Library › Devices and mechanical support › Durable LVADs
Durable left ventricular assist devices: candidacy, implant and management
Durable left ventricular assist devices in adults with advanced heart failure, from candidate selection and risk assessment through implantation, long-term device management, and outcomes.
A durable left ventricular assist device (LVAD) is an implanted continuous-flow pump that takes blood from the left ventricular apex and returns it to the ascending aorta, supporting the circulation for years in advanced (stage D) heart failure with reduced ejection fraction. The fully magnetically levitated HeartMate 3 has been the only durable LVAD implanted commercially in the United States since the HVAD left the market in 2021. Implantation is Class 1 in selected New York Heart Association (NYHA) class IV patients who depend on intravenous inotropes or temporary mechanical support and Class 2a in selected NYHA class IV patients despite guideline-directed therapy. Contemporary registry survival with the HeartMate 3 is about 86% at 1 year and 60% at 5 years. Long-term care includes a warfarin-only antithrombotic regimen for most HeartMate 3 recipients, blood pressure control, and echocardiographic and invasive speed optimization. It also requires driveline care and early recognition of right heart failure, bleeding, infection, stroke, and pump dysfunction.
Durable left ventricular assist devices in advanced heart failure
A durable LVAD is an implantable rotary blood pump that unloads the left ventricle and provides systemic flow for months to years. Its inflow cannula sits in the left ventricular apex, the pump lies against the apex in the pericardial space, and an outflow graft carries blood to the ascending aorta. A percutaneous driveline exits the abdominal wall and connects the pump to an external controller and power source. The device works in series with the native left ventricle: blood that enters the ventricle from the left atrium leaves through the pump and, when ventricular pressure is high enough, through the aortic valve.
The 2023 International Society for Heart and Lung Transplantation (ISHLT) guidelines for mechanical circulatory support use the term durable mechanical circulatory support for all long-term implantable pumps, including left ventricular, biventricular, and total artificial heart configurations. "Durable" distinguishes these systems from temporary devices such as the intra-aortic balloon pump, microaxial flow pumps, and venoarterial extracorporeal membrane oxygenation (VA-ECMO), which are covered on the cardiogenic shock page. Temporary support is designed for days to weeks in hospital; a durable LVAD is designed for discharge home and life outside the hospital.
The indication is advanced heart failure with reduced ejection fraction (HFrEF). The 2022 American Heart Association (AHA), American College of Cardiology (ACC), and Heart Failure Society of America (HFSA) guideline defines stage D as marked symptoms that interfere with daily life and recurrent hospitalizations despite attempts to optimize guideline-directed medical therapy (GDMT). Most candidates have a left ventricular ejection fraction (LVEF) of 25% or less, NYHA class IIIb to IV symptoms, intolerance of GDMT, and objective evidence of low output or end-organ hypoperfusion. Stage D recognition and the overall advanced heart failure pathway are summarized on the heart failure with reduced ejection fraction page.
Implant volumes and the population receiving durable LVADs
The Society of Thoracic Surgeons Interagency Registry for Mechanically Assisted Circulatory Support (STS Intermacs) collects data on durable devices implanted in the United States. Its 2025 annual report includes 28,029 continuous-flow LVADs implanted between 2015 and 2024. Annual implant volume has been flat for several years despite improving outcomes. The report calls for revised eligibility criteria to close the gap between the patients who could benefit and those who receive a device (Nayak 2026). The 2024 report, built on 29,634 continuous-flow devices from 2014 to 2024, showed the transition to magnetically levitated pumps and the resulting survival gain (Meyer 2025).
Recipients are predominantly men. In MOMENTUM 3, about 61% of the 1,020 patients were implanted as destination therapy. In the 5-year extended-phase cohort, 18% were women and the median age was 62 years (Mehra 2022). Ischemic and nonischemic dilated cardiomyopathy account for most implants. Restrictive and hypertrophic cardiomyopathies, in which a small left ventricular cavity limits pump filling, are a minority and carry higher risk.
Acuity at implant has shifted. In the STS Intermacs cohort of 5,787 HeartMate 3 recipients implanted from 2020 to 2024 in Intermacs profiles 1 to 3, two-thirds (66.7%) were on temporary mechanical support at the time of implantation. Use of high-flow microaxial pumps as a bridge rose from 19.1% to 35.1% over the period (Kanwar 2025). Revisions to the US heart allocation system, which raised the priority of patients on temporary support, account for part of this trend.
From pulsatile pumps to magnetic levitation: generations of durable LVADs
Durable LVAD engineering has progressively reduced pump size, mechanical wear, and blood trauma. Three generations are recognized.
First generation: pulsatile volume-displacement pumps. The HeartMate XVE (vented electric) and similar devices contained a pusher plate that filled and ejected a blood sac, with inflow and outflow valves, producing a physiologic pulse. They were large, required a big preperitoneal or intra-abdominal pocket, and often failed mechanically within 2 years. In the Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) trial, 129 transplant-ineligible patients with NYHA class IV heart failure were randomized to an LVAD or optimal medical management. The device reduced the risk of death from any cause by 48% (relative risk 0.52). Survival at 1 year was 52% with the device and 25% with medical therapy, and at 2 years 23% and 8%. Serious adverse events were 2.35 times as frequent with the device, mostly infection, bleeding, and device malfunction (Rose 2001). REMATCH established destination therapy and remains the only randomized comparison of a durable LVAD with medical therapy.
Second generation: continuous-flow axial pumps. The HeartMate II used a single rotor suspended on blood-washed mechanical bearings, usually operating at about 8,000 to 10,000 revolutions per minute (rpm). It was a fraction of the size of the pulsatile pumps and far more durable. In the bridge-to-transplant study of 133 listed patients, 75% reached the principal outcome of transplantation, recovery, or ongoing support while eligible at 180 days, and survival during support was 68% at 12 months (Miller 2007). In the destination therapy trial, 200 transplant-ineligible patients were randomized 2:1 to the HeartMate II or the HeartMate XVE. The primary end point, survival at 2 years free of disabling stroke and reoperation to repair or replace the device, was met by 46% with the continuous-flow pump and 11% with the pulsatile pump. Two-year survival was 58% against 24% (Slaughter 2009). The contact bearings generated heat and shear, and pump thrombosis became the defining failure of the axial era.
Third generation: centrifugal pumps with noncontact bearings. The HeartWare HVAD was a small centrifugal pump with a hydrodynamically and magnetically suspended impeller, implanted entirely within the pericardium. In ENDURANCE, 446 transplant-ineligible patients were randomized 2:1 to the HVAD or the HeartMate II. The HVAD was noninferior for survival free of disabling stroke or device removal for malfunction at 2 years (55.4% against 59.1%), with fewer device replacements (8.8% against 16.2%) but more strokes (29.7% against 12.1%) (Rogers 2017). The ENDURANCE Supplemental trial of 465 patients then tested an enhanced blood pressure protocol. The protocol lowered mean arterial pressure and reduced hemorrhagic stroke by 50.5% compared with the HVAD arm of ENDURANCE, but the prespecified noninferiority end point for stroke was not met (14.7% against 12.1%) (Milano 2018). In 2021, the US Food and Drug Administration (FDA) announced that the manufacturer had stopped distribution and sale of the HVAD. The announcement cited higher rates of neurologic adverse events and mortality than with other durable pumps, and reports of pumps that failed to restart. Patients already supported by the HVAD continue to be followed.
The HeartMate 3 is a centrifugal pump whose rotor is fully suspended by magnetic levitation, with wide blood-flow gaps and no mechanical contact. It was approved by the FDA as a bridge to transplant in 2017 and for destination therapy in 2018, after the MOMENTUM 3 trial. Implantation through a lateral thoracotomy is also approved.
| Feature | Pulsatile (HeartMate XVE) | Axial continuous flow (HeartMate II) | Centrifugal, hydrodynamic and magnetic suspension (HVAD) | Centrifugal, full magnetic levitation (HeartMate 3) |
|---|---|---|---|---|
| Mechanism | Pusher plate fills and empties a blood sac; inflow and outflow valves | Single rotor on blood-washed contact bearings | Impeller suspended without contact bearings | Rotor fully levitated by magnets; wide flow gaps |
| Pump position | Abdominal or preperitoneal pocket | Preperitoneal pocket below the diaphragm | Intrapericardial | Intrapericardial |
| Arterial pulse | Physiologic | Diminished or absent | Diminished or absent | Diminished; programmed artificial pulse |
| Defining weakness | Mechanical failure, often within 2 years; size | Pump thrombosis; bleeding | Stroke; pump restart failures | Bleeding, infection, and right heart failure |
| Pivotal evidence | REMATCH (2001) | Bridge study (2007); destination therapy trial (2009) | ENDURANCE (2017); ENDURANCE Supplemental (2018) | MOMENTUM 3 (2019, 5-year 2022); ARIES-HM3 (2023) |
| US status | Retired | Superseded; implanted patients still followed | Withdrawn from sale in 2021 | Current device |
Device generations of durable LVADs. Other centrifugal pumps used outside the United States or in trials are not shown.
MOMENTUM 3 and the HeartMate 3 evidence base
The Multicenter Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy with HeartMate 3 (MOMENTUM 3) randomized 1,028 patients with advanced heart failure to the HeartMate 3 or the HeartMate II, irrespective of the intended goal of therapy. The primary end point was survival at 2 years free of disabling stroke or reoperation to replace or remove a malfunctioning device. It was reached by 76.9% of the centrifugal-flow group and 64.8% of the axial-flow group (relative risk 0.84, P<0.001 for superiority). Pump replacement was required in 2.3% against 11.3% (relative risk 0.21). Stroke of any severity, major bleeding, and gastrointestinal hemorrhage were all less frequent per patient-year with the HeartMate 3 (Mehra 2019).
The extended phase followed patients still on support at 2 years (477 of 536 contributed data), and 5-year outcomes were estimated in the full per-protocol population of 1,020 patients. The 5-year Kaplan-Meier estimate of survival to transplant, recovery, or ongoing support free of disabling stroke or pump replacement was 54.0% with the HeartMate 3 and 29.7% with the HeartMate II (hazard ratio 0.55). Overall survival was 58.4% against 43.7% (hazard ratio 0.72). Over 5 years, pump thrombosis occurred at 0.010 against 0.108 events per patient-year, stroke at 0.050 against 0.136, and bleeding at 0.430 against 0.765. Infection, arrhythmia, and right ventricular failure did not differ between pumps (Mehra 2022). Transplantation occurred in 51.5% of HeartMate 3 recipients implanted as a bridge to transplant or candidacy and in 17.0% of those implanted as destination therapy.
The European ELEVATE registry of 540 patients confirmed the trial in practice: survival in the primary implant cohort was 83% at 2 years, with stroke in 10.2%, gastrointestinal bleeding in 9.7%, pump thrombosis in 1.5%, and outflow graft twist in 3.5% (Zimpfer 2020). MOMENTUM 3 showed a marked reduction in pump thrombosis, the defining problem of axial pumps, but bleeding, driveline infection, and right heart failure remained the main long-term limitations.
