000000000002473
- Namdaran P, Zikos TA, Pan JY, Banerjee D. Thalidomide use reduces risk of refractory gastrointestinal bleeding in patients with continuous flow left ventricular assist devices. ASAIO J. 2020;66:645–651. doi:10.1097/MAT.0000000000001054
- Patel SR, Madan S, Saeed O, et al. Association of nasal mucosal vascular alterations, gastrointestinal arteriovenous malformations, and bleeding in patients with continuous-flow left ventricular assist devices. JACC Heart Fail. 2016;4:962–970. doi:10.1016/j.jchf.2016.08.005
- Starling RC, Moazami N, Silvestry SC, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370:33–40. doi:10.1056/NEJMoa1313385
- Cowger JA, Romano MA, Shah P, et al. Hemolysis: a harbinger of adverse outcome after left ventricular assist device implant. J Heart Lung Transplant. 2014;33:35–43. doi:10.1016/j.healun.2013.08.021
- Thenappan T, Stulak JM, Agarwal R, et al. Early intervention for lactate dehydrogenase elevation improves clinical outcomes in patients with the HeartMate II left ventricular assist device: insights from the PREVENT study. J Heart Lung Transplant. 2018;37:25–32. doi:10.1016/j.healun.2017.10.017
- Goldstein DJ, John R, Salerno C, et al. Algorithm for the diagnosis and management of suspected pump thrombus. J Heart Lung Transplant. 2013;32:667–670. doi:10.1016/j.healun.2013.05.002
- Uriel N, Morrison KA, Garan AR, et al. Development of a Novel Echocardiography Ramp Test for Speed Optimization and Diagnosis of Device Thrombosis in Continuous-Flow Left Ventricular Assist Devices. J Am Coll Cardiol. 2012;60:1764–1775. doi:10.1016/j.jacc.2012.07.052
- Stulak JM, Dunlay SM, Sharma S, et al. Treatment of device thrombus in the HeartWare HVAD: success and outcomes depend significantly on the initial treatment strategy. J Heart Lung Transplant. 2015;34:1535–1541. doi:10.1016/j.healun.2015.10.020
- Wert L, Stewart GC, Mehra MR, et al. A multicenter evaluation of external outflow graft obstruction with a fully magnetically levitated left ventricular assist device. J Thorac Cardiovasc Surg. 2024;167:1322–1330.e6. doi:10.1016/j.jtcvs.2022.09.051
- Goldstein DJ, Kanwar M, Cowger J, et al. Extrinsic outflow graft obstruction of the HeartMate 3 LVAD: a state-of-the-art review. J Heart Lung Transplant. 2024;43:1873–1880. doi:10.1016/j.healun.2024.07.005
- Mehra MR, Salerno C, Naka Y, et al. A tale of the twist in the outflow graft: an analysis from the MOMENTUM 3 trial. J Heart Lung Transplant. 2018;37:1281–1284. doi:10.1016/j.healun.2018.08.011
- Frontera JA, Starling R, Cho SM, et al. Risk factors, mortality, and timing of ischemic and hemorrhagic stroke with left ventricular assist devices. J Heart Lung Transplant. 2017;36:673–683. doi:10.1016/j.healun.2016.12.010
- Milano CA, Rogers JG, Tatooles AJ, et al. HVAD: The ENDURANCE Supplemental Trial. JACC Heart Fail. 2018;6:792–802. doi:10.1016/j.jchf.2018.05.012
- Al-Mufti F, Bauerschmidt A, Claassen J, et al. Neuroendovascular interventions for acute ischemic strokes in patients supported with left ventricular assist devices: a single-center case series and review of the literature. World Neurosurg. 2016;88:199–204. doi:10.1016/j.wneu.2015.12.061
- Kusne S, Mooney M, Danziger-Isakov L, et al. An ISHLT consensus document for prevention and management strategies for mechanical circulatory support infection. J Heart Lung Transplant. 2017;36:1137–1153. doi:10.1016/j.healun.2017.06.007
- Shah P, Birk SE, Cooper LB, et al. Stroke and death risk in ventricular assist device patients varies by ISHLT infection category: an INTERMACS analysis. J Heart Lung Transplant. 2019;38:721–730. doi:10.1016/j.healun.2019.02.006
- Patel CB, Blue L, Cagliostro B, et al. Left ventricular assist systems and infection-related outcomes: a comprehensive analysis of the MOMENTUM 3 trial. J Heart Lung Transplant. 2020;39:774–781. doi:10.1016/j.healun.2020.03.002
- Tam MC, Patel VN, Weinberg RL, et al. Diagnostic accuracy of FDG PET/CT in suspected LVAD infections: a case series, systematic review, and meta-analysis. JACC Cardiovasc Imaging. 2020;13:1191–1202. doi:10.1016/j.jcmg.2019.04.024
- Kormos RL, Teuteberg JJ, Pagani FD, et al. Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg. 2010;139:1316–1324. doi:10.1016/j.jtcvs.2009.11.020
- Anderson MB, Goldstein J, Milano C, et al. Benefits of a novel percutaneous ventricular assist device for right heart failure: the prospective RECOVER RIGHT study of the Impella RP device. J Heart Lung Transplant. 2015;34:1549–1560. doi:10.1016/j.healun.2015.08.018
- Wang Y, Vu T, Farag J, et al. Incidence and outcomes of late right heart failure following left ventricular assist device implantation. ASAIO J. 2026. Published online. doi:10.1097/MAT.0000000000002788
- Truby LK, Garan AR, Givens RC, et al. Aortic insufficiency during contemporary left ventricular assist device support: analysis of the INTERMACS registry. JACC Heart Fail. 2018;6:951–960. doi:10.1016/j.jchf.2018.07.012
- Hinkov H, Lee CB, Pitts L, et al. Transcatheter management of pure native aortic valve regurgitation in patients with left ventricular assist device. Eur J Cardiothorac Surg. 2024;65:ezae028. doi:10.1093/ejcts/ezae028
- Hinkov H, Yousefzada S, Just IA, et al. Early multicenter outcomes of dedicated transcatheter aortic valve replacement for left ventricular assist device-associated pure aortic regurgitation. Artif Organs. 2026;50:1427–1433. doi:10.1111/aor.70226
- Anderson RD, Lee G, Virk S, et al. Catheter ablation of ventricular tachycardia in patients with a ventricular assist device: a systematic review of procedural characteristics and outcomes. JACC Clin Electrophysiol. 2019;5:39–51. doi:10.1016/j.jacep.2018.08.009
- Kormos RL, McCall M, Althouse A, et al. Left ventricular assist device malfunctions: it is more than just the pump. Circulation. 2017;136:1714–1725. doi:10.1161/CIRCULATIONAHA.117.027360
- Thongprayoon C, Lertjitbanjong P, Cheungpasitporn W, et al. Incidence and impact of acute kidney injury on patients with implantable left ventricular assist devices: a meta-analysis. Ren Fail. 2020;42:495–512. doi:10.1080/0886022X.2020.1768116
- Peberdy MA, Gluck JA, Ornato JP, et al. Cardiopulmonary Resuscitation in Adults and Children With Mechanical Circulatory Support: A Scientific Statement From the American Heart Association. Circulation. 2017;135:e1115–e1134. doi:10.1161/cir.0000000000000504
- Estep JD, Nicoara A, Cavalcante J, et al. Recommendations for Multimodality Imaging of Patients With Left Ventricular Assist Devices and Temporary Mechanical Support: Updated Recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2024;37:820–871. doi:10.1016/j.echo.2024.06.005