Library › Coronary artery disease › IVUS and OCT
Intravascular imaging: IVUS and OCT
Coronary intravascular ultrasound and optical coherence tomography in adults, from physics, acquisition, and image interpretation through lesion assessment, stent optimization, stent failure, randomized trials, and guideline recommendations.
Intravascular ultrasound (IVUS) and optical coherence tomography (OCT) image the coronary wall in cross-section from inside the artery. IVUS penetrates 4–8 mm, has an axial resolution of about 100–150 μm, and needs no blood clearance. OCT resolves 10–20 μm but penetrates only 1–3 mm and requires a contrast flush. In acute coronary syndromes, intravascular imaging to guide percutaneous coronary intervention (PCI) is a Class 1, Level of Evidence A recommendation in the 2025 ACC/AHA guideline. In a network meta-analysis of 22 randomized trials, imaging-guided stent implantation reduced target-lesion failure, cardiac death, stent thrombosis, and all-cause death compared with angiography alone.
Intravascular ultrasound: principles, catheters, and resolution
IVUS forms an image from ultrasound reflected at interfaces of differing acoustic impedance. A transducer at the catheter tip transmits and receives across 360°, and the console displays a cross-sectional image of the lumen and wall at the transducer position together with a longitudinal reconstruction of the pullback. Most coronary IVUS catheters pass through 5–6 French guide catheters (Velagapudi 2021).
Catheter designs. Two designs are in clinical use. Mechanical (rotational) catheters carry a single transducer on a flexible high-torque drive shaft that rotates at about 30 revolutions per second. They give good image quality and signal strength. Their limitations are a drive shaft that reduces flexibility in tortuous vessels, a sheath that requires saline flushing to remove air, and binding of the drive cable, which produces nonuniform rotational distortion. Electronic (phased-array or solid-state) catheters arrange multiple small crystals around the catheter tip; they need no rotation and no continuous flushing, and a central guidewire stays in the vessel during imaging. The short distance from tip to sensor in some phased-array designs is useful in chronic total occlusion PCI (Velagapudi 2021).
Frequency, resolution, and penetration. Frequency determines the trade-off between resolution and field of view. Frequencies of 8–12 MHz suit the aorta and inferior vena cava, about 20 MHz suits the carotid and femoropopliteal arteries, and 40–60 MHz is used in coronary arteries (Velagapudi 2021). Axial resolution ranges from 38 to 170 μm, depending on the manufacturer (Velagapudi 2021). Conventional coronary IVUS has an axial resolution of 100–150 μm and a lateral resolution of 150–300 μm (Ono 2020). Reported penetration ranges are 5–6 mm (Ono 2020) and 4–8 mm (Tsigkas 2023), enough to image the external elastic membrane (EEM) and the full plaque thickness in most coronary segments. High-definition systems at 60 MHz reach an axial resolution below 40 μm, which narrows the resolution gap with OCT while retaining penetration (Ono 2020). Ultrasound passes through blood, so imaging needs no flush. Calcium reflects ultrasound almost completely: the leading edge of calcium is bright and the tissue behind it is lost in an acoustic shadow.
Pullback. Images are acquired by motorized pullback at a fixed speed, classically 0.5 or 1.0 mm/s, or by manual pullback; motorized pullback allows measurement of lesion and stent length from the longitudinal view (Velagapudi 2021).
Optical coherence tomography: principles, platforms, and blood clearance
OCT measures the echo delay and intensity of backscattered near-infrared light, at a wavelength of about 1.3 μm, to build cross-sectional and volumetric images with an axial resolution of 10–20 μm and a lateral resolution of 20–90 μm (Ono 2020). Reported penetration ranges for light are 1–2 mm (Ono 2020) and 1–3 mm (Tsigkas 2023), so the EEM is often invisible behind thick or lipid-rich plaque (Ben-Assa 2021).
Platforms. Contemporary systems are frequency-domain (Fourier-domain) OCT, which replaced time-domain systems (van der Sijde 2017). Optical frequency domain imaging (OFDI) is the frequency-domain platform of a second manufacturer (Ono 2020). A typical frequency-domain platform uses a 2.7 French rapid-exchange catheter on a 0.014-inch guidewire and offers two pullback modes: a 75 mm survey pullback at 36 mm/s and a 54 mm high-resolution pullback at 18 mm/s (Ben-Assa 2021). A complete acquisition takes about 5 seconds (Ben-Assa 2021).
Blood clearance. Red blood cells scatter light, so blood must be displaced from the lumen during the pullback. Clearance is achieved with an injection of contrast through a coaxially seated guide catheter, usually 3–4 mL/s for a total bolus of about 15 mL in the right coronary artery and 3–5 mL/s for 15–20 mL in the left. An automated injector delivers about 14 mL at 4 mL/s. Diluted contrast or low-molecular-weight dextran reduces the contrast load in kidney disease (Ben-Assa 2021; Ono 2020). Pullback starts automatically when the lumen clears. Incomplete clearance is the most frequent cause of a nondiagnostic OCT run, and aorto-ostial and very large vessels clear poorly (Ben-Assa 2021).
| Feature | Intravascular ultrasound | Optical coherence tomography |
|---|---|---|
| Energy source | Ultrasound, 40–60 MHz in coronary arteries | Near-infrared light, wavelength about 1.3 μm |
| Axial resolution | 100–150 μm (below 40 μm with 60 MHz high-definition systems) | 10–20 μm |
| Lateral resolution | 150–300 μm | 20–90 μm |
| Penetration | 4–8 mm; full wall and EEM visible in most segments | 1–3 mm; EEM often obscured behind lipid or thick plaque |
| Blood clearance | Not required | Contrast (or dextran or saline) flush required |
| Pullback | 0.5–1.0 mm/s motorized, or manual | 18–36 mm/s; acquisition about 5 seconds |
| Minimum guide catheter | 5 French | 6 French preferred |
| Calcium | Bright leading edge with acoustic shadow; arc and length measurable, thickness usually not | Sharply delineated, low attenuation; arc, length, and thickness measurable |
| Lipid | Attenuation; necrotic core inferred | Signal-poor region with diffuse border and high attenuation; cap thickness measurable |
| Thrombus | Echolucent, layered or pedunculated mass; often hard to identify | Intraluminal mass; red and white thrombus distinguished by attenuation |
| Stent struts | Bright reflectors; apposition assessed with lower sensitivity | Bright line with shadow; malapposition, protrusion, and coverage measured at strut level |
| Ostial left main | Imaged well | Imaged poorly (incomplete clearance, large diameter) |
| Three-dimensional reconstruction | Limited | Online reconstruction available |
Technical characteristics of coronary IVUS and OCT. Data from Ono 2020, Tsigkas 2023, Velagapudi 2021, and Ben-Assa 2021. EEM, external elastic membrane.

Matched OCT and high-definition IVUS frames. Panel A: normal coronary artery; the white dotted line traces the external elastic membrane, the red dotted line the internal elastic membrane, and the solid white line the lumen, best defined by OCT; the yellow wedge marks the guidewire artifact. Panel B: fibrocalcific plaque; IVUS shows the external elastic membrane behind the plaque, which OCT loses through attenuation at 3–4 o'clock, while OCT outlines the calcium (yellow line) and the fibrous tissue over it. Panel C: spontaneous coronary artery dissection; OCT shows intima-media separation from the external elastic membrane (i), and IVUS shows the depth of the intramural hematoma (ii'). Credit: Alasnag M, Ahmed W, Al-Bawardy R, Al Shammeri O, Biswas S, Johnson TW, Front Cardiovasc Med 2022;9:878801, Figure 1; CC BY 4.0.
Indications and contraindications for intracoronary imaging
The diagnostic uses of intravascular imaging are the severity of an angiographically indeterminate stenosis, above all in the left main coronary artery; the mechanism of an acute coronary syndrome when the culprit is uncertain; the cause of ambiguous angiographic findings such as haziness after stenting; the mechanism of restenosis and stent thrombosis; and allograft vasculopathy after heart transplantation (Velagapudi 2021; Räber 2018). Imaging identifies the cause of haziness after stenting: edge dissection, intramural hematoma, thrombus, plaque protrusion, or underexpansion. Its procedural uses are vessel sizing; landing-zone selection; assessment of calcium to choose lesion preparation; wire position in chronic total occlusion; confirmation of stent expansion and apposition; detection of edge dissection, tissue protrusion, and geographic miss; and contrast-sparing PCI (Velagapudi 2021; Ben-Assa 2021).
The contraindications are those of coronary angiography (Velagapudi 2021). Practical limits are a vessel too small or tortuous to accept the catheter, a subtotal stenosis that the catheter would occlude (managed by small-balloon predilation), and, for OCT, a contrast load the kidneys cannot tolerate and an aorto-ostial target (Ben-Assa 2021). The 2011 PCI guideline gave routine lesion assessment with IVUS a Class III (no benefit) recommendation when revascularization is not contemplated (Velagapudi 2021). Imaging is performed during the same catheterization, under the same sedation and anticoagulation, and an OCT acquisition adds a contrast injection lasting about 5 seconds (Ben-Assa 2021).
Image acquisition: anticoagulation, nitroglycerin, catheter position, and pullback
Imaging is performed under full anticoagulation, usually with unfractionated heparin to a therapeutic activated clotting time, because the catheter occupies the artery and can promote thrombus (Velagapudi 2021; Ben-Assa 2021). Intracoronary nitroglycerin is given before the catheter enters the artery, unless contraindicated, to prevent spasm and to standardize vessel tone for measurement (Velagapudi 2021).
