Library › Devices and mechanical support › Temporary cardiac pacing
Temporary cardiac pacing: transcutaneous, transvenous, epicardial, and active-fixation
Temporary cardiac pacing in adults by the transcutaneous, transvenous, epicardial, and active-fixation routes, from indications, capture, thresholds, and generator settings through complications, pacing after cardiac surgery and TAVR, and transition to a permanent device.
Temporary pacing supports the heart rate until a reversible cause resolves or a permanent device is implanted. Transcutaneous pacing is available within seconds but captures unreliably; capture is confirmed by a femoral pulse or an arterial waveform, never by the monitor alone. Transvenous pacing is reasonable for unstable bradycardia refractory to medical therapy, and pooled series report serious complications in about 10% of patients. An active-fixation lead connected to an externalized generator is the stable alternative when pacing must last days or weeks.
Routes of temporary cardiac pacing compared
Temporary pacing delivers electrical stimuli from an external generator to atrial or ventricular myocardium. Five routes are used: adhesive pads on the chest wall (transcutaneous); a catheter in the right ventricle (transvenous); wires sutured to the epicardium at surgery (epicardial); a permanent screw-in lead connected to an externalized generator (active-fixation temporary pacing, also called a temporary-permanent pacemaker); and, during structural procedures, a stiff guidewire in the left ventricle. Transesophageal atrial pacing is a sixth route with limited use. The routes differ in time to pacing, reliability of capture, patient comfort, mobility, and complication profile, and the choice follows the clinical setting and the expected duration of pacing.
| Route | Time to pacing | Reliability of capture | Expected duration | Main limitations |
|---|---|---|---|---|
| Transcutaneous | Seconds | Low and often intermittent; electrical capture does not guarantee mechanical capture | Minutes to hours, as a bridge | Pain from skeletal muscle stimulation; failure with obesity, hypoxemia, or high thresholds; skin injury with prolonged use |
| Transvenous, passive catheter | Minutes (about 20 to 45 minutes from decision in one prospective series) | Good when the tip contacts the right ventricular endocardium | Hours to a few days | Dislodgement, failure to capture or sense, perforation, infection, thrombosis, immobility |
| Active-fixation lead with externalized generator | Longer; requires fluoroscopy and a device operator | High; screw-in fixation and low, stable thresholds | Days to weeks | Off-label use of permanent hardware; accidental traction on the external system; skin-exit care |
| Epicardial wires | Immediate after cardiac surgery | Good at first; thresholds rise and electrogram amplitudes fall within days | Days after surgery | Loss of function over time; bleeding or tamponade at removal |
| Left ventricular guidewire | Immediate during transcatheter aortic valve replacement (TAVR) | High for rapid pacing during valve deployment | Procedure only | No backup after the guidewire is withdrawn |
| Transesophageal | Minutes | Atrial capture only; ventricular capture unreliable | Short | Requires intact atrioventricular conduction; discomfort |
Comparison of temporary pacing routes in adults. Time to transvenous pacing is from a prospective series of emergency insertions (median 22 minutes with echocardiographic guidance and 43 minutes with fluoroscopic guidance).
The 2018 American College of Cardiology (ACC)/American Heart Association (AHA)/Heart Rhythm Society (HRS) bradycardia guideline places temporary pacing after correction of reversible causes and after drug therapy has failed. For bradycardia from sinus node dysfunction or atrioventricular (AV) block with symptoms or hemodynamic compromise refractory to medical therapy, temporary transvenous pacing is reasonable (Class 2a), and transcutaneous pacing may be considered (Class 2b) until a transvenous or permanent pacemaker is placed. Temporary pacing, transvenous or transcutaneous, is not indicated for minimal or infrequent symptoms without hemodynamic compromise (Class 3: Harm). The 2021 European Society of Cardiology (ESC) pacing guideline recommends temporary transvenous pacing for hemodynamically compromising bradyarrhythmia refractory to intravenous chronotropic drugs (Class I). Its text advises that temporary transvenous pacing be avoided where possible and kept as short as possible because of its complications.
Capture, threshold, sensing, and impedance
Capture. Capture is depolarization of the paced chamber by the stimulus. Electrical capture is seen on the electrocardiogram (ECG) as a stimulus artifact followed by a wide QRS complex (ventricular pacing) or a P wave (atrial pacing) and its repolarization wave. Mechanical capture is the contraction that the electrical event produces, shown by a pulse, an arterial pressure waveform, a plethysmographic waveform timed to the paced beats, or ventricular wall motion on echocardiography. Electrical capture without mechanical capture is pulseless electrical activity, and an artifact without any depolarization is not capture at all.
Capture threshold. The capture threshold is the smallest stimulus that depolarizes the chamber consistently. Temporary external generators deliver current and are programmed in milliamperes (mA); permanent generators, including those used for active-fixation temporary pacing, are programmed in volts (V) at a stated pulse width. The threshold depends on the distance between electrode and excitable myocardium, the electrode surface area, and the state of the tissue. Electrode contact with the endocardium gives a low threshold; a free-floating catheter tip, edema or fibrosis at the interface, ischemic or infarcted myocardium, and a perforated tip outside the ventricle raise it. Hyperkalemia raises thresholds throughout the myocardium, and pacing often fails in severe hyperkalemia until the potassium is lowered; severe hypoxemia also predicts failure of transcutaneous pacing. Drugs contribute: in a crossover study of transcutaneous pacing, patients taking beta-blockers needed substantially more current.
Safety margin. Output is programmed above the threshold so that small rises do not cause loss of capture. A common convention for temporary transvenous and epicardial systems is an output at least twice the threshold, a twofold margin; higher outputs waste battery, stimulate the phrenic nerve or diaphragm, and accelerate fibrosis at the electrode interface. A threshold that approaches the maximum output of the generator leaves no safety margin, and the system is revised in a pacing-dependent patient.
Sensing. Sensing is detection of the intrinsic electrogram by the generator. The sensing threshold is the largest sensitivity value, in millivolts (mV), at which all intrinsic beats are still detected. Sensitivity works inversely: a lower millivolt setting makes the generator more sensitive. Sensitivity is usually programmed at half the sensing threshold, which gives a twofold margin for the small signals that occur with ectopy and for the fall in electrogram amplitude over days. Undersensing (a sensitivity value that is too high, or a small electrogram) causes competitive pacing and, in the ventricle, stimuli falling on the T wave. Oversensing (a sensitivity value that is too low) inhibits pacing in response to noise, T waves, myopotentials, or far-field signals.
Impedance. Impedance is the opposition to current flow in the lead and tissue. A sudden fall suggests insulation failure, and a sudden rise suggests conductor fracture or a loose connection. Temporary external generators do not report impedance; permanent generators used for active-fixation pacing do, which is one reason these systems are easier to monitor.
Indications for temporary pacing
In a scoping review of 32 consecutive-patient series (4,546 adults), AV block of any degree was the indication for temporary transvenous pacing in 62.7%, acute myocardial infarction without a specified rhythm in 11.4%, sinus node disease in 6.7%, other bradycardia in 4.9%, permanent pacemaker failure in 4.4%, prophylactic or periprocedural pacing in 2.7%, drug toxicity in 2.4%, overdrive suppression in 2.3%, and asystole or cardiac arrest in 1.1%. Acute myocardial infarction, alone or with AV block, accounted for 40.8% of cases, a share that has fallen with primary reperfusion.
| Setting | Temporary pacing role | Preferred route |
|---|---|---|
| Symptomatic or unstable bradycardia refractory to atropine and chronotropic drugs | Bridge to recovery or to permanent pacing; transvenous pacing reasonable (2018 ACC/AHA/HRS, Class 2a), transcutaneous pacing may be considered (Class 2b) | Transcutaneous at once, then transvenous |
| Infranodal AV block with a slow or unreliable wide escape rhythm | Atropine is usually ineffective; pacing is prepared without waiting for a drug response | Transvenous |
| AV block in acute myocardial infarction | Medically refractory symptomatic or hemodynamically significant bradycardia; infranodal high-grade block in anterior infarction | Transvenous; atrioventricular sequential pacing in right ventricular infarction |
| Drug toxicity (beta-blocker, calcium channel blocker, digoxin, antiarrhythmic) | Supports rate while antidotes act and the drug clears; contractility and vascular tone may remain the limiting factors | Transcutaneous, then transvenous |
| Hyperkalemia with bradycardia | Pacing often fails to capture until potassium falls; calcium and potassium-lowering therapy come first | Transvenous if bradycardia persists after treatment |
| Lyme carditis and other reversible inflammatory block | Bridge while antibiotics or immunosuppression restore conduction | Active-fixation lead with externalized generator when pacing is expected to last days |
| Device infection or bacteremia in a pacing-dependent patient | Bridge from extraction to reimplantation after the infection clears | Active-fixation lead, usually from the internal jugular vein |
| After cardiac surgery | Postoperative sinus node dysfunction, AV block, and overdrive pacing | Epicardial wires |
| During and after transcatheter aortic valve replacement | Rapid pacing for valve deployment; backup pacing for procedural AV block; bridge after the procedure | Left ventricular guidewire or right ventricular catheter; active-fixation lead for prolonged monitoring |
| Pause-dependent torsades de pointes and recurrent monomorphic ventricular tachycardia | Overdrive pacing at a rate that suppresses pauses or ectopy | Transvenous atrial or ventricular |
| Permanent pacemaker failure in a pacing-dependent patient | Rescue pacing while the system is revised | Transcutaneous, then transvenous |
Indications for temporary pacing in adults. Guideline language is from the 2018 ACC/AHA/HRS bradycardia guideline. Drug, potassium, and toxin management are covered in the next section.
