Cardiac device infection and lead extraction

Infection of pacemakers, defibrillators, and resynchronization devices in adults, from incidence, risk, and prevention through diagnosis, complete system removal, transvenous lead extraction, noninfectious lead management, and reimplantation.

Definite infection of a pacemaker, defibrillator, or resynchronization device, whether confined to the pocket or involving the leads, is treated by complete removal of the generator and all leads, ideally within 3 days of diagnosis, combined with pathogen-directed antibiotics. Staphylococcus aureus bacteremia in a patient with a device is accompanied by device infection in about 45% of cases. Transvenous extraction at an experienced center achieves clinical success in about 97%, with procedure-related death near 0.5%.

Governing documents for device infection and lead management

Cardiac implantable electronic device (CIED) infection and lead extraction are governed by consensus statements and scientific statements more than by randomized evidence, and the documents agree on the central principles. The 2017 Heart Rhythm Society (HRS) expert consensus statement on lead management and extraction, written with US and international societies, defined the indications, facilities, training, and outcome terms that most centers use. A 2026 update by HRS, the American Heart Association (AHA), and six other societies replaces it. The update addresses leadless and extravascular systems, prevention and diagnosis of infection, lead management before transcatheter tricuspid intervention, and standardization of extraction facilities, with a lifetime approach to device and lead selection. The AHA published scientific statements on CIED infection in 2010 and 2024. The European Heart Rhythm Association (EHRA) international consensus of 2020, endorsed by US infectious disease and cardiac societies, contains the most detailed recommendations on prevention, diagnosis, and antibiotic therapy. Endocarditis involving a device is also addressed in the 2020 American College of Cardiology (ACC)/AHA valvular heart disease guideline, the 2023 European Society of Cardiology (ESC) endocarditis guideline, and the 2023 Duke-International Society for Cardiovascular Infectious Diseases (ISCVID) diagnostic criteria.

DocumentScopePrincipal content used in practice
2017 HRS expert consensus (lead management and extraction)Lead management, extraction indications, facilities, training, outcomesComplete removal for definite infection; definitions of removal, explant, and extraction; procedural and clinical success; surgical backup
2026 HRS/AHA/APHRS/EHRA/IDSA/LAHRS/PACES/STS updateReplaces the 2017 statementLifetime lead strategy, leadless and extravascular systems, infection prevention and diagnosis, tricuspid intervention, standardized extraction programs
2024 AHA scientific statement (CIED infection)Prevention, diagnosis, managementHematoma prevention, cefazolin prophylaxis, envelope in high-risk patients, removal for valvular endocarditis without proven device involvement, reimplantation timing
2020 EHRA international consensusPrevention, diagnosis, antibiotics, reimplantation, center volume2019 International CIED Infection Criteria, organism-specific removal, antibiotic durations, operator volume
2018 EHRA expert consensus (lead extraction)Definitions and research end pointsStandard terms for success, failure, and complications
2020 ACC/AHA valvular guideline; 2023 ESC endocarditis guidelineEndocarditisDevice removal in device-related and selected valvular endocarditis
2023 Duke-ISCVID criteriaEndocarditis case definitionLead vegetation and intracardiac lead fluorodeoxyglucose (FDG) uptake as major imaging criteria

Documents that govern device infection and lead extraction. APHRS, Asia Pacific Heart Rhythm Society; EHRA, European Heart Rhythm Association; IDSA, Infectious Diseases Society of America; LAHRS, Latin American Heart Rhythm Society; PACES, Pediatric and Congenital Electrophysiology Society; STS, Society of Thoracic Surgeons.

Endocarditis as a whole, including the Duke-ISCVID criteria, organism-specific antibiotic regimens, and valve surgery, is covered on the infective endocarditis page. Implantation complications other than infection are covered under pacemaker complications, implantation and lead extraction in bradycardia, and complications of defibrillator systems.

Clinical classification of device infection

Device infection is classified by the extent of tissue involved, because extent determines whether hardware must be removed and how long antibiotics are given. The EHRA consensus and the AHA statements use overlapping categories.

CategoryDefinitionHardware
Superficial incisional infectionSkin and subcutaneous tissue of the incision, without communication with the pocket, usually early after implantation, generator change, or revisionRetained; oral antibiotics and close observation
Isolated pocket infectionErythema, warmth, swelling, pain, fluctuance, purulent drainage, or sinus tract at the generator site, with negative blood culturesComplete system removal
Pocket erosionGenerator or proximal lead exposed through the skin, with or without inflammation; threatened erosion (adherent, thinned, discolored skin) carries the same meaningComplete system removal; the system is infected regardless of culture results
Pocket infection with bacteremiaLocal signs with positive blood cultures and no lead or valve vegetationComplete system removal
Lead infection or device-related endocarditisVegetation on a lead, the tricuspid valve, or another valve, or abnormal FDG uptake along the intracardiac lead, with or without pocket signsComplete system removal
Occult bacteremiaBacteremia without pocket findings, vegetation, or another sourceRemoval depends on organism and persistence or relapse
Bacteremia from another sourceBacteremia with an identified source and no evidence of device involvementRetained for most gram-negative and streptococcal organisms; removal favored for S. aureus

Clinical categories of CIED infection and their implication for the hardware, from the 2020 EHRA consensus and the 2024 AHA statement.

Infection is also classified by timing. Early infection, within weeks to months of a procedure, usually reflects contamination at operation and presents in the pocket. Late infection, a year or more after the last procedure, more often follows hematogenous seeding of the leads and presents as bacteremia or endocarditis with a normal pocket.

Incidence and trends of device infection

Reported infection rates depend on definitions, follow-up, and study design. Prospective trials and registries report 0.6% to 1.3% in the first year after a procedure, and retrospective series report 2.3% to 3.4%. The difference is attributed partly to closer adherence to preventive measures in prospective studies. In a US Nationwide Inpatient Sample analysis, infection after device procedures rose from 1.53% in 2004 to 2.41% in 2008, coinciding with a rise in renal failure, heart failure, respiratory failure, and diabetes among recipients. A second inpatient database analysis found an increase from 1.45% to 3.41% between 2000 and 2012, most marked for resynchronization devices. Infections have increased faster than implantations.

The complete Danish device cohort gives the most reliable lifetime estimates. Among 97,750 patients followed for 566,275 device-years, infection causing device removal occurred in 1.19% of pacemaker recipients over the device lifetime and in 3.35% of resynchronization defibrillator recipients. Reinterventions, complex devices, prior device infection, male sex, and younger age were associated with higher risk. In a cohort of more than 1 million US Medicare beneficiaries with a first device, 1.1% developed infection more than a year after implantation, a mean of 3.7 years after the procedure.

DeviceLifetime incidence of infectionIncidence rate after de novo implantation (per 1,000 device-years)
Pacemaker1.19%2.04
Implantable cardioverter-defibrillator (ICD)1.91%3.84
Resynchronization pacemaker (CRT-P)2.18%4.38
Resynchronization defibrillator (CRT-D)3.35%6.76

Device-related infection requiring removal in the complete Danish device cohort, 1982 to 2018 (97,750 patients). In the older Danish pacemaker register, the rate was 4.82 per 1,000 device-years after a first implantation and 12.12 after a replacement.

Pathogenesis: contamination, biofilm, and hematogenous seeding

Two mechanisms account for most infections. The more common is contamination of the generator or leads at implantation or at a later procedure, from the patient's skin flora, the air of the procedure room, or the hands of anyone who handles the device. Contamination produces pocket infection that can track along the leads into the venous system and the heart. The second mechanism is hematogenous seeding of the intravascular leads during bacteremia from a distant focus, such as a vascular catheter, skin or soft tissue infection, osteomyelitis, pneumonia, or the urinary or gastrointestinal tract. Erosion through the skin can result from an indolent infection and can also introduce infection.

Bacteria adhere to the device and form a biofilm, a matrix in which organisms are protected from host defenses and are tolerant of antibiotic concentrations that kill free-living bacteria. Adherence is greater on irregular and hydrophobic surfaces and differs among polymers and metals. The fibrous capsule around a chronic generator further impairs host defense and antibiotic penetration, and cultures taken at routine generator change frequently grow organisms even without clinical infection. Biofilm explains why antibiotics alone rarely cure an infected device and why all components, including abandoned leads and fragments, are removed.

Microbiology of device infection

Gram-positive organisms cause 70% to 90% of infections. In pooled cohorts cited by the EHRA consensus, coagulase-negative staphylococci accounted for 37.6% of isolates and S. aureus for 30.8%. Methicillin-resistant staphylococci were found in 33.8% of infections, nearly half of staphylococcal infections, with wide variation by country and hospital. Gram-negative bacilli accounted for 8.9%, and streptococci, enterococci, anaerobes such as Cutibacterium acnes, fungi, and nontuberculous mycobacteria were less frequent. In a single-center series of 189 infections, coagulase-negative staphylococci and S. aureus caused 42% and 29%. S. aureus dominates bacteremic and early pocket infections; coagulase-negative staphylococci dominate indolent pocket infections and erosions. Cultures are negative in a minority, usually after prior antibiotics.

The 2023 Duke-ISCVID criteria treat coagulase-negative staphylococci, Corynebacterium striatum and C. jeikeium, Serratia marcescens, Pseudomonas aeruginosa, C. acnes, nontuberculous mycobacteria, and Candida as typical organisms when intracardiac prosthetic material, including leads, is present. This change recognizes that organisms often dismissed as contaminants are true pathogens on a device.

Risk factors for device infection

Risk factors are patient-related, procedure-related, and device-related. A meta-analysis summarized in the EHRA consensus quantified their effect, and the strongest associations were with end-stage kidney disease, prior device infection, pocket hematoma, lead repositioning, and epicardial leads. Age and sex were not consistent predictors in the meta-analysis, although younger age and male sex were associated with infection in the Danish cohort, where younger patients accumulate more procedures over a lifetime.