IVUS. A mechanical catheter is flushed meticulously to remove air, which degrades the image and risks air embolism; a phased-array catheter needs only a flush of the wire port. The catheter is advanced over the guidewire well beyond the segment of interest under fluoroscopy and is never pushed against resistance; when resistance is met, the catheter is withdrawn and the vessel is predilated. Pullback is motorized at 0.5 or 1.0 mm/s, or manual, and a final angiogram after removal of the catheter excludes injury (Velagapudi 2021).
OCT. The catheter is purged with undiluted contrast, connected to the drive motor, and calibrated. The lens marker is placed distal to the target, a guide catheter seated coaxially in the ostium allows a selective flush, and the pullback is triggered when the lumen clears. Problems with clearance are addressed by repositioning the guide, raising the flush rate, or predilating a tight lesion with a 1.5–2.0 mm balloon, which modifies the lesion before it is imaged (Ben-Assa 2021).
Co-registration. Angiographic co-registration maps the imaging frame to its position on the angiogram, which places stent edges in the landing zones chosen on imaging and reduces geographic miss (Ben-Assa 2021).
Safety and complications of intracoronary imaging
Complications are uncommon. In early multicenter IVUS experience, the major complication rate was 0.1–0.5%, and coronary spasm, reported in 2.9% of cases, was the most frequent event (Velagapudi 2021). In a single-center registry of 1,142 OCT and 2,476 IVUS procedures, imaging-related complications occurred in 0.6% with OCT and 0.5% with IVUS. All were self-limiting after withdrawal of the catheter or treatable in the laboratory, and none caused permanent harm (van der Sijde 2017). In ILUMIEN IV, OCT-related adverse events occurred in 1 of 1,233 patients assigned to OCT (Ali 2023). In OCTIVUS, no imaging-related complications were observed (Kang 2023). Device-related events occurred in 0.4% of patients in the near-infrared spectroscopy IVUS natural-history study (Waksman 2019).
The recognized events are spasm, coronary dissection or perforation, air embolism from an inadequately flushed mechanical catheter, arrhythmia during the OCT flush, and contrast-associated acute kidney injury with OCT (Velagapudi 2021; Ben-Assa 2021). In spontaneous coronary artery dissection (SCAD), catheter passage and the hydraulic force of the OCT flush can extend the dissection, which limits imaging to cases in which the result will change management (imaging in SCAD). Complications of the PCI itself and of vascular access are distinct from imaging complications.
The normal coronary wall and IVUS and OCT measurements
Normal wall. On IVUS, the normal artery shows a bright inner layer (intima and internal elastic lamina), a dark middle layer (media, made of smooth muscle), and a bright, layered outer zone (adventitia); the interface between media and adventitia marks the EEM (Velagapudi 2021). On OCT, the intima and adventitia are signal-rich and the media is a signal-poor band; the internal and external elastic laminae lie below the resolution of OCT and are located by the transitions between these layers (Ben-Assa 2021).
Measurement conventions. IVUS boundaries are traced at the leading edge of the echo (Velagapudi 2021). The lumen area is the area within the luminal border; the EEM area (vessel area) is the area within the EEM; plaque-plus-media area is the EEM area minus the lumen area, because the media cannot be separated reliably from plaque on IVUS. Plaque burden is the plaque-plus-media area divided by the EEM area, expressed as a percentage. The reference segments are the most normal-looking cross-sections proximal and distal to the lesion (Ben-Assa 2021). Remodeling compares the EEM area at the lesion with that at the reference: in positive (expansive) remodeling the vessel enlarges around the plaque and the lumen is preserved, and in negative remodeling the vessel shrinks and the lumen narrows.
| Measurement | Definition | Main use |
|---|---|---|
| Minimal lumen area (MLA) | Smallest lumen cross-sectional area in the lesion | Severity of an intermediate stenosis; deferral of left main revascularization |
| Lumen area stenosis | Reference lumen area minus minimal lumen area, divided by reference lumen area | Severity relative to the vessel |
| EEM (vessel) area | Area within the external elastic membrane | Stent sizing by the vessel; remodeling |
| Plaque burden | (EEM area − lumen area) ÷ EEM area × 100 | Landing-zone selection; high-risk plaque (70% or more in PROSPECT) |
| Minimal stent area (MSA) | Smallest area within the stent struts | Absolute stent expansion; predictor of restenosis and thrombosis |
| Stent expansion | MSA ÷ reference lumen area (distal, or mean of proximal and distal) × 100 | Relative stent expansion |
| Malapposition distance | Distance from the strut surface to the vessel wall | Apposition (OCT, measured at the strut level) |
| Fibrous cap thickness | Thinnest cap over a lipid core | Thin-cap fibroatheroma (OCT, below 65 μm) |
Standard quantitative measurements in intracoronary imaging. Definitions from Velagapudi 2021, Ben-Assa 2021, Stone 2011, and Ono 2020. EEM, external elastic membrane.
Reproducibility. Lumen and stent measurements are highly reproducible; in a study of 100 stents, intraobserver and interobserver correlations were 0.92–0.97 for reference and stent lumen areas. Derived stent expansion was less reproducible (correlations 0.80 and 0.70), and two observers agreed on whether expansion reached 90% in only 77% of stents (Blessing 1999). OCT measures lumen dimensions more reproducibly than IVUS because the lumen border is sharp after flushing. In diseased segments, OCT often cannot see the EEM and reports smaller areas than IVUS, so IVUS sizing values are not transferred to OCT (Ben-Assa 2021).
Image artifacts on IVUS and OCT
Artifacts create features that do not exist or hide features that do, and their recognition prevents false measurements.
| Artifact | Modality | Mechanism and appearance | Management |
|---|---|---|---|
| Nonuniform rotational distortion | IVUS (mechanical) | Binding of the drive cable in tortuous vessels, a kinked sheath, or a tight guide; part of the image is stretched and part compressed | Straightening of the system; software correction on some consoles |
| Ring-down | IVUS | Bright halos around the catheter from high-amplitude signals near the transducer | Time-gain compensation; digital subtraction of a reference mask |
| Blood speckle | IVUS | Echogenic blood at high frequency and low flow obscures the lumen border | Contrast or saline flush; time-gain compensation |
| Side lobe | IVUS | Bright arcs beside a strong reflector (calcium, struts) from beams outside the main lobe | Recognition; measurement away from the arc |
| Guidewire shadow | IVUS and OCT | Sector of signal loss behind the wire; may mimic a side branch | Recognition; measurement across adjacent frames |
| Residual blood | OCT | Signal-rich swirls in the lumen, attenuation, and shadowing; can mimic thrombus and distort struts | Repeat run after correction of the flush |
| Thrombus shadow | OCT | Red thrombus attenuates light and hides the wall and its own back surface | Recognition; thrombus size not measured from the shadowed side |
| Fold-over | OCT | Aliasing when the vessel or a side branch exceeds the field of view; tissue appears folded back | Frame excluded from geometric measurement |
| Merry-go-round, blooming, ghost struts | OCT | Stent strut distortion from residual blood or catheter position | Recognition; repeat flush |
Common artifacts in intracoronary imaging and their correction. Data from Velagapudi 2021 and Ben-Assa 2021.
Metal casts a shadow on OCT because light cannot penetrate it: a round guidewire appears as a crescent and a stent strut as a bright line, and the shadow widens as the metal lies closer to the catheter and farther from the wall (Ben-Assa 2021).
Plaque characterization: fibrous, lipid-rich, and calcified plaque
IVUS. Plaque components are classified by echogenicity compared with the adventitia. Soft (lipid-rich or necrotic) plaque has low echogenicity, fibrous plaque intermediate echogenicity, and calcium high echogenicity with acoustic shadowing (Velagapudi 2021). Attenuated plaque, in which the signal fades behind a noncalcified plaque, corresponds to a large necrotic core or lipid pool and thin-cap fibroatheroma (Jia 2016). Gray-scale IVUS cannot measure a fibrous cap.
OCT. OCT characterization has been validated against histology. Fibrous tissue is homogeneous and signal-rich with low attenuation; lipid is signal-poor with high attenuation and a diffuse border that hides deeper structures; calcium is a heterogeneous, low-attenuation region with a sharp border (Ben-Assa 2021). A thin-cap fibroatheroma is a necrotic core covered by a fibrous cap thinner than 65 μm (Ono 2020). Macrophages appear as bright spots or bands at the boundary between cap and lipid core; cholesterol crystals are bright, signal-rich structures; and layered plaque shows layers of fibrous tissue with differing signal (Gupta 2022; Alasnag 2022). Superficial macrophages attenuate the signal and can make a fibrous plaque look like a thin-cap fibroatheroma (Ono 2020).

OCT plaque morphology. (A) Normal artery with intima, media, and adventitia. (B) Lipid-rich plaque (asterisks). (C) Fibrous plaque. (D) Calcified nodule. (E) Calcium with an arc near 360°. (F) Deep calcium, 1.20 mm from the lumen. (G) Thin-cap fibroatheroma with a cap below 60 μm. (H) Red thrombus with shadowing. (I) White thrombus. (J) Macrophages as bright spots at the boundary between cap and lipid core. (K) Cholesterol crystals. (L) Thin cap over a large lipid pool with macrophages. Credit: Gupta A, Shrivastava A, Vijayvergiya R, et al., Front Cardiovasc Med 2022;9:854554, Figure 1; CC BY 4.0.