Prophylactic transvenous pacing has a narrow role. It is used before procedures that carry a high risk of AV block in patients with preexisting conduction disease (alcohol septal ablation, some transcatheter valve procedures in patients with right bundle branch block) and before generator changes or lead revision in a pacing-dependent patient. Transcutaneous pads placed on standby, without pacing, serve the same purpose for many lower-risk procedures; in the original clinical trials of noninvasive pacing, 38 of 40 patients kept on standby for up to a month never needed stimulation.
Temporary pacing in bradycardia emergencies
The periarrest bradycardia algorithm, drug doses, and the classes of the AHA adult advanced life support guideline are set out in adult cardiac arrest and advanced life support. The evaluation of conduction disease, the drug table, and permanent pacing indications are set out in bradycardia and atrioventricular block. The emergency is defined by hypoperfusion (hypotension, altered mentation, shock, ischemic chest pain, or acute heart failure) and not by a heart rate number. Reversible causes are treated at the same time as the rate is supported. Atropine acts on vagally mediated sinus and nodal bradycardia and is often ineffective in infranodal block and after heart transplantation. Dopamine or epinephrine infusions, and isoproterenol when ischemia is unlikely, support the rate while pacing is prepared.
Pacing is the step for bradycardia that persists despite these measures, and for infranodal block in which drugs are unlikely to work. Transcutaneous pads are applied as soon as the rhythm is recognized as unstable, because pacing can start within seconds; transvenous pacing follows when transcutaneous capture is unreliable, the patient cannot tolerate it, or pacing will be needed for more than a short bridge. In a prehospital feasibility trial (PrePACE) of 82 patients with unstable bradycardia unresponsive to fluid and up to 3 mg of atropine, survival to discharge or 30 days was 69% with transcutaneous pacing and 70% with dopamine; a definitive trial would need about 690 patients per group, and none has been done.
Pacing in asystolic cardiac arrest. Transcutaneous pacing does not improve survival in asystole. In an emergency medical services trial, pacing started as early as possible in 112 of 278 patients with primary asystole; 4% were discharged alive, against 2% of 259 control patients, a difference that was not significant, and pacing after defibrillation-induced asystole was no more effective. Routine pacing in cardiac arrest is therefore not part of resuscitation.
Hyperkalemia. Bradycardia, junctional rhythm, and AV block are manifestations of hyperkalemia that are easily attributed to conduction disease. Atropine and electrical pacing are often ineffective until calcium stabilizes the membrane and potassium is lowered, and a temporary pacing catheter placed for unexplained bradycardia without a potassium measurement can fail to capture. The ECG changes of hyperkalemia are described in ECG in electrolyte disorders, drugs, and systemic disease.
Drug toxicity. In beta-blocker and calcium channel blocker poisoning, pacing treats bradycardia but not the vasodilation and myocardial depression that accompany it, and capture frequently fails or restores rhythm without restoring blood pressure. Calcium, high-dose insulin, glucagon, vasopressors, and venoarterial extracorporeal membrane oxygenation remain the mainstays, as covered in cardiovascular toxicology and poisoning. Digoxin toxicity is treated with digoxin-specific antibody fragments, and transvenous catheter manipulation in a digoxin-toxic ventricle can provoke ventricular arrhythmia. Digoxin toxicity management is covered in digoxin.
Lyme carditis and other reversible block. Lyme carditis produces fluctuating AV block, often at the level of the AV node, which usually resolves with intravenous antibiotics. Temporary pacing supports patients with syncope or an inadequate escape rhythm, and permanent pacing is deferred during the reversible phase. Because block can persist for days after antibiotics begin, one Canadian group places an active-fixation lead with an externalized generator from the outset in patients who need pacing. Of 21 patients diagnosed with Lyme carditis at that center over 5 years, 4 needed such a system for symptomatic bradycardia. Pacing lasted a mean of 10.5 days without procedure-related complications, and conduction recovered in all 4 without permanent pacing. Before the system is removed, 1:1 AV conduction is confirmed during exercise at a heart rate above 120 beats/min, at least 10 to 14 days after antibiotics begin. Immune checkpoint inhibitor myocarditis, giant cell myocarditis, and cardiac sarcoidosis can also present with AV block that needs a bridge while immunosuppression starts.
Heart transplant recipients. The denervated heart does not respond to atropine, and atropine can provoke paradoxical AV block. Beta-agonists and, when needed, temporary pacing through epicardial wires or a transvenous catheter support the rate while rejection and other causes are assessed.
Temporary pacing in acute myocardial infarction
The site of block determines the need for pacing in acute infarction. The degree of block alone does not. In inferior infarction, bradycardia and AV block arise in the AV node from vagal reflexes and nodal ischemia; the escape rhythm is junctional and usually adequate, atropine often restores conduction, and the block resolves within days of reperfusion. Pacing is needed only when bradycardia causes hypoperfusion despite atropine. The 2018 ACC/AHA/HRS guideline recommends temporary pacing in acute infarction for medically refractory symptomatic or hemodynamically significant bradycardia from sinus node dysfunction or AV block (Class 1). In anterior infarction, block is infranodal, results from septal necrosis, and is often preceded by new bundle branch or bifascicular block. The escape is wide and unreliable, mortality is high because the infarct is large, and transvenous pacing is placed for high-grade block. Reperfusion is the treatment that changes outcome; temporary pacing supports the rate. Indications and timing of permanent pacing after the infarct are covered in bradycardia and atrioventricular block, and arrhythmia management in ST-elevation myocardial infarction.
Right ventricular infarction. Right ventricular involvement accompanies nearly half of inferior ST-elevation infarctions. The ischemic right ventricle is stiff, dilated, and dependent on atrial contraction for filling. When the right atrial branches are spared, augmented right atrial contraction is an important compensation, and loss of AV synchrony from heart block or atrial ischemia worsens right ventricular filling and output. Single-chamber ventricular pacing restores rate without restoring the atrial contribution, and dual-chamber (atrioventricular sequential) temporary pacing, with an atrial catheter or screw-in atrial lead in addition to the ventricular catheter, is used when hypotension persists after rate correction and volume loading.
Anticoagulation and fibrinolysis. Many patients who need pacing during infarction have received antiplatelet drugs, heparin, or a fibrinolytic. A compressible site and ultrasound guidance reduce bleeding risk; the femoral vein is compressible, and the right internal jugular vein under ultrasound guidance is an alternative with a lower infection risk. Subclavian puncture, which cannot be compressed, is avoided after fibrinolysis.
Catheter-induced arrhythmia during infarction. Catheter manipulation in an ischemic ventricle provokes ventricular ectopy and ventricular tachycardia or fibrillation, and a defibrillator is connected during insertion. Undersensing by a catheter that sits in infarcted myocardium allows stimuli to fall in the vulnerable period of the T wave, which can trigger ventricular fibrillation in ischemic myocardium.
Transcutaneous pacing: equipment, pad position, and settings
Transcutaneous pacing uses large adhesive electrodes, usually the same multifunction pads used for defibrillation, connected to a monitor-defibrillator with a pacing module. The current passes through skin, chest wall muscle, and lung, so far more current is needed than with an endocardial electrode, and skeletal muscle is stimulated with all paced beats. The technique was made clinically practical by a pacemaker-monitor that used a long-duration pulse through large electrodes. In the clinical trials that introduced it, 134 patients were paced; stimulation was tolerated by 73 of 82 conscious patients, and electrocardiographic responses were obtained in 105. Failures occurred in patients with prolonged hypoxia or severe discomfort.
Pad position. In the anteroposterior position, the anterior (negative) pad is placed over the cardiac apex to the left of the lower sternum and the posterior (positive) pad on the left back between the scapula and the spine, so that the heart lies between them. In the anterolateral position, one pad sits below the right clavicle and the other over the left lower lateral chest, as for defibrillation. The anteroposterior position has traditionally been preferred for lower thresholds. In a crossover study of 16 patients with symptomatic bradycardia paced in both positions under sedation, median thresholds were 125 mA anteroposterior and 140 mA anterolateral, a difference that was not significant. The position that allows rapid, secure application without delay is therefore acceptable, and the alternative position is tried when capture fails. Pads are kept away from an implanted generator, on dry skin with good adhesion, and chest hair is clipped when adhesion is poor.
Settings. The generator is set to demand mode, which inhibits pacing when an intrinsic beat is sensed, unless monitor artifact causes oversensing, in which case asynchronous (fixed-rate) mode is used. The rate is set above the intrinsic rate. Output starts low and is increased until electrical capture appears, then raised by a small margin; in profound hypotension, output is started high and reduced after capture is confirmed. Capture typically requires tens to more than 100 mA: the median maximum current in 299 prehospital patients was 90 mA, and median thresholds in the crossover study exceeded 120 mA.