HeartMate 3 system components
The implanted system consists of the pump with its integrated inflow cannula, the outflow graft, and the percutaneous driveline. The external system consists of a wearable controller, two batteries, and a power module that connects to a wall outlet for sleep.
| Component | Description | Clinical relevance |
|---|---|---|
| Inflow cannula | Short titanium cannula integrated with the pump, inserted through a sewing ring cored into the left ventricular apex | Directed toward the mitral orifice, away from the septum and free wall; malposition causes suction, low flow, thrombus, and ventricular ectopy |
| Pump | Centrifugal pump with a fully magnetically levitated rotor, seated in the pericardium against the apex | No bearing wear; low rate of pump thrombosis; speed set between 3,000 and 9,000 rpm |
| Outflow graft | Polyester graft with a bend relief, anastomosed end to side to the ascending aorta | Can twist, kink, or be compressed from outside by accumulated material, causing late low flow; assessed by Doppler and CT angiography |
| Driveline | Percutaneous lead carrying power and data, tunneled to an abdominal exit site | Permanent breach of the skin barrier and the site of the most common device infection; requires anchoring and sterile dressing care |
| System controller | Wearable computer that drives the pump, displays parameters, and issues alarms | A backup controller travels with the patient at all times |
| Power | Two 14-volt lithium-ion batteries (about 10 to 17 hours per pair), a power module (wall outlet), or a mobile power unit | Patients connect to wall power for sleep; loss of both power sources stops the pump |
Component details from the manufacturer's clinician guide to HeartMate 3 pump parameters and the 2024 American Society of Echocardiography recommendations.
Physiology of continuous-flow support
A rotary pump generates flow in proportion to its speed and in inverse relation to the pressure difference it must overcome, which is aortic pressure minus left ventricular pressure. During systole the ventricle raises its own pressure, the pressure difference falls, and pump flow rises; during diastole the difference widens and flow falls. The pump therefore produces a continuous flow with a modest systolic peak, and the size of that peak reflects the contractility and filling of the native ventricle. Three consequences govern daily care.
Preload dependence. The pump can move only the blood that reaches the left ventricle. Hypovolemia, right ventricular failure, tamponade, pulmonary embolism, and arrhythmias that impair right ventricular output all reduce left ventricular filling and pump flow. When filling is low and speed high, the ventricular walls collapse toward the inflow cannula, producing suction events, ventricular ectopy, and transient flow drops.
Afterload sensitivity. At any given speed, a higher arterial pressure lowers pump flow. Hypertension therefore reduces forward output, and it also raises the risk of stroke, particularly hemorrhagic stroke.
Loss of pulsatility. Arterial pulse pressure narrows, the radial pulse may be absent, and automated cuffs and pulse oximeters may fail. The aortic valve opens only when the ventricle generates more pressure than the pump-supported aorta; at higher speeds it may stay closed. The HeartMate 3 adds a programmed artificial pulse. At 2-second intervals the rotor speed falls 2,000 rpm below the set speed for 0.15 seconds, then rises 2,000 rpm above it for 0.2 seconds before returning. The cycle is asynchronous with the heartbeat. The pulse is active at set speeds of 4,000 rpm and above. It is designed to wash the pump and reduce stasis; it does not restore physiologic pulsatility.
The loss of pulsatility and high shear inside any rotary pump contribute to the long-term complications. Shear cleaves the largest von Willebrand factor multimers, producing an acquired von Willebrand syndrome, and low pulsatility promotes gastrointestinal angiodysplasia. A closed aortic valve permits commissural fusion and aortic regurgitation, and regurgitant blood re-enters the pump in a closed loop. These mechanisms are summarized under complications of durable LVAD support.
Goals of therapy: bridge to transplant, bridge to candidacy, destination therapy, and bridge to recovery
The intended goal of support is recorded at implant and frequently changes during support.
| Strategy | Definition | Typical patient |
|---|---|---|
| Bridge to transplant (BTT) | Support of a listed transplant candidate who is unlikely to survive or remain stable until a donor heart is available | Listed candidate with deteriorating end-organ function, long expected wait (blood group O, high body weight, sensitization), or inotrope intolerance |
| Bridge to candidacy (BTC) | Support of a patient with a potentially reversible barrier to transplant, with reassessment on support | Elevated pulmonary vascular resistance that may fall with unloading; kidney dysfunction from low output; obesity; recent cancer; need for psychosocial stabilization |
| Destination therapy (DT) | Permanent support in a patient who is not a transplant candidate | Older age, comorbidity, or patient choice; now the largest indication |
| Bridge to recovery | Support with the intention of device explant after myocardial recovery | Younger patients with nonischemic cardiomyopathy and short heart failure duration |
| Bridge to decision | Temporary support in shock until neurologic and end-organ status and the next step are known | Patient in shock, usually supported with a temporary device before a durable implant |
Definitions follow the 2021 European Society of Cardiology (ESC) heart failure guideline and the 2023 ISHLT mechanical circulatory support guidelines.
The distinction between bridge and destination therapy has weakened. MOMENTUM 3 enrolled both groups under one protocol and found the same benefit irrespective of the goal of therapy, and the HeartMate 3 is approved for both short-term and long-term support. A patient implanted as destination therapy may become a transplant candidate after kidney function, weight, or pulmonary pressure improves, and a listed patient may become ineligible after a stroke or infection. In the STS Intermacs 2024 report, recipients under 50 years had 1-year survival of 91.6% and 5-year survival of 72.6%.
Guideline recommendations for durable LVAD implantation
The 2022 AHA/ACC/HFSA guideline makes two recommendations for durable mechanical circulatory support in advanced HFrEF. In select patients with NYHA class IV symptoms who are dependent on continuous intravenous inotropes or temporary mechanical support, durable LVAD implantation is effective to improve functional status, quality of life, and survival (Class 1, level B-R). In select patients with NYHA class IV symptoms despite GDMT, durable mechanical support can be beneficial to improve symptoms and functional class and to reduce mortality (Class 2a, level B-R). The same guideline recommends timely referral of patients with advanced heart failure to a heart failure specialty team when consistent with the patient's goals (Class 1). Continuous intravenous inotropic support is reasonable as a bridge in patients eligible for and awaiting mechanical support or transplantation (Class 2a). Palliative care, including symptom management and goals-of-care discussions, belongs throughout the course of advanced disease (Class 1).
The 2023 ISHLT guidelines are the governing device-specific document. They update the 2013 guidelines through nine task forces. The topics are candidate selection, preoperative optimization and timing, intraoperative and immediate postoperative management, inpatient management, outpatient management, devices in adult congenital heart disease, myocardial recovery, biventricular support and the total artificial heart, and center quality metrics and staffing.
The 2021 ESC heart failure guideline gives long-term mechanical support a Class IIa recommendation as destination therapy (level A) in patients with advanced HFrEF despite optimal therapy who are not transplant candidates and have no severe right ventricular dysfunction. It gives the same class as a bridge to transplantation (level B). It differs from US guidance in listing explicit eligibility criteria, which include a stable psychosocial background and the absence of severe right ventricular dysfunction.
Timing of referral for advanced therapies
Late referral is the most common reason a patient who would have benefited from an LVAD does not receive one. Transplantation and durable support both require adequate kidney, liver, and pulmonary function and an absence of fixed pulmonary hypertension, and all of these deteriorate as heart failure advances. A patient referred after irreversible multiorgan injury has developed may no longer be a candidate. Progressive intolerance of GDMT, often attributed to drug side effects, is one of the earliest markers that the window is narrowing.
The mnemonic I NEED HELP summarizes triggers for referral to an advanced heart failure center.
| Letter | Trigger | Clinical significance |
|---|---|---|
| I | Intravenous inotropes required | Inotrope dependence places the patient in Intermacs profile 3 or worse and, with NYHA class IV symptoms, meets the 2022 AHA/ACC/HFSA Class 1 indication |
| N | NYHA class IIIb to IV or persistently elevated natriuretic peptides | Persistent severe symptoms despite therapy |
| E | End-organ dysfunction | Worsening kidney or liver function from low output or congestion |
| E | Ejection fraction 35% or less | Most candidates have an LVEF of 25% or less |
| D | Defibrillator shocks | Recurrent appropriate shocks indicate progression |
| H | Hospitalizations: more than one in the past year | Successive admissions mark steps down in prognosis |
| E | Edema despite escalating diuretics | Diuretic resistance and rising doses |
| L | Low systolic blood pressure (90 mm Hg or less) and high resting heart rate | Low output and sympathetic activation |
| P | Progressive intolerance or down-titration of GDMT | Hypotension or kidney dysfunction forcing dose reduction of neurohormonal antagonists |
The I NEED HELP mnemonic (Baumwol 2017), as adopted in the 2017 ACC expert consensus decision pathway for heart failure with reduced ejection fraction (Yancy 2018). One trigger is enough to prompt a discussion with an advanced heart failure team.
Intermacs profiles and modifiers
The Intermacs profiles stratify NYHA class III to IV heart failure into seven clinical profiles that describe acuity and the time frame for intervention (Stevenson 2009). They predict outcome after implantation: patients implanted in profile 1 have the highest early mortality, and stabilization with temporary support before durable implantation is preferred when possible.
| Profile | Short name | Description | Time frame for intervention |
|---|---|---|---|
| 1 | Critical cardiogenic shock ("crash and burn") | Life-threatening hypotension despite rapidly escalating inotropes, with critical organ hypoperfusion and rising lactate | Within hours |
| 2 | Progressive decline on inotropes ("sliding on inotropes") | Declining function despite intravenous inotropes: worsening kidney function, nutrition, or congestion; or inability to restore perfusion | Within days |
| 3 | Stable but inotrope dependent | Stable blood pressure, organ function, nutrition, and symptoms on continuous inotropes, with repeated failure to wean because of hypotension, symptoms, or kidney dysfunction | Elective over weeks to a few months |
| 4 | Resting symptoms | Symptoms at rest or with daily activities on oral therapy at home; congestion often present | Elective over weeks to a few months |
| 5 | Exertion intolerant ("housebound") | Comfortable at rest, unable to engage in other activity, living mostly within the house | Variable; depends on nutrition, organ function, and activity |
| 6 | Exertion limited ("walking wounded") | No congestion, comfortable at rest and with daily activities; fatigues after the first few minutes of meaningful activity | Variable; depends on nutrition, organ function, and activity |
| 7 | Advanced NYHA class III | Clinically stable with a reasonable level of comfortable activity and no recent unstable fluid balance | Transplant or support may not currently be indicated |
Modifiers: TCS, temporary circulatory support (profiles 1 to 3, in hospital); A, arrhythmia (any profile), recurrent ventricular tachyarrhythmias with frequent shocks or external defibrillation; FF, frequent flyer (profile 3 at home and profiles 4 to 6), at least two emergency visits or admissions in 3 months or three in 6 months. Adapted from Stevenson 2009 and the 2021 ESC guideline.
Most current implants are in profiles 2 and 3. The ESC guideline recommends that long-term support be considered in profiles 2 to 4. Implantation in profiles 5 to 7 is uncommon; the risks of surgery and device complications are weighed against a lower short-term mortality without a device, and the decision depends on objective evidence of progression and on patient preference.