Thrombus. OCT identifies thrombus as a mass attached to the wall or floating in the lumen. Red thrombus, rich in red blood cells, is signal-rich with high attenuation and casts a shadow; white thrombus, made mainly of platelets and white cells, is signal-rich with low attenuation (Ben-Assa 2021). On IVUS, thrombus is relatively echolucent and often layered or pedunculated, and it can be difficult to separate from soft plaque or blood (Velagapudi 2021).
| Tissue | Gray-scale IVUS | OCT |
|---|---|---|
| Fibrous plaque | Intermediate echogenicity | Homogeneous, signal-rich, low attenuation |
| Lipid-rich plaque or necrotic core | Low echogenicity; attenuation behind noncalcified plaque | Signal-poor, diffuse border, high attenuation |
| Calcium | Bright leading edge with acoustic shadow | Heterogeneous, low attenuation, sharp border |
| Thin-cap fibroatheroma | Not measurable; inferred on radiofrequency analysis | Lipid core with cap below 65 μm |
| Macrophages | Not resolved | Bright spots or bands at the boundary between cap and lipid core with shadowing |
| Red thrombus | Echolucent mass, layered or pedunculated | Signal-rich mass with high attenuation and shadow |
| White thrombus | Difficult to identify | Signal-rich mass with low attenuation |
| Neointimal hyperplasia | Low echogenicity early, high echogenicity late | Homogeneous or heterogeneous tissue over struts |
Appearance of coronary tissue types on gray-scale IVUS and OCT. Data from Velagapudi 2021, Ben-Assa 2021, Ono 2020, Gupta 2022, and Jia 2016.
Radiofrequency IVUS, virtual histology, and near-infrared spectroscopy
Virtual histology IVUS. Radiofrequency analysis of the backscattered ultrasound signal produces color-coded maps of plaque composition: dense calcium in white, fibrous tissue in dark green, fibrofatty tissue in light green, and necrotic core in red, reported as percentages of plaque area and volume. The technique contributed to the identification of high-risk nonculprit lesions in PROSPECT and other natural-history studies. Its limitations are substantial: in a porcine model, accuracy against histology was 58% for fibrous, 38% for fibrofatty, and 38% for necrotic core plaque; necrotic core size did not correlate with histology. Dense calcium, thrombus, and stent struts generate artifacts. Virtual histology has limited clinical use on its own (Velagapudi 2021).
Near-infrared spectroscopy. Combined near-infrared spectroscopy (NIRS) and IVUS catheters measure the chemical signature of lipid in the vessel wall and report it as a lipid core burden index; the maximum value in any 4 mm segment (maxLCBI4mm) summarizes the most lipid-rich segment. In the Lipid-Rich Plaque (LRP) study of 1,271 patients, a 100-unit increase in maxLCBI4mm raised the adjusted hazard of nonculprit events by 18% at the patient level, and segments with a maxLCBI4mm above 400 carried an adjusted hazard ratio of 3.39 (Waksman 2019). Natural-history results are summarized under vulnerable plaque.
Calcium assessment and lesion preparation
Calcium prevents stent expansion, and stent expansion is the most validated metric of PCI optimization (Ben-Assa 2021). Imaging grades calcium by arc, thickness, length, depth, and the presence of a calcified nodule.
OCT calcium score. In a derivation and validation study, maximum calcium arc, maximum thickness, and length predicted stent expansion. The score assigns 2 points for a maximum arc above 180°, 1 point for a maximum thickness above 0.5 mm, and 1 point for a length above 5 mm. Lesions with a score of 0–3 achieved a mean stent expansion of 96%, and lesions with a score of 4 achieved 78% (Fujino 2018). A score of 4 identifies lesions that need calcium modification before stenting.
IVUS. IVUS measures the arc and length of calcium and shows whether it is superficial or deep, but the acoustic shadow usually prevents measurement of thickness. OCT measures calcium thickness, which distinguishes a thin superficial plate that balloon pressure can fracture from thick calcium that requires atherectomy or lithotripsy (Ben-Assa 2021).
Choice of device. Fibrous or lipid-rich lesions are treated by direct stenting or undersized balloon predilation. Thin calcium is often managed with noncompliant balloons; deeper calcium with cutting or scoring balloons; and thick superficial calcium with atherectomy (Ben-Assa 2021). Intravascular lithotripsy delivers acoustic pressure waves that fracture calcium. Disrupt CAD III enrolled 431 patients with severely calcified lesions (mean calcium angle 292.5°, maximum thickness 0.96 mm). OCT showed calcium fractures after lithotripsy in 67.4% of lesions, and the final minimal stent area was 6.5 ± 2.1 mm² whether or not fractures were visible (Hill 2020). Repeat imaging after modification confirms fracture and adequate lumen gain before the stent is placed.
| Imaging finding | Expected stent expansion | Lesion preparation |
|---|---|---|
| Fibrous or lipid-rich plaque without calcium | Adequate | Direct stenting or predilation with an undersized compliant balloon |
| OCT calcium score 0–3 | Mean 96% in validation | Noncompliant balloon when calcium is thin |
| OCT calcium score 4 (arc above 180°, thickness above 0.5 mm, length above 5 mm) | Mean 78% in validation | Calcium modification before stenting |
| Deeper calcium | Limited by arc and depth | Cutting or scoring balloon |
| Thick superficial calcium | Limited | Atherectomy |
| Severe calcium (Disrupt CAD III population) | MSA 6.5 ± 2.1 mm² after lithotripsy | Intravascular lithotripsy |
Calcium on intracoronary imaging and lesion preparation. Score from Fujino 2018; device selection from Ben-Assa 2021; lithotripsy result from Hill 2020. MSA, minimal stent area.
Culprit mechanisms in acute coronary syndromes: rupture, erosion, and calcified nodule
OCT distinguishes the three pathologic substrates of coronary thrombosis. Plaque rupture is a discontinuity of the fibrous cap over a lipid-rich plaque, with thrombus rich in red blood cells. Plaque erosion is thrombus, typically platelet-rich white thrombus, over an intact fibrous cap. Calcified nodule is calcium protruding into the lumen as the substrate of thrombosis (Jia 2013; Tsigkas 2023). In 126 patients with acute coronary syndromes imaged before intervention, the culprit was rupture in 43.7%, erosion in 31.0%, and calcified nodule in 7.9%. Patients with erosion were younger (mean 53.8 years), more often presented with non-ST-elevation syndromes, had thicker caps (169 μm against 60 μm with rupture), and had less severe stenosis (Jia 2013).
Treatment by mechanism. In the single-arm EROSION study, 60 patients with OCT-defined erosion and residual diameter stenosis below 70% received antithrombotic therapy without stenting; 78.3% met the primary end point of a thrombus volume reduction of more than 50% at 1 month (Jia 2017). Deferral of stenting in erosion remains investigational and is applied only within protocols.
Ambiguous culprits and MINOCA. In non-ST-elevation acute coronary syndrome (NSTE-ACS) without an obvious culprit, imaging identifies rupture, erosion, or thrombus (angiography and imaging in NSTE-ACS). In myocardial infarction with nonobstructive coronary arteries (MINOCA), OCT or IVUS at the index procedure, followed by cardiac MRI, identifies plaque disruption, layered plaque, or intramural hematoma in a large share of women (MINOCA algorithm).
Intermediate non-left main stenosis: minimal lumen area and its limits
The IVUS minimal lumen area correlates only moderately with fractional flow reserve (FFR) (Waksman 2013). In 700 left anterior descending artery lesions, independent predictors of FFR were age, body surface area, lesion length, angiographic diameter stenosis, IVUS MLA, and plaque burden. When left ventricular mass was included, it replaced body surface area. IVUS MLA agreed with FFR in 71% of men and 64% of women (Kang 2013).
- In 236 lesions, an MLA below 2.4 mm² best predicted an FFR below 0.80, with 90% sensitivity and 60% specificity; 96% of lesions with an MLA of 2.4 mm² or more had an FFR of 0.80 or more, while only 37% of lesions below 2.4 mm² had an FFR below 0.80 (Kang 2011a).
- In the FIRST registry of 367 lesions at 10 US and European sites, the best overall MLA threshold was below 3.07 mm² (sensitivity 64.0%, specificity 64.9%, area under the curve 0.65). Thresholds rose with vessel size: below 2.4 mm² for reference diameters under 3.0 mm, below 2.7 mm² for 3.0–3.5 mm, and below 3.6 mm² for over 3.5 mm (Waksman 2013).
The high negative predictive value and low positive predictive value mean that a large MLA supports deferral, while a small MLA does not establish ischemia (Kang 2011a; Velagapudi 2021). Physiology is the standard for the decision to revascularize an intermediate non-left main stenosis (FFR principle and threshold).
FLAVOUR. FLAVOUR randomized 1,682 patients with intermediate stenoses (40–70% by visual estimate) to FFR-guided or IVUS-guided decisions. PCI was performed for an FFR of 0.80 or less, or for an IVUS MLA of 3 mm² or less, or 3–4 mm² with plaque burden above 70%. PCI was performed in 44.4% of the FFR group and 65.3% of the IVUS group. The 24-month rate of death, myocardial infarction, or revascularization was 8.1% and 8.5%, respectively, and FFR guidance met the criterion for noninferiority (Koo 2022). Over a median of 6.3 years, the composite was 23.1% with FFR and 20.9% with IVUS (hazard ratio 1.15, nonsignificant). Revascularization after the index procedure was more frequent with FFR (14.9% against 11.8%), driven by late revascularization of deferred vessels. Total target-vessel PCI, including the index procedure, remained lower with FFR (38.8% against 60.5%) (Yang 2025).