Electrical and mechanical capture during transcutaneous pacing
Electrical capture. True electrical capture is a broad QRS complex with a distinct T wave of opposite polarity after the stimulus, at the programmed rate. The large stimulus artifact and the electrical recovery of the chest wall after it can imitate a wide QRS complex, especially when the monitor filters and blanks the stimulus. An apparent QRS without a following T wave, a "QRS" that does not change morphology as output is increased, and an underlying rhythm that continues independently of the stimuli indicate false capture.
Mechanical capture. Mechanical capture is confirmed by a pulse that matches the paced rate, a pulsatile arterial line or plethysmographic waveform timed to paced beats, or ventricular contraction on point-of-care echocardiography. Palpation is unreliable during transcutaneous pacing because the chest wall and arm muscles twitch with the stimuli, which is why the femoral pulse, or better an arterial waveform, is used. In a prehospital cohort of 56 patients reviewed beat by beat, clinicians reported a palpable pulse 26 times during pacing, and only 4 of these reports (15%) coincided with electrical capture.
How often capture is achieved. Recent prehospital data show that transcutaneous capture is less reliable than once assumed. In the same 56-patient cohort, electrical capture occurred in 32% and, when present, was complete for a mean of only 27% of the pacing time; systolic pressure rose by a median of 22 mm Hg in patients with capture and did not change in those without it. In a four-agency cohort of 299 patients, sustained electrical capture occurred in 9.7%. Sustained capture was defined as a wide QRS complex with a T wave after at least 90% of stimuli over 60 seconds and after at least 90% for the rest of the attempt. Lower body weight, lower prepacing heart rate, and higher delivered current were associated with capture. In a national registry of 13,270 prehospital transcutaneous pacing attempts, 20.4% of patients who were not in arrest before pacing progressed to cardiac arrest, and body weight above 100 kg and an oxygen saturation below 80% before pacing were associated with failure.
| Study | Patients | Finding |
|---|---|---|
| Zoll, 1985 (multicenter clinical trials) | 134 | Electrocardiographic response in 105; tolerated by 73 of 82 conscious patients; failures with prolonged hypoxia or severe discomfort |
| Cummins, 1993 (asystolic arrest) | 278 vs 259 | Discharge 4% with early pacing vs 2% without; no significant difference |
| PrePACE, 2008 (randomized feasibility) | 82 | Survival 69% with pacing vs 70% with dopamine; 690 per group needed for a definitive trial |
| Smida, 2025 (national prehospital registry) | 13,270 | Progression to cardiac arrest after pacing in 20.4%; failure associated with weight above 100 kg and severe hypoxemia |
| Smida, 2026 (four agencies) | 299 | Sustained electrical capture in 9.7%; median maximum current 90 mA |
| Hershkowitz, 2026 (beat-by-beat review) | 56 | Electrical capture in 32%, complete for 27% of pacing time; 15% of reported pulses coincided with capture |
| Goldschmied, 2026 (crossover, pad position) | 16 | Median threshold 125 mA anteroposterior vs 140 mA anterolateral, not significant; higher thresholds with beta-blockers and lower myocardial mass |
Clinical studies of transcutaneous pacing in adults. Electrical capture definitions differ between studies.
Analgesia, sedation, and the limits of transcutaneous pacing
The stimuli contract the pectoral, intercostal, and back muscles, and conscious patients experience pain and anxiety that rise with output. Analgesia with a short-acting opioid and anxiolysis or sedation are given in titrated doses while airway, ventilation, and blood pressure are monitored; sedation does not delay pacing in a patient with profound hypotension or altered consciousness. A patient who pulls at pads during uncontrolled pain is at risk of losing pacing altogether.
Transcutaneous pacing is a bridge. Thresholds can rise during prolonged pacing, so capture is reassessed regularly. During long use, the skin under the pads is inspected and the pads are repositioned periodically because of the risk of skin injury. Pacing is less effective in obesity, emphysema, large pericardial or pleural effusions, and severe hypoxemia or acidosis. When transcutaneous pacing is required for more than a short period, or capture is unreliable, transvenous pacing is placed. Transcutaneous pacing does not interfere with transvenous catheter placement, and pads left in place provide backup during and after insertion.
Transvenous pacing: venous access
The choice of vein balances speed, operator experience, bleeding risk, infection risk, catheter stability, and the need to preserve the left subclavian and axillary veins for a later permanent device.
| Vein | Advantages | Disadvantages |
|---|---|---|
| Right internal jugular | Most direct course to the right ventricle; stable catheter position; ultrasound-guided puncture; preserves the left side for a permanent device | Arterial puncture and, rarely, pneumothorax without imaging; neck movement |
| Left internal jugular | Ultrasound-guided puncture | Tortuous course; more difficult to reach the right ventricle; may encounter a persistent left superior vena cava |
| Subclavian or axillary | Stable course to the right ventricle; patent in hypovolemia; comfortable | Pneumothorax; noncompressible bleeding; left side reserved for a permanent device |
| Femoral | Fast; compressible; no pneumothorax; distant from the airway during resuscitation | Least stable catheter position; harder to advance to the right ventricle and usually placed under fluoroscopy; immobility; higher infection and thrombosis risk |
| Brachial or antecubital | Compressible; used after fibrinolysis | Catheter moves with the arm; high dislodgement rate; phlebitis |
Venous access for temporary transvenous pacing.
Practice varies widely. In the scoping review, the femoral vein was used in 47.2% of procedures, the subclavian in 25.4%, the internal jugular in 12.0%, and arm veins in about 14%; in a Spanish series of 530 procedures, 99% were femoral. A 2026 meta-analysis of six observational studies (2,267 patients) found catheter-related infection markedly less frequent with jugular than femoral access (relative risk 0.25), with no significant difference in bleeding (relative risk 0.54), lead repositioning (0.73), or cardiac perforation (0.50). The authors favored jugular access when prolonged pacing or infection risk is expected and considered femoral access reasonable for short-term and emergency use.
Ultrasound-guided puncture of the internal jugular, axillary, or femoral vein improves first-pass success and reduces arterial puncture and pneumothorax, and it is the standard for central venous access unless resuscitation leaves no time. In the scoping review, complicated venous access (arterial puncture, pneumothorax, access-site bleeding, or nerve injury) occurred in 2.0% of patients. When a permanent pacemaker is likely, left-sided access is avoided so that the left subclavian and axillary veins remain free of thrombus and of infection from a recent temporary catheter.
Pacing catheters and methods of guidance
Catheters. Temporary transvenous pacing catheters are bipolar, with a distal (negative) electrode at the tip and a proximal (positive) ring electrode behind it, connected to the generator by cable. Balloon-tipped (flotation) catheters, usually 5 to 7 French, are inflated in the right atrium and carried by blood flow across the tricuspid valve; they can be placed without fluoroscopy and are used for bedside and emergency insertion. Semirigid catheters are steered under fluoroscopy and are favored by many operators for femoral access, where flow does not carry a balloon toward the right ventricle. The balloon is deflated before the tip is advanced to the endocardium, because an inflated balloon floats into the pulmonary artery.
Fluoroscopic guidance. Fluoroscopy remains the reference method. It shows the course of the catheter, the crossing of the tricuspid valve, and the tip position, and it detects catheters that enter the coronary sinus, the inferior vena cava, or the pulmonary artery. Its limitations are the need to move an unstable patient and the delay that causes.
Echocardiographic guidance. Transthoracic echocardiography shows the catheter as a bright linear echo in the right atrium and right ventricle. The subcostal four-chamber view is used because it does not interfere with the sterile field at the neck, with the apical four-chamber view as an alternative. The subcostal inferior vena cava view detects the most common misplacement, passage of the catheter from the superior into the inferior vena cava. Echocardiography also detects pericardial effusion, septal perforation, and misplacement into the pulmonary artery. In a prospective series of 203 emergency insertions, the median time from decision to pacing was 22 minutes with echocardiography-guided jugular placement and 43 minutes with fluoroscopy-guided femoral placement. Overall complications fell from 28.8% to 15.0%, mainly because infections (2.7% vs 11.1%) and puncture-site hematomas were less frequent. The groups were not randomized and differed in clinical setting. Transesophageal echocardiography, through midesophageal bicaval and right ventricular inflow-outflow views, guides insertion when transthoracic windows are inadequate in obesity, emphysema, or after sternotomy.
Intracavitary electrogram. When fluoroscopy is not available, the distal electrode of the catheter is connected through a cable with alligator clips to a precordial lead of a standard electrocardiograph, so that the V lead records from the catheter tip. The changing electrogram as the catheter advances from the superior vena cava through the atrium into the ventricle indicates its position, although the coronary sinus, inferior vena cava, and a perforated position can produce confusing patterns. Contact with the right ventricular endocardium produces marked ST-segment elevation, a current of injury, which identifies a tip in contact with myocardium and predicts low thresholds. From the right internal jugular vein, this pattern usually appears at a catheter depth of about 35 to 40 cm; advancing beyond 40 to 45 cm risks coiling and atypical positions. Ultrasound shows that the catheter is in the ventricle and the electrogram shows that it is in contact; combined, they reduce reliance on blind insertion. In one series, only 62% of catheters seen in the right ventricle by echocardiography achieved adequate capture.