Evaluation of the LVAD candidate: hemodynamics, exercise testing, and imaging
The evaluation answers three questions: is the heart failure severe and refractory enough to justify a device, is the patient likely to survive the operation and benefit from support, and are there barriers that make long-term device care unsafe? It is performed by a multidisciplinary team of heart failure cardiologists, cardiac surgeons, coordinators, social workers, pharmacists, dietitians, and palliative care specialists.
Confirmation of refractory disease. Reversible causes are excluded first: ischemia amenable to revascularization, severe valve disease amenable to repair or replacement, tachycardia-mediated cardiomyopathy, uncontrolled hypertension, thyroid disease, alcohol and drug toxicity, and myocarditis. GDMT is optimized to the maximal tolerated doses, cardiac resynchronization therapy is considered when the QRS criteria are met, and transcatheter edge-to-edge mitral repair is considered for severe secondary mitral regurgitation in suitable anatomy.
Cardiopulmonary exercise testing. Peak oxygen consumption (VO2) and the ventilatory efficiency slope quantify functional reserve and prognosis in ambulatory candidates. The ESC guideline lists a peak VO2 below 12 mL/kg/min or below 50% of predicted as one marker of eligibility for an LVAD. Peak VO2 is interpreted with the respiratory exchange ratio as a marker of effort, with beta-blocker therapy, and with age, sex, and body size; no single value determines candidacy. Thresholds and their use in transplant and device selection are described on the cardiopulmonary exercise testing page.
Right heart catheterization. Invasive hemodynamics define right atrial pressure (RAP), pulmonary capillary wedge pressure (PCWP), cardiac index, pulmonary artery pressures, pulmonary vascular resistance (PVR), and transpulmonary gradient. They identify low output that justifies support, quantify right ventricular reserve, and measure pulmonary vascular load. A RAP/PCWP ratio above about 0.6 identifies a right ventricle that is failing relative to the left and is incorporated in the HeartMate 3 survival risk score. The pulmonary artery pulsatility index (PAPi), pulmonary artery pulse pressure divided by RAP, quantifies right ventricular function; thresholds associated with right heart failure after LVAD differ from those used in acute right ventricular infarction. Measurement technique is covered on the right heart catheterization page. When PVR is elevated, testing with vasodilators or after unloading separates fixed from reversible pulmonary hypertension; a high PVR that falls with unloading is the classic reason for bridge to candidacy.
Echocardiography before implantation. The pre-implant study seeks findings that change the operation or the risk. Right ventricular size and function, tricuspid regurgitation, and estimated RAP bear on the risk of right heart failure after unloading. More than mild aortic regurgitation worsens on support because the pump holds aortic pressure above left ventricular pressure through most of the cardiac cycle. Mitral stenosis limits pump filling, and left ventricular apical thrombus threatens the inflow site. A patent foramen ovale or atrial septal defect can reverse to right-to-left flow once left atrial pressure falls, producing hypoxemia and paradoxical embolism, so a saline contrast study is performed. A mechanical aortic prosthesis, aortic root disease, and a small left ventricular cavity are documented. Transesophageal echocardiography in the operating room confirms these findings before bypass (covered on the transesophageal echocardiography page).
Other testing. Chest CT defines the sternum, prior grafts, and the aorta in redo surgery. Coronary angiography is performed when ischemic disease has not been defined. Pulmonary function tests, carotid and peripheral vascular imaging when indicated, age-appropriate cancer screening, dental evaluation, and infectious disease screening complete the evaluation.
Frailty, nutrition, and end-organ function
The noncardiac condition of the candidate strongly influences early mortality.
Frailty. The ISHLT consensus statement on frailty in advanced heart failure describes frailty across five domains: physical performance, sarcopenia, disability, comorbidity, and cognitive and psychological impairment. It also proposes a standard definition and assessment (Denfeld 2024). Frailty is common in advanced heart failure and predicts death and complications after LVAD implantation. Some of it is caused by low cardiac output and congestion and improves after support, while frailty from age, sarcopenia, and neurologic or musculoskeletal disease does not. Standardized assessment of physical performance, muscle mass, and cognition separates the two, and prehabilitation is used where time allows.
Nutrition. Cardiac cachexia, low serum albumin, and low muscle mass predict mortality and infection. Albumin is one of the three laboratory components of the HeartMate II risk score, with creatinine and the international normalized ratio (INR). Severe obesity raises the risk of driveline infection, wound complications, and thromboembolism, and it is a common reason for bridge to candidacy in patients who need weight loss to become transplant candidates. Dietitian assessment and nutritional repletion before elective implantation are part of standard evaluation.
Kidney function. Kidney dysfunction in advanced heart failure reflects low output, venous congestion, and intrinsic kidney disease in varying proportions. Function from the first two causes often improves after LVAD implantation, whereas intrinsic disease does not. Creatinine is a component of the HeartMate II risk score, and blood urea nitrogen is a component of the HeartMate 3 risk score. Dialysis dependence before implantation is associated with high mortality. The 2023 ISHLT guidelines treat anticipated permanent dialysis as a contraindication to destination therapy (Class III, level B). The trajectory of creatinine during inotrope or temporary support separates reversible from fixed injury.
Liver function. Congestive hepatopathy and ischemic hepatitis raise bilirubin and transaminases and impair synthetic function. An elevated INR off anticoagulation, a high bilirubin, and an elevated aspartate aminotransferase predict bleeding, right heart failure, and death. Cirrhosis with portal hypertension is a contraindication to implantation. The Model for End-Stage Liver Disease score excluding INR (MELD-XI) is used in some programs to quantify combined kidney and liver risk.
Neurologic and other systems. Recent stroke, cognitive impairment severe enough to prevent device self-care, active infection, uncontrolled diabetes with end-organ damage, severe chronic lung disease, and cancer with limited expected survival all reduce the likelihood of benefit.
Psychosocial and caregiver evaluation and shared decision making
An LVAD requires daily self-care: dressing changes, battery management, response to alarms, anticoagulation monitoring, and frequent clinic visits. Patients must be able to understand the device and act when it alarms, and most programs require a committed caregiver, especially in the first months after discharge. The 2021 ESC guideline lists a stable psychosocial background among its eligibility conditions for long-term support. The 2023 ISHLT guidelines require that the primary designated caregiver demonstrate competency in device care before discharge. Structured instruments such as the Stanford Integrated Psychosocial Assessment for Transplantation are used to quantify risk from substance use, psychiatric illness, adherence history, and social support.
Shared decision making is central, particularly for destination therapy, in which the device is the last disease-modifying therapy and complications can lead to a life the patient would not have chosen. In the DECIDE-LVAD stepped-wedge trial of 248 patients at six US centers, a decision aid with pamphlet, video, and clinician education modestly improved patient knowledge and concordance between stated values and treatment choice. The adjusted rate of LVAD implantation by 6 months was 53.9% with the intervention and 79.9% in the control period, indicating that informed patients choose a device less often (Allen 2018). Preparedness planning, which records in advance the patient's wishes if a disabling stroke or other catastrophic complication occurs, is part of the consent process.
Contraindications to durable LVAD implantation
Contraindications are graded as generally prohibitive or relative and are weighed against the expected survival without a device. Most are relative and may improve with treatment or time.
| Domain | Generally prohibitive | Relative; weighed by the team |
|---|---|---|
| Cardiac | Severe biventricular failure unsuitable for left-sided support alone (biventricular support or a total artificial heart is considered instead) | Severe right ventricular dysfunction or severe tricuspid regurgitation; small left ventricular cavity (restrictive or hypertrophic cardiomyopathy); ventricular septal defect; complex congenital anatomy |
| Anticoagulation | Inability to take a vitamin K antagonist (VKA); active bleeding that cannot be controlled | Prior gastrointestinal bleeding from angiodysplasia; thrombophilia; heparin-induced thrombocytopenia (alternative anticoagulants are used at implant) |
| Neurologic | Irreversible neurologic injury; dementia preventing self-care | Recent stroke; severe cerebrovascular disease |
| Kidney and liver | Cirrhosis with portal hypertension; chronic dialysis | Kidney or liver dysfunction expected to improve with unloading |
| Infection and cancer | Uncontrolled systemic infection; cancer with life expectancy shorter than the expected benefit | Treated cancer with good prognosis; chronic infections such as HIV under control |
| Lung | Severe obstructive or restrictive lung disease | Moderate lung disease; pulmonary hypertension from lung disease |
| Body habitus and frailty | Severe frailty not attributable to heart failure | Severe obesity; cachexia; low muscle mass |
| Psychosocial | Inability to manage the device without a caregiver; active substance use; refusal of care | Limited social support that may be strengthened before implant; untreated psychiatric illness |
Adapted from the 2023 ISHLT mechanical circulatory support guidelines and the 2021 ESC heart failure guideline (Table 16). Programs differ in how they apply relative contraindications.
Survival risk scores before implantation
Risk models estimate survival with and without a device and frame the discussion with the patient. The 2022 AHA/ACC/HFSA guideline considers validated multivariable risk scores reasonable to estimate mortality in heart failure. None replaces clinical judgment, and all were derived in populations that differ from current candidates.
| Score | Population and outcome | Components | Performance and use |
|---|---|---|---|
| Seattle Heart Failure Model (Levy 2006) | Ambulatory heart failure; projected survival on medical therapy | Age, sex, NYHA class, weight, LVEF, systolic pressure, ischemic cause, medications and doses, devices, sodium, cholesterol, hemoglobin, lymphocytes, uric acid | Estimates expected survival without a device; was not derived in inotrope-dependent patients |
| HeartMate II risk score (Cowger 2013) | 1,122 HeartMate II trial patients; 90-day mortality | Age, albumin, creatinine, INR, and implanting center volume (fewer than 15 trial patients) | 90-day mortality 4%, 16%, and 29% in low, medium, and high groups of the derivation cohort; area under the curve 0.71 |
| HeartMate 3 survival risk score (Mehra, JACC Heart Fail 2022) | 2,200 HeartMate 3 recipients in MOMENTUM 3 and its continued access protocol; 1- and 2-year survival | Age, prior cardiac surgery (bypass or valve), lower sodium, higher blood urea nitrogen, small left ventricular size, RAP/PCWP ratio above 0.6 | Area under the curve 0.76 at 1 year and 0.71 at 2 years; observed mortality doubles from one tertile to the next |
| Intermacs profile | Registry populations | Clinical acuity at implant | Profile 1 carries the highest early mortality; used only in combination with the scores |
Components and performance from the derivation publications. The HeartMate 3 score is the only one derived exclusively in recipients of the current device.
The HeartMate 3 score adds two hemodynamic and anatomic variables to laboratory markers of end-organ function. One is a small left ventricle, which limits pump filling and predisposes to suction. The other is a high RAP/PCWP ratio, which marks a right ventricle that cannot match the output of the pump. Prior cardiac surgery adds operative risk. Sodium and blood urea nitrogen reflect neurohormonal activation and kidney perfusion.