Left main coronary artery assessment by IVUS
Angiographic measurements are unreliable in intermediate left main lesions (de la Torre Hernandez 2011). IVUS is the intravascular method used for left main severity; OCT images the ostial left main poorly because the aorto-ostial junction does not clear (Ben-Assa 2021).
Thresholds. In the LITRO study of 354 patients at 22 centers, revascularization was performed for an IVUS MLA below 6 mm² and deferred for 6 mm² or more. Revascularization was deferred in 96% of patients with an MLA of 6 mm² or more. At 2 years, cardiac death-free survival was 97.7% in deferred and 94.5% in revascularized patients, and only 4.4% of deferred patients needed later left main revascularization, none with infarction (de la Torre Hernandez 2011). In Korean patients with isolated ostial or shaft left main stenosis, an MLA of 4.5 mm² or less best identified an FFR of 0.80 or less (sensitivity 77%, specificity 82%), and correction for body surface area, body mass index, or left ventricular mass did not improve accuracy (Park 2014). An earlier Korean series of 55 patients found a best cutoff below 4.8 mm² (Kang 2011b). These lower cutoffs were derived in Korean patients and are applied with physiology and clinical context. Plaque rupture lowered FFR independently of MLA (Park 2014; Kang 2011b).
Guidelines. The 2021 ACC/AHA/SCAI revascularization guideline states that IVUS is reasonable to help define lesion severity in patients with intermediate stenosis of the left main artery (Class 2a, Level of Evidence B-NR) (Lawton 2022).
| Setting | IVUS threshold | Study and population | Interpretation |
|---|---|---|---|
| Left main, deferral | MLA 6 mm² or more | LITRO, 354 patients, 22 centers, prospective (de la Torre Hernandez 2011) | Deferral safe at 2 years |
| Left main, FFR 0.80 or less | MLA 4.5 mm² or less | 112 Korean patients, ostial or shaft lesions (Park 2014) | Sensitivity 77%, specificity 82% |
| Left main, FFR below 0.80 | MLA below 4.8 mm² | 55 Korean patients (Kang 2011b) | Sensitivity 89%, specificity 83% |
| Non-left main, FFR below 0.80 | MLA below 2.4 mm² | 236 lesions, Korea (Kang 2011a) | Negative predictive value high; specificity 60% |
| Non-left main, FFR below 0.80 | MLA below 3.07 mm² overall; 2.4, 2.7, and 3.6 mm² by vessel size | FIRST, 367 lesions, US and Europe (Waksman 2013) | Area under the curve 0.65; accuracy vessel-dependent |
| Non-left main, PCI trigger in FLAVOUR | MLA 3 mm² or less, or 3–4 mm² with plaque burden above 70% | FLAVOUR, 1,682 patients (Koo 2022) | Outcomes similar to FFR guidance; more PCI |
IVUS minimal lumen area thresholds for intermediate coronary stenoses. FFR, fractional flow reserve; MLA, minimal lumen area.
Myocardial bridging on intravascular imaging
Myocardial bridging is a segment of coronary artery, most often the left anterior descending artery, that runs within the myocardium. Angiography shows systolic narrowing ("milking") in 0.5–2.5% of patients, compared with 15–85% at autopsy (Ge 1999). In all 62 patients with angiographic milking studied by IVUS, an echolucent "half-moon" band of muscle surrounded the bridged segment throughout the cardiac cycle, 0.47 mm thick in diastole and 0.52 mm in systole. Systolic compression reduced the lumen by 36.4%. Intracoronary Doppler showed an early diastolic "fingertip" velocity peak in 87% and reduced or absent systolic antegrade flow, and nitroglycerin provoked retrograde systolic flow in the proximal segment in 77%. IVUS found atherosclerosis in the segment proximal to the bridge in 88% of patients and none within the bridge or distal to it (Ge 1999). The half-moon sign and the plaque-free bridged segment with proximal atherosclerosis are the intravascular signature of bridging.
Pre-PCI planning: vessel sizing, stent length, and landing zones
Morphology. Pre-PCI imaging identifies the plaque type, calcium burden, thrombus, and mechanism, which determine lesion preparation (Ben-Assa 2021).
Diameter. IVUS sizing by the vessel uses the mean media-to-media (EEM) diameter at the distal reference, rounded down to the nearest stent size; when the EEM cannot be defined, the mean distal reference lumen diameter is rounded up (Paul 2022). OCT sizing by the vessel, used in ILUMIEN III, takes the smaller of the mean proximal and distal EEM diameters and rounds it down to the nearest 0.25 mm; a reference measuring 2.7 by 2.9 mm (mean 2.8 mm) leads to a 2.75 mm stent. Upsizing is not used with EEM-based sizing and is permitted to a limited extent with lumen-based sizing (Ben-Assa 2021). With this protocol, ILUMIEN III achieved a minimal stent area with OCT guidance that was noninferior to IVUS guidance (median 5.79 against 5.89 mm²) (Ali 2016).
Length and landing zones. Landing zones are chosen in segments with a plaque burden below 50% (Paul 2022) and away from lipid-rich plaque and calcium at the edges. Stent edges in lipid pools are associated with periprocedural infarction, edge dissection, and edge restenosis (Ben-Assa 2021). Motorized pullback or co-registration measures the distance between landing zones and avoids geographic miss: a stent that leaves part of the diseased segment uncovered or lands its edge on residual disease. Post-dilation is performed separately in the proximal and distal stent segments; with vessel-based sizing, the balloon does not exceed the EEM diameter of the corresponding reference segment (Ben-Assa 2021).
Stent optimization criteria after implantation
Post-PCI imaging assesses expansion, apposition, edge dissection, tissue protrusion, and residual disease at the edges. Stent expansion is the most validated optimization metric linked to clinical outcome (Ben-Assa 2021).
Expansion. Absolute expansion is the minimal stent area. Relative expansion is the MSA divided by a reference lumen area: the distal reference, the smaller reference, or the mean of proximal and distal references (Ben-Assa 2021; Paul 2022). In early IVUS studies, an MSA of 6.4–6.5 mm² predicted freedom from restenosis with bare-metal stents, and the corresponding threshold for drug-eluting stents was 5.3–5.5 mm² (Velagapudi 2021). The SCAI algorithm for IVUS targets an MSA of at least 80%, and ideally above 90%, of the distal reference lumen area, or an MSA above 5.0–5.5 mm² outside the left main and above 8 mm² in the left main (Paul 2022). For OCT, expansion against the mean reference lumen area is graded as optimal at 95% or more, acceptable at 90% to below 95%, and unacceptable below 90% (Ben-Assa 2021). In a multicenter OCT registry of 900 lesions, an MSA below 5.0 mm² in a drug-eluting stent and irregular tissue protrusion were the independent predictors of 1-year device-oriented events (Soeda 2015).
Left main. In 403 patients after left main stenting with 9-month angiographic follow-up, the MSA cutoffs that best predicted restenosis were 5.0 mm² at the ostial circumflex, 6.3 mm² at the ostial left anterior descending artery (LAD), 7.2 mm² at the polygon of confluence, and 8.2 mm² in the left main above the confluence. Underexpansion of at least one segment occurred in 33.8% of patients and was associated with restenosis in 24.1% against 5.4% and with an adjusted hazard ratio of 5.56 for major adverse cardiac events (Kang 2011c).
Optimal procedure in ULTIMATE. In the IVUS arm of ULTIMATE, target-vessel failure at 1 year was 1.6% when all IVUS optimization criteria were met and 4.4% when any criterion failed (hazard ratio 0.349), and the advantage of an optimal procedure persisted at 3 years (Zhang 2018; Gao 2021).
Malapposition. Malapposition is a separation of struts from the wall in a segment without a side branch. Acute malapposition has no established link to clinical events, although pathology studies associate it with delayed healing. OCT malapposition below 200–250 μm from strut to wall is not considered clinically relevant; consoles color-code struts above 200 μm and above 300 μm, and a long malapposed segment is treated by post-dilation (Ben-Assa 2021). The SCAI IVUS algorithm treats gross malapposition longer than 3 mm with post-dilation using a larger semicompliant balloon at nominal pressure (Paul 2022). In OCT studies of stent thrombosis, malapposition is a leading finding in both early and late events (stent thrombosis).
Edge dissection. OCT detects many small intimal edge dissections that need no treatment. Features that identify a major dissection are a circumferential extent of 60° or more, a length of 3 mm or more, extension into the media or adventitia, and a residual lumen area within the dissection below 90% of the reference (Ben-Assa 2021). On IVUS, a medial dissection with an arc above 60° or a length above 3 mm from the stent edge is covered with an additional stent (Paul 2022). Post-stent OCT found edge dissection in 28.7% of lesions in one registry (Soeda 2015) and in 37.5% of patients in DOCTORS (Meneveau 2016).
Tissue and thrombus protrusion. Protrusion is tissue at least 0.2 mm in dimension projecting through the struts into the lumen; it is smooth, disrupted fibrous, or irregular, and it is major when the protrusion area is 10% or more of the stent area (Ben-Assa 2021). Irregular protrusion predicted 1-year events (Soeda 2015).