Risk factorPooled odds ratio (95% confidence interval)Comment
End-stage kidney disease8.73 (3.42–22.31)GFR 15 mL/min or less, or dialysis
Pocket hematoma8.46 (4.01–17.86)9.33 in prospective studies only
Epicardial leads8.09 (3.46–18.92)Retrospective data
Prior device infection7.84 (1.94–31.60)
Lead repositioning6.37 (2.93–13.82)7.03 in prospective studies
Fever within 24 hours before the procedure4.27 (1.13–16.12)
Abdominal generator pocket4.01 (2.48–6.49)The only device factor significant in prospective studies
Corticosteroid use3.44 (1.62–7.32)
Reduced kidney function (GFR below 60 mL/min)3.02 (1.38–6.64)
Chronic obstructive pulmonary disease2.95 (1.78–4.90)
Inexperienced operator (fewer than 100 procedures)2.85 (1.23–6.58)
Temporary pacing before implantation2.31 (1.36–3.92)3.29 in prospective studies
Malignancy2.23 (1.26–3.95)
Diabetes mellitus2.08 (1.62–2.67)Consistent across 18 studies
Two or more leads2.02 (1.11–3.69)
Device replacement, revision, or upgrade1.98 (1.46–2.70)Generator change alone 1.74
Heparin bridging1.87 (1.03–3.41)
Oral anticoagulation1.59 (1.01–2.48)
Antibiotic prophylaxis0.32 (0.18–0.55)Protective

Pooled estimates for risk factors predisposing to device infection, from the meta-analysis summarized in the 2020 EHRA consensus (Table 3). GFR, glomerular filtration rate.

Procedure duration and device complexity. In the Danish registry, the relative risk of infection, compared with procedures lasting less than 30 minutes, was 1.54 for procedures lasting 60 to 90 minutes, 1.85 for 90 to 120 minutes, and 2.42 for longer than 120 minutes. Compared with pacemakers, the hazard ratio was 1.26 for defibrillators, 1.67 for resynchronization pacemakers, and 2.22 for resynchronization defibrillators.

Hematoma. In BRUISE CONTROL INFECTION, a prospective analysis of 659 patients from the BRUISE CONTROL trial, clinically significant hematoma was the only independent predictor of infection requiring hospitalization within a year. A hematoma was clinically significant when it required reoperation, interruption of anticoagulation, or prolonged hospitalization. Infection occurred in 11% of patients with a hematoma and 1.5% without (hazard ratio 7.7). Empiric antibiotics given when a hematoma developed did not reduce later infection.

Early reintervention. Early reoperation for hematoma or lead dislodgement is among the strongest predictors of infection in registry data matched with Medicare claims, which is one reason careful lead fixation and hemostasis are infection-prevention measures.

Risk scores for device infection

Several scores combine risk factors into a single estimate. They differ in derivation population, weighting, and outcome, and none has been widely adopted for routine treatment decisions. The EHRA consensus considered the evidence behind them too weak for a recommendation. The Prevention of Arrhythmia Device Infection Trial (PADIT) score is the only score derived prospectively, from the 19,603 patients of that trial. It has been validated in more than 51,000 US patients from healthcare claims. Among patients without prior device infection, the risk of major infection rose by 28% per 1-point increase, and prior device infection added independent predictive value (hazard ratio 5.66). When the PADIT, RI-AIAC, Kolek, and Shariff scores were compared in the Italian RI-AIAC registry, only PADIT and the RI-AIAC infection score were significantly associated with infection (C-index 0.64 for both). In the external validation cohort of 1,017 patients, none of the four scores predicted infection (PADIT C-index 0.53).

ScoreDerivationComponents and pointsRisk strata
PADIT19,603 patients in a cluster crossover trialPrior procedures (1 = 1, 2 or more = 4); age (under 60 = 2, 60–69 = 1); eGFR below 30 mL/min (1); immunocompromised (3); procedure type (ICD 2, CRT 4, revision or upgrade 5)1-year hospitalization for infection: 0–4 low (0.51%), 5–6 intermediate (1.42%), 7 or more high (3.41%)
Shariff1,467 patients, retrospectiveOne point per factor: diabetes, heart failure, oral anticoagulation, chronic corticosteroids, creatinine above 1.5 mg/dL, prior device infection, more than 2 leads, epicardial lead, temporary pacing, replacement or upgradeBelow 3: 1.0%; 3 or more: 2.4% at 6 months
Kolek899 patients, retrospectiveOne point per factor: diabetes, creatinine 1.5 mg/dL or more, anticoagulation, corticosteroids, preimplant fever or leukocytosis, prior device infection, 3 or more leads, pacemaker dependence, pocket reentry within 2 weeks2 or more: high risk
Mittal2,891 ICD and CRT-D recipients, single centerEarly pocket reintervention 11, male sex 6, diabetes 3, upgrade 2, heart failure 1, hypertension 1, eGFR below 60 mL/min 10–7: 1%; 8–14: 3.4%; 15 or more: 11.1%
PACE DRAPICD and CRT proceduresValve prosthesis, blood pressure 160/100 mm Hg or more, cancer within 5 years, age 75 or older, CRT or ICD surgery, upgrade, clopidogrel (2 per factor), ticagrelor (3), eGFR below 60 mL/min (1)Below 6: 0.7%; 6 or more: 4.6%
RI-AIAC infection score2,675 patients, 18 Italian centers, with external validationRevision, upgrade, or reimplantation 2; replacement 1; diabetes 1; hospital-acquired infection 11 or more: higher risk (sensitivity 36%, specificity 90%)

Published risk scores for CIED infection. Immunocompromised in PADIT means therapy that suppresses resistance to infection (immunosuppressants, chemotherapy, radiation, long-term or recent high-dose corticosteroids) or a disease advanced enough to do so (leukemia, lymphoma, HIV infection). eGFR, estimated glomerular filtration rate; CRT, cardiac resynchronization therapy; ICD, implantable cardioverter-defibrillator.

PADIT infection risk calculator

PADIT score

One-year risk of hospitalization for device infection
PADIT score—
Risk group—
One-year infection hospitalization—
—

PADIT = Prior procedures + Age + Depressed renal function + Immunocompromised + procedure Type, range 0 to 15. Event rates are those of the derivation cohort, in which overall one-year hospitalization for infection was 0.90% and the optimism-corrected C-index was 0.704. The score was derived in patients who received cefazolin prophylaxis, and in the PADIT trial it did not identify patients who benefited from incremental antibiotics.

Prevention before the procedure

Indication and timing. The first preventive measure is confirming that a device is needed and that the chosen device is the right one. Implantation is deferred in a patient with fever or active infection. Temporary transvenous pacing and central venous lines are avoided when possible and removed before new hardware is introduced. In the PEOPLE study, temporary pacing before implantation was independently associated with infection, a finding summarized on the temporary pacing page.

Device selection. A leadless pacemaker avoids a pocket and transvenous leads, and a subcutaneous defibrillator avoids intravascular leads. Both are options for patients at high risk of infection, including those on dialysis, those with prior infection, and those with limited venous access. In clinical trials enrolling more than 3,000 patients, no leadless pacemaker infections were reported.

Anticoagulation and antiplatelet therapy. In BRUISE CONTROL, continued warfarin produced clinically significant hematoma in 3.5% of patients compared with 16.0% with heparin bridging. The AHA statement considers implantation with an international normalized ratio (INR) of 2 to 3.5 reasonable in patients taking warfarin. In BRUISE CONTROL-2, continued and interrupted direct oral anticoagulant (DOAC) therapy both produced clinically significant hematoma in 2.1%, so either approach is acceptable. Heparin bridging is avoided, and antiplatelet agents other than aspirin are stopped when the indication allows. Perioperative anticoagulant management is covered on the anticoagulant reversal and periprocedural management page.

Systemic antibiotic prophylaxis. A double-blind trial of patients undergoing implantation or generator replacement was stopped early after 649 of 1,000 planned patients. In it, 1 g of intravenous cefazolin before the procedure reduced infection from 3.28% to 0.63% (relative risk 0.19). Prophylaxis is completed within 1 hour before incision. Cefazolin (1 to 2 g intravenously) is the standard agent. Vancomycin (15 mg/kg) is used for beta-lactam allergy and, because it is infused over about an hour, is started 90 to 120 minutes before incision. Routine coverage for methicillin-resistant S. aureus is guided by local prevalence and patient risk.

Skin preparation. Hair is removed with electric clippers; razor shaving is avoided. Topical S. aureus decolonization and a preprocedural skin wash may be performed, although neither has been tested in device-specific trials.

Intraprocedural and postprocedural prevention

Environment and technique. The EHRA consensus recommends that devices be implanted in a room with a sterile environment equivalent to that for other surgical implants. Alcoholic 2% chlorhexidine is superior to povidone-iodine for surgical and vascular catheter procedures, and it is allowed to dry completely before incision because pooled alcohol is a fire hazard with electrocautery. Minimal tissue trauma, meticulous hemostasis, vigorous saline irrigation of the pocket, and layered closure without tension reduce infection. Many operators change gloves before handling the generator. Capsulectomy at generator change is not routine, because it increases bleeding.

Pocket antibiotic wash and postoperative antibiotics: the PADIT trial. PADIT randomized 28 centers, in a cluster crossover design, to conventional prophylaxis (preprocedural cefazolin) or incremental prophylaxis (preprocedural cefazolin plus vancomycin, an intraprocedural bacitracin pocket wash, and 2 days of oral cephalexin). All device procedures were included, 19,603 patients in all. Hospitalization for infection within a year occurred in 1.03% with conventional and 0.78% with incremental treatment (odds ratio 0.77, 95% confidence interval 0.56 to 1.05), and in the high-risk group in 1.23% and 1.01% (odds ratio 0.82; P = 0.26). Antibiotic or antiseptic instillation into the pocket and postoperative antibiotics are therefore not recommended, and the AHA statement confirms that the incremental regimen did not reduce infection.