Right ventricular failure risk before implantation
Right heart failure after LVAD implantation is the leading cause of early morbidity. It results from increased venous return to a right ventricle that may already be impaired, from leftward shift of the interventricular septum when the left ventricle is unloaded, and from perioperative injury, transfusion, and arrhythmia. Severe early right heart failure is commonly defined as a need for right-sided mechanical support, inotropes for more than 14 days, or inhaled nitric oxide for 48 hours or more; registry definitions differ in detail. Predicting right heart failure is part of candidate selection.
| Predictor or score | Variables | Findings |
|---|---|---|
| Michigan right ventricular failure risk score (Matthews 2008) | Vasopressor requirement (4 points), aspartate aminotransferase ≥80 IU/L (2 points), bilirubin ≥2.0 mg/dL (2.5 points), creatinine ≥2.3 mg/dL (3 points) | Right ventricular failure in 35% of 197 implants; 180-day survival 90%, 80%, and 66% for scores ≤3.0, 4.0 to 5.0, and ≥5.5 |
| EUROMACS right-sided heart failure score (Soliman 2018) | Intermacs profile 1 to 3, multiple inotropes, severe right ventricular dysfunction on echocardiography, RAP/PCWP ratio, hemoglobin | Early severe right heart failure in 21.7%; risk 11% at scores of 0 to 2 and 43.1% above 4; area under the curve 0.70 in derivation and 0.67 in validation |
| Hemodynamic indices | High RAP, RAP/PCWP ratio, low PAPi, low right ventricular stroke work index | All identify a right ventricle with little reserve; thresholds vary between cohorts |
| Echocardiographic indices | Right ventricular dilation, low tricuspid annular plane systolic excursion, low right ventricular free wall strain, severe tricuspid regurgitation, plethoric inferior vena cava | Assessed with loading conditions and hemodynamics |
Score definitions and outcomes from the derivation publications. All scores have modest discrimination, and none has been validated prospectively in HeartMate 3 recipients alone.
When the predicted risk is high, the team optimizes the right ventricle before surgery with inotropes, treatment of pulmonary vascular load, and diuresis or ultrafiltration to lower RAP. It also plans for temporary right ventricular support at implant. When severe biventricular failure makes isolated left-sided support unlikely to succeed, biventricular support or a total artificial heart is considered.
Medicare (CMS) coverage criteria for durable LVADs
The Centers for Medicare and Medicaid Services (CMS) national coverage determination 20.9.1 (version 2) covers FDA-approved LVADs used as short-term support (bridge to recovery or bridge to transplant) or long-term support (destination therapy) under the same patient criteria. The current version removed the earlier separate requirements for destination therapy, including the peak VO2 threshold, and the requirement that bridge patients be listed for transplantation.
| Requirement | Criterion |
|---|---|
| Symptoms | NYHA class IV heart failure |
| Ejection fraction | LVEF 25% or less |
| Severity: option 1 | Inotrope dependent |
| Severity: option 2 | Cardiac index below 2.2 L/min/m² while not on inotropes, and either (a) on optimal medical management based on current heart failure guidelines for at least 45 of the last 60 days and failing to respond, or (b) advanced heart failure for at least 14 days and dependent on an intra-aortic balloon pump or similar temporary mechanical support for at least 7 days |
| Team | An explicitly identified multidisciplinary team based at the facility: at least one cardiothoracic surgeon who has implanted at least 10 durable intracorporeal LVADs in the previous 36 months with activity in the last year; at least one cardiologist trained in advanced heart failure; a ventricular assist device (VAD) program coordinator; a social worker; and a palliative care specialist |
| Facility | Credentialed by an organization approved by CMS |
From the CMS national coverage determination for ventricular assist devices (20.9.1). Post-cardiotomy support is covered separately when the device is FDA approved for that purpose.
The coverage criteria contain no age limit. The presence of a palliative care specialist on the team is a mandatory coverage requirement.
LVAD referral and candidacy guide
The guide compares a patient's Intermacs profile, NYHA class, ejection fraction, severity, and transplant candidacy with the profile definitions, the CMS patient criteria, the 2022 AHA/ACC/HFSA recommendations, and the referral triggers above.
Criteria from the Intermacs profile table (Stevenson 2009), the I NEED HELP table (Baumwol 2017), the CMS national coverage determination 20.9.1, and the 2022 AHA/ACC/HFSA guideline recommendations quoted on this page.
Older adults and age limits
Age is a continuous risk factor for death after implantation and enters the HeartMate II and HeartMate 3 risk scores, but neither guidelines nor CMS set an upper age limit. Many older patients are not transplant candidates, so destination therapy is the usual intent. The STS Intermacs 2025 report focused on recipients aged 65 years and older: 1-year survival was 80.6% and 5-year survival 51.1%, compared with 85.9% and 60.2% for all HeartMate 3 recipients. Survival in older adults was consistent across acuity, temporary support use, and psychosocial risk, and the report concluded that 5-year survival above 50% even in septuagenarians argues for revising eligibility criteria (Nayak 2026). In the 2024 report, recipients younger than 50 years had 1-year survival of 91.6% and 5-year survival of 72.6% (Meyer 2025).
The decision in an older adult depends on frailty, cognition, kidney function, caregiver support, and the patient's goals more than on chronologic age. Older recipients have more gastrointestinal bleeding and need more help with device care. Structured frailty and cognitive assessment, and an explicit discussion of what life with a driveline involves, are especially important in this group.
Preoperative optimization and bridging with temporary support
Elective implantation in a stable patient gives the best results. Preoperative optimization lowers RAP and PCWP with diuretics, inotropes, or ultrafiltration, restores kidney and liver function, treats infection, corrects anemia and nutrition, and stops antiplatelet drugs when possible. Inotrope bridging with milrinone or dobutamine is reasonable in patients awaiting durable support (covered on the vasopressors and inotropes page).
Patients in Intermacs profile 1 or deteriorating profile 2 are often stabilized with temporary mechanical support before durable implantation. In the STS Intermacs analysis of 5,787 HeartMate 3 recipients in profiles 1 to 3, 1-year and 2-year survival were 82.9% and 77.9% overall. Survival was lowest after VA-ECMO (70.6% and 67.1%), while patients bridged with an intra-aortic balloon pump (85.7% and 81.2%) or a high-flow microaxial pump (82.9% and 77.5%) had survival similar to those without temporary support. In adjusted analysis, both the balloon pump and the microaxial pump were associated with better survival than no temporary support, with similar stroke rates (Kanwar 2025). Temporary support allows assessment of neurologic status, right ventricular function, and end-organ recovery before committing to a durable device.
Surgical implantation: sternotomy and thoracotomy
The standard operation is performed through a median sternotomy on cardiopulmonary bypass, usually with the heart beating. A sewing ring is attached to the left ventricular apex, a core of myocardium is removed, and the pump with its integrated inflow cannula is inserted and oriented toward the mitral orifice. The outflow graft is cut to length, routed along the right side of the heart, and anastomosed end to side to the ascending aorta. The driveline is tunneled through the abdominal wall, often with the entire velour segment kept beneath the skin to promote tissue ingrowth. Transesophageal echocardiography confirms deairing, cannula position, septal position, right ventricular function, and aortic valve competence as bypass is weaned and speed is increased.
Less invasive approaches avoid a full sternotomy. In the lateral thoracotomy approach, the pump is inserted through a left anterolateral thoracotomy over the apex and the outflow graft is anastomosed to the ascending aorta through a small upper hemisternotomy or right anterior thoracotomy. The HeartMate 3 was approved for this approach on the basis of the ELEVATE registry and a single-arm feasibility study that reported less bleeding requiring surgery, fewer infections, and fewer arrhythmias. Less invasive implantation preserves the sternum for a later transplant. Its disadvantages are limited access for concomitant valve surgery and a learning curve.
The early postoperative course centers on right ventricular function, bleeding, and speed. Inotropes such as milrinone, dobutamine, or epinephrine and inhaled pulmonary vasodilators support the right ventricle. Heparin is started once surgical bleeding has settled, followed by warfarin. Speed is set with echocardiographic guidance to unload the left ventricle with the septum near the midline and without suction. Patients are mobilized early, and discharge requires that patient and caregiver demonstrate competence with equipment changes, dressing care, and alarm responses.
Concomitant procedures at implantation
Several lesions are corrected at the time of implantation because they become harder to treat later or would undermine support.
| Lesion | Effect on LVAD support | Usual management at implant |
|---|---|---|
| Aortic regurgitation more than mild | Worsens on support; creates a closed loop through the pump and reduces effective systemic flow | Repair (central coaptation stitch), bioprosthetic replacement, or patch closure of the outflow tract |
| Mechanical aortic prosthesis | Thrombosis when the prosthesis rarely opens | Replacement with a bioprosthesis or closure of the outflow tract |
| Patent foramen ovale or atrial septal defect | Right-to-left shunt and hypoxemia once left atrial pressure falls; paradoxical embolism | Closure |
| Mitral stenosis | Limits left ventricular filling and pump flow | Mitral replacement or commissurotomy |
| Mechanical mitral prosthesis | Can remain in place; some surgeons remove the leaflets | Individualized; anticoagulation continues |
| Severe tricuspid regurgitation | Contributes to right heart failure and congestion | Tricuspid repair at some centers; benefit is uncertain and evidence is observational |
| Left ventricular apical thrombus | Embolism and inflow obstruction | Removal during apical coring |
| Secondary mitral regurgitation | Usually falls with unloading | Not usually repaired |
| Left atrial appendage in atrial fibrillation | Possible thromboembolic source | Occlusion at some centers; evidence is limited |
Management reflects the 2023 ISHLT guidelines and the 2024 American Society of Echocardiography imaging recommendations. ISHLT recommendations: more than mild aortic regurgitation addressed at implant, with bioprosthetic replacement if necessary (Class I, level C), and valve closure techniques in selected patients (Class IIb); mechanical aortic prosthesis replaced with a bioprosthesis (Class I, level B); routine replacement of a functioning mechanical mitral prosthesis (Class III); closure of a significant interatrial shunt (Class I); tricuspid intervention for more than moderate regurgitation (Class IIb, level B). Practice varies most for tricuspid and appendage procedures.
HeartMate 3 pump parameters: speed, power, flow, and pulsatility index
The controller displays four parameters and detects pulsatility index events. No single value is a surrogate for clinical status, and a patient's values are interpreted against that patient's own trend.
| Parameter | What it is | Interpretation |
|---|---|---|
| Speed (rpm) | Fixed rotor speed set by the clinical team; range 3,000 to 9,000 rpm; displayed value within ±100 rpm of actual speed | Changed only with the clinical system monitor; the artificial pulse operates at 4,000 rpm and above |
| Power (watts) | Direct measurement of motor voltage and current | Rises with speed, flow, and friction. A sustained rise suggests thrombus in the pump or aortic regurgitation; a gradual fall suggests obstruction of flow. Power of 10.0 W or more is reported to the implanting center. |
| Flow (L/min) | Estimate calculated from speed, power, and the patient's hematocrit; flow pulses are averaged over 15-second intervals | Not a measured value. A rise in power from thrombus or friction produces a falsely high flow estimate. An incorrect hematocrit entry distorts the estimate. A low flow alarm sounds below a preset limit of 2.5 or 2.0 L/min. |
| Pulsatility index (PI) | Magnitude of flow pulses produced by the native left ventricle; observed clinical range 1 to 10 | Reflects left ventricular filling and contractility. A fall at stable speed and power suggests lower preload (hypovolemia, bleeding, right ventricular failure, tamponade) or lower contractility; a rise suggests more filling or recovery. |
| PI event | Sudden PI drop detected by the controller as possible ventricular collapse | Speed falls to the low speed limit and ramps back to the set speed; no audible alarm. Frequent events indicate suction, hypovolemia, arrhythmia, or a speed that is too high. |
Parameter definitions and limits from the manufacturer's clinician guide to HeartMate 3 pump parameters. Causes of PI events include sudden changes in power or speed, coughing, arrhythmias, suction, beat-to-beat changes in ventricular volume, and intra-aortic balloon pumps.