Residual reference disease. Untreated disease at the stent edges, defined in ILUMIEN III as an MLA of 60% or less of the adjacent reference lumen area within 10 mm of the stent edge, is associated with stent thrombosis and is avoided by stent length selection (Ben-Assa 2021).
| Criterion | IVUS | OCT | Source |
|---|---|---|---|
| Absolute expansion, non-left main | MSA above 5.0–5.5 mm² | MSA 5.0 mm² or more (below 5.0 mm² predicted events) | Paul 2022; Soeda 2015 |
| Relative expansion | MSA 80% or more, ideally above 90%, of distal reference lumen area | Optimal 95% or more; acceptable 90% to below 95%; unacceptable below 90% | Paul 2022; Ben-Assa 2021 |
| Left main segments | Ostial LCX 5.0, ostial LAD 6.3, POC 7.2, LM above POC 8.2 mm² | IVUS preferred for the ostial left main | Kang 2011c |
| Landing zone | Plaque burden below 50% | No lipid-rich plaque or calcium at edges | Paul 2022; Ben-Assa 2021 |
| Edge dissection requiring treatment | Medial dissection with arc above 60° or length above 3 mm | Arc 60° or more, length 3 mm or more, medial or adventitial extension, or lumen below 90% of reference | Paul 2022; Ben-Assa 2021 |
| Malapposition | Gross malapposition over more than 3 mm: post-dilation | Below 200–250 μm not clinically relevant; long segments post-dilated | Paul 2022; Ben-Assa 2021 |
| Tissue protrusion | Recognized less often | Major when 10% or more of stent area; irregular protrusion predicts events | Ben-Assa 2021; Soeda 2015 |
| Residual edge disease | Plaque burden 50% or more at edge | MLA 60% or less of reference within 10 mm of edge | Paul 2022; Ben-Assa 2021 |
Stent optimization criteria for IVUS and OCT. LAD, left anterior descending artery; LCX, left circumflex artery; LM, left main; MLA, minimal lumen area; MSA, minimal stent area; POC, polygon of confluence.
Stent optimization calculator
The calculator applies the criteria above to measurements entered from a post-stent IVUS or OCT pullback. For IVUS outside the left main, it applies the SCAI expansion criteria, which are met by either the relative or the absolute threshold. For IVUS in the left main, it applies the segment cutoffs derived from 403 patients. For OCT, it grades relative expansion against the mean reference lumen area and flags an MSA below 5.0 mm².
IVUS non-left main criteria from the SCAI quality algorithm (Paul 2022): MSA at least 80%, ideally above 90%, of the distal reference lumen area, or above 5.0–5.5 mm²; landing-zone plaque burden below 50%; medial edge dissection with an arc above 60° or length above 3 mm covered; gross malapposition over more than 3 mm post-dilated. IVUS left main cutoffs from Kang 2011c. OCT expansion categories from Ben-Assa 2021, using the mean of proximal and distal reference lumen areas; MSA below 5.0 mm² from Soeda 2015; dissection and malapposition features from Ben-Assa 2021.
The default entry (IVUS, MSA 5.8 mm², distal reference 6.2 mm², landing-zone plaque burden 42%) gives an expansion of 93.5% of the distal reference and meets the criteria. A non-left main IVUS result with an MSA of 4.3 mm² against a distal reference of 6.0 mm² (71.7%) and a medial edge dissection of 75° and 4 mm fails the expansion and dissection criteria. An OCT result with an MSA of 4.6 mm² against references of 6.5 and 5.5 mm² (76.7% of the mean) is unacceptable.
Stent failure: restenosis, underexpansion, neoatherosclerosis, and fracture
Imaging defines the mechanism of stent failure, and the mechanism determines treatment. The 2021 ACC/AHA/SCAI guideline states that IVUS or OCT is reasonable to determine the mechanism of stent failure (Class 2a, Level of Evidence C-LD) (Lawton 2022), and the European interventional consensus recommends routine intracoronary imaging in restenosis and stent thrombosis (Räber 2018).
Mechanical causes. Underexpansion, a minimal stent area well below the reference, is associated with restenosis and thrombosis and may justify further imaging-guided balloon dilation (Velagapudi 2021). Geographic miss and edge restenosis arise from uncovered disease at the stent edges (Ben-Assa 2021). Stent fracture, a complete or partial separation of the stent on fluoroscopy or IVUS, occurred in 2.9% of lesions after everolimus-eluting stent implantation. Ostial location, hinge motion, tortuosity, and calcification predicted it, and fracture was associated with target-lesion revascularization of 25.6% against 2.0% and stent thrombosis of 5.1% against 0.4% (Kuramitsu 2012).
Biologic causes. Neointimal hyperplasia appears on IVUS as low-echogenicity tissue early and higher echogenicity late (Velagapudi 2021). OCT separates in-stent restenosis caused by neointimal hyperplasia from restenosis caused by neoatherosclerosis (Gupta 2022). Neoatherosclerosis is new lipid-laden or calcified plaque inside the stent, best seen on OCT; ruptured neoatherosclerosis is a cause of very late stent thrombosis (Souteyrand 2016; Gupta 2022). Treatment of restenosis follows the mechanism found on imaging.
Bioresorbable scaffolds. Polymeric scaffold struts are barely visible on IVUS, so OCT is the modality for scaffold sizing, apposition, discontinuity, and late scaffold restenosis (Ono 2020; Gupta 2022).
Stent thrombosis on intracoronary imaging
In the French PESTO registry of 120 patients with confirmed stent thrombosis studied by OCT, an underlying abnormality was found in 97%. The findings were strut malapposition in 34%, neoatherosclerosis in 22%, major underexpansion in 11%, coronary evagination in 8%, isolated uncovered struts in 8%, edge-related disease progression in 8%, and neointimal hyperplasia in 4%. Acute and subacute thrombosis was most often due to malapposition (48%) and underexpansion (26%), and late and very late thrombosis to malapposition (31%) and neoatherosclerosis (28%). Ruptured neoatherosclerosis was more frequent with bare-metal stents (36% against 14%), and evaginations with drug-eluting stents (12% against 2%) (Souteyrand 2016). OCT in PESTO was performed after initial treatment of the culprit lesion, at a deferred procedure in 69% of cases, because thrombus obscures the struts. Imaging-guided implantation reduces stent thrombosis in randomized trials (trials of OCT guidance; meta-analyses).
Strut coverage and healing. OCT resolves the thin tissue layer over individual struts and is the method used to assess coverage and healing after stent implantation; uncovered struts are a recognized finding in late stent thrombosis (Souteyrand 2016; Ono 2020).
Imaging in bifurcation, chronic total occlusion, and left main PCI
Bifurcations. OCTOBER randomized 1,201 patients with complex bifurcation lesions at 38 European centers to OCT-guided or angiography-guided PCI; 18.5–19.3% had left main bifurcations. At 2 years, death from cardiac causes, target-lesion infarction, or ischemia-driven target-lesion revascularization occurred in 10.1% with OCT and 14.1% with angiography (hazard ratio 0.70; 95% CI 0.50–0.98). Procedure-related complications occurred in 6.8% and 5.7% (Holm 2023). OCT shows the carina, side-branch ostium, jailing struts, and struts floating across the side-branch ostium after two-stent techniques (Gupta 2022).
Chronic total occlusion. IVUS identifies the proximal cap of a stumpless occlusion from a side branch. During antegrade dissection re-entry and reverse controlled antegrade and retrograde tracking (reverse CART), it confirms whether the wire is within plaque or in the subintimal space. It also sizes balloons and reduces contrast and radiation. Calcium limits identification of the true lumen (Velagapudi 2021). IVUS needs no contrast flush, which suits it to occlusion crossing. In the CTO-IVUS trial, 402 patients randomized after successful wire crossing had lower 12-month major adverse cardiac events with IVUS guidance (2.6% against 7.1%; hazard ratio 0.35) and fewer cardiac deaths or myocardial infarctions (0% against 2.0%), without a difference in cardiac death alone (Kim 2015).
Left main PCI. IVUS guides left main sizing, the choice between single- and two-stent strategies, and expansion at the four left main segments (left main cutoffs).
Slow flow and no-reflow prediction
Attenuated plaque on IVUS, in which echo signal fades behind a noncalcified plaque, marks a necrotic core or lipid pool. In a meta-analysis of 3,833 patients in 9 studies, attenuated plaque did not affect Thrombolysis in Myocardial Infarction (TIMI) flow before PCI, but after balloon dilation and stenting TIMI 0–2 flow was more frequent with attenuated plaque (relative risk 4.73; 95% CI 3.03–7.40). Major adverse events within 3 years did not differ (Jia 2016). Ultrasound attenuation 5 mm or more in length has been associated with no-reflow in ST-elevation myocardial infarction, and in patients imaged with both modalities, OCT lipid index and IVUS plaque burden predicted no-reflow (Tsigkas 2023). Treatment of no-reflow is described under no-reflow treatment.
Contrast-sparing and zero-contrast PCI in kidney disease
IVUS images without contrast and can replace angiographic runs for vessel sizing, lesion length, and the final result. MOZART randomized 83 patients to a meticulous IVUS-guided strategy or angiographic guidance. Median contrast volume was 20.0 mL with IVUS and 64.5 mL with angiography, with no difference in in-hospital or 4-month outcomes (Mariani 2014). OCT uses about 14 ± 3 mL of contrast per acquisition (Ben-Assa 2021); diluted contrast, low-molecular-weight dextran, or saline flush reduces the load, and dextran flushing has been used for zero-contrast PCI with OCT (Ono 2020). Coronary dimensions measured with a saline flush were comparable to those measured with contrast (Gupta 2022). In OCTIVUS, contrast-induced nephropathy occurred in 1.4% with OCT and 1.5% with IVUS (Kang 2023), and in DOCTORS acute kidney injury was 1.6% in both arms despite more contrast with OCT (Meneveau 2016). Contrast-associated kidney injury and its prevention are described with iodinated and gadolinium contrast agents.