Hematoma management. Aspiration of a pocket hematoma is contraindicated because it can inoculate the pocket. Evacuation is reserved for uncontrolled pain or threatened wound integrity and is performed in an operating environment.

Wound care. A dressing covers the incision for 2 to 10 days, a pressure dressing may be used for the first 24 hours, and patients avoid soaking the wound until it has healed, usually about a month.

Operator volume. Operators with fewer than about 100 procedures work under supervision. In a Danish nationwide cohort, annual operator volume below 50 procedures was associated with more complications and more infections (1.7% compared with 0.5%). The EHRA consensus recommends an annual minimum of about 50 device procedures per operator.

MeasureStatusEvidence
Preprocedural cefazolin within 1 hour of incision (vancomycin within 120 minutes)RecommendedRandomized trial and meta-analysis
Continued warfarin or continued or interrupted DOAC; no heparin bridgingRecommendedBRUISE CONTROL, BRUISE CONTROL-2
Clipping, chlorhexidine-alcohol skin preparationRecommendedSurgical and catheter trials
Absorbable antibacterial envelope in high-risk proceduresRecommended for selected patientsWRAP-IT
Avoidance of temporary pacing and central lines before implantationRecommended when possibleObservational
Antibiotic or antiseptic pocket washNot recommendedPADIT
Postoperative antibioticsNot recommendedPADIT
Routine capsulectomy at generator changeNot recommendedExpert opinion
Hematoma aspirationContraindicatedExpert opinion
Antibiotic prophylaxis for dental, respiratory, gastrointestinal, or genitourinary procedures in device recipientsNot recommendedNo evidence of device infection from these procedures

Preventive measures for device infection, from the 2020 EHRA consensus and the 2024 AHA statement. DOAC, direct oral anticoagulant.

The antibacterial envelope and the WRAP-IT trial

The absorbable antibacterial envelope is a mesh that holds the generator, elutes minocycline and rifampin locally for at least 7 days, and is resorbed in about 9 weeks. WRAP-IT randomized 6,983 patients undergoing pocket revision, generator replacement, or system upgrade, or initial implantation of a resynchronization defibrillator, to the envelope or no envelope, on top of standard prevention. The primary end point, infection leading to system extraction or revision, long-term antibiotic therapy with recurrence, or death within 12 months, occurred in 0.7% with the envelope and 1.2% without (hazard ratio 0.60, 95% confidence interval 0.36 to 0.98; P = 0.04). Procedure- or system-related complications occurred in 6.0% and 6.9%, meeting noninferiority. Over a mean 20.7 months, major infections occurred in 32 patients with the envelope and 51 without (hazard ratio 0.63).

The absolute reduction was small because infection rates were low in a trial population that excluded patients on dialysis, receiving immunosuppression, or with long-term vascular access, so the number needed to treat to prevent one infection was high. Cost-effectiveness improves where baseline infection risk is higher. The AHA statement considers the envelope for patients at high risk of infection, including those at high risk of hematoma, and the EHRA consensus recommends it for WRAP-IT-type procedures and other high-risk patients, adjusted to local infection rates. Bioscaffold or pericardial patches and antibiotic-soaked gauze have not been tested and are not recommended.

Clinical presentation of device infection

Pocket infection. Erythema, warmth, swelling, tenderness, fluctuance, wound dehiscence, purulent drainage, or a sinus tract at the generator site define pocket infection. Signs may be subtle and fluctuating, especially with coagulase-negative staphylococci. Pocket deformation, skin adherent to the generator, and thinned or discolored skin are signs of indolent infection and threatened erosion. Pain alone can be the first sign. Exposure of the generator or any lead through the skin means the system is infected whatever the cultures show.

Left infraclavicular pacemaker pocket with thinned, violaceous skin stretched over the generator and a slit-like wound at its upper edge through which metal hardware is visible, marked by an arrow.
Pacemaker pocket erosion. The skin over the generator is thinned and discolored, and hardware is exposed through a breach at the upper edge of the pocket (arrow). Exposure of any component establishes infection of the whole system and is an indication for complete removal. Credit: McGuire C, Naitore J, Ramu V, Cureus 2024;16:e75581, PubMed Central, CC BY 4.0.

Systemic infection. Lead infection and device-related endocarditis present with fever, chills, night sweats, malaise, or weight loss, often months to years after the last procedure, and frequently without pocket signs. In reviews, 20% to 50% of patients with device-related endocarditis lack systemic signs. In a prospective cohort of device recipients with S. aureus bacteremia, 60% of confirmed device infections had no local sign of pocket infection. Septic pulmonary emboli from right-sided vegetations present as pneumonia, pleural effusion, or lung abscess and are often misdiagnosed as primary pulmonary infection. Vertebral osteomyelitis and discitis can be the presenting feature. Recurrent fever that responds to antibiotics and relapses after they stop is characteristic.

Physical examination. The pocket is inspected and palpated, the incision examined for dehiscence, and the venous course of the leads examined for thrombophlebitis. A new tricuspid regurgitant murmur, signs of right heart failure, or embolic phenomena indicate valvular or lead involvement. Left-sided endocarditis can coexist, especially with S. aureus.

Distinguishing superficial incisional infection from pocket infection

A superficial incisional infection involves only the skin and subcutaneous tissue of the incision, as a stitch abscess or cellulitis around the wound, and does not communicate with the pocket. It does not require removal of the device. The EHRA consensus treats it with oral antibiotics covering S. aureus for 7 to 10 days after a wound culture. In US practice, cephalexin or dicloxacillin is used, with amoxicillin-clavulanate as an alternative. Trimethoprim-sulfamethoxazole, clindamycin, doxycycline, or linezolid is used where methicillin resistance is prevalent. Close follow-up is needed because early recurrence signals a deeper pocket infection. Inflammation of the incision in the first days after implantation is managed case by case; British and European documents allow the device to be left in place initially, with antibiotics decided individually.

The distinction from pocket infection is clinical and can be difficult. Fluctuance, purulence from the pocket, dehiscence over the generator, and persistence after treatment indicate pocket involvement. FDG positron emission tomography with computed tomography (PET/CT) can distinguish superficial from deep infection when the examination is equivocal. Needle aspiration of an intact pocket is avoided because it can introduce infection; when fluid is sampled for microbiology, it is taken through intact skin away from any sinus, to identify the organism and not to establish the diagnosis.

Blood cultures and microbiologic evaluation

All patients with suspected device infection, including those with an apparently isolated pocket infection, have blood cultures before antibiotics. The EHRA consensus specifies at least three sets drawn at least 30 minutes apart; in a patient with sepsis, two sets are drawn and empiric therapy starts without delay. In a stable patient who has already received antibiotics, a 2- to 3-day antibiotic-free interval may improve the diagnostic yield. Blood is drawn with aseptic technique and the bottles are filled completely, because skin organisms are frequent causes of device infection. A single positive bottle of coagulase-negative staphylococci in a patient with a device is evaluated actively as a possible pathogen. When cultures remain negative, extended incubation for 10 to 14 days and molecular methods identify fastidious organisms.

Swabs from a draining sinus are discouraged because they grow colonizing flora. At removal, pocket tissue, proximal and distal lead segments, and any vegetation are sent for culture; sonication of hardware increases microbial recovery but is interpreted with caution. Blood cultures are repeated 48 to 72 hours after device removal to confirm clearance, and the date of the first negative culture determines antibiotic duration and reimplantation timing.

C-reactive protein is nonspecific. Procalcitonin has high specificity for systemic infection and may help in suspected endocarditis with embolism and in S. aureus device endocarditis.

Bacteremia in a patient with a device: organism-specific evaluation

The probability that bacteremia involves the device depends on the organism, the timing after the last procedure, and whether bacteremia persists or relapses. Patients with a device and bacteremia have the pocket examined and is evaluated for an alternative source.

S. aureus. In a prospective cohort of 33 device recipients with S. aureus bacteremia, 45.4% had confirmed device infection. Device infection was confirmed in 75% when bacteremia occurred within a year of implantation, and was confirmed or possible in 71.5% of later bacteremias. Among 125 patients whose devices were retained after S. aureus bacteremia without evidence of device infection, 20% relapsed within 6 months, and half of those relapses involved the device. Retention of the device after S. aureus bacteremia was associated with higher 1-year mortality. A pacemaker (compared with a defibrillator), more than one prior device procedure, and bacteremia lasting 4 days or more predict device involvement when the pocket is normal. The EHRA consensus indicates complete removal for bacteremia with S. aureus, coagulase-negative staphylococci, Cutibacterium, or Candida, even without a vegetation. It also recommends transesophageal echocardiography and FDG PET/CT in device recipients with S. aureus bacteremia.

Gram-negative bacilli. In a cohort of 49 device recipients with gram-negative bacteremia, only 6% had definite or possible device infection, both definite cases with obvious pocket infection, and relapse among the 34 patients whose devices were retained was rare. Bacteremia with gram-negative organisms other than Pseudomonas and Serratia, or with pneumococci, is treated first with antibiotics, and the device is removed when bacteremia persists or relapses without another source.

Streptococci and enterococci. For viridans and beta-hemolytic streptococci and enterococci, removal may be first-line or may follow persistent or recurrent bacteremia despite appropriate therapy, with prolonged intravenous treatment for about 4 weeks when the device is retained.