Blind insertion. Insertion without imaging or an electrogram, advancing the catheter with the generator pacing until capture appears, has higher failure and complication rates and is limited to arrest situations in which no other method is available.
Catheter position and confirmation of placement
The traditional target is the right ventricular apex, where trabeculation stabilizes a passive tip. The apical septum and the right ventricular outflow tract are alternatives but are less stable for passive catheters. From the apex, ventricular pacing produces a left bundle branch block pattern with a superior frontal axis (negative QRS in II, III, and aVF); from the outflow tract it produces left bundle branch block with an inferior axis.
Right bundle branch block pattern of the paced QRS. A paced QRS that is predominantly positive in V1 indicates left ventricular activation first and calls for confirmation of the catheter position. The causes are perforation of the interventricular septum or the right ventricular free wall, placement in the coronary sinus or a cardiac vein (more likely from the left side or with a persistent left superior vena cava), entry into the left ventricle through a patent foramen ovale or atrial septal defect, and, uncommonly, a correctly placed right ventricular catheter in a patient whose paced QRS is positive in V1 for other reasons. Echocardiography, a lateral chest radiograph, and the clinical setting distinguish them: a catheter in the coronary sinus points posteriorly on the lateral radiograph, and a catheter tip beyond the cardiac silhouette or a new effusion indicates perforation.
Chest radiograph. A frontal and, when possible, lateral radiograph after insertion shows the catheter course and tip, excludes pneumothorax after jugular or subclavian puncture, and serves as a baseline for later comparison. On the frontal view an apical tip lies near the left heart border above the diaphragm; on the lateral view the tip points anteriorly. A catheter with redundant loops in the atrium or ventricle is more likely to migrate, and a catheter pulled taut across the tricuspid valve is more likely to dislodge with inspiration or patient movement.
Fixation and documentation. The catheter is secured at the skin and the sheath valve, and a sterile sleeve allows small adjustments without contamination. Catheter length at the skin, threshold, sensing threshold, and settings are documented after insertion and at subsequent checks, because a change in length or threshold is often the first sign of migration.
Temporary generator settings and pacing modes
Temporary external generators provide single-chamber or dual-chamber pacing with programmable rate, output in milliamperes per chamber, sensitivity in millivolts, and, in dual-chamber models, the AV interval, upper rate, and atrial refractory period. Mode notation follows the generic pacemaker code: the first letter is the paced chamber, the second the sensed chamber, and the third the response to sensing (I for inhibit, T for trigger, D for both, O for none). The code, mode trials, and permanent device programming are set out in bradycardia and atrioventricular block.
| Mode | Behavior | Use in temporary pacing | Hazard |
|---|---|---|---|
| VVI | Ventricular pacing, inhibited by sensed ventricular beats | Default backup for transvenous catheters and ventricular epicardial wires; atrial fibrillation with slow ventricular response | Loss of AV synchrony; pacemaker syndrome; ventriculoatrial conduction |
| AAI | Atrial pacing, inhibited by sensed atrial beats | Sinus node dysfunction with intact AV conduction after cardiac surgery; atrial overdrive pacing | No ventricular backup if AV block develops; lost with atrial fibrillation |
| DDD | Atrial and ventricular sensing and pacing, with atrial tracking | AV block after cardiac surgery; right ventricular infarction; low output in which AV synchrony matters | Tracking of atrial arrhythmias; crosstalk; pacemaker-mediated tachycardia |
| DVI | Atrial and ventricular pacing, ventricular sensing only | Epicardial pacing when the atrial electrogram is too small to sense reliably | Competitive atrial pacing; atrial arrhythmia |
| VOO, AOO, DOO (asynchronous) | Fixed-rate pacing without sensing | Electromagnetic interference during surgery; oversensing that inhibits pacing in a pacing-dependent patient | Competition with the intrinsic rhythm; ventricular stimulus on the T wave |
Pacing modes used with temporary generators.
Rate. The rate is set to restore perfusion: above the intrinsic rate for bradycardia support, higher when cardiac output is rate dependent after surgery or in a small, stiff ventricle, and higher still for overdrive suppression. A backup rate below the intrinsic rate keeps the generator in standby without competition.
Output and sensitivity. Output is set after the capture threshold is measured and sensitivity after the sensing threshold is measured, as described below. In a pacing-dependent patient with no measurable intrinsic rhythm, the sensing threshold cannot be measured, sensitivity is left at a standard value, and the rhythm strip is monitored for oversensing.
Asynchronous pacing. Asynchronous modes are used when electrocautery or other electromagnetic interference would inhibit a demand generator in a pacing-dependent patient. Asynchronous ventricular pacing competes with intrinsic beats, so it is used only for the duration of the interference and only when the intrinsic rate is below the pacing rate.
Threshold testing and the safety margin
Capture threshold. With the generator pacing at a rate above the intrinsic rate (often 10 beats/min above), output is reduced gradually until capture is lost; the lowest output that captures all beats is the threshold. In a pacing-dependent patient, output is reduced in small steps for only a few beats at a time, with transcutaneous pads in place, because loss of capture leaves no escape rhythm. Output is then programmed at least twice the threshold; for example, a ventricular threshold of 2.5 mA leads to an output of 5 mA. Thresholds are measured at least daily, after any change in clinical state, and whenever pacing behaves unexpectedly.
Sensing threshold. With the pacing rate reduced below the intrinsic rate (often about 20 beats/min below) and the output turned down, sensitivity is changed from the most sensitive value toward less sensitive values until the sense indicator no longer flashes with all intrinsic beats. The last value at which all beats are sensed is the sensing threshold, and sensitivity is programmed at half that value. The sensing threshold is not measured if lowering the pacing rate would leave the patient without a perfusing rhythm. After cardiac surgery, the pacing rate is not reduced below 30 beats/min during the check, because a patient without an intrinsic rhythm at that rate has sinus node dysfunction or high-grade AV block.
Transcutaneous thresholds. For transcutaneous pacing, the threshold is the lowest current with electrical capture confirmed by mechanical capture, and the output is set modestly above it. Because thresholds of 100 mA or more are common and the range of the device is limited, the margin is smaller than with an endocardial electrode.
Pacing output, sensitivity, and threshold trend calculator
The calculator applies the twofold output margin and the half-threshold sensitivity convention to measured values for a transvenous catheter or an epicardial wire. When the previous threshold is entered, it reports the change, and when the generator's maximum output is entered, it reports whether a twofold margin can be programmed.
Temporary pacing output and sensitivity
Twofold output and sensing marginsProgrammed output = 2 × capture threshold; programmed sensitivity = sensing threshold ÷ 2; threshold change = (current − previous) ÷ previous. The maximum output varies between generator models and between atrial and ventricular channels. Transcutaneous pacing thresholds are not entered, because the device range does not allow a twofold margin.
Troubleshooting temporary pacing
Malfunction falls into four electrical patterns: no stimulus when one is expected (failure to pace or output failure), stimulus without depolarization (failure to capture), pacing despite intrinsic beats (undersensing), and absent pacing caused by sensed signals that are not cardiac depolarizations (oversensing). These are diagnosed from the rhythm strip and the generator settings and corrected in a fixed sequence, with transcutaneous pads in place for a pacing-dependent patient. The table also lists diaphragmatic stimulation, an extracardiac problem.
| Pattern | ECG appearance | Causes | Correction |
|---|---|---|---|
| Failure to pace (output failure) | No stimulus artifact when the rate falls below the programmed rate | Generator off or battery depleted; loose or disconnected cable; conductor fracture; oversensing; rate set below intrinsic rate | Connections and battery checked; cable replaced; sensitivity value raised (less sensitive) or asynchronous mode used in a pacing-dependent patient |
| Failure to capture | Stimulus artifacts without following QRS or P waves | Catheter dislodgement or migration; perforation; rising threshold from edema, fibrosis, ischemia, hyperkalemia, acidosis, or drugs; output set too low | Output increased; threshold remeasured; catheter repositioned under imaging; electrolytes, pH, and drugs corrected; echocardiography for effusion |
| Undersensing | Stimuli at the programmed interval despite intrinsic beats; stimuli on the T wave | Sensitivity value too high; small or changing electrogram; catheter tip in infarcted or scarred myocardium; ectopic beats from another site | Sensitivity value lowered (more sensitive); catheter repositioned; asynchronous mode avoided |
| Oversensing | Pauses longer than the pacing interval without intrinsic beats | Sensitivity value too low; T-wave or far-field sensing; myopotentials; electrical noise from loose connections or other equipment | Sensitivity value raised (less sensitive); connections secured; asynchronous mode used in a pacing-dependent patient while the cause is corrected |
| Diaphragmatic or phrenic stimulation | Hiccups or abdominal twitching at the pacing rate | High output; catheter near the diaphragm or perforated; epicardial wire near the phrenic nerve | Output reduced if the margin allows; catheter repositioned; echocardiography for perforation |
Troubleshooting temporary transvenous and epicardial pacing.