Randomized trials of IVUS-guided versus angiography-guided PCI
IVUS-XPL. In 1,400 Korean patients with long lesions (stent length 28 mm or more), major adverse cardiac events at 1 year were 2.9% with IVUS and 5.8% with angiography (hazard ratio 0.48), driven by target-lesion revascularization (Hong 2015). At 5 years, the rates were 5.6% and 10.7% (hazard ratio 0.50), with benefit accruing both in the first year and between 1 and 5 years (Hong 2020).
ULTIMATE. In 1,448 all-comer patients in China, target-vessel failure at 12 months was 2.9% with IVUS and 5.4% with angiography (hazard ratio 0.53) (Zhang 2018). At 3 years, target-vessel failure was 6.6% and 10.7%, and definite or probable stent thrombosis 0.1% and 1.1% (Gao 2021).
RENOVATE-COMPLEX-PCI. In 1,639 Korean patients with complex lesions randomized 2:1, with IVUS or OCT at the operator's choice, the composite of cardiac death, target-vessel myocardial infarction, or clinically driven target-vessel revascularization at a median of 2.1 years was 7.7% with imaging and 12.3% with angiography (hazard ratio 0.64; 95% CI 0.45–0.89). Cardiac death was 1.7% and 3.8%, target-vessel infarction 3.7% and 5.6%, and target-vessel revascularization 3.4% and 5.5%, without a difference in procedural safety events (Lee 2023).
IVUS-ACS. In 3,505 patients with acute coronary syndromes at 58 centers in China, Italy, Pakistan, and the United Kingdom, target-vessel failure at 1 year was 4.0% with IVUS and 7.3% with angiography (hazard ratio 0.55; 95% CI 0.41–0.74), driven by target-vessel infarction and revascularization, without differences in all-cause death or stent thrombosis (Li 2024).
| Trial | Population | Comparison | Primary end point and result |
|---|---|---|---|
| IVUS-XPL (Hong 2015; Hong 2020) | 1,400; long lesions | IVUS vs angiography | MACE 1 year 2.9% vs 5.8% (HR 0.48); 5 years 5.6% vs 10.7% (HR 0.50) |
| ULTIMATE (Zhang 2018; Gao 2021) | 1,448; all comers | IVUS vs angiography | TVF 1 year 2.9% vs 5.4% (HR 0.53); 3 years 6.6% vs 10.7% |
| CTO-IVUS (Kim 2015) | 402; chronic total occlusion after crossing | IVUS vs angiography | Cardiac death 0% vs 1.0% (NS); MACE 2.6% vs 7.1% (HR 0.35) |
| RENOVATE-COMPLEX-PCI (Lee 2023) | 1,639; complex lesions | IVUS or OCT vs angiography | TVF median 2.1 years 7.7% vs 12.3% (HR 0.64) |
| IVUS-ACS (Li 2024) | 3,505; acute coronary syndromes | IVUS vs angiography | TVF 1 year 4.0% vs 7.3% (HR 0.55) |
| FLAVOUR (Koo 2022; Yang 2025) | 1,682; intermediate stenoses | FFR vs IVUS for decision and optimization | Death, MI, or revascularization 2 years 8.1% vs 8.5% (FFR noninferior) |
Randomized trials of IVUS-guided PCI. FFR, fractional flow reserve; HR, hazard ratio; MACE, major adverse cardiac events; MI, myocardial infarction; NS, not significant; TVF, target-vessel failure.
Randomized trials of OCT-guided PCI
DOCTORS. In 240 patients with NSTE-ACS, OCT guidance changed the procedural strategy in 50%, increased post-dilation (43% against 12.5%), and produced a higher post-PCI FFR (0.94 against 0.92), the primary end point. OCT found underexpansion in 42%, malapposition in 32%, incomplete lesion coverage in 20%, and edge dissection in 37.5% (Meneveau 2016).
ILUMIEN III. In 450 patients randomized to OCT, IVUS, or angiography, the EEM-based OCT protocol produced a median MSA of 5.79 mm², noninferior to IVUS (5.89 mm²) and numerically larger than angiography (5.49 mm², superiority not shown). Left main, ostial right coronary, bypass graft, chronic total occlusion, planned two-stent bifurcation, and in-stent restenosis lesions were excluded (Ali 2016).
ILUMIEN IV. In 2,487 patients with medication-treated diabetes or complex lesions at 80 sites in 18 countries, OCT guidance produced a larger MSA (5.72 against 5.36 mm²), but 2-year target-vessel failure did not differ (7.4% against 8.2%; hazard ratio 0.90; 95% CI 0.67–1.19). Stent thrombosis within 2 years occurred in 0.5% and 1.4% (Ali 2023).
OCTOBER. In 1,201 patients with complex bifurcation lesions, OCT guidance reduced the 2-year composite of cardiac death, target-lesion infarction, or ischemia-driven target-lesion revascularization from 14.1% to 10.1% (hazard ratio 0.70) (Holm 2023).
OCCUPI. In 1,604 Korean patients with at least one complex lesion, the 1-year composite of cardiac death, myocardial infarction, stent thrombosis, or ischemia-driven target-vessel revascularization was 5% with OCT and 7% with angiography (hazard ratio 0.62; 95% CI 0.41–0.93), without differences in stroke, bleeding, or contrast-induced nephropathy (Hong 2024).
OCTIVUS. In 2,008 Korean patients randomized to OCT-guided or IVUS-guided PCI, the 1-year composite of cardiac death, target-vessel infarction, or ischemia-driven target-vessel revascularization was 2.5% with OCT and 3.1% with IVUS, meeting noninferiority. Major procedural complications were lower with OCT (2.2% against 3.7%), and no imaging-related complications occurred. The authors noted the selected population and lower-than-expected event rates (Kang 2023).
| Trial | Population | Comparison | Primary end point and result |
|---|---|---|---|
| DOCTORS (Meneveau 2016) | 240; NSTE-ACS | OCT vs angiography | Post-PCI FFR 0.94 vs 0.92 |
| ILUMIEN III (Ali 2016) | 450; native lesions 2.25–3.50 mm, below 40 mm long | OCT vs IVUS vs angiography | Median MSA 5.79 vs 5.89 vs 5.49 mm²; OCT noninferior to IVUS |
| ILUMIEN IV (Ali 2023) | 2,487; diabetes or complex lesions | OCT vs angiography | MSA 5.72 vs 5.36 mm²; TVF 2 years 7.4% vs 8.2% (HR 0.90, NS) |
| OCTOBER (Holm 2023) | 1,201; complex bifurcations | OCT vs angiography | MACE 2 years 10.1% vs 14.1% (HR 0.70) |
| OCCUPI (Hong 2024) | 1,604; complex lesions | OCT vs angiography | MACE 1 year 5% vs 7% (HR 0.62) |
| OCTIVUS (Kang 2023) | 2,008; significant lesions | OCT vs IVUS | TVF 1 year 2.5% vs 3.1%; OCT noninferior |
Randomized trials of OCT-guided PCI. FFR, fractional flow reserve; HR, hazard ratio; MACE, major adverse cardiac events; MSA, minimal stent area; NS, not significant; TVF, target-vessel failure.
Meta-analyses of imaging-guided PCI
A network meta-analysis included 22 trials and 15,964 patients followed for a weighted mean of 24.7 months. Imaging guidance, with IVUS or OCT, reduced target-lesion failure (relative risk 0.71; 95% CI 0.63–0.80), cardiac death (0.55), target-vessel infarction (0.82), target-lesion revascularization (0.72), stent thrombosis (0.52), all myocardial infarction (0.83), and all-cause death (0.75) compared with angiography. OCT and IVUS gave similar results (Stone 2024). A second network meta-analysis of 24 trials (15,489 patients) found reduced target-lesion revascularization with IVUS against angiography (odds ratio 0.69), no significant difference for OCT against angiography (0.83) or OCT against IVUS, and no significant reduction in myocardial infarction with any strategy. In pairwise analysis of 25 trials, imaging reduced target-lesion revascularization, cardiac death, and stent thrombosis (Giacoppo 2024). The two analyses agree that imaging reduces repeat revascularization and stent thrombosis; they differ on myocardial infarction and on the strength of the evidence for OCT alone.
Guideline recommendations for intravascular imaging
2025 ACC/AHA/ACEP/NAEMSP/SCAI acute coronary syndromes guideline. In patients with acute coronary syndromes, intravascular imaging to guide PCI is a Class 1, Level of Evidence A recommendation (Rao 2025; ACC 2025). The 2021 revascularization guideline gave IVUS guidance of stent implantation a Class 2a recommendation (Lawton 2022). Imaging in NSTE-ACS and during primary PCI for ST-elevation myocardial infarction is summarized with those conditions.
2021 ACC/AHA/SCAI coronary revascularization guideline (Lawton 2022):
- In patients undergoing coronary stent implantation, IVUS can be useful for procedural guidance, particularly in left main or complex coronary stenting, to reduce ischemic events (Class 2a, Level of Evidence B-R).
- In patients undergoing coronary stent implantation, OCT is a reasonable alternative to IVUS for procedural guidance, except in ostial left main disease (Class 2a, B-R).
- In patients with stent failure, IVUS or OCT is reasonable to determine the mechanism of stent failure (Class 2a, C-LD).