Organism in bloodApproach to the device without pocket signs or vegetation
S. aureusComplete removal; transesophageal echocardiography (TEE) and FDG PET/CT
Coagulase-negative staphylococci (true bacteremia)Complete removal
Cutibacterium speciesComplete removal
Candida speciesComplete removal
Viridans or beta-hemolytic streptococci, enterococciRemoval first-line, or after persistent or relapsing bacteremia despite therapy
Pseudomonas aeruginosa, SerratiaIndividualized; typical organisms on prosthetic material under Duke-ISCVID
Other gram-negative bacilli, pneumococciAntibiotics first; removal for persistent or relapsing bacteremia without another source

Device management in bacteremia without pocket infection or echocardiographic vegetation, from the 2020 EHRA consensus. Any bacteremia with pocket signs, a lead or valve vegetation, or abnormal intracardiac lead uptake is a device infection.

Echocardiography: transthoracic, transesophageal, and intracardiac

Transthoracic echocardiography is the first imaging study. It assesses ventricular function, pericardial effusion, tricuspid regurgitation, and pulmonary pressure, and can show large vegetations, but it images leads poorly. The EHRA consensus recommends transesophageal echocardiography in suspected device infection, whether blood cultures are positive or negative and whatever the transthoracic result, before extraction. Transesophageal imaging is more sensitive for lead vegetations, especially in the superior vena cava and high right atrium, and defines vegetation size, mobility, attachment, and valve involvement. The full intracardiac course of all leads, the tricuspid valve, and the left-sided valves are examined. When the initial study is negative and suspicion remains high, transthoracic or transesophageal echocardiography is repeated within 5 to 7 days.

Echocardiographic findings are interpreted with the clinical and microbiologic picture. Fibrin strands and thrombi on leads are common in patients without infection, and lead masses in asymptomatic recipients do not predict later endocarditis. A lead mass by itself therefore does not establish infection; bacteremia, pocket findings, septic pulmonary emboli, and comparison with prior studies do. The AHA statement warns that misclassifying lead echodensities leads to removal of uninfected devices. Conversely, a normal echocardiogram does not exclude device endocarditis, since small vegetations may not be resolved.

Intracardiac echocardiography has high sensitivity for lead vegetations and may be used when transthoracic and transesophageal studies are negative and blood cultures remain positive; a vegetation seen on intracardiac imaging counts as a major criterion in the EHRA criteria. Echocardiographic technique in endocarditis is covered under echocardiography in endocarditis.

After extraction, transthoracic echocardiography before discharge identifies retained lead segments, tricuspid valve injury, right ventricular function, and pulmonary hypertension. Transesophageal imaging is used when sepsis persists after extraction to identify a ghost (a residual tubular sheath of fibrous and infected tissue left in the right atrium or superior vena cava after the lead is withdrawn), retained material, or valve complications.

FDG PET/CT, leukocyte scintigraphy, and CT in device infection

FDG PET/CT and radiolabeled leukocyte single-photon emission computed tomography (SPECT)/CT are complementary to echocardiography, most useful in possible infection, in bacteremia with a negative echocardiogram, and in distinguishing superficial from pocket infection. The EHRA consensus recommends FDG PET/CT, leukocyte scintigraphy, or contrast CT when device endocarditis is suspected with positive blood cultures and negative echocardiography, and FDG PET/CT in device recipients with S. aureus bacteremia. In pooled studies cited by EHRA, FDG PET/CT diagnosed local (generator pocket) infection with a sensitivity of 98% and a specificity of 93%; accuracy is lower for lead infection. Leukocyte SPECT/CT had a sensitivity of 94% and specificity of 100% in the largest study.

Interpretation. FDG accumulates in metabolically active inflammation, whether infectious or sterile. Postimplantation inflammation usually resolves within 6 weeks, so uptake early after a procedure is interpreted with caution. Myocardial glucose uptake is suppressed by a high-fat, low-carbohydrate diet and prolonged fasting. Failed suppression, from diabetes, insulin, recent carbohydrate intake, or dextrose infusions, conceals intracardiac lead activity, and such a study is reported as nondiagnostic because it cannot exclude lead infection. Uptake in the generator pocket, along the extracardiac lead course, and along the intracardiac lead is reported separately. The pocket lies away from myocardial and blood-pool background and is assessed reliably; the intracardiac lead is small, moving, and surrounded by blood pool, and sensitivity for isolated intracardiac lead infection is lower. Under the Duke-ISCVID criteria, abnormal uptake along an intracardiac lead is a major imaging criterion, while isolated pocket uptake is not. A negative study does not override persistent bacteremia or other strong evidence of device infection.

Axial fused FDG PET/CT image of the chest showing intense hypermetabolic uptake surrounding a defibrillator generator in the left anterior chest wall, with an arrow, and no comparable uptake in the heart or mediastinum.
Defibrillator pocket infection on fused FDG PET/CT in a man whose generator had extruded through the skin 5 months after replacement. Intense uptake surrounds the generator in the left anterior chest wall (arrow). Credit: Beladel W, Aabbar S, Hasni K, El Minaoui M, Cureus 2025;17:e90392, PubMed Central, CC BY 4.0.

Whole-body findings. Whole-body imaging identifies septic pulmonary emboli, vertebral osteomyelitis, splenic lesions, mycotic aneurysms, and the portal of entry, findings that change antibiotic duration and source control. Patients with definite device infection without pocket uptake have had worse outcomes, consistent with endovascular infection from an unrecognized distant source.

After incomplete removal. When hardware cannot be removed completely, FDG PET/CT can identify residual infection and guide the decision on chronic suppressive antibiotics.

Computed tomography. Contrast CT shows pocket collections, septic pulmonary emboli, abscesses, and mycotic aneurysms, and CT pulmonary angiography is recommended for recurrent pneumonia in a device recipient. A filling defect on a lead at CT is not called a vegetation from CT alone; echocardiography and FDG PET/CT are more sensitive for infection. Advanced imaging in endocarditis is covered under cardiac CT, FDG PET/CT, and leukocyte scintigraphy.

Diagnostic criteria: International CIED Infection Criteria and Duke-ISCVID

The modified Duke criteria were not designed for device infection. The EHRA consensus merged the modified Duke and 2015 ESC criteria with device-specific findings into the 2019 International CIED Infection Criteria. The 2023 Duke-ISCVID criteria, used in the United States and summarized on the endocarditis page, also incorporate devices. An endovascular electronic device is a predisposing condition; a lead vegetation on echocardiography or CT and abnormal FDG uptake along an intracardiac lead are major imaging criteria; organisms typical of prosthetic material count as typical organisms; and microorganisms identified in an extracted device satisfy the pathologic criterion.

Element2019 International CIED Infection Criteria
Definite pocket or generator infectionSwelling, erythema, warmth, pain, and purulent discharge or sinus; or pocket deformation, adherence, and threatened erosion; or exposed generator or proximal leads
Major microbiologicBlood cultures positive for typical device organisms (coagulase-negative staphylococci, S. aureus); organisms consistent with endocarditis in 2 separate cultures; persistently positive cultures; Coxiella burnetii criteria
Major imagingEchocardiography, including intracardiac, showing pocket infection, lead vegetation, or valve endocarditis; FDG PET/CT or leukocyte SPECT/CT uptake at the pocket, along the leads, or at a valve; definite paravalvular leak on CT
MinorPredisposition (including new tricuspid regurgitation) or injection drug use; temperature above 38 °C; vascular phenomena, including septic pulmonary emboli seen only on imaging; microbiologic evidence short of a major criterion, including pocket or lead culture from a removal through a noninfected pocket
Definite device endocarditis2 major, or 1 major and 3 minor criteria
Possible device endocarditis1 major and 1 minor, or 3 minor criteria

The 2019 International CIED Infection Criteria from the 2020 EHRA consensus, which merged the modified Duke and 2015 ESC criteria with device-specific findings.

Complete system removal as the standard of care

Complete removal of the generator and all leads, including abandoned and epicardial leads and lead fragments, combined with antibiotic therapy is the treatment of definite device infection, whether confined to the pocket or systemic. The 2017 HRS consensus, the 2020 ACC/AHA valvular guideline (Class 1 for documented infection of the device or leads), the 2020 EHRA consensus, and the 2024 AHA statement agree. Antibiotics alone, generator removal without leads, and pocket debridement with device retention are followed by high relapse rates. In a cohort of 416 patients, antibiotic therapy without device removal was associated with a sevenfold increase in 30-day mortality (hazard ratio 6.97). Vascular ports and long-term dialysis catheters are removed at the same time where feasible.

Valvular endocarditis without proven device involvement. The 2024 AHA statement recommends complete device removal in patients with valvular endocarditis even without definite device or lead involvement, because leads are frequently seeded and the risk of relapse with retained hardware is high. The 2020 ACC/AHA guideline rates complete removal reasonable (Class 2a) in valvular endocarditis caused by S. aureus or fungi and in patients undergoing valve surgery for endocarditis.

Timing. The EHRA consensus specifies removal ideally within 3 days of diagnosis, because transvenous extraction within 3 days of admission was associated with lower in-hospital mortality and shorter hospitalization. In the cohort of 416 patients, immediate removal, compared with initial antibiotics and delayed or no removal, was associated with a threefold reduction in 1-year mortality (hazard ratio 0.35). In US Medicare data, extraction within 6 days of diagnosis was associated with lower mortality than no extraction (adjusted hazard ratio 0.69), as was any extraction (adjusted hazard ratio 0.82). The AHA statement summarizes these data as a threefold lower 1-year mortality with removal within 7 days.

Method. Percutaneous transvenous extraction is the method of choice, because major complications and mortality at 1 and 12 months are lower than with open surgery. Epicardial leads require surgical removal. When a pocket infection does not involve the distal portion of an epicardial lead, the lead may be cut through a separate incision away from the pocket, the proximal portion removed, and the distal portion left in place, a decision supported by FDG PET/CT.