Failure to capture in a pacing-dependent patient is treated first by maximizing output and starting transcutaneous pacing, and the cause is sought once a perfusing rhythm is restored. A change in paced QRS morphology, a rising threshold, and diaphragmatic stimulation together suggest perforation. Malfunction is distinguished from normal behavior: a ventricular stimulus that falls within an intrinsic QRS complex (pseudofusion) or that partly depolarizes the ventricle (fusion) is expected when the intrinsic and paced rates are close and does not indicate undersensing.
Complications of temporary transvenous pacing
Reported complication rates vary from below 1% to more than 80%, reflecting differences in definitions, patients, operators, and access. The most comprehensive synthesis, a scoping review of 32 series with 4,546 patients, found a weighted complication rate of 36.7%, of which 10.2% were serious (sepsis, cardiac perforation, major bleeding at the access site, pulmonary embolism, or procedure-related death). The rate fell from 31.2% in series published before 1980 to 22.9% in those published from 2010 to 2019. Malfunction dominates: failure to pace or sense, dislodgement, and reintervention accounted for 21.1% of patients.
| Complication | Scoping review (4,546 patients) | US National Inpatient Sample (360,223 patients) |
|---|---|---|
| Any complication | 36.7% | 3.9% (coded pericardial, pneumothorax, and bleeding events only) |
| Pacing failure (malsensing or malpacing) | 9.5% | Not coded |
| Lead dislodgement | 4.6% | Not coded |
| Reintervention | 5.1% | Not coded |
| Cardiac perforation or tamponade | 1.6% | 0.6% tamponade |
| Infection (local, phlebitis, sepsis, pericarditis) | 4.8% | Not coded |
| Ventricular tachycardia during insertion | 1.7% | Not coded |
| Arterial puncture | 0.5% | Not coded |
| Pneumothorax | 0.1% | 0.9% |
| Bleeding at the access site | 1.2% | 2.4% (nonpericardial bleeding) |
| Deep venous thrombosis or pulmonary embolism | 0.3% | Not coded |
| Death attributed to pacing | 0.2% | Not coded |
| In-hospital mortality (all causes) | 14.5% | 14.1% |
| Subsequent permanent pacemaker | 64.2% | 37.9% |
Complications of temporary transvenous pacing in the scoping review (weighted means of consecutive-patient series, 1967–2019) and in the National Inpatient Sample analysis of administrative data (2004–2014), which captured only coded events and therefore underestimates malfunction and infection.
The US administrative analysis estimated 10 to 12 temporary transvenous procedures per 100,000 population per year, stable over the decade. Female sex (odds ratio 1.33), in-hospital cardiac arrest (3.52), and treatment at a teaching hospital (1.91) were associated with tamponade. Prior coronary bypass surgery, which leaves pericardial adhesions, was associated with a lower risk (0.26). Tamponade was associated with fivefold higher odds of in-hospital death (odds ratio 5.00). In a Spanish series of 530 femoral procedures with a mean duration of 4.2 days, serious complications occurred in 22% of patients, and 9% needed repositioning for loss of capture or sensing.
Arrhythmia. Ventricular ectopy and nonsustained ventricular tachycardia occur as the catheter crosses the tricuspid valve and touches the right ventricle, and sustained ventricular tachycardia or fibrillation can follow, particularly in acute ischemia, hypokalemia, digoxin toxicity, and QT prolongation. A defibrillator is connected during insertion. Undersensing with stimuli on the T wave is the other mechanism.
Venous thrombosis and infection. Catheter-associated thrombosis is underdiagnosed and can involve the subclavian or jugular veins that a permanent device would use. Infection risk rises with duration, femoral access, and emergency insertion, and temporary pacing before permanent implantation is an independent risk factor for later device infection (described under infection and catheter duration).
Perforation and cardiac tamponade
The right ventricular apex is thin, and a semirigid catheter or a balloon-tipped catheter advanced with force can pass through the myocardium into the pericardium; septal perforation carries the tip into the left ventricle. Perforation may occur at insertion or days later as the catheter erodes through the wall with repeated contractions.
Signs are a rising threshold or loss of capture, a change in paced QRS morphology (often to a right bundle branch block pattern), diaphragmatic or chest wall stimulation, pleuritic chest pain, a pericardial rub, and hypotension. Echocardiography establishes the diagnosis by showing an effusion, the catheter tip outside the right ventricle, or the tip across the septum. Many perforations seal when the catheter is withdrawn into the ventricular cavity, but withdrawal can unmask bleeding, so it is done with pericardiocentesis and surgical support available. In a pacing-dependent patient, a new catheter is positioned before or as the old one is withdrawn. Tamponade is treated by pericardiocentesis, and persistent bleeding by surgical repair. Hemodynamic assessment and drainage are covered in pericardial effusion and cardiac tamponade.
Infection, catheter duration, and the risk to a future permanent device
Temporary passive catheters are designed for hours to a few days. Dislodgement, failure to sense or capture, infection, and thrombosis accumulate with time, and immobility adds delirium, pressure injury, pneumonia, and venous thromboembolism, especially in older patients. In the series of the scoping review that reported it, hospital stay averaged 11.2 days.
Temporary pacing also threatens the permanent device that often follows. In the prospective PEOPLE study of 6,319 patients receiving pacemakers or defibrillators, device infection occurred in 0.68% at 12 months. Temporary pacing before the implantation procedure was independently associated with infection (adjusted odds ratio 2.46), as were fever within 24 hours before implantation (5.83) and early reintervention (15.04). This association is one reason the ESC guideline recommends that temporary transvenous pacing be avoided or kept short. It is also one reason to place a permanent device without delay once the indication is established, or to use an active-fixation temporary system when the wait will be long.
Infection is reduced by chlorhexidine skin preparation (which lowers catheter-related bacteremia more than povidone-iodine or alcohol), full barrier precautions, ultrasound-guided jugular access, a closed sterile sleeve, daily inspection of the site, and removal of the catheter as soon as it is no longer needed.
Active-fixation temporary leads with an externalized generator
An active-fixation temporary pacing system uses a permanent bipolar screw-in lead, placed transvenously and fixed into the right ventricular myocardium, connected to a permanent pulse generator that lies on the skin and is secured with sutures, adhesive dressings, or a dedicated holder. The system is also called a temporary-permanent pacemaker or an externalized permanent pacemaker. The lead is placed under fluoroscopy, usually through the right internal jugular vein so that the left side remains free for a permanent device. The helix is extended into the septum or apex, and the lead is either tunneled a short distance or brought out through the skin without a tunnel and is connected to a resterilized or new generator. Removal is by unscrewing the helix and withdrawing the lead at the bedside.
The system addresses the principal weaknesses of the passive catheter. Fixation prevents dislodgement, thresholds are low and stable, the generator reports thresholds, sensing, and lead impedance and stores arrhythmia episodes, and patients can sit, walk, and in selected cases leave the hospital. Its weaknesses are three. The permanent lead and generator are used off-label, because neither is labeled for externalized temporary pacing. The system requires an operator skilled in permanent lead placement and fluoroscopy. The external generator can also be pulled, with loss of pacing or lead dislodgement. Dedicated temporary leads with a helix or distal stabilizers, used with a temporary external generator, are cleared by the US Food and Drug Administration for temporary pacing of no more than 7 days. Many bridges, such as those after device extraction or TAVR, last longer.
| Study | Setting and design | Patients | Duration | Principal results |
|---|---|---|---|---|
| Braun, 2006 | Systemic infection with bradycardia; prospective, controlled | 23 active vs 26 passive | About 8 days in both groups | Pacing-related adverse events 1 vs 24; median threshold 0.6 V vs 1.0 V |
| Kawata, 2013 | Pacing-dependent after extraction of an infected device | 23 | Mean 19.4 days (range 3–45) | No loss of capture or sensing failure; 12 discharged home or to a nursing facility with the system |
| Beneyto, 2022 | Critical cardiac care; retrospective | 46 | Median 9 days | At least one complication in 22%; 56.5% received a permanent device and 15 had the system removed after recovery |
| Parekh, 2024 | TAVR; before the procedure for high-risk conduction disease or after procedural block | 89 | Median 6 days | One lead dislodgement (1.1%); no infection, perforation, or pneumothorax; 44.9% avoided a permanent pacemaker; median stay 3 days |
| Chang, 2024 | TAVR with high-grade or complete block or new LBBB with first-degree AV block; prospective, 13 centers | 70 | 1-month bridge | 75.7% free of a pacing indication at 1 month; 1-month pacemaker rate 2.47% of all TAVR patients vs 8.28% at 48 hours |
| TEMPACE, 2026 | Mixed indications with a dedicated external holder; prospective | 30 | Mean 7.8 days (maximum 22) | One complication (3%) from traction during agitation; 79.2% of pacing time spent on the cardiology ward |
| Fischer, 2026 | TAVR in patients with baseline RBBB; screw-in vs conventional temporary lead; multicenter observational | 124 vs 420 | 2.4 vs 1.2 days | ICU use 23% vs 99%; dislodgement 0.8% vs 3.3%; hospital stay 5.2 vs 7.4 days; 30-day permanent pacemaker 50.0% vs 46.2% |
Studies of active-fixation temporary pacing with an externalized generator. Thresholds in volts reflect the use of permanent generators. TAVR, transcatheter aortic valve replacement; LBBB, left bundle branch block; RBBB, right bundle branch block; ICU, intensive care unit.