- In patients with intermediate stenosis of the left main artery, IVUS is reasonable to help define lesion severity (Class 2a, B-NR).
| Indication | 2011 ACCF/AHA/SCAI PCI | 2021 ACC/AHA/SCAI revascularization | 2025 ACC/AHA ACS |
|---|---|---|---|
| Guidance of stent implantation | IIb (B), particularly left main | 2a (B-R) for IVUS, particularly left main or complex; 2a (B-R) for OCT except ostial left main | 1 (A) in ACS |
| Intermediate left main stenosis | IIa (B) | 2a (B-NR) for IVUS | — |
| Intermediate non-left main stenosis | IIb (B) | — | — |
| Mechanism of stent failure | IIa (C) restenosis; IIb (C) thrombosis | 2a (C-LD), IVUS or OCT | — |
| Transplant vasculopathy | IIa (B) at 4–6 weeks and 1 year | — | — |
| Routine assessment without planned revascularization | III (no benefit) | — | — |
Class of recommendation and level of evidence for intravascular imaging across US guidelines. 2011 classes as reproduced in the ACC Coronary Interventions Handbook (Velagapudi 2021); 2021 classes from the AHA guideline slide set (Lawton 2022); 2025 class from the ACC press release (ACC 2025). A dash indicates no recommendation in that document on that question. ACS, acute coronary syndromes.
American and European guidelines endorse intracoronary imaging to guide left main bifurcation PCI in both acute and chronic coronary syndromes (Johnson 2026). For chronic coronary syndromes, the applicable US recommendation is Class 2a in the 2021 revascularization guideline.
Choosing IVUS or OCT
The randomized evidence gives similar clinical results for the two modalities: OCT achieved a minimal stent area noninferior to IVUS in ILUMIEN III (Ali 2016), clinical outcomes were noninferior in OCTIVUS (Kang 2023), and outcomes were similar across modalities in the network meta-analysis (Stone 2024). The choice therefore follows the question, the anatomy, and the patient.
| Clinical situation | Preferred modality | Reason |
|---|---|---|
| Ostial or intermediate left main stenosis | IVUS | OCT cannot clear the aorto-ostial junction; left main thresholds derived with IVUS |
| Advanced kidney disease or contrast limits | IVUS | No flush; supports zero-contrast strategies |
| Large vessel, large plaque burden, or vessel sizing by EEM | IVUS | Penetration shows the EEM behind plaque |
| Chronic total occlusion crossing and re-entry | IVUS | Wire position within plaque or subintimal space; no flush |
| Calcium thickness and lesion preparation | OCT | Measures thickness; calcium score |
| ACS mechanism (rupture, erosion, nodule) and thrombus | OCT | Resolves the cap and thrombus type |
| Malapposition, protrusion, edge dissection, strut coverage | OCT | Strut-level resolution |
| Stent failure mechanism, neoatherosclerosis | OCT (IVUS acceptable) | Neointimal tissue characterization |
| Bioresorbable scaffolds | OCT | Polymeric struts barely visible on IVUS |
| SCAD with uncertain angiographic diagnosis | OCT, or IVUS for hematoma depth | OCT defines the membrane; IVUS avoids flush pressure |
Choice between IVUS and OCT by clinical situation. Data from Lawton 2022, Ben-Assa 2021, Velagapudi 2021, Ono 2020, and Alasnag 2022. ACS, acute coronary syndrome; EEM, external elastic membrane; SCAD, spontaneous coronary artery dissection.
Intravascular imaging and coronary physiology
Imaging describes anatomy and physiology describes flow. For intermediate non-left main stenoses, the decision to revascularize rests on FFR or a nonhyperemic ratio, and imaging thresholds are a second-line guide (coronary physiology). FLAVOUR showed that an IVUS-guided decision strategy leads to more PCI than FFR with similar long-term hard outcomes (Koo 2022; Yang 2025). Imaging and physiology answer complementary questions after PCI as well: in DOCTORS, OCT-guided optimization raised the final FFR (Meneveau 2016). A negative FFR does not exclude high-risk plaque: in COMBINE OCT-FFR, among 390 patients with diabetes and at least one FFR-negative lesion, the 25% with an OCT thin-cap fibroatheroma had an 18-month event rate of 13.3% against 3.1% (hazard ratio 4.65) (Kedhi 2021).
Imaging-derived physiology. Computational methods estimate FFR from the lumen geometry of an imaging pullback. The optical flow ratio, computed from OCT, agreed with wire FFR in 90% of 626 vessels in an individual-patient meta-analysis (sensitivity 84%, specificity 94%); slow pullback speed was associated with overestimation (Hu 2023). Angiography-derived FFR is covered with coronary physiology.
Co-registration, three-dimensional reconstruction, hybrid catheters, and automated analysis
Co-registration. Angiographic co-registration displays the imaging frame on the angiogram and is used to place stent edges in the landing zones chosen on imaging (Ben-Assa 2021).
Three-dimensional reconstruction. OCT consoles reconstruct the stent and lumen in three dimensions online. The reconstruction displays strut position across a side-branch ostium after bifurcation stenting and shows malapposed struts in color. Three-dimensional IVUS reconstruction requires laborious segmentation (Ono 2020; Gupta 2022).
Hybrid IVUS-OCT catheters. Combined catheters acquire IVUS and OCT in one pullback, avoiding the landmark matching required between separate pullbacks. One system (Novasight Hybrid) arranges the transducer and lens collinearly, with a 1.7 French tip, a 6 mm IVUS field-of-view radius, a 100 mm maximum pullback, and US Food and Drug Administration 510(k) clearance. Another (Dual Sensor) places the IVUS transducer and lens in sequence 0.3 mm apart, with axial resolutions of 120 μm for IVUS and 20 μm for OCT and pullbacks up to 40 mm/s (Ono 2020). These catheters combine depth (vessel size, plaque burden, remodeling) with surface detail (cap thickness, thrombus, strut apposition) in co-registered frames.
Automated analysis and artificial intelligence. OCT consoles detect the lumen automatically and color-code strut apposition (Ben-Assa 2021). Deep-learning segmentation of IVUS, trained on 1,063 patients, delineated lumen and EEM with Dice coefficients of 0.93 and 0.97 and correlated with expert measurements (r = 0.92 for MLA and 0.88 for EEM area). Segmentation of a 20 mm lesion took 13.2 seconds against an average of 187.5 minutes manually (Jeong 2024). Automated analysis shortens measurement and may reduce the interobserver variability of derived indices such as stent expansion (Blessing 1999).
Vulnerable plaque: natural-history studies and preventive PCI
PROSPECT. In 697 patients with acute coronary syndromes who underwent three-vessel gray-scale and radiofrequency IVUS after PCI, 3-year major adverse events were 20.4%, attributed equally to culprit (12.9%) and nonculprit (11.6%) lesions. Nonculprit lesions causing events were angiographically mild (mean diameter stenosis 32.3%) but more often had a plaque burden of 70% or more (hazard ratio 5.03), an MLA of 4.0 mm² or less (hazard ratio 3.21), or a thin-cap fibroatheroma on radiofrequency IVUS (hazard ratio 3.35) (Stone 2011).
LRP and PROSPECT II. The LRP study related NIRS lipid content to nonculprit events in 1,271 patients (NIRS) (Waksman 2019). PROSPECT II used combined NIRS and IVUS to image 3,629 nonculprit lesions in 898 patients after myocardial infarction. Over a median of 3.7 years, 8.0% had events from untreated nonculprit lesions with a mean angiographic stenosis of 46.9%. Highly lipidic lesions (adjusted odds ratio 7.83 at the lesion level) and large plaque burden were independent predictors, and lesions with both features had a 4-year event rate of 7.0% (Erlinge 2021).
OCT studies. In CLIMA, OCT of the untreated proximal LAD in 1,003 patients identified MLA below 3.5 mm² (hazard ratio 2.1), fibrous cap below 75 μm (4.7), lipid arc above 180° (2.4), and OCT-defined macrophages (2.7) as predictors of cardiac death or target-segment infarction at 1 year; all four together carried a hazard ratio of 7.54 (Prati 2020). COMBINE OCT-FFR linked thin-cap fibroatheroma to events in patients with diabetes despite normal FFR (Kedhi 2021).
Preventive PCI. PREVENT randomized 1,606 patients with non-flow-limiting (FFR above 0.80) vulnerable plaques identified by intracoronary imaging to PCI plus optimal medical therapy or medical therapy alone. At 2 years, cardiac death, target-vessel infarction, ischemia-driven target-vessel revascularization, or hospitalization for unstable or progressive angina occurred in 0.4% and 3.4% (absolute difference −3.0 percentage points). Death (0.5% against 1.3%) and myocardial infarction (1.1% against 1.7%) did not differ significantly (Park 2024).
| Study | Imaging | Population | High-risk features and result |
|---|---|---|---|
| PROSPECT (Stone 2011) | Gray-scale and radiofrequency IVUS | 697, ACS | Plaque burden 70% or more (HR 5.03), MLA 4.0 mm² or less (HR 3.21), thin-cap fibroatheroma (HR 3.35) |
| LRP (Waksman 2019) | NIRS-IVUS | 1,271, catheterization | maxLCBI4mm above 400: segment HR 3.39 (adjusted) |
| PROSPECT II (Erlinge 2021) | NIRS-IVUS | 898, recent MI | Highly lipidic lesions (lesion-level OR 7.83) and large plaque burden; both features 7.0% at 4 years |
| CLIMA (Prati 2020) | OCT, proximal LAD | 1,003 | MLA below 3.5 mm², cap below 75 μm, lipid arc above 180°, macrophages; all four HR 7.54 |
| COMBINE OCT-FFR (Kedhi 2021) | OCT after FFR | 390 with diabetes and FFR-negative lesions | Thin-cap fibroatheroma: 13.3% vs 3.1% at 18 months (HR 4.65) |
| PREVENT (Park 2024) | Intracoronary imaging | 1,606, FFR above 0.80 | Preventive PCI 0.4% vs medical therapy 3.4% at 2 years |
Natural-history and treatment studies of high-risk coronary plaque. ACS, acute coronary syndrome; FFR, fractional flow reserve; HR, hazard ratio; LAD, left anterior descending artery; maxLCBI4mm, maximum lipid core burden index in 4 mm; MI, myocardial infarction; MLA, minimal lumen area; NIRS, near-infrared spectroscopy; OR, odds ratio.