Patterns in the parameters localize problems. Low flow with low PI suggests inadequate left ventricular filling: hypovolemia, bleeding, right ventricular failure, tamponade, inflow malposition, or hypertension. High power with apparently high flow and hemolysis suggests pump thrombosis. Low flow with persistent aortic valve opening and a dilated ventricle suggests outflow obstruction from graft twist, kink, or external compression. These patterns prompt echocardiography and, depending on the finding, CT angiography and laboratory markers of hemolysis.
Controller alarms and their causes
Controller alarms are grouped by priority. Hazard alarms, shown with a red heart symbol and a continuous tone, indicate a condition that requires immediate action: the pump has stopped, flow is low, or power is critically low. Advisory alarms, shown with a yellow symbol, indicate a problem that requires attention but not emergency action: a low battery, a disconnected power cable, a controller fault, or a change in a parameter. Patients are trained to respond to all alarms and to call the implanting center for any hazard alarm or any advisory alarm they cannot resolve.
| Alarm or finding | Common causes | Immediate assessment |
|---|---|---|
| Pump off or pump stopped | Loss of both power sources; driveline fracture or disconnection; controller failure | Restoration of power, reconnection of the driveline, or exchange of the controller; auscultation for the pump hum; evaluation of perfusion |
| Low flow | Hypovolemia, bleeding, right ventricular failure, tamponade, arrhythmia, hypertension, inflow or outflow obstruction, pump thrombosis | Volume status, blood pressure, rhythm, hemoglobin, echocardiography; CT angiography when obstruction is suspected |
| High power | Pump thrombosis or ingested material; aortic regurgitation | Lactate dehydrogenase, plasma free hemoglobin, hemoglobinuria, echocardiography |
| Low battery or power cable disconnected | Depleted batteries; loose connection | Replacement with charged batteries or connection to wall power |
| Frequent PI events (no alarm) | Suction from hypovolemia or excessive speed; arrhythmia | Volume status, rhythm, echocardiography of cannula and septum |
| Driveline fault | Damage to conductors in the driveline | Inspection, controller log review, device team assessment |
Alarm categories and responses summarized from manufacturer and implanting-center guides. Exact alarm wording differs between controller versions.
Blood pressure measurement and blood pressure targets
The narrow pulse pressure of continuous-flow support makes noninvasive blood pressure measurement difficult. Automated oscillometric cuffs often fail or give inaccurate values. The standard noninvasive method is Doppler: a manual cuff is inflated over the brachial artery, a Doppler probe is placed at the radial or brachial artery, and the pressure at which flow first returns during deflation is recorded as the opening pressure.
The meaning of the Doppler opening pressure depends on pulsatility. In a study of 60 HeartMate II recipients, inpatients with arterial lines provided the intra-arterial comparison. The mean Doppler pressure was 4.1 mm Hg lower than intra-arterial systolic pressure and 9.5 mm Hg higher than intra-arterial mean pressure, so Doppler more closely approximated systolic than mean pressure (Lanier 2013). When pulse pressure is very narrow, the Doppler pressure, systolic pressure, and mean pressure converge. The measurement method is therefore recorded with the value, and an automated cuff that measures both systolic and diastolic pressures can be used when a palpable pulse is present. An arterial line is used whenever precise control is needed.
Hypertension reduces pump flow and increases stroke risk. In the ENDURANCE Supplemental trial, an enhanced blood pressure protocol lowered mean arterial pressure and was associated with a 50.5% reduction in hemorrhagic stroke compared with the original HVAD cohort (Milano 2018). The manufacturer's guidance for the HeartMate 3 specifies a mean pressure below 90 mm Hg. The 2023 ISHLT guidelines recommend a mean arterial pressure goal of 75 to 90 mm Hg in recipients of continuous-flow LVADs (Class IIa, level B). They advise against pharmacologic lowering of mean pressure below 75 mm Hg (Class III, level C), because hypotension carries its own risks of low perfusion, suction, and death. Early after implantation, arterial line monitoring is recommended (Class I, level C). The optimal target in HeartMate 3 recipients is not established by randomized evidence.
Antihypertensive therapy uses the drugs that also treat heart failure. The 2023 ISHLT guidelines prefer neurohormonal antagonists (angiotensin-converting enzyme inhibitors, angiotensin receptor blockers, sacubitril-valsartan, beta blockers, and mineralocorticoid receptor antagonists) as first-line agents (Class I, level B). Hydralazine with isosorbide nitrate is second-line therapy (Class IIb, level C), and dihydropyridine calcium channel blockers, clonidine, and alpha-1 antagonists are third-line agents for resistant hypertension.
Medical therapy during LVAD support
The LVAD unloads the ventricle but does not remove the neurohormonal activation, arrhythmia substrate, or right ventricular disease of heart failure. Medical therapy during support has four goals: control blood pressure, maintain euvolemia, support the right ventricle, and limit complications.
Neurohormonal antagonists. Renin-angiotensin-aldosterone system inhibitors lower blood pressure and may reduce bleeding and right heart failure. In a secondary analysis of 1,855 clinically stable recipients in the MOMENTUM 3 trial portfolio, renin-angiotensin-aldosterone system inhibition at 1 month after implantation was associated with fewer hemocompatibility-related adverse events at 2 years (adjusted rate ratio 0.83). The association was driven by less nonsurgical bleeding and was independent of blood pressure and anticoagulation intensity (Isath 2026). The finding is observational. Beta blockers are used for blood pressure, rate control, and ventricular arrhythmia, with attention to right ventricular function early after implantation. Mineralocorticoid receptor antagonists are used for blood pressure, hypokalemia, and congestion. Sodium-glucose cotransporter 2 inhibitors have been studied only in observational LVAD cohorts.
Diuretics. Loop diuretics manage congestion from right ventricular dysfunction and tricuspid regurgitation. Overdiuresis lowers left ventricular filling and causes suction events and low flow.
Pulmonary vasodilators. Phosphodiesterase-5 inhibitors are used in some programs for residual pulmonary hypertension and right ventricular dysfunction after implantation, mainly as a bridge to candidacy. Their benefit is not established in randomized trials.
Drugs used with caution. Nonsteroidal anti-inflammatory drugs increase bleeding and sodium retention. Drugs that interact with warfarin (amiodarone, many antibiotics, azole antifungals) require INR monitoring. Antiarrhythmic drugs follow standard indications.
Residual heart failure on support, defined as persistent congestion, low output, or hospitalization despite a functioning pump, affects a substantial minority of recipients. Its evaluation includes echocardiography, an invasive hemodynamic ramp study, and assessment for aortic regurgitation, outflow obstruction, and right ventricular failure.
Antithrombotic therapy: warfarin target and the omission of aspirin
Long-term antithrombotic therapy after HeartMate 3 implantation is a VKA, almost always warfarin, with an INR goal of 2.0 to 3.0 in the manufacturer's guidance. The 2023 ISHLT guidelines recommend warfarin with the INR range specified by the manufacturer (Class I, level B). Their aspirin recommendation was written before ARIES-HM3. Heparin bridges the early postoperative period until the INR is therapeutic. Warfarin dosing and INR management are described on the warfarin page.
Aspirin was historically added to warfarin for all continuous-flow LVADs without randomized evidence. The Antiplatelet Removal and Hemocompatibility Events with the HeartMate 3 Pump (ARIES-HM3) trial randomized 628 HeartMate 3 recipients at 51 centers in 9 countries to aspirin 100 mg/day or placebo in addition to a VKA. The primary analysis population included 589 patients. The primary end point was survival free of major nonsurgical hemocompatibility-related adverse events (stroke, pump thrombosis, major bleeding, or arterial peripheral thromboembolism) at 12 months. It was reached by 74% with placebo and 68% with aspirin, which met the noninferiority criterion. Omitting aspirin reduced nonsurgical bleeding events (relative risk 0.66) without increasing stroke or other thromboembolic events, consistently across subgroups (Mehra 2023). After ARIES-HM3, the FDA approved a HeartMate 3 labeling change that removes aspirin from routine management.
The result applies to the HeartMate 3. It does not extend automatically to patients still supported by older axial-flow or HVAD pumps, or to patients with a separate indication for antiplatelet therapy, such as recent coronary stenting. Direct oral anticoagulants are not an established replacement for warfarin in LVAD recipients. The LVAD team changes the antithrombotic regimen after bleeding, stroke, or pump thrombosis and before procedures.
Echocardiographic surveillance during LVAD support
Echocardiography is performed before discharge, at intervals during follow-up, and whenever flow, symptoms, or laboratory values change. The 2024 American Society of Echocardiography (ASE) recommendations describe a standardized study that begins with the device settings and ends with the pericardium (Estep 2024).
| Step | Target | Findings recorded |
|---|---|---|
| 1 | Device and patient | Model, speed, flow, power, PI, blood pressure and its method |
| 2 | Left ventricle | Diastolic dimension, septal position, mitral regurgitation, spontaneous contrast or thrombus |
| 3 | Aortic valve | Opening frequency by M-mode (all beats, intermittent, or none); aortic regurgitation through the cycle |
| 4 | Inflow cannula | Direction, wall contact, color and pulsed Doppler velocity |
| 5 | Outflow graft | Color flow and peak velocity compared with the patient's baseline |
| 6 | Right ventricle and veins | Size and function, tricuspid regurgitation, inferior vena cava and estimated RAP |
| 7 | Pericardium | Effusion and regional compression |
A systematic LVAD echocardiogram based on the 2024 ASE recommendations. Device components cause shadowing, reverberation, and electrical interference, so a suspected thrombus or jet is confirmed in orthogonal views.
Inflow and outflow velocities. Unobstructed inflow is laminar, generally 1 to 2 m/s across continuous-flow devices and typically below 1.0 m/s with the HeartMate 3, rising to about 1.2 m/s with the artificial pulse. Higher, turbulent inflow suggests malposition, wall contact, thrombus, or a small cavity. The outflow graft is usually imaged from a high right parasternal window. Baseline peak systolic velocity is below 2 m/s, and the artificial pulse can produce intermittent peaks above 2 m/s. The value describes a normal baseline and is not a universal diagnostic threshold; a focal, turbulent, sustained rise from the patient's own baseline suggests graft obstruction and prompts CT angiography.
Left ventricular unloading and septal position. Adequate support produces a smaller left ventricle, less secondary mitral regurgitation, and a septum near the midline. The septum bows toward the left ventricle with excessive unloading or a pressure-loaded right ventricle and toward the right ventricle with inadequate unloading. Septal position guides the differential diagnosis and is not a speed target in isolation.