A 2025 synthesis of prospective series found 0.35 complications per procedure with passive temporary catheters (670 patients, mean 3.7 days) and 0.03 with active-fixation systems (186 patients, mean 11.8 days), despite the threefold longer duration. The comparison is between series and did not come from a randomized trial. The active-fixation series came from experienced centers. For patients who need prolonged temporary pacing, the 2018 ACC/AHA/HRS guideline considers an externalized active-fixation permanent lead reasonable in preference to a passive-fixation temporary lead (Class 2a). The 2021 ESC pacing guideline gives the same approach a Class IIa recommendation (level of evidence C).
Indications for active-fixation temporary pacing. Active-fixation temporary pacing is used when pacing is expected to last more than a few days or when a passive catheter would confine the patient to intensive care: pacing dependence after extraction of an infected device until bloodstream infection clears, a permanent pacing indication during active infection, reversible block in Lyme carditis or myocarditis that may need a week or more to recover, AV block after TAVR or alcohol septal ablation during an observation period, and pacing dependence from other causes that requires a long bridge. The decision rests on expected duration, the availability of an implanting operator, and the patient's ability to protect the external system.
Care of the active-fixation system. The generator is programmed like a permanent device, usually VVI with a lower rate and output set by threshold testing in volts. The exit site is covered with an antiseptic dressing, the lead and generator are fixed so that traction does not reach the helix, and thresholds, sensing, and impedance are checked through the programmer at least daily in hospital and at all outpatient visits. Patients discharged with the system are selected for reliability and proximity to the hospital and are instructed to protect the generator during dressing changes and sleep.
Epicardial pacing after cardiac surgery
Temporary epicardial wires are placed at the end of cardiac surgery, before chest closure. Ventricular wires are sutured to the diaphragmatic surface of the right ventricle and atrial wires to the right atrium; by convention, atrial wires exit the skin to the right of the sternum and ventricular wires to the left. Wires are unipolar, with a separate skin electrode as the anode, or bipolar, with both electrodes on the heart; bipolar systems have lower capture thresholds, better sensing, and less susceptibility to interference, and they suit dual-chamber pacing. Wires held against the heart without sutures reduce bleeding at removal but have less reliable contact and higher thresholds.
Patients who need epicardial wires. Postoperative bradyarrhythmia is most likely after valve surgery, particularly aortic and tricuspid procedures, multiple valve operations, surgery for atrial fibrillation, and transplantation. Risk is also higher in older patients with reduced ejection fraction, diabetes, elevated pulmonary artery pressure, preoperative arrhythmia, preoperative beta-blocker or digoxin therapy, long cross-clamp time, and pacing needed to separate from cardiopulmonary bypass. After isolated coronary bypass surgery, few patients need pacing: in a prospective study of 222 patients, 8.6% required postoperative pacing, and only 2.6% of patients without diabetes, preoperative arrhythmia, or pacing at separation from bypass did so. These figures support selective use of wires after coronary surgery. The guideline recommendations for routine wires after valve and atrial fibrillation surgery, and for permanent epicardial leads during tricuspid surgery, are summarized in bradycardia and atrioventricular block.
Function over time. Epicardial wire function deteriorates daily as edema, inflammation, and fibrosis develop at the electrode interface. In a prospective study of 60 patients followed for about 5 days, atrial and ventricular capture thresholds rose significantly by the fourth postoperative day, and P-wave and R-wave amplitudes fell significantly by the second day. P-wave amplitudes were consistently less than half of R-wave amplitudes. Atrial wires fail first. Thresholds and sensing are therefore checked at least daily, and a patient who depends on epicardial pacing for more than a few days is evaluated for a transvenous or permanent system before the wires fail.
Modes after surgery. VVI is the safest backup mode when the wires are working and the patient is not pacing-dependent. AAI preserves the native conduction sequence and the atrial contribution in sinus node dysfunction with intact AV conduction but provides no ventricular backup if AV block or atrial fibrillation develops. DDD maintains AV synchrony in AV block and is preferred when cardiac output depends on the atrial contribution, as after surgery for hypertrophied or stiff ventricles. DVI is used when the small atrial electrogram cannot be sensed reliably. Asynchronous modes are reserved for electromagnetic interference in a pacing-dependent patient.
Diagnostic and therapeutic uses. Atrial epicardial wires record a large atrial electrogram that clarifies the diagnosis of narrow-complex tachycardias and of wide-complex rhythms with hidden P waves. Rapid atrial pacing through the atrial wires, at a rate above the flutter rate, can terminate typical atrial flutter without cardioversion; it does not terminate atrial fibrillation. Ventricular overdrive pacing through the ventricular wires can suppress pause-dependent ventricular arrhythmias. A ventricular stimulus without sensing can fall on the T wave, so asynchronous ventricular pacing after surgery is avoided unless needed.
Removal of epicardial wires and the transition to permanent pacing after surgery
Wires are removed when the patient no longer needs pacing and the risk of significant conduction disease is low, usually before discharge. Removal is a sterile bedside procedure: sutures are cut and the wire is withdrawn with gentle, steady traction. Resistance suggests that the wire is caught by a suture or fixed to the heart; it is then cut at the skin under tension so that the end retracts beneath the surface. Retained wire fragments are usually harmless but have been reported to migrate.
The serious complications of wire removal are bleeding from the epicardial puncture site, tamponade, and, rarely, injury to a coronary graft or ventricular arrhythmia, so patients are observed with blood pressure and rhythm monitoring after removal. Anticoagulation is managed around removal. Warfarin is commonly continued, with removal accepted at an international normalized ratio up to about 3.0. In patients at high thrombotic risk, such as those with a mechanical mitral valve, the anticoagulant is kept at the lower end of the therapeutic range. Direct oral anticoagulants are started after wire removal. Thrombocytopenia is allowed to recover to a platelet count of at least 100 × 10⁹/L. Postoperative effusion and late tamponade are covered in pericardial effusion and cardiac tamponade.
Timing of permanent pacing. Most postoperative conduction disturbances improve as edema resolves, so the decision on permanent pacing follows a waiting period. Transition is reasonable at about 4 to 5 days, and many centers wait 7 to 14 days when the patient is stable on temporary pacing. Persistent high-grade AV block and symptomatic sinus node dysfunction are the usual indications. Because epicardial wire function deteriorates, a patient who remains pacing-dependent during the waiting period may need a transvenous catheter or an active-fixation system to bridge the remaining days.
Temporary pacing during transcatheter aortic valve replacement
Rapid ventricular pacing reduces stroke volume and transvalvular flow during balloon valvuloplasty, deployment of balloon-expandable valves, and postdilation, preventing valve migration. Temporary pacing also provides backup when the procedure causes high-grade AV block. The conventional method is a temporary catheter in the right ventricular apex through the jugular or femoral vein, which adds venous access and the risk of right ventricular perforation and tamponade.
An alternative uses the stiff guidewire already in the left ventricle: the cathode of an external generator is attached to the wire with an alligator clip, the delivery system or balloon catheter insulates the shaft, and the anode is attached to the skin, a subcutaneous needle near the arterial sheath, or a venous wire. In the randomized EASY TAVI trial of 303 transfemoral TAVR procedures with a balloon-expandable valve, left ventricular guidewire pacing shortened mean procedure duration (48.4 vs 55.6 minutes) and fluoroscopy time and lowered cost. Effective stimulation (84.9% vs 87.1%), procedural success, and 30-day major adverse events were similar. Dedicated insulated pacing guidewires have since been cleared by the US Food and Drug Administration for rapid left ventricular pacing during TAVR. Guidewire pacing ends when the wire is removed, so patients at high risk of AV block keep a right ventricular catheter or receive one when block appears.
Conduction block after TAVR. The mechanism, incidence, predictors, the ACC expert consensus pathway, permanent pacing indications, and the risk of delayed block are set out in conduction disturbances after TAVR. For temporary pacing, the pathway removes the temporary catheter in the procedure room when no new conduction disturbance appears, keeps a secure temporary lead through internal jugular access for new left bundle branch block, PR or QRS prolongation, or transient complete heart block, and maintains pacing until a permanent device is implanted for persistent complete heart block. Preexisting right bundle branch block identifies the patients most likely to need prolonged temporary pacing.
Active-fixation leads after TAVR. Conduction disturbances after TAVR often recover, and pressure to discharge early can lead to permanent pacemakers that are later rarely used. An active-fixation temporary system extends the observation window without confining the patient to intensive care. In the prospective Chinese multicenter study, using the system as a 1-month bridge left 75.7% of 70 patients with high-grade block or new left bundle branch block with first-degree AV block free of a pacing indication at 1 month. In the Iowa experience, 44.9% of 89 patients who received the system before or after TAVR ultimately did not need a permanent pacemaker, with one dislodgement and a median stay of 3 days. In patients with baseline right bundle branch block, screw-in temporary leads reduced intensive care use from 99% to 23% and shortened hospital stay without more complications. These are observational data. The RECOVER trial is designed to randomize 160 patients with conduction block after TAVR to a 1-month active-fixation bridge or conventional temporary pacing for 24 to 48 hours, with permanent pacemaker implantation at 6 months as the primary end point.