Pocket management. After removal, the pocket is debrided, with excision of the fibrous capsule and removal of all nonabsorbable suture, and irrigated with saline; antibiotic irrigation offers no advantage. The wound is closed primarily, with or without a drain, or left for delayed closure with negative pressure wound therapy. No part of the removed system, including the extraction sheath, is reused, and central and peripheral lines are changed where feasible.

Evaluation by an experienced team. The 2017 HRS consensus recommends evaluation of all patients with documented device infection by a physician experienced in device infection and lead extraction. A multidisciplinary endocarditis team that reviewed imaging was associated with a reduction in 1-year mortality from 18.5% to 8.2% in a study cited by the EHRA consensus.

Underuse of extraction in device infection

Despite the consistency of recommendations, most patients with device infection do not undergo timely extraction. Among 11,304 US Medicare beneficiaries with device infection more than a year after implantation, only 18.6% underwent extraction within 30 days of diagnosis: 13.4% within 6 days and 5.2% between days 7 and 30. One-year survival in the whole cohort was 68.3%. In a US readmissions database of 25,303 admissions of device recipients with endocarditis between 2016 and 2019, only 11.5% were managed with extraction, a proportion that rose from 7.6% to 14.9%. Procedural complications occurred in 2.7%, index mortality was lower with extraction (6.0% compared with 9.5%), and extraction was independently associated with lower mortality (adjusted odds ratio 0.47). Extraction was less likely in older patients, women, and those with dementia or kidney disease.

Hospital extraction volume influences use. Among admissions for device-related endocarditis at extraction centers, extraction was performed in 6.9%, 19.3%, and 26% at low-volume (1 to 17 per year), medium-volume (18 to 45), and high-volume (more than 45) centers, with similar complication rates. These observational data cannot remove confounding by frailty, but they indicate that many patients who meet guideline criteria are not referred, and that referral to an extraction center is part of guideline-directed care.

Antibiotic therapy and duration after removal

Antibiotics complement removal; they treat associated systemic infection and residual infection in native tissue. Empiric therapy covers methicillin-resistant staphylococci, usually with vancomycin (or daptomycin), with gram-negative coverage when the patient is septic. Definitive therapy follows culture and susceptibility, with a beta-lactam for methicillin-susceptible staphylococci. Rifampin is not added for biofilm when the device is removed, unless another nonremovable foreign body, such as a prosthetic valve, is infected. Organism-specific regimens for endocarditis are covered on the antibiotic therapy for endocarditis page.

SyndromeEmpiric therapyDuration
Superficial incisional infectionOral antistaphylococcal agent (cephalexin, dicloxacillin, or amoxicillin-clavulanate; trimethoprim-sulfamethoxazole, clindamycin, doxycycline, or linezolid for MRSA)7–10 days; device retained
Isolated pocket infection, negative blood culturesVancomycin (or daptomycin); gram-negative coverage if systemic signs10–14 days after removal; partial oral therapy often used
Bacteremia without lead or valve vegetation, with or without pocket infectionVancomycin (or daptomycin) plus gram-negative coverageAt least 2 weeks after removal (4 weeks in the EHRA consensus; 2 weeks considered when blood cultures clear promptly)
Isolated lead vegetation, no valve involvementAs for endocarditis2 weeks after removal may suffice if cultures clear, post-removal TEE shows no valve vegetation, and there is no lung abscess; a total course of at least 4 weeks (S. aureus excepted)
Native valve endocarditis with a deviceAs for endocarditis4 weeks after removal
Prosthetic valve endocarditis with a deviceAs for prosthetic valve endocarditis; rifampin added for staphylococci after 5–7 days4–6 weeks
Salvage when removal is not possibleAs for prosthetic valve endocarditis4–6 weeks intravenous, then observation or individualized long-term oral suppression

Antibiotic therapy by syndrome, from the 2020 EHRA consensus (Table 9). Vancomycin 30 to 60 mg/kg/day in 2 to 3 doses and daptomycin 8 to 10 mg/kg once daily were the listed doses, adjusted for kidney function. Durations are counted from device removal. MRSA, methicillin-resistant S. aureus; TEE, transesophageal echocardiography.

In a single-center series in which 98% of patients had complete removal, median antibiotic duration after removal was 18 days for pocket infection and 28 days for endocarditis, and 96% of patients were cured. Oral switch is reasonable for isolated pocket infection after removal, because residual infection involves skin and soft tissue, but not when bacteremia was present. When a pocket has eroded with minimal inflammation, antibiotics can be deferred until removal so that pocket cultures are not sterilized.

Retained hardware: salvage and chronic suppressive therapy

Some patients cannot undergo removal because of frailty, limited life expectancy, prohibitive procedural risk, or refusal. Antibiotic therapy alone is not recommended for device infection, and salvage succeeds in a minority. When removal is not possible, the EHRA consensus recommends intravenous antibiotics, as for prosthetic valve endocarditis, for 4 to 6 weeks, after which antibiotics are stopped under close follow-up or continued as individualized oral suppression. Oral options depend on susceptibility: trimethoprim-sulfamethoxazole, clindamycin, or doxycycline for methicillin-resistant staphylococci, with combination suppression generally avoided and linezolid unsuitable for long-term use. The 2024 AHA statement supports chronic suppression in selected patients ineligible for removal, and describes continuous in situ high-concentration antibiotic delivery for non-S. aureus infection as investigational. Patients with retained infected hardware have high in-hospital and subsequent mortality.

Lead extraction: definitions and outcome terms

The HRS consensus documents and the 2018 EHRA consensus define the terms used to compare extraction studies.

TermDefinition
Lead removalRemoval of a pacing or defibrillator lead by any technique
Lead explantRemoval, by simple traction through the implant vein and without specialized extraction tools, of leads that were all implanted less than 1 year earlier
Lead extractionRemoval of at least one lead implanted for more than 1 year, or removal of any lead that requires specialized equipment not part of a standard implant package, or removal through a route other than the implant vein
Complete procedural successRemoval of all targeted leads and lead material from the vascular space without procedure-related death or permanently disabling complication
Clinical successRemoval of all targeted leads, or retention of a small fragment that does not compromise the goals of the procedure, without procedure-related death or permanently disabling complication
Partial radiological successLess than 4 cm of a targeted lead remaining
Radiological failureMore than 4 cm of a targeted lead abandoned after an attempt at removal
Major complicationDeath, or a complication that is life-threatening or requires significant surgical intervention, such as vascular or cardiac tear requiring pericardiocentesis, chest tube, or surgical repair

Standard terms for lead removal procedures and outcomes, from the HRS consensus documents, the 2018 EHRA consensus, and the ELECTRa registry. The fragment allowed under clinical success is one that does not increase the risk of further complications, and for infection it must not be a source of persistent infection.

The 1-year threshold reflects fibrosis: a lead in place for less than a year is usually removable by traction, while older leads are bound to the vein, the superior vena cava, the tricuspid apparatus, the ventricular wall, and one another by fibrous and sometimes calcified tissue.

Noninfectious indications for lead extraction

Noninfectious extraction is individualized against procedural risk, the patient's age and expected device lifetime, and the alternatives of abandonment, venoplasty, or a contralateral or leadless system.

Abandoned leads, MRI, and future infection

Lead management is a lifetime decision. When a lead is nonfunctional or no longer needed, abandonment and extraction are weighed using lead age and design, the number of leads already present, venous patency, infection risk, interaction with the tricuspid valve, the likely need for future leads, the need for MRI, the patient's age and comorbidity, and the expected extraction risk. Abandonment defers risk; it does not remove it, because older leads are harder to extract and an infection years later requires removal of all leads, abandoned ones included.

Factors favoring extraction include intravascular crowding (four or more leads in one vessel or five or more in the superior vena cava), electrical interaction between abandoned and active leads causing oversensing, young age with a long future of device procedures, and a high risk of future infection, such as immunosuppression or prior endocarditis. In a 5% Medicare sample of 6,859 patients who received a new lead at least a year after implantation, 16.2% had the old lead extracted and 83.8% had it capped and abandoned. Mortality, lead revision, and later extraction at 5 years did not differ, and extraction was associated with less device infection at 5 years (adjusted hazard ratio 0.78). When a lead is abandoned, it is capped and left in a condition that permits future extraction.

MRI. The 2017 HRS MRI consensus considers MRI reasonable in a patient with a non-MRI-conditional system under three conditions: no fractured, epicardial, or abandoned leads; the scan is the best test for the clinical question; and an institutional protocol is followed. Abandoned and epicardial leads can heat at their tips in the radiofrequency field. Their presence excludes the scan from that pathway, so scanning such patients is reserved for experienced centers with a specific risk assessment. MRI in patients with devices is covered under MRI in patients with pacemakers and MRI in patients with an ICD.

Planning a lead extraction

Planning begins with the indication, the patient's pacing and defibrillation needs, and a procedural risk estimate. The device and lead models, implant dates, fixation mechanisms, and number of leads are recorded. Older leads, defibrillator leads (particularly dual-coil leads with a superior vena cava coil), passive-fixation leads, and leads with coils exposed to tissue ingrowth are harder to remove.

Predictors of complications. In ELECTRa, procedure-related major complications were more frequent in women (odds ratio 2.11), with lead dwell time longer than 10 years (odds ratio 3.54), with powered sheaths used alone, and with the femoral approach. In LExICon, major adverse events were associated with body mass index below 25 kg/m², and procedural failure was more frequent with leads implanted more than 10 years and at low-volume centers. Risk scores combining such factors, including the SAFeTY TLE score, identify patients for whom a high-volume center with surgical backup is appropriate.

Imaging. Chest radiography identifies leads, abandoned fragments, and lead course. CT venography maps subclavian, innominate, and superior vena cava stenosis or occlusion and collaterals; a patent-appearing superior vena cava does not exclude upstream occlusion. Gated CT can show lead tips beyond the myocardium, lead-to-vessel adhesion, and calcification. Transesophageal echocardiography measures vegetation size and mobility and documents baseline tricuspid function and pericardial fluid.