Aortic valve opening. Opening depends on speed and native contractility. Intermittent opening is a secondary goal of speed setting, and relief of heart failure takes priority. A valve that never opens promotes commissural fusion, aortic root thrombus, and aortic regurgitation, so persistent closure is reported.
Aortic regurgitation on support. When the valve stays closed, regurgitation can be continuous through systole and diastole, because aortic pressure exceeds left ventricular pressure throughout the cycle. Pressure half-time and descending aortic flow reversal are distorted by continuous pump flow. Vena contracta width, jet duration, and the outflow graft Doppler signal are integrated with left ventricular size; an outflow graft systolic-to-diastolic velocity ratio below 5.0 supports at least moderate regurgitation (Estep 2024).
Filling pressure. The ASE diastolic function pathway for LVAD recipients uses a mitral E/A ratio above 2, an estimated RAP above 10 mm Hg, and a pulmonary artery systolic pressure above 40 mm Hg as support for elevated left-sided filling pressure. Left atrial volume, E/e′, and septal position are read together with these. General echocardiographic methods are described on the transthoracic echocardiography page.
Loading patterns: suction, right ventricular failure, and inadequate unloading
Three hemodynamic patterns explain most low-flow states and poor support. Echocardiography separates them, and they respond differently to a change in speed.
| Finding | Excessive unloading (suction) | Right ventricular failure | Inadequate unloading |
|---|---|---|---|
| Left ventricular size | Small | Small | Dilated |
| Septum | Bowed toward the left ventricle | Bowed toward the left ventricle | Bowed toward the right ventricle |
| Aortic valve | Closed | Closed or rarely opening | Opens frequently |
| Mitral regurgitation | Minimal | Minimal | Persistent |
| Right ventricle and tricuspid regurgitation | Right ventricle usually adequate | Dilated, hypokinetic right ventricle; severe tricuspid regurgitation | Variable |
| Inferior vena cava | Small and collapsing | Dilated, little respiratory variation | Variable |
| Response to higher speed | Worsens | Worsens septal shift and right ventricular geometry | May improve after obstruction is excluded |
| Management | Volume; lower speed | Inotropes, diuresis, pulmonary vasodilators, rhythm control; temporary right-sided support if severe | Higher speed; afterload reduction; search for outflow obstruction or aortic regurgitation |
Echocardiographic loading patterns during continuous-flow support, based on the 2024 ASE recommendations. Raising speed because flow is low worsens both suction and right ventricular failure.
Speed optimization and echocardiographic ramp studies
A ramp study records left ventricular end-diastolic dimension, septal position, aortic valve opening, mitral and tricuspid regurgitation, right ventricular function, blood pressure, and pump parameters as speed is raised in steps. It is used to set the speed before discharge and to investigate suspected obstruction or heart failure on support.
The Columbia ramp protocol, developed for the HeartMate II, increased speed in 400-rpm steps from 8,000 to 12,000 rpm. Among 28 studies performed for speed optimization, speed was changed in 61% of patients. All patients with confirmed device thrombosis, and none of the others, had a slope of left ventricular end-diastolic dimension against speed greater (flatter) than −0.16 (Uriel 2012). With a patent circuit, left ventricular dimension falls progressively and the aortic valve opens less often as speed rises. A flat dimension response with persistent valve opening suggests obstruction of the inflow cannula, the pump, or the outflow graft.
HeartMate 3 ramp protocols change speed in 100-rpm steps; the typical operating range is 4,800 to 5,600 rpm, and the chosen speed is usually about 100 rpm below the highest speed that gives optimal unloading (Estep 2024). Pump thrombosis is uncommon with this device. A flat ramp in a HeartMate 3 recipient is more often caused by outflow graft obstruction, which is confirmed by CT angiography. Protocols and criteria are device and center specific, and no single variable is used alone.
Invasive hemodynamic optimization
Echocardiographic findings do not reliably predict filling pressures during LVAD support. In 35 recipients of HeartMate II and HVAD pumps who underwent combined invasive and echocardiographic ramp studies, only 42.9% had a normal central venous pressure and PCWP at their original speed. Raising speed from the lowest to the highest setting increased cardiac output by a mean of 1.28 L/min and lowered PCWP by 9.9 mm Hg, while central venous pressure did not change significantly. Speed adjustment achieved near-normal pressures in 56%, and in the rest the study indicated which medical therapy to pursue (Uriel 2016).
In 60 stable HeartMate 3 outpatients who underwent routine invasive ramp studies, 58% had optimized filling pressures (central venous pressure below 12 mm Hg and PCWP below 18 mm Hg) at baseline speed, and 65% after speed adjustment. Patients with optimized pressures had higher 6-month and 12-month freedom from readmission, mainly from fewer cardiac admissions (Rubinstein 2024). Invasive ramp studies are performed when symptoms persist on support, before listing, and when right ventricular function and pulmonary pressures need precise definition. A failure of pressures to normalize at any speed points to right ventricular failure, aortic regurgitation, or outflow obstruction.
Imaging beyond echocardiography
Chest radiography. The radiograph shows the pump position, the angle of the inflow cannula relative to the apex, the driveline course, and pulmonary congestion. A change in pump angle on serial films suggests cannula malposition, and a discontinuity in the driveline suggests conductor damage.
CT angiography. Contrast CT defines outflow graft twist, kink, external compression, and anastomotic stenosis; cannula position; and apical or aortic root thrombus. The scan is timed for the aorta and the graft, and metal artifact from the pump obscures the inflow region. CT is also used for driveline and pocket infection.
Nuclear imaging. Fluorodeoxyglucose positron emission tomography (PET) combined with CT shows the extent of driveline and pump infection and guides surgical debridement.
MRI. The HeartMate 3 is not compatible with MRI. Patients who need brain or spine imaging undergo CT or other modalities (covered on the cardiac MRI safety page).
Arrhythmias and defibrillators during LVAD support
Atrial fibrillation and ventricular arrhythmias are common during support. Ventricular tachycardia and fibrillation are often tolerated for minutes to hours because the pump maintains systemic output while the right ventricle can still deliver preload to the left ventricle. Sustained arrhythmia eventually causes right ventricular failure, low flow, and syncope. Suction events and inflow cannula contact with the septum provoke ventricular ectopy and must be excluded before antiarrhythmic treatment is escalated. Management of ventricular arrhythmias, including catheter ablation, is described on the ventricular tachycardia and sudden cardiac death page.
The benefit of an implantable cardioverter-defibrillator (ICD) during continuous-flow support is uncertain. In a propensity-matched Intermacs analysis of 4,418 patients, the presence of an ICD was associated with higher mortality (hazard ratio 1.20; 95% confidence interval 1.04 to 1.39), more treated ventricular arrhythmias, and more rehospitalization; the observational design leaves residual confounding (Clerkin 2017). The 2023 ISHLT guidelines recommend reactivating an existing ICD after implantation (Class I, level A) and programming it conservatively to favor antitachycardia pacing and minimize shocks (Class IIa, level B). They also recommend deferring a new primary-prevention ICD in patients with no history of ventricular arrhythmia (Class IIa, level B). A subcutaneous ICD is not recommended for secondary prevention after implantation (Class III, level C). Cardioversion and defibrillation are performed without disconnecting the pump.
Complications of durable LVAD support
Complications are the main limitation of LVAD therapy and the main reason for readmission.
| Complication | Mechanism | Recognition |
|---|---|---|
| Gastrointestinal bleeding | Acquired von Willebrand syndrome from shear; angiodysplasia from low pulsatility; anticoagulation | Most common long-term complication; anemia, melena, falling PI |
| Driveline infection | Percutaneous lead breaches the skin barrier; trauma to the exit site | Most common device infection; drainage, erythema, fluid tracking along the driveline on CT |
| Right heart failure | Increased venous return and leftward septal shift; progressive right ventricular disease | Early or late; congestion, low flow, small left ventricle |
| Stroke | Embolism, hemorrhage, infection; linked to blood pressure and anticoagulation intensity | New deficit; immediate CT |
| Pump thrombosis and hemolysis | Thrombus in the pump; rare with the HeartMate 3 | Rising power, lactate dehydrogenase, plasma free hemoglobin, dark urine |
| Outflow graft obstruction | Twist, kink, external compression, or anastomotic stenosis | Late low flow; focal high outflow velocity; CT angiography |
| Aortic regurgitation | Commissural fusion of a closed valve | Progressive; dilated left ventricle with high flow and power |
| Ventricular arrhythmias | Scar substrate, suction, cannula contact | Low flow, palpitations, ICD therapies |
Five-year freedom from these events with the HeartMate 3 in the STS Intermacs 2025 report was 73.5% for gastrointestinal bleeding, 81.8% for device malfunction, and 87.1% for stroke.
Emergencies and cardiac arrest in LVAD recipients
The unresponsive LVAD recipient may have no palpable pulse and no measurable noninvasive blood pressure while adequately perfused. Assessment therefore relies on signs of perfusion: mental status, skin color and capillary refill, end-tidal carbon dioxide, and mean pressure by Doppler or arterial line. The device is assessed by auscultation over the upper abdomen and left chest for the pump hum, by the controller display and alarms, and by inspection of the driveline and power connections. The implanting center's emergency line is called early, and the patient's backup controller and batteries travel with the patient.
The 2017 AHA scientific statement on resuscitation in patients with mechanical circulatory support provides an algorithm built on this assessment, and the 2023 ISHLT guidelines adopt it. A mean arterial pressure of 75 mm Hg or more usually indicates adequate perfusion (Class I, level C). In an unresponsive LVAD recipient, a Doppler opening pressure below 50 mm Hg or an end-tidal carbon dioxide below 20 mm Hg indicates markedly reduced perfusion. Unless advance directives state otherwise, chest compressions are advised in an unconscious patient on isolated LVAD support without adequate perfusion (Class IIa, level B), and attempts to restore pump function proceed in parallel. The risk of cannula dislodgement from compressions has proved small. Defibrillation and cardioversion are performed without disconnecting the device (Class I, level B). The full resuscitation sequence is described on the resuscitation in special circumstances page.
Common emergencies with a running pump are gastrointestinal bleeding; hypovolemia from vomiting or diarrhea, with suction and ventricular arrhythmia; stroke; sepsis from driveline infection; and right heart failure. Volume is given in small boluses for suspected hypovolemia; vasopressors are used for vasodilatory shock; and blood pressure is lowered promptly in intracranial hemorrhage. Anticoagulation reversal for life-threatening bleeding balances the risk of pump thrombosis and is coordinated with the LVAD team.
Noncardiac surgery and procedures during LVAD support
Noncardiac procedures are common in LVAD recipients: endoscopy for bleeding, general and orthopedic surgery, and dental work. Elective procedures are performed at, or in consultation with, the implanting center. Planning addresses anticoagulation, monitoring, volume, and the device.
Anticoagulation. Minor procedures such as dental extraction and diagnostic endoscopy are often performed with a therapeutic INR. For procedures with higher bleeding risk, warfarin is withheld and heparin bridging is individualized according to the device, history of thrombosis, and bleeding risk (covered on the periprocedural anticoagulation page). Vitamin K and prothrombin complex concentrate are used for urgent reversal.