Temporary pacing after transcatheter tricuspid valve replacement and other structural procedures
Transcatheter tricuspid valve replacement complicates temporary pacing. In TRISCEND II, new permanent pacemakers were implanted in 17.4% of the replacement group and 2.3% of controls (Hahn 2025). A review of bradycardia after transcatheter replacement estimates that AV block requiring a new pacemaker affects about one in four recipients of the EVOQUE valve (Hain 2025) (conduction block after transcatheter tricuspid replacement). The valve frame narrows the orifice and deflects catheters toward the atrial walls. A balloon-tipped catheter placed without fluoroscopy may fail to cross the prosthesis, and leads that already cross the tricuspid valve are trapped by the frame. Backup pacing during implantation uses the stiff wire that stabilizes the delivery system in the right ventricle or a second wire in the left ventricle. The permanent pacing strategy for patients who develop block (leadless device, coronary sinus lead, epicardial lead, or a lead placed before the valve) is planned before the procedure, with fluoroscopy available for any urgent temporary pacing.
After alcohol septal ablation for hypertrophic cardiomyopathy, AV block is common in the first days, and a temporary catheter or active-fixation lead is kept in place during monitoring; septal reduction and its conduction outcomes are covered in hypertrophic cardiomyopathy. During percutaneous coronary intervention, transient AV block can be managed with a temporary right ventricular catheter or with pacing through the coronary guidewire.
Overdrive pacing for torsades de pointes and ventricular tachycardia
Pacing suppresses ventricular arrhythmias by removing the pauses and short-long-short sequences that trigger torsades de pointes, by shortening and homogenizing repolarization, and by suppressing ectopic foci. In pause-dependent torsades from drug-induced or acquired QT prolongation, temporary atrial or ventricular pacing at a rate that abolishes pauses prevents recurrence until the drug clears and electrolytes are corrected, with intravenous magnesium and isoproterenol as bridges. The sequence and doses are covered in long QT syndrome and QT prolongation.
In recurrent monomorphic ventricular tachycardia and electrical storm, overdrive pacing at a rate above the native rate suppresses some episodes, and burst pacing through a temporary catheter can terminate tachycardia; acceleration to faster tachycardia or fibrillation is a known risk, and a defibrillator is connected. Overdrive pacing in electrical storm, through an implanted device or a temporary lead, is described in electrical storm. Overdrive suppression accounted for 2.3% of temporary transvenous pacing in the scoping review.
Alternative temporary routes: transesophageal, coronary guidewire, and coronary sinus pacing
Transesophageal pacing. An electrode in the esophagus behind the left atrium captures the atrium, but ventricular capture is unreliable. The route is limited to atrial pacing in sinus bradycardia with intact AV conduction and to overdrive termination of atrial flutter, and its use in adults is now uncommon.
Coronary guidewire pacing. During percutaneous coronary intervention, unipolar pacing through the coronary guidewire, insulated by the balloon or microcatheter, with the anode on the skin, provides backup pacing for transient AV block without venous access. The principle is the same as left ventricular guidewire pacing during TAVR.
Coronary sinus and atrial pacing. A catheter in the coronary sinus paces the left atrium and, from distal veins, the left ventricle; it is an option when the tricuspid valve cannot be crossed, as after mechanical tricuspid replacement or transcatheter tricuspid replacement. Temporary atrial pacing catheters, J-shaped for the right atrial appendage, provide AV sequential pacing with a ventricular catheter, but passive atrial catheters are unstable, and an active-fixation atrial lead is used when atrial pacing must be reliable.
Monitoring, nursing care, and transport of the paced patient
Patients with temporary transvenous pacing are monitored continuously with an ECG that displays stimulus artifacts, because pacing malfunction can be fatal in a pacing-dependent patient. Transcutaneous pads remain in place and connected for immediate backup. Generator settings, battery status, connections, and catheter length at the skin are recorded on all shifts, thresholds are measured at least daily, and the access site is inspected for bleeding, hematoma, and infection. Exposed terminals at the connector are insulated, because the catheter conducts directly to the endocardium.
Patients with a femoral catheter remain in bed with the leg straight, which adds the risks of immobility. Patients with jugular or subclavian catheters can sit up but avoid sudden movements of the head and arm. Before transport within the hospital or between hospitals, the catheter is secured, the generator is fixed to the patient or bed, a spare battery is available, thresholds and sensing are confirmed, and transcutaneous pads and a defibrillator travel with the patient. Electrocautery during surgery or procedures can inhibit a demand generator, so pacing-dependent patients are switched to an asynchronous mode, or the pacing system is supervised by someone able to change settings, for the duration of the interference. Magnetic resonance imaging is avoided with conventional temporary catheters and epicardial wires connected to a generator.
Active-fixation systems with permanent generators allow care outside the intensive care unit. Patients are managed on a telemetry ward, mobilize, and, in selected cases, are discharged with weekly clinic review, dressing changes, and device interrogation.
Transition from temporary to permanent pacing
Most patients who need temporary transvenous pacing go on to need a permanent device: 64.2% in the scoping review, 69.6% in the Spanish series, and 37.9% during the index admission in US administrative data. The decision follows the permanent pacing indications in the 2018 ACC/AHA/HRS guideline, applied after reversible causes are excluded: drug toxicity treated and the drug cleared, potassium corrected, ischemia treated, infection controlled, and a waiting period observed after infarction, cardiac surgery, and TAVR. The indications for permanent pacing in sinus node dysfunction and AV block are set out in bradycardia and atrioventricular block, and bundle branch block indications in bundle branch and fascicular blocks.
The timing balances two risks. Early implantation in a patient whose block would have recovered commits the patient to a lifelong device, and conduction often recovers in patients who receive pacemakers after TAVR. Delayed implantation prolongs exposure to the complications of temporary pacing, and temporary pacing before implantation increases the risk of device infection. When the indication is clear and reversible causes are excluded, the permanent device is implanted promptly and the temporary system removed. When recovery is uncertain, an active-fixation temporary system allows a longer observation period with fewer complications. In a patient with an infected device, reimplantation waits for negative blood cultures and, often, a leadless device or a contralateral system is chosen.
Sources of error in temporary pacing
- Electrical capture assumed from the monitor during transcutaneous pacing: stimulus artifact and chest wall recovery imitate a wide QRS, and muscle twitching imitates a pulse. Capture is confirmed by a QRS with a T wave at the paced rate and by an arterial or plethysmographic waveform, a femoral pulse, or echocardiography.
- Waiting for atropine in infranodal block: a wide escape rhythm, Mobitz II, or complete block in anterior infarction does not respond, and pacing is prepared at once.
- Pacing before measuring potassium: hyperkalemic bradycardia often fails to capture, and calcium and potassium-lowering therapy come first.
- Electrical capture accepted as success in drug toxicity: a paced rhythm without an adequate blood pressure indicates myocardial depression or vasodilation that pacing does not treat.
- A catheter seen in the right ventricle on echocardiography assumed to be in contact: a free-floating tip gives a high threshold and unstable capture; the intracavitary current of injury or a low threshold confirms contact.
- A right bundle branch block paced pattern accepted without investigation: septal or free wall perforation and coronary sinus placement are excluded with echocardiography and a lateral radiograph.
- Output set without measuring the threshold, or the threshold measured only once: thresholds rise with edema, fibrosis, migration, and metabolic change, and a fixed high output masks the rise until the margin is gone.
- Sensitivity confused in direction: raising the millivolt value makes the generator less sensitive and can produce undersensing and stimuli on the T wave; lowering it can produce oversensing and pauses.
- Asynchronous ventricular pacing left in place after electrocautery: competition with the intrinsic rhythm can deliver a stimulus in the vulnerable period.
- Left subclavian access in a patient likely to need a permanent pacemaker: the vein needed for the device is exposed to thrombosis and contamination.
- Single-chamber ventricular pacing in right ventricular infarction with persistent hypotension: loss of the atrial contribution limits right ventricular filling, and AV sequential pacing restores it.
- A passive catheter kept for a week or more: dislodgement, infection, and immobility accumulate, and an active-fixation system or a timely permanent device is the alternative.
- Epicardial wires relied on beyond the first days after surgery: atrial and ventricular thresholds rise and electrograms shrink, and a pacing-dependent patient is left without reliable backup.
- Epicardial wires removed without monitoring or with anticoagulation unmanaged: bleeding and tamponade can follow removal.