Pacing and defibrillation needs. Pacing dependence determines whether temporary pacing is needed during and after extraction. Patients with a defibrillator indication are monitored with external defibrillation available, and the need for protection after extraction is planned in advance.

Consent and center data. The 2017 HRS consensus states that programs share their volume, success rates, and complication rates with patients before extraction.

Extraction tools and techniques

Extraction proceeds in steps of increasing force and invasiveness, usually through the implant vein from the superior approach, with femoral, internal jugular, or surgical approaches when needed.

TechniqueMechanismPrincipal use and limitsReported success
Simple tractionTraction on the lead, with a standard stylet and unscrewing of an active-fixation helixRecent leads (under about 2 years) and some infected leads; risks lead fracture and venous obstruction with older leadsUp to 85% in selected leads
Locking styletStylet locked inside the conductor lumen to transmit traction to the tip and act as a railUsed with almost all sheath techniques; limited by sharp bends and damaged conductorsUp to 93%
Mechanical telescoping sheathsInner and outer polypropylene, polymer, or steel sheaths advanced over the lead with rotation and countertractionFibrous adhesions; less effective against dense calcification; long proceduresAbout 98% in reported series
Excimer laser sheathUltraviolet laser energy ablates fibrous tissue circumferentially at the sheath tipFibrous binding sites; little effect on calcium; stiff sheath increases perforation risk94% complete removal (PLEXES)
Rotational mechanical sheathHand-powered bidirectional rotating threaded tip cuts through fibrous and calcified tissueDense or calcified adhesions and venous entry sites95.2% complete removal of the 1,552 leads extracted with a rotational sheath, median dwell 106 months (PROMET)
Electrosurgical dissection sheathRadiofrequency energy between two tip electrodesFocal dissection; largely withdrawn from use—
Femoral snares and basketsFree lead end pulled into the inferior vena cava and grasped from the femoral veinCut or fractured leads, free-floating fragments, failed superior approachMore than 90%
Surgical and hybrid extractionSternotomy, thoracotomy, or minithoracotomy, with or without cardiopulmonary bypass, alone or combined with transvenous toolsFailed transvenous extraction, concomitant cardiac surgery, very large vegetations, epicardial leads—

Tools and techniques for transvenous lead extraction, with success rates reported in reviews and trials.

Laser sheaths. In PLEXES, the only randomized trial of extraction technique, 301 patients with 465 chronic pacemaker leads were assigned to extraction with an excimer laser sheath or the same tools without the laser. Complete removal was achieved in 94% with the laser and 64% without (P = 0.001), and failed nonlaser extractions were completed with the laser in 88%. Potentially life-threatening complications occurred in three laser patients, one fatal, and in no nonlaser patients, a difference that was not significant.

Rotational mechanical sheaths. The PROMET multicenter study included 2,205 patients and 3,849 leads (median dwell time 74 months; infection the indication in 46%). Clinical success was 97.0%, complete lead removal 96.5%, major complications 1%, and procedure-related mortality 0.18%. In the 1,552 leads extracted with a rotational sheath (median dwell 106 months), complete removal was 95.2% and procedure-related mortality 0.4%.

Femoral and internal jugular approaches. When the proximal end of a lead cannot be reached, has retracted into the vein, or breaks during extraction, a femoral approach with a snare, a deflectable catheter, or a basket grasps the free end in the inferior vena cava or right atrium and removes it through a large femoral sheath. The femoral route is also used for fragments in the heart or pulmonary arteries. It is effective but involves longer procedures and more fluoroscopy, and in ELECTRa it was associated with more complications. An internal jugular approach is an alternative superior route when the implant vein cannot be used.

Surgical and hybrid extraction

Open surgical extraction is preferred when transvenous extraction has failed, when the patient needs cardiac surgery for another reason (such as valve surgery for endocarditis), when epicardial leads must be removed, and, in some centers, when lead vegetations are very large. The 2009 HRS consensus on transvenous lead extraction, as summarized in a 2022 review, lists two situations in which open surgical extraction or preemptive aspiration is considered. These are leads placed through abnormal structures such as the left ventricle, the aorta, or the subclavian artery, and vegetations larger than about 25 mm. A previous sternotomy or bypass grafts influence the surgical approach; a lateral thoracotomy can avoid internal mammary grafts. Hybrid procedures combine transvenous tools with a minimally invasive surgical approach to the heart or great veins.

Facility, team, and operator requirements for lead extraction

Vascular tear, cardiac perforation, and tricuspid injury can be fatal within minutes, so extraction is performed where surgical repair is immediately possible. The procedure room, whether an electrophysiology laboratory, a hybrid room, or an operating room, matters less than immediate access to cardiothoracic surgery; a 2022 review of transvenous extraction cites cardiothoracic surgical intervention available no later than 10 minutes after a complication. The team includes the extracting physician, a cardiothoracic surgeon available on standby, anesthesia, imaging (usually transesophageal echocardiography), and staff trained in the equipment. Arterial pressure monitoring, large-bore venous access, blood products, and a pericardiocentesis set are prepared, and many centers place a wire in the inferior vena cava to allow rapid deployment of an occlusion balloon.

Training and volume. A 2022 review, citing the 2018 EHRA expert consensus, gives the training requirement as at least 40 leads extracted in at least 30 procedures. Maintenance of competence requires at least 15 procedures and 20 leads extracted per year. In ELECTRa, centers performing 30 or more extractions a year had higher radiological and clinical success than lower-volume centers. All-cause in-hospital major complications were lower (2.4% compared with 4.1%), as were deaths (1.2% compared with 2.5%). Procedure-related complications did not differ significantly. In LExICon, low center volume was associated with procedural failure. The 2026 HRS update addresses standardization of the extraction approach, protocol, and facilities.

Outcomes of transvenous lead extraction

Large registries show high success and low procedural mortality at experienced centers. Overall mortality after extraction for infection is driven by the infection and comorbidity more than by the procedure.

StudyDesign and populationSuccessMajor complicationsMortality
PLEXES, 1999Randomized, 301 patients, 465 pacemaker leads; laser vs nonlaser sheathsComplete removal 94% vs 64%3 life-threatening events with laser, 0 without1 procedural death (laser)
LExICon, 201013 US centers, 1,449 consecutive patients, 2,405 leads (median dwell 82 months), laser-assistedComplete removal 96.5%; clinical success 97.7%1.4%Procedural 0.28%; in-hospital all-cause 1.86%
ELECTRa, 2017Prospective, 73 European centers, 3,510 patients; infection about 53% of indicationsClinical 96.7%; radiological 95.7%Procedure-related 1.7%Procedure-related 0.5%
PROMET, 20206 European centers, 2,205 patients, 3,849 leads; mechanical and rotational toolsClinical 97.0%; complete lead removal 96.5%1%Procedure-related 0.18%
Mayo Clinic infection cohort, 2011416 patients with device infectionComplete procedural 81.9%; clinical 93.9%Removal complications 12.0%Removal-related death 0.8%; 30-day 5.5%; 1-year 14.6%

Principal studies of lead extraction outcomes. Definitions of success differed among studies.

In LExICon, in-hospital all-cause mortality was 4.3% when extraction was performed for endocarditis, 7.9% with endocarditis and diabetes, and 12.4% with endocarditis and a creatinine of 2.0 mg/dL or more; pocket infection and device-related endocarditis were independent predictors of death. In ELECTRa, two-thirds of procedure-related major complications occurred during the procedure and one-third afterward, and systemic infection predicted all-cause mortality (odds ratio 4.93).

Complications of lead extraction: vascular tear and tamponade

The most feared complications are tears of the superior vena cava or brachiocephalic vein and perforation of the right atrium or ventricle. In ELECTRa, complications requiring pericardiocentesis, a chest tube, or surgical repair occurred in 1.4%. A superior vena cava tear bleeds into the right pleural space or mediastinum and presents with sudden hypotension, sometimes without pericardial effusion; a cardiac perforation presents with tamponade. Recognition depends on continuous arterial pressure monitoring and immediate echocardiography.

Endovascular occlusion balloon. A compliant balloon placed over a wire from the femoral vein and inflated across the superior vena cava tamponades a tear and buys time for sternotomy and repair. Among 116 confirmed superior vena cava tears reported in the United States between 2016 and 2018, 88.2% of patients survived when the balloon was used properly, compared with 56.9% when it was not used or used improperly (P = 0.0002). Proper deployment was an independent predictor of survival (odds ratio for death 0.13). In a series of 21 high-risk extractions, prophylactic positioning of the balloon was not associated with adverse events.

ComplicationFeaturesManagement
Superior vena cava or brachiocephalic tearSudden hypotension, hemothorax or mediastinal hematomaOcclusion balloon, volume, emergency sternotomy and repair
Right atrial or ventricular perforationTamponade, new effusionPericardiocentesis; surgical repair if bleeding continues
Tricuspid valve injuryNew or worse tricuspid regurgitation after extraction of leads adherent to the valveFollow-up imaging; repair or replacement when severe and symptomatic
Pulmonary embolism of vegetation or thrombusHypoxemia, fever, septic pulmonary infarctsSupportive care, antibiotics; aspiration before extraction in selected large vegetations
Lead fragment retentionResidual segment on radiographyFemoral snare; left in place if small and noninfectious
Hemothorax, pneumothoraxAccess or vascular injuryChest tube
Arrhythmia and loss of pacingAsystole in pacing-dependent patients, ventricular arrhythmiasTemporary pacing, external defibrillation
Pocket hematomaMost frequent minor complicationCompression; evacuation if tense
Arteriovenous fistula, vascular access injuryFemoral or subclavian approachVascular repair

Complications of transvenous lead extraction and their management. Pericardiocentesis technique is described on the pericardiocentesis page.