Anesthesia. Arterial pressure monitoring is standard for procedures under general anesthesia, because noninvasive cuffs are unreliable. Induction agents and positive-pressure ventilation reduce preload and increase right ventricular afterload; volume, vasopressors, and inotropes maintain left ventricular filling and mean pressure. Pulse oximetry may fail, and cerebral oximetry or arterial blood gases are used.
Device management. The pump is never stopped. The controller is connected to wall power, and the LVAD coordinator or a trained clinician is present. Electrocautery can interfere with the controller and electrocardiographic monitoring. ICD tachycardia therapies are suspended during cautery with external pads in place. Planning for abdominal procedures accounts for the driveline course, which is identified before incision.
Daily life with an LVAD
Power and equipment. The controller is connected to batteries or wall power at all times. A pair of batteries supports 10 to 17 hours, and patients connect to wall power whenever they sleep or may fall asleep.
Driveline care. The exit site is dressed with a sterile technique on a schedule set by the implanting center, and the driveline is anchored to prevent pulling and twisting. Trauma to the exit site is the main trigger for driveline infection. The driveline is not twisted, kinked, or sharply bent.
Showering and swimming. Showering is permitted once the exit site has healed and the program approves, with the controller and batteries in a waterproof shower bag. Swimming and bathing that submerge the equipment are prohibited.
Activity and exercise. Walking and cardiac rehabilitation improve functional capacity after implantation and are encouraged. Contact sports, activities with a risk of driveline trauma, and activities near water that risk submersion are avoided.
Driving and travel. Driving rules differ between programs and jurisdictions, and many programs restrict driving for a period after implantation and after syncope or arrhythmia. Air travel is possible with equipment, documentation, and contact information for implanting centers along the route.
Electrical and imaging hazards. MRI is contraindicated. Static electricity discharge and strong magnetic fields are avoided according to the manufacturer's instructions. The pump may interfere with surface electrocardiographic monitoring, and lead repositioning reduces interference.
Pregnancy and contraception. Pregnancy during LVAD support carries maternal and fetal risk from warfarin and hemodynamic load, and effective contraception is part of counseling for women of childbearing potential.
Survival and outcomes in trials and registries
Survival has improved with successive device generations, while adverse events continue to affect quality of life.
| Study | Design and population | Survival | Other results |
|---|---|---|---|
| REMATCH (2001) | Randomized; 129 transplant-ineligible NYHA IV patients; pulsatile LVAD against medical therapy | 1 year 52% vs 25%; 2 years 23% vs 8% | Death reduced by 48%; adverse events 2.35 times as frequent with the device |
| HeartMate II bridge (2007) | Single arm; 133 listed patients | 68% at 12 months on support | 75% transplanted, recovered, or supported and eligible at 180 days |
| HeartMate II destination therapy (2009) | Randomized 2:1; 200 patients; axial against pulsatile | 2 years 58% vs 24% | Primary composite 46% vs 11% |
| ENDURANCE (2017) | Randomized 2:1; 446 patients; HVAD against HeartMate II | Not the primary end point | Noninferior composite; stroke 29.7% vs 12.1% |
| MOMENTUM 3 (2019) | Randomized; 1,028 patients; HeartMate 3 against HeartMate II | Not the primary end point | Primary composite at 2 years 76.9% vs 64.8%; pump replacement 2.3% vs 11.3% |
| MOMENTUM 3 extended (2022) | 5-year follow-up of patients on support at 2 years | 5 years 58.4% vs 43.7% | Composite 54.0% vs 29.7%; fewer strokes, bleeding, and pump thrombosis |
| ELEVATE (2020) | Registry; 540 patients in Europe and the Middle East | 83% at 2 years (primary implants) | Stroke 10.2%; gastrointestinal bleeding 9.7%; pump thrombosis 1.5% |
| STS Intermacs 2024 report | 29,634 continuous-flow LVADs, 2014 to 2024 | Magnetic levitation: 1 year 85.7%, 5 years 59.7%; other pumps 78.4% and 43.7% | Recipients under 50 years: 91.6% and 72.6% |
| STS Intermacs 2025 report | 28,029 continuous-flow LVADs, 2015 to 2024 | HeartMate 3: 1 year 85.9%, 5 years 60.2%, 7 years 51.8% | Age 65 or older: 80.6% and 51.1% |
Survival figures are Kaplan-Meier estimates from the primary publications. Trial populations were selected and may have better outcomes than the general implant population.
The causes of death during support have shifted with device technology. Pump thrombosis has become rare, while multiorgan failure early after implantation, stroke, infection, and right heart failure remain important. Transplantation removes patients from the survival denominator in bridge cohorts, so registry survival is reported with transplantation and recovery as competing outcomes.
Quality of life and functional capacity
All major LVAD trials have shown improvement in symptoms, functional class, and quality of life. In REMATCH, quality of life improved significantly at 1 year with the device. In the HeartMate II bridge study, NYHA class, 6-minute walk distance, and scores on the Minnesota Living with Heart Failure and Kansas City Cardiomyopathy Questionnaires improved by 3 months. ELEVATE recorded sustained improvement in 6-minute walk distance and the EQ-5D visual analog quality-of-life score over 2 years. The STS Intermacs 2025 report found favorable quality-of-life measures in recipients aged 65 years and older.
Peak VO2 after implantation remains well below normal in most recipients, because the pump does not increase its output much with exercise and native ventricular reserve is limited. Patients nevertheless return to walking, light work, travel, and many daily activities. The burdens of support are the driveline and its care, dependence on equipment, frequent clinic visits and INR checks, readmissions for bleeding and infection, and the risk of disabling stroke. These burdens shape the choice of destination therapy and are presented during shared decision making.
Myocardial recovery and device explant
Unloading reverses part of the cellular and structural remodeling of heart failure, and a small proportion of patients recover enough function to have the device removed. Recovery is more likely in younger patients with nonischemic cardiomyopathy, short heart failure duration, and smaller ventricles.
The RESTAGE-HF study tested whether a protocol of optimized unloading with aggressive neurohormonal therapy and regular testing of myocardial function at low speed would increase explantation. It enrolled 40 patients with nonischemic cardiomyopathy supported by the HeartMate II, with a mean age of 35 years. Forty percent of all enrolled patients reached the primary end point of explantation within 18 months with sustained remission from heart failure at 12 months. Of the 36 patients who received the protocol, 18 (50%) were explanted within 18 months. Before explant, the mean LVEF was 57% at a speed producing no net pump flow (Birks 2020).
Assessment for recovery uses echocardiography and invasive hemodynamics at reduced speed, sometimes with exercise. Explant techniques range from complete pump removal to driveline division with outflow graft ligation and an apical plug. Patients are followed after explant for recurrence, and continuation of GDMT is standard.
LVAD support and heart transplantation
Heart transplantation offers longer survival and better quality of life than any device and is the preferred therapy for eligible candidates. LVADs support listed patients and make some ineligible patients eligible. Unloading can lower PVR over months in patients whose pulmonary hypertension is driven by left-sided filling pressure, converting prohibitive physiology into transplantable physiology. Reassessment of pulmonary hemodynamics by right heart catheterization during support establishes whether the pulmonary vascular load is now acceptable for a donor right ventricle (covered on the right heart catheterization and pulmonary hypertension pages). Kidney function, weight, and treated cancer may also improve during support.
Allocation rules affect LVAD use. Under the current US adult heart allocation system, stable patients with a durable LVAD have lower priority than patients on temporary mechanical support. A limited discretionary period of higher priority follows implant, and device complications carry higher priority. Use of a durable LVAD as a bridge has therefore declined in favor of transplantation from temporary support. Transplantation after LVAD support is a reoperation with adhesions, and sensitization from transfusion during support may lengthen the wait.
Biventricular support, the total artificial heart, and investigational pumps
Patients with severe biventricular failure who cannot be supported by an LVAD alone need right-sided support as well. Options include a temporary right ventricular assist device after LVAD implantation (often a centrifugal pump with or without an oxygenator), a durable biventricular configuration with two continuous-flow pumps, or a total artificial heart.
The SynCardia total artificial heart (formerly CardioWest) replaces both ventricles and all four valves with a pneumatically driven pulsatile pump, removing the problems of right heart failure, valvular regurgitation, arrhythmia, and intracardiac thrombus. In a nonrandomized study of 81 transplant-eligible patients with irreversible biventricular failure, 79% survived to transplantation, compared with 46% of 35 historical controls, and 1-year survival was 70% against 31% (Copeland 2004). It is used as a bridge to transplant in biventricular failure, restrictive or infiltrative cardiomyopathy with small ventricles, massive ventricular thrombus, complex congenital anatomy, and failed prior grafts. Total artificial hearts with magnetically levitated rotary pumps are in early clinical study. The 2023 ISHLT guidelines include a dedicated task force on biventricular support and the total artificial heart.
The driveline remains the principal unsolved problem: it causes infection, limits bathing and swimming, and requires daily care. Fully implantable systems use transcutaneous energy transfer, in which an external coil transmits power across intact skin to an implanted receiving coil and battery. These systems are in early clinical study. Other centrifugal pumps are in clinical trials against existing technology in the United States, and pumps with improved hemocompatibility, sensor-based speed adjustment, and wireless monitoring are in development. None has replaced the HeartMate 3 in US practice.
Palliative care and device deactivation
Palliative care is part of LVAD care from the evaluation onward. The specialist is a required member of the implanting team under CMS coverage, and the 2022 AHA/ACC/HFSA guideline recommends palliative care throughout advanced heart failure. Before implantation, preparedness planning records the patient's goals and the limits of acceptable outcomes. During support, palliative care treats symptoms, supports caregivers, and helps with decisions after major complications.
Withdrawal of LVAD support at the request of a patient or surrogate is ethically and legally accepted as withdrawal of a life-sustaining treatment, a category distinct from euthanasia or physician-assisted death. Deactivation is planned with the patient, family, LVAD team, and palliative care. Symptoms of dyspnea and anxiety are anticipated and treated before the pump is stopped, because death usually follows within minutes to hours in a pump-dependent patient. Alarms are silenced, and any ICD is deactivated before or at the same time (covered on the implantable cardioverter-defibrillator page). Hospice care is available to LVAD recipients, though some hospices lack the experience to manage the device.
Special populations
Women. Women are a minority of recipients (18% of the MOMENTUM 3 extended cohort). Smaller body size and ventricle, higher rates of nonischemic cardiomyopathy, and later referral influence their outcomes, and sex-specific risk models have been proposed.
Restrictive, infiltrative, and hypertrophic cardiomyopathy. A small, thick-walled left ventricle limits pump filling and predisposes to suction, and right ventricular involvement is common. Selected patients with amyloid or hypertrophic cardiomyopathy and an adequate cavity receive LVADs, while others are bridged with a total artificial heart or transplanted without a bridge (covered on the cardiac amyloidosis and hypertrophic cardiomyopathy pages).
Chronic kidney disease. Kidney function from low output often improves after implantation, while intrinsic disease does not. Dialysis dependence before implantation carries high mortality.