Plaque progression and regression: percent atheroma volume
Serial IVUS of a matched coronary segment measures atheroma volume and is the standard end point of plaque progression trials. Percent atheroma volume is the total plaque-plus-media volume divided by the total EEM volume across the analyzed segment. In GLAGOV, 968 statin-treated patients were randomized to evolocumab or placebo for 76 weeks. Mean low-density lipoprotein (LDL) cholesterol was 36.6 mg/dL with evolocumab and 93.0 mg/dL with placebo, and percent atheroma volume decreased by 0.95% with evolocumab and increased by 0.05% with placebo (difference −1.0%) (Nicholls 2016). GLAGOV showed regression of coronary atheroma at very low LDL cholesterol; the clinical benefit of the drugs rests on outcome trials (LDL-lowering therapy).
Intravascular imaging in spontaneous coronary artery dissection, MINOCA, and transplant vasculopathy
Spontaneous coronary artery dissection. OCT shows the intimal-medial membrane, false lumen, and intramural hematoma, and IVUS shows hematoma depth without a flush; imaging is reserved for uncertain diagnoses because it can extend the dissection (SCAD imaging).
MINOCA. OCT or IVUS of the involved territory at the index angiogram identifies rupture, erosion, calcified nodule, layered plaque, or intramural hematoma, and cardiac MRI completes the evaluation (MINOCA algorithm).
Cardiac allograft vasculopathy. IVUS detects intimal thickening before luminal narrowing. In a multicenter study of 125 recipients, an increase in maximal intimal thickness of 0.5 mm or more in the first year after transplantation was associated with death or graft loss at 5 years (20.8% against 5.9%) (Kobashigawa 2005). The 2011 ACCF/AHA/SCAI PCI guideline gave a Class IIa recommendation to IVUS at 4–6 weeks and 1 year after transplantation to exclude donor coronary disease, detect rapidly progressive vasculopathy, and provide prognostic information (Velagapudi 2021) (invasive surveillance after heart transplantation).
Peripheral and venous intravascular ultrasound
Low-frequency IVUS (8–20 MHz) images the aorta, inferior vena cava, iliac veins, carotid arteries, and femoropopliteal arteries (Velagapudi 2021). In the VIDIO trial of 100 patients with advanced chronic venous disease (clinical class C4–C6) at 14 US and European sites, venography identified iliofemoral obstruction in 51 patients and IVUS in 81. Venography underestimated diameter stenosis by 11% on average, IVUS found significant lesions missed by three-view venography in 26.3%, and IVUS changed the treatment plan in 57 patients (Gagne 2017). An international expert consensus rated IVUS appropriate during all phases of iliofemoral venous intervention and of tibial arterial intervention, and during most iliac and femoropopliteal arterial scenarios (Secemsky 2022).
Utilization of intravascular imaging in the United States
Among 1,189,470 Medicare beneficiaries undergoing PCI, intravascular imaging was used in 10.5% of procedures overall, rising from 9.5% in 2013 to 15.4% in 2019. Use varied widely among operators (median operator use 3.92%; interquartile range 0.36–12.82%). Imaging use was associated with lower adjusted 1-year mortality (hazard ratio 0.96), myocardial infarction (0.97), repeat PCI (0.74), and major adverse events (0.85), with no association with falsification (negative-control) end points (Fazel 2023).
Sources of error in intravascular imaging
- Measuring a nondiagnostic OCT run. Residual blood mimics thrombus, attenuates the wall, and distorts struts; the acquisition is repeated after correcting guide position or flush.
- Applying IVUS sizing values to OCT. OCT often cannot see the EEM in diseased segments and reports smaller areas; thresholds are modality-specific.
- Using OCT for the ostial left main. The aorto-ostial junction does not clear, and the 2021 guideline excludes ostial left main disease from OCT guidance.
- Treating a small MLA as proof of ischemia. MLA thresholds have high negative and low positive predictive value; physiology decides revascularization of intermediate non-left main lesions.
- Applying Korean left main cutoffs universally. Thresholds of 4.5–4.8 mm² were derived in Korean patients; 6 mm² was the prospectively validated deferral value in LITRO.
- Imaging without nitroglycerin. Spasm narrows the lumen and EEM and leads to undersizing.
- Sizing from the proximal reference in a tapering vessel. The distal reference determines stent diameter, and the proximal segment is post-dilated to its own reference.
- Measuring calcium thickness on IVUS. The acoustic shadow hides the depth of calcium; OCT measures thickness.
- Accepting angiographic expansion. A smooth angiographic lumen can hide underexpansion; expansion is confirmed by MSA and relative expansion.
- Measuring stent expansion against a single frame. Expansion depends on the chosen reference; observers disagree on adequacy at the 90% threshold in about a quarter of stents.
- Treating all OCT edge dissections. Small intimal flaps are common; treatment is reserved for medial extension, arcs of 60° or more, lengths of 3 mm or more, or a compromised lumen.
- Post-dilating any malapposed strut. Isolated acute malapposition below 200–250 μm has no established clinical significance; long malapposed segments are treated.
- Misreading the guidewire shadow as a side branch. The shadow is recognized by its position behind the wire and its continuity across frames.
- Misreading superficial macrophages as a thin cap. Macrophage attenuation can make a fibrous plaque resemble a thin-cap fibroatheroma.
- Mistaking nonuniform rotational distortion for eccentric plaque. Distortion from drive-cable binding stretches and compresses sectors of the image.
- Imaging a SCAD lesion that will be managed conservatively. Catheter passage and the OCT flush can extend the dissection.
- Reporting stent thrombosis without imaging the stent. OCT identifies a mechanism in nearly all cases, and the mechanism changes treatment.
- Advancing an imaging catheter against resistance. The catheter is withdrawn and the lesion predilated to avoid dissection or entrapment.
- Interpreting a negative imaging trial as a negative modality. ILUMIEN IV enlarged the stent without reducing target-vessel failure, while OCTOBER, OCCUPI, and pooled analyses showed clinical benefit in complex lesions.
References
- Rao SV, O'Donoghue ML, Ruel M, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI Guideline for the Management of Patients With Acute Coronary Syndromes: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2025;85:2135–2237. Also published in Circulation 2025;151:e771–e862 (corrections: J Am Coll Cardiol 2025;85:1800 and 2025;86:2723). doi:10.1016/j.jacc.2024.11.009
- Lawton JS, Tamis-Holland JE, Bangalore S, et al. 2021 ACC/AHA/SCAI Guideline for Coronary Artery Revascularization: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2022;79:e21–e129. Also published in Circulation 2022;145:e18–e114 (corrections: J Am Coll Cardiol 2022;79:1547 and 2024;84:771). Recommendation text verified in the AHA guideline slide set. doi:10.1016/j.jacc.2021.09.006
- Räber L, Mintz GS, Koskinas KC, et al. Clinical use of intracoronary imaging. Part 1: guidance and optimization of coronary interventions. An expert consensus document of the European Association of Percutaneous Cardiovascular Interventions. Eur Heart J. 2018;39:3281–3300. Correction: Eur Heart J 2019;40:308. doi:10.1093/eurheartj/ehy285
- American College of Cardiology. ACC, AHA issue new acute coronary syndromes guideline (press release). 2025. acc.org
- Secemsky EA, Mosarla RC, Rosenfield K, et al. Appropriate Use of Intravascular Ultrasound During Arterial and Venous Lower Extremity Interventions. JACC Cardiovasc Interv. 2022;15:1558–1568. doi:10.1016/j.jcin.2022.04.034
- Hong SJ, Kim BK, Shin DH, et al. Effect of Intravascular Ultrasound-Guided vs Angiography-Guided Everolimus-Eluting Stent Implantation: The IVUS-XPL Randomized Clinical Trial. JAMA. 2015;314:2155–2163. doi:10.1001/jama.2015.15454
- Hong SJ, Mintz GS, Ahn CM, et al. Effect of Intravascular Ultrasound-Guided Drug-Eluting Stent Implantation: 5-Year Follow-Up of the IVUS-XPL Randomized Trial. JACC Cardiovasc Interv. 2020;13:62–71. doi:10.1016/j.jcin.2019.09.033
- Zhang J, Gao X, Kan J, et al. Intravascular Ultrasound Versus Angiography-Guided Drug-Eluting Stent Implantation: The ULTIMATE Trial. J Am Coll Cardiol. 2018;72:3126–3137. doi:10.1016/j.jacc.2018.09.013
- Gao XF, Ge Z, Kong XQ, et al. 3-Year Outcomes of the ULTIMATE Trial Comparing Intravascular Ultrasound Versus Angiography-Guided Drug-Eluting Stent Implantation. JACC Cardiovasc Interv. 2021;14:247–257. doi:10.1016/j.jcin.2020.10.001
- Lee JM, Choi KH, Song YB, et al. Intravascular Imaging-Guided or Angiography-