The EHRA consensus reports in-hospital extraction complications of 2% to 3% in infection series, including emergency thoracotomy, arteriovenous fistula, and tricuspid damage, and septic pulmonary emboli in about 1%.

Large vegetations and percutaneous aspiration

Vegetation size influences the procedural plan because extraction can dislodge vegetation into the pulmonary arteries. Transvenous extraction remains preferred for vegetations larger than 10 mm, and small series report good short-term outcomes despite a high frequency of asymptomatic pulmonary embolism. The EHRA consensus states that for vegetations of approximately 20 mm or more, percutaneous aspiration before and during transvenous extraction, or surgical extraction, may be considered. Friability and mobility on transesophageal imaging matter as much as size, and vegetation size is difficult to measure accurately.

Percutaneous aspiration uses large-bore catheters, either a veno-venous extracorporeal circuit with an in-line filter or mechanical aspiration systems, to debulk vegetations before or during extraction. In a six-center registry of 53 patients who underwent aspiration of 63 lead-related masses during extraction between 2016 and 2024 (mean vegetation size 28 mm), a mechanical aspiration system (Penumbra Indigo) was used in 75%. Extraction achieved complete success in 92% and clinical success in 98%, aspiration-related major complications occurred in 4%, and 19% of patients died during a median follow-up of 139 days. Aspiration is an adjunct to source control; it does not replace complete removal of the hardware or treatment of a destroyed valve. Whether it improves survival has not been tested in a controlled study.

Bridging pacing and arrhythmia protection after extraction

In patients who are pacing dependent, temporary pacing bridges the interval until infection is controlled. An active-fixation lead connected to an externalized permanent generator is preferred to a passive temporary catheter for bridges longer than a few days, because it dislodges less, allows mobilization, and has fewer complications. The EHRA consensus suggests placing it ipsilateral to the extraction, preferably not through the vein from which the infected lead was removed, to preserve the contralateral side for the permanent system. Studies of active-fixation bridging after extraction of infected devices are tabulated under active-fixation temporary leads. Temporary pacing in patients who are not pacing dependent adds risk without benefit.

A leadless pacemaker can be implanted at the time of extraction or shortly afterward in a pacing-dependent patient. In the Micra postapproval registry, 105 patients had a leadless pacemaker implanted within 30 days of explant of an infected system, a median of 6 days after explant and on the same day in 37%. Implantation succeeded in 99%, and during a mean follow-up of 8.5 months no device was removed for infection, although two patients died of sepsis.

For patients with a defibrillator indication and no pacing requirement, a wearable cardioverter-defibrillator provides protection until reimplantation, as noted in the EHRA consensus. The 2017 AHA/ACC/HRS ventricular arrhythmia guideline considers it reasonable after removal of an ICD for infection in a patient with prior sudden cardiac arrest or sustained ventricular arrhythmia (Class 2a), as described under wearable defibrillator indications. Defibrillator selection after extraction, including the subcutaneous and extravascular systems, is covered under choosing the defibrillator system.

Reimplantation: indication, timing, site, and device

Indication. The need for a device is reassessed after any removal. A substantial minority of patients no longer meet an indication: the pacing indication may have resolved, a primary prevention defibrillator may no longer be warranted after reassessment of ejection fraction and prognosis, and a resynchronization device may be downgraded or upgraded. A different device may serve better than the original.

Timing. Reimplantation is delayed until local and systemic signs of infection have resolved. The 2024 AHA statement and the 2020 EHRA consensus rate reimplantation reasonable after blood cultures have been negative for at least 72 hours after removal; in valvular endocarditis the AHA statement advises waiting 14 days. Persistent bacteremia after extraction prompts a search for retained hardware, a ghost, valve infection, septic thrombophlebitis, or metastatic infection before any new implant.

Site. The new system is implanted away from the infected site: the contralateral pectoral region, the iliac vein, or an epicardial position. Ipsilateral transvenous reimplantation is avoided.

Device choice. A leadless pacemaker is a reasonable choice for single-chamber ventricular pacing in patients at high risk of reinfection; leadless pacing with atrioventricular synchrony has broadened its use. A subcutaneous defibrillator, for patients who need no pacing, antitachycardia pacing, or resynchronization, reduces the risk of new endovascular infection, and its removal is simpler if infection recurs. In the EFFORTLESS registry of 985 patients, infection requiring removal occurred in 2.4% over 3 years. Epicardial systems have long been used in patients at very high risk of reinfection. Resynchronization candidates are reassessed for the indication and for conduction system pacing, as described under upgrade to resynchronization.

SituationReimplantation approach
No continuing device indicationNo reimplantation
Isolated pocket infection, cultures negativeContralateral system once the pocket is controlled and blood cultures are negative for at least 72 hours
Bacteremia or lead infection without valve involvementNew system after at least 72 hours of negative cultures and resolution of systemic signs
Valvular endocarditisDelay of 14 days after removal advised by the AHA statement
Pacing dependent during the delayActive-fixation lead with externalized generator, or leadless pacemaker
Defibrillator indication without pacing need during the delayWearable cardioverter-defibrillator
High risk of reinfectionLeadless pacemaker, subcutaneous or extravascular defibrillator, or epicardial system

Reimplantation after removal of an infected device, from the 2024 AHA statement and the 2020 EHRA consensus. No part of the removed system is reused.

Device infection in older adults, women, kidney disease, and adult congenital heart disease

Older and frail adults. Age is not an independent predictor of infection after adjustment for comorbidity, but frailty, loss of subcutaneous tissue, and muscle wasting increase erosion. A submuscular generator position is recommended in selected older patients with little subcutaneous tissue. Older age, female sex, dementia, and kidney disease are associated with lower use of extraction, and the risk of extraction is weighed against the high mortality of retained infected hardware.

Kidney disease and dialysis. End-stage kidney disease carries the highest pooled risk of infection, and dialysis catheters provide a continuing source of bacteremia. Leadless pacemakers and subcutaneous defibrillators preserve veins needed for dialysis access and avoid endovascular leads.

Women. Female sex was associated with procedure-related major complications in ELECTRa (odds ratio 2.11), attributed in part to smaller and more fragile veins, and with lower use of extraction in US data.

Adult congenital heart disease. Patients with congenital heart disease accumulate procedures, lead revisions, and epicardial leads over a lifetime, all of which raise infection risk. In adults with congenital heart disease, acute infection after defibrillator implantation was low (0.2%) in the US national defibrillator registry. Most infections appear later. Over long-term follow-up, serious infection occurs in about 1% to 5% of patients, a range similar to that in other adults. Venous anomalies, baffles, and systemic venous atria complicate extraction, and leads placed through a systemic venous atrium or systemic ventricle are a relative contraindication to transvenous extraction. The subcutaneous defibrillator is an alternative in patients without a pacing need.

Ventricular assist devices. Driveline and pump infections differ from electronic device infection and are summarized on the endocarditis page.

Prognosis and mortality after device infection

Device infection carries in-hospital or 30-day mortality of 5% to 8%, most of it from ongoing sepsis and comorbidity. The extraction itself carries a procedure-related mortality near 0.5%. Long-term mortality is 1.5 to 2.4 times that of uninfected device recipients, and the excess persists for years. In Medicare beneficiaries, 3-year mortality was 47.7% to 53.8% after device infection, depending on device type, compared with 31.6% to 36.5% without infection. In a later Medicare cohort with infection more than 1 year after implantation (median age 75 years), 1-year survival was 68.3%, so about 32% died within a year. Preliminary ELECTRa data showed 1-year mortality of 15.1% after extraction for systemic infection, 6.9% for local infection, and 3.0% without infection.

Mortality is higher with device endocarditis than with pocket infection, with heart failure, kidney failure, diabetes, or corticosteroid use, in women, and in patients who do not undergo complete removal or whose removal is delayed. Patients cured by complete removal and a full course of antibiotics may have a prognosis similar to that of patients never infected, which suggests that much of the excess mortality reflects the comorbidities that predispose to infection and the consequences of incomplete treatment.

Follow-up after treatment of device infection

After removal and antibiotic therapy, follow-up confirms cure and protects the new system. Patients are examined for fever, pocket changes, and signs of relapse; recurrent bacteremia with the same organism suggests relapse and raises the possibility of retained material, a ghost, valve infection, or metastatic infection, which transesophageal echocardiography and FDG PET/CT evaluate. Wound checks after reimplantation may be conducted in clinic or by video consultation. Remote monitoring of the new device continues as for any device recipient. Antibiotic prophylaxis is not recommended for device recipients undergoing dental, respiratory, gastrointestinal, genitourinary, or cardiac procedures; it is given for incision and drainage of infection at other sites and for replacement of an infected device. Lifetime lead strategy, including the choice of leadless and extravascular systems and the management of future abandoned leads, is documented in the record.