References
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2019;140:e382–e482. Correction: Circulation. 2019;140:e506–e508. doi:10.1161/CIR.0000000000000628
- Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021;42:3427–3520. doi:10.1093/eurheartj/ehab364
- Fonseca M, Cheng E, Do D, et al. Bradyarrhythmias in cardio-oncology. South Asian J Cancer. 2021;10:195–210. doi:10.1055/s-0041-1731907
- Tjong FVY, de Ruijter UW, Beurskens NEG, Knops RE. A comprehensive scoping review on transvenous temporary pacing therapy. Neth Heart J. 2019;27:462–473. doi:10.1007/s12471-019-01307-x
- Metkus TS, Schulman SP, Marine JE, Eid SM. Complications and outcomes of temporary transvenous pacing: an analysis of >360,000 patients from the National Inpatient Sample. Chest. 2019;155:749–757. doi:10.1016/j.chest.2018.11.026
- López Ayerbe J, Villuendas Sabaté R, García García C, et al. Temporary pacemakers: current use and complications [in Spanish]. Rev Esp Cardiol. 2004;57:1045–1052. doi:10.1016/s0300-8932(04)77240-1
- Pereira RA, Barcellos G, Lenz G, et al. Complications of temporary transvenous cardiac pacing by access site: a systematic review and meta-analysis. Pacing Clin Electrophysiol. 2026;49:1100–1107. doi:10.1111/pace.70295
- Ferri LA, Farina A, Lenatti L, et al. Emergent transvenous cardiac pacing using ultrasound guidance: a prospective study versus the standard fluoroscopy-guided procedure. Eur Heart J Acute Cardiovasc Care. 2016;5:125–129. doi:10.1177/2048872615572598
- Blanco P. Temporary transvenous pacing guided by the combined use of ultrasound and intracavitary electrocardiography: a feasible and safe technique. Ultrasound J. 2019;11:8. doi:10.1186/s13089-019-0122-y
- Adi O, Fong CP, Tze MKH, et al. Transesophageal echocardiography (TEE)-guided transvenous pacing (TVP) in emergency department. Ultrasound J. 2023;15:35. doi:10.1186/s13089-023-00332-7
- Klug D, Balde M, Pavin D, et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation. 2007;116:1349–1355. doi:10.1161/CIRCULATIONAHA.106.678664
- Braun MU, Rauwolf T, Bock M, et al. Percutaneous lead implantation connected to an external device in stimulation-dependent patients with systemic infection: a prospective and controlled study. Pacing Clin Electrophysiol. 2006;29:875–879. doi:10.1111/j.1540-8159.2006.00454.x
- Kawata H, Pretorius V, Phan H, et al. Utility and safety of temporary pacing using active fixation leads and externalized re-usable permanent pacemakers after lead extraction. Europace. 2013;15:1287–1291. doi:10.1093/europace/eut045
- Beneyto M, Seguret M, Taranzano M, et al. Externalized reusable permanent pacemaker for prolonged temporary cardiac pacing in critical cardiac care units: an observational monocentric retrospective study. J Clin Med. 2022;11:7206. doi:10.3390/jcm11237206
- Vicente-Miralles R, Rivas-Gándara N, Jordán Marchite P, et al. A novel device for temporary cardiac stimulation using permanent pacemaker generators and active-fixation leads: the TEMPACE multicentre prospective study. Eur Heart J Acute Cardiovasc Care. 2026;14:732–738. doi:10.1093/ehjacc/zuaf111
- Vicente-Miralles R, Rivas-Gándara N, Antón Pascual JL, et al. Temporary pacing with active-fixation leads: clinical and economic impact versus conventional temporary transvenous pacing. J Clin Med. 2025;15:125. doi:10.3390/jcm15010125
- Parekh J, Sharma V, Robl J, et al. Temporary-permanent pacemakers in the management of conduction abnormalities in patients undergoing transcatheter aortic valve replacement. J Soc Cardiovasc Angiogr Interv. 2024;3:101310. doi:10.1016/j.jscai.2024.101310
- Chang S, Jiang Z, Liu X, et al. Permanent pacemaker reduction using temporary-permanent pacemaker as a 1-month bridge after transcatheter aortic valve replacement: a prospective, multicentre, single-arm, observational study. EClinicalMedicine. 2024;72:102603. doi:10.1016/j.eclinm.2024.102603
- Jiang Z, Chang S, Tao L, et al. The randomized controlled trial to compare temporary permanent pacemaker vs temporary pacemaker in patients with conduction block after transcatheter aortic valve replacement: rationale and design of the RECOVER trial. Am Heart J. 2025;287:41–49. doi:10.1016/j.ahj.2025.04.017
- Fischer Q, Urena M, Asmarats L, et al. Temporary active fixation leads in transcatheter aortic valve replacement patients with baseline right bundle branch block. Struct Heart. 2026;10:101119. doi:10.1016/j.shj.2026.101119
- Wang CN, Baranchuk A. An ambulatory approach to atrioventricular block in Lyme carditis using temporary permanent pacemakers [letter]. J Innov Card Rhythm Manag. 2023;14:5369–5371. doi:10.19102/icrm.2023.14033
- Zoll PM, Zoll RH, Falk RH, et al. External noninvasive temporary cardiac pacing: clinical trials. Circulation. 1985;71:937–944. doi:10.1161/01.cir.71.5.937
- Cummins RO, Graves JR, Larsen MP, et al. Out-of-hospital transcutaneous pacing by emergency medical technicians in patients with asystolic cardiac arrest. N Engl J Med. 1993;328:1377–1382. doi:10.1056/NEJM199305133281903
- Morrison LJ, Long J, Vermeulen M, et al. A randomized controlled feasibility trial comparing safety and effectiveness of prehospital pacing versus conventional treatment: 'PrePACE'. Resuscitation. 2008;76:341–349. doi:10.1016/j.resuscitation.2007.08.008
- Smida T, Voges L, Crowe R, et al. Prehospital transcutaneous cardiac pacing in the United States: treatment epidemiology, predictors of treatment failure, and associated outcomes. Prehosp Emerg Care. 2025;29:578–585. doi:10.1080/10903127.2024.2393768
- Smida T, Kimbrell JM, Kreinbrook JA, et al. The prevalence of sustained electrical capture during prehospital transcutaneous pacing: a multicenter observational study. Resuscitation. 2026;219:110934. doi:10.1016/j.resuscitation.2025.110934
- Hershkowitz MN, Kniest LJ, Davis-Allen P, et al. Transcutaneous pacing in the prehospital setting: a retrospective cohort study of electrical capture. Prehosp Emerg Care. 2026:1–16. doi:10.1080/10903127.2026.2756599
- Goldschmied A, Sigle M, Toskas I, et al. Anteroposterior versus anterolateral pacer pad position in patients with symptomatic bradycardia. Int J Cardiol Heart Vasc. 2026;62:101857. doi:10.1016/j.ijcha.2025.101857
- Mattu A, Hayes BD, Martinez JP, et al. Hyperkalemia-induced bradydysrhythmias. J Acute Med. 2025;15:43–51. doi:10.6705/j.jacme.202506_15(2).0001
- Goldstein JA. Hemodynamic complications of right ventricular infarction: role of the right atrium. JACC Case Rep. 2021;3:1174–1176. doi:10.1016/j.jaccas.2021.02.045
- Waqanivavalagi SWFR. Temporary pacing following cardiac surgery: a reference guide for surgical teams. J Cardiothorac Surg. 2024;19:115. doi:10.1186/s13019-024-02619-9
- Elmi F, Tullo NG, Khalighi K. Natural history and predictors of temporary epicardial pacemaker wire function in patients after open heart surgery. Cardiology. 2002;98:175–180. doi:10.1159/000067312
- Bethea BT, Salazar JD, Grega MA, et al. Determining the utility of temporary pacing wires after coronary artery bypass surgery. Ann Thorac Surg. 2005;79:104–107. doi:10.1016/j.athoracsur.2004.06.087
- Manuel L. Temporary epicardial pacing wires post-cardiac surgery: a literature review. Gen Thorac Cardiovasc Surg. 2022;70:595–601. doi:10.1007/s11748-022-01831-5
- Reade MC. Temporary epicardial pacing after cardiac surgery: a practical review. Part 1: general considerations in the management of epicardial pacing. Anaesthesia. 2007;62:264–271. doi:10.1111/j.1365-2044.2007.04950.x
- Reade MC. Temporary epicardial pacing after cardiac surgery: a practical review. Part 2: selection of epicardial pacing modes and troubleshooting. Anaesthesia. 2007;62:364–373. doi:10.1111/j.1365-2044.2007.04951.x
- Faurie B, Souteyrand G, Staat P, et al. Left ventricular rapid pacing via the valve delivery guidewire in transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2019;12:2449–2459. doi:10.1016/j.jcin.2019.09.029
- Berman E, Abu-Much A, Reisman M, et al. Contemporary review of the methods for rapid ventricular pacing during transcatheter aortic valve replacement. Struct Heart. 2025;9:100306. doi:10.1016/j.shj.2024.100306
- Hain A, Schmermund BN, Kriechbaum SD, et al. Managing bradycardia in patients after transcatheter tricuspid valve replacement. Struct Heart. 2025;9:100710. doi:10.1016/j.shj.2025.100710
- Hahn RT, Makkar R, Thourani VH, et al. Transcatheter valve replacement in severe tricuspid regurgitation. N Engl J Med. 2025;392:115–126. doi:10.1056/NEJMoa2401918