Adult congenital heart disease. Systemic right ventricles, prior surgery, and unusual anatomy complicate implantation. Experienced centers support selected patients with a systemic right ventricle or failing biventricular repair.
Obesity. Severe obesity increases infection and thrombosis and is a common indication for bridge to candidacy.
Sources of error in the care of durable LVAD recipients
- Referring after irreversible kidney, liver, or pulmonary vascular injury, when the patient no longer qualifies for a device or transplant.
- Treating chronologic age as a contraindication: neither guidelines nor CMS set an upper age limit, and 5-year survival in recipients aged 65 and older exceeds 50%.
- Implanting an isolated LVAD in a patient with severe biventricular failure without a plan for right-sided support.
- Leaving more than mild aortic regurgitation, a patent foramen ovale, or a mechanical aortic prosthesis untreated at implant.
- Reading displayed flow as a measured value: flow is calculated from power, speed, and hematocrit, and pump thrombosis produces a falsely high estimate.
- Raising speed in response to low flow before excluding hypovolemia, suction, and right ventricular failure, all of which worsen at higher speed.
- Treating Doppler opening pressure as mean arterial pressure in a patient with residual pulsatility; it lies closer to systolic pressure.
- Continuing aspirin by habit in a HeartMate 3 recipient without a separate indication after ARIES-HM3, or applying the aspirin-free result to an older pump.
- Substituting a direct oral anticoagulant for warfarin.
- Escalating antiarrhythmic therapy for ventricular ectopy caused by suction or cannula contact.
- Withholding chest compressions in an unresponsive patient without perfusion whose pump cannot be restarted.
- Ordering MRI in an LVAD recipient.
- Postponing goals-of-care and preparedness planning until a catastrophic complication occurs.
References
- Heidenreich PA, Bozkurt B, Aguilar D, et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2022;145:e895–e1032. doi:10.1161/cir.0000000000001063
- Saeed D, Feldman D, El Banayosy A, et al. The 2023 International Society for Heart and Lung Transplantation Guidelines for Mechanical Circulatory Support: a 10-year update. J Heart Lung Transplant. 2023;42:e1–e222. doi:10.1016/j.healun.2022.12.004
- McDonagh TA, Metra M, Adamo M, et al. 2021 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J. 2021;42:3599–3726. doi:10.1093/eurheartj/ehab368
- Nayak A, Cascino TM, DeFilippis EM, et al. The Society of Thoracic Surgeons Intermacs 2025 Annual Report: Focus on Outcomes in Older Adults. Ann Thorac Surg. 2026;121:15–38. doi:10.1016/j.athoracsur.2025.10.005
- Meyer DM, Nayak A, Wood KL, et al. The Society of Thoracic Surgeons Intermacs 2024 Annual Report: Focus on Outcomes in Younger Patients. Ann Thorac Surg. 2025;119:34–58. doi:10.1016/j.athoracsur.2024.10.003
- Rose EA, Gelijns AC, Moskowitz AJ, et al. Long-Term Use of a Left Ventricular Assist Device for End-Stage Heart Failure. N Engl J Med. 2001;345:1435–1443. doi:10.1056/nejmoa012175
- Miller LW, Pagani FD, Russell SD, et al. Use of a Continuous-Flow Device in Patients Awaiting Heart Transplantation. N Engl J Med. 2007;357:885–896. doi:10.1056/nejmoa067758
- Slaughter MS, Rogers JG, Milano CA, et al. Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist Device. N Engl J Med. 2009;361:2241–2251. doi:10.1056/nejmoa0909938
- Rogers JG, Pagani FD, Tatooles AJ, et al. Intrapericardial Left Ventricular Assist Device for Advanced Heart Failure. N Engl J Med. 2017;376:451–460. doi:10.1056/nejmoa1602954
- Milano CA, Rogers JG, Tatooles AJ, et al. HVAD: The ENDURANCE Supplemental Trial. JACC Heart Fail. 2018;6:792–802. doi:10.1016/j.jchf.2018.05.012
- Mehra MR, Uriel N, Naka Y, et al. A Fully Magnetically Levitated Left Ventricular Assist Device – Final Report. N Engl J Med. 2019;380:1618–1627. doi:10.1056/nejmoa1900486
- Mehra MR, Goldstein DJ, Cleveland JC, et al. Five-Year Outcomes in Patients With Fully Magnetically Levitated vs Axial-Flow Left Ventricular Assist Devices in the MOMENTUM 3 Randomized Trial. JAMA. 2022;328:1233–1242. doi:10.1001/jama.2022.16197
- Zimpfer D, Gustafsson F, Potapov E, et al. Two-year outcome after implantation of a full magnetically levitated left ventricular assist device: results from the ELEVATE Registry. Eur Heart J. 2020;41:3801–3809. doi:10.1093/eurheartj/ehaa639
- Mehra MR, Netuka I, Uriel N, et al. Aspirin and Hemocompatibility Events With a Left Ventricular Assist Device in Advanced Heart Failure: the ARIES-HM3 randomized clinical trial. JAMA. 2023;330:2171–2181. doi:10.1001/jama.2023.23204
- Stevenson LW, Pagani FD, Young JB, et al. INTERMACS Profiles of Advanced Heart Failure: The Current Picture. J Heart Lung Transplant. 2009;28:535–541. doi:10.1016/j.healun.2009.02.015
- Denfeld QE, Jha SR, Fung E, et al. Assessing and managing frailty in advanced heart failure: An International Society for Heart and Lung Transplantation consensus statement. J Heart Lung Transplant. 2024;43:1–27. doi:10.1016/j.healun.2023.09.013
- Allen LA, McIlvennan CK, Thompson JS, et al. Effectiveness of an Intervention Supporting Shared Decision Making for Destination Therapy Left Ventricular Assist Device: the DECIDE-LVAD randomized clinical trial. JAMA Intern Med. 2018;178:520–529. doi:10.1001/jamainternmed.2017.8713
- Levy WC, Mozaffarian D, Linker DT, et al. The Seattle Heart Failure Model: prediction of survival in heart failure. Circulation. 2006;113:1424–1433. doi:10.1161/circulationaha.105.584102
- Cowger J, Sundareswaran K, Rogers JG, et al. Predicting Survival in Patients Receiving Continuous Flow Left Ventricular Assist Devices: the HeartMate II Risk Score. J Am Coll Cardiol. 2013;61:313–321. doi:10.1016/j.jacc.2012.09.055
- Mehra MR, Nayak A, Morris AA, et al. Prediction of Survival After Implantation of a Fully Magnetically Levitated Left Ventricular Assist Device. JACC Heart Fail. 2022;10:948–959. doi:10.1016/j.jchf.2022.08.002
- Matthews JC, Koelling TM, Pagani FD, et al. The Right Ventricular Failure Risk Score: a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol. 2008;51:2163–2172. doi:10.1016/j.jacc.2008.03.009
- Soliman OII, Akin S, Muslem R, et al. Derivation and Validation of a Novel Right-Sided Heart Failure Model After Implantation of Continuous Flow Left Ventricular Assist Devices: the EUROMACS Right-Sided Heart Failure Risk Score. Circulation. 2018;137:891–906. doi:10.1161/circulationaha.117.030543
- Lanier GM, Orlanes K, Hayashi Y, et al. Validity and Reliability of a Novel Slow Cuff-Deflation System for Noninvasive Blood Pressure Monitoring in Patients With Continuous-Flow Left Ventricular Assist Device. Circ Heart Fail. 2013;6:1005–1012. doi:10.1161/circheartfailure.112.000186
- Isath A, Vukicevic M, Goldstein D, et al. Renin–angiotensin–aldosterone system inhibition and haemocompatibility-related adverse events in patients with durable left ventricular assist devices: the MOMENTUM-3 trial. Eur Heart J. 2026. Published online. doi:10.1093/eurheartj/ehag386
- Estep JD, Nicoara A, Cavalcante J, et al. Recommendations for Multimodality Imaging of Patients With Left Ventricular Assist Devices and Temporary Mechanical Support: Updated Recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2024;37:820–871. doi:10.1016/j.echo.2024.06.005
- Uriel N, Morrison KA, Garan AR, et al. Development of a Novel Echocardiography Ramp Test for Speed Optimization and Diagnosis of Device Thrombosis in Continuous-Flow Left Ventricular Assist Devices. J Am Coll Cardiol. 2012;60:1764–1775. doi:10.1016/j.jacc.2012.07.052
- Uriel N, Sayer G, Addetia K, et al. Hemodynamic Ramp Tests in Patients With Left Ventricular Assist Devices. JACC Heart Fail. 2016;4:208–217. doi:10.1016/j.jchf.2015.10.001
- Rubinstein G, Moeller CM, Lotan D, et al. Hemodynamic Optimization by Invasive Ramp Test in Patients Supported With HeartMate 3 Left Ventricular Assist Device. ASAIO J. 2024;70:641–650. doi:10.1097/mat.0000000000002167
- Clerkin KJ, Topkara VK, Demmer RT, et al. Implantable Cardioverter-Defibrillators in Patients With a Continuous-Flow Left Ventricular Assist Device: an analysis of the INTERMACS registry. JACC Heart Fail. 2017;5:916–926. doi:10.1016/j.jchf.2017.08.014
- Birks EJ, Drakos SG, Patel SR, et al. Prospective Multicenter Study of Myocardial Recovery Using Left Ventricular Assist Devices (RESTAGE-HF [Remission from Stage D Heart Failure]): medium-term and primary end point results. Circulation. 2020;142:2016–2028. doi:10.1161/circulationaha.120.046415
- Copeland JG, Smith RG, Arabia FA, et al. Cardiac Replacement with a Total Artificial Heart as a Bridge to Transplantation. N Engl J Med. 2004;351:859–867. doi:10.1056/nejmoa040186
- Kanwar MK, Shah P, Cascino T, et al. Impact of Preoperative Temporary Mechanical Circulatory Support on Durable LVAD Outcomes: an analysis of the Society of Thoracic Surgeons National Intermacs Database. JACC Heart Fail. 2025;13:102671. doi:10.1016/j.jchf.2025.102671
- Peberdy MA, Gluck JA, Ornato JP, et al. Cardiopulmonary Resuscitation in Adults and Children With Mechanical Circulatory Support: A Scientific Statement From the American Heart Association. Circulation. 2017;135:e1115–e1134. doi:10.1161/cir.0000000000000504
- Centers for Medicare & Medicaid Services. National coverage determination for ventricular assist devices (20.9.1), version 2, effective December 1, 2020. cms.gov
- Abbott. HeartMate 3 left ventricular assist device: pump parameter overview for clinicians. cardiovascular.abbott
- Baumwol J. "I Need Help"—a mnemonic to aid timely referral in advanced heart failure. J Heart Lung Transplant. 2017;36:593–594. doi:10.1016/j.healun.2017.02.010
- Yancy CW, Januzzi JL Jr, Allen LA, et al. 2017 ACC Expert Consensus Decision Pathway for Optimization of Heart Failure Treatment: Answers to 10 Pivotal Issues About Heart Failure With Reduced Ejection Fraction. J Am Coll Cardiol. 2018;71:201–230. doi:10.1016/j.jacc.2017.11.025