Sources of error in device infection and lead extraction

References

  1. Kusumoto FM, Schoenfeld MH, Wilkoff BL, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14:e503–e551. doi:10.1016/j.hrthm.2017.09.001
  2. Cha YM, El-Chami MF, Liu CF, et al. 2026 HRS/AHA/APHRS/EHRA/IDSA/LAHRS/PACES/STS expert consensus statement update on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2026;23:e1716–e1792. doi:10.1016/j.hrthm.2026.04.015
  3. Baddour LM, Esquer Garrigos Z, Rizwan Sohail M, et al. Update on cardiovascular implantable electronic device infections and their prevention, diagnosis, and management: a scientific statement from the American Heart Association. Circulation. 2024;149:e201–e216. doi:10.1161/CIR.0000000000001187
  4. Baddour LM, Epstein AE, Erickson CC, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010;121:458–477. doi:10.1161/CIRCULATIONAHA.109.192665
  5. Blomström-Lundqvist C, Traykov V, Erba PA, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections. Europace. 2020;22:515–549. doi:10.1093/europace/euz246
  6. Otto CM, Nishimura RA, Bonow RO, et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2021;143:e72–e227. doi:10.1161/CIR.0000000000000923
  7. Delgado V, Ajmone Marsan N, de Waha S, et al. 2023 ESC guidelines for the management of endocarditis. Eur Heart J. 2023;44:3948–4042. doi:10.1093/eurheartj/ehad193
  8. Fowler VG Jr, Durack DT, Selton-Suty C, et al. The 2023 Duke-International Society for Cardiovascular Infectious Diseases criteria for infective endocarditis: updating the modified Duke criteria. Clin Infect Dis. 2023;77:518–526. doi:10.1093/cid/ciad271
  9. Bongiorni MG, Burri H, Deharo JC, et al. 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS. Europace. 2018;20:1217. doi:10.1093/europace/euy050
  10. Indik JH, Gimbel JR, Abe H, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017;14:e97–e153. doi:10.1016/j.hrthm.2017.04.025
  11. Olsen T, Jørgensen OD, Nielsen JC, et al. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982–2018). Eur Heart J. 2019;40:1862–1869. doi:10.1093/eurheartj/ehz316
  12. Greenspon AJ, Patel JD, Lau E, et al. 16-year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States 1993 to 2008. J Am Coll Cardiol. 2011;58:1001–1006. doi:10.1016/j.jacc.2011.04.033
  13. Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015;17:767–777. doi:10.1093/europace/euv053
  14. Krahn AD, Longtin Y, Philippon F, et al. Prevention of Arrhythmia Device Infection Trial: the PADIT trial. J Am Coll Cardiol. 2018;72:3098–3109. doi:10.1016/j.jacc.2018.09.068
  15. Birnie DH, Wang J, Alings M, et al. Risk factors for infections involving cardiac implanted electronic devices. J Am Coll Cardiol. 2019;74:2845–2854. doi:10.1016/j.jacc.2019.09.060
  16. Malagù M, Donazzan L, Capanni A, et al. Risk scores for cardiac implantable electronic device infection: which one to believe in? J Clin Med. 2022;11:6556. doi:10.3390/jcm11216556
  17. Mittal S, Shaw RE, Michel K, et al. Cardiac implantable electronic device infections: incidence, risk factors, and the effect of the AigisRx antibacterial envelope. Heart Rhythm. 2014;11:595–601. doi:10.1016/j.hrthm.2013.12.013
  18. Tarakji KG, Mittal S, Kennergren C, et al. Antibacterial envelope to prevent cardiac implantable device infection. N Engl J Med. 2019;380:1895–1905. doi:10.1056/NEJMoa1901111
  19. de Oliveira JC, Martinelli M, Nishioka SA, et al. Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial. Circ Arrhythm Electrophysiol. 2009;2:29–34. doi:10.1161/CIRCEP.108.795906
  20. Birnie DH, Healey JS, Wells GA, et al. Pacemaker or defibrillator surgery without interruption of anticoagulation. N Engl J Med. 2013;368:2084–2093. doi:10.1056/NEJMoa1302946
  21. Essebag V, Verma A, Healey JS, et al. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION study. J Am Coll Cardiol. 2016;67:1300–1308. doi:10.1016/j.jacc.2016.01.009
  22. Birnie DH, Healey JS, Wells GA, et al. Continued vs. interrupted direct oral anticoagulants at the time of device surgery, in patients with moderate to high risk of arterial thrombo-embolic events (BRUISE CONTROL-2). Eur Heart J. 2018;39:3973–3979. doi:10.1093/eurheartj/ehy413
  23. Chamis AL, Peterson GE, Cabell CH, et al. Staphylococcus aureus bacteremia in patients with permanent pacemakers or implantable cardioverter-defibrillators. Circulation. 2001;104:1029–1033. doi:10.1161/hc3401.095097
  24. Uslan DZ, Sohail MR, Friedman PA, et al. Frequency of permanent pacemaker or implantable cardioverter-defibrillator infection in patients with gram-negative bacteremia. Clin Infect Dis. 2006;43:731–736. doi:10.1086/506942
  25. Axell-House DB, Khalil S, Sohail MR. Clinical approach to evaluation of underlying cardiac device infection in patients hospitalized with bacteremia. Methodist Debakey Cardiovasc J. 2023;19:48–57. doi:10.14797/mdcvj.1271
  26. Sohail MR, Uslan DZ, Khan AH, et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol. 2007;49:1851–1859. doi:10.1016/j.jacc.2007.01.072
  27. Toriello F, Saviano M, Faggiano A, et al. Cardiac implantable electronic devices infection assessment, diagnosis and management: a review of the literature. J Clin Med. 2022;11:5898. doi:10.3390/jcm11195898
  28. Le KY, Sohail MR, Friedman PA, et al. Impact of timing of device removal on mortality in patients with cardiovascular implantable electronic device infections. Heart Rhythm. 2011;8:1678–1685. doi:10.1016/j.hrthm.2011.05.015
  29. Pokorney SD, Zepel L, Greiner MA, et al. Lead extraction and mortality among patients with cardiac implanted electronic device infection. JAMA Cardiol. 2023;8:1165–1173. doi:10.1001/jamacardio.2023.3379
  30. Sciria CT, Kogan EV, Mandler AG, et al. Low utilization of lead extraction among patients with infective endocarditis and implanted cardiac electronic devices. J Am Coll Cardiol. 2023;81:1714–1725. doi:10.1016/j.jacc.2023.02.042
  31. Mandler AG, Sciria CT, Kogan EV, et al. Impact of hospital lead extraction volume on management of cardiac implantable electronic device-associated infective endocarditis. Europace. 2025;27:euae308. doi:10.1093/europace/euae308
  32. Pokorney SD, Mi X, Lewis RK, et al. Outcomes associated with extraction versus capping and abandoning pacing and defibrillator leads. Circulation. 2017;136:1387–1395. doi:10.1161/CIRCULATIONAHA.117.027636
  33. Boarescu PM, Roşian AN, Roşian ŞH. Transvenous lead extraction procedure: indications, methods, and complications. Biomedicines. 2022;10:2780. doi:10.3390/biomedicines10112780
  34. Wilkoff BL, Byrd CL, Love CJ, et al. Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial. J Am Coll Cardiol. 1999;33:1671–1676. doi:10.1016/S0735-1097(99)00074-1
  35. Wazni O, Epstein LM, Carrillo RG, et al. Lead extraction in the contemporary setting: the LExICon study: an observational retrospective study of consecutive laser lead extractions. J Am Coll Cardiol. 2010;55:579–586. doi:10.1016/j.jacc.2009.08.070
  36. Bongiorni MG, Kennergren C, Butter C, et al. The European Lead Extraction ConTRolled (ELECTRa) study: a European Heart Rhythm Association (EHRA) registry of transvenous lead extraction outcomes. Eur Heart J. 2017;38:2995–3005. doi:10.1093/eurheartj/ehx080
  37. Auricchio A, Regoli F, Conte G, Caputo ML. Key lessons from the ELECTRa registry in the modern era of transvenous lead extraction. Arrhythm Electrophysiol Rev. 2017;6:111–113. doi:10.15420/aer.2017.25.1
  38. Starck CT, Gonzalez E, Al-Razzo O, et al. Results of the Patient-Related Outcomes of Mechanical lead Extraction Techniques (PROMET) study: a multicentre retrospective study on advanced mechanical lead extraction techniques. Europace. 2020;22:1103–1110. doi:10.1093/europace/euaa103
  39. Azarrafiy R, Tsang DC, Wilkoff BL, Carrillo RG. Endovascular occlusion balloon for treatment of superior vena cava tears during transvenous lead extraction: a multiyear analysis and an update to best practice protocol. Circ Arrhythm Electrophysiol. 2019;12:e007266. doi:10.1161/CIRCEP.119.007266
  40. Tsang DC, Azarrafiy R, Pecha S, et al. Long-term outcomes of prophylactic placement of an endovascular balloon in the vena cava for high-risk transvenous lead extractions. Heart Rhythm. 2017;14:1833–1838. doi:10.1016/j.hrthm.2017.08.003
  41. Yogasundaram H, Mekary W, El-Chami MF, et al. Safety and outcomes of percutaneous vegetation aspiration during transvenous lead extraction: a multicenter experience. J Cardiovasc Electrophysiol. Published online October 15, 2026. doi:10.1111/jce.70530
  42. El-Chami MF, Bonner M, Holbrook R, et al. Leadless pacemakers reduce risk of device-related infection: review of the potential mechanisms. Heart Rhythm. 2020;17:1393–1397. doi:10.1016/j.hrthm.2020.03.019
  43. El-Chami MF, Johansen JB, Zaidi A, et al. Leadless pacemaker implant in patients with pre-existing infections: results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019;30:569–574. doi:10.1111/jce.13851
  44. Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management. Heart Rhythm. 2009;6:1085–1104. doi:10.1016/j.hrthm.2009.05.020
  45. Al-Khatib SM, Stevenson WG, Ackerman MJ, et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2018;138:e272–e391. doi:10.1161/CIR.0000000000000549
  46. Ahmed FZ, Blomström-Lundqvist C, Bloom H, et al. Use of healthcare claims to validate the Prevention of Arrhythmia Device Infection Trial cardiac implantable electronic device infection risk score. Europace. 2021;23:1446–1455. doi:10.1093/europace/euab028
  47. Rizwan Sohail M, Henrikson CA, Braid-Forbes MJ, Forbes KF, Lerner DJ. Increased long-term mortality in patients with cardiovascular implantable electronic device infections. Pacing Clin Electrophysiol. 2015;38:231–239. doi:10.1111/pace.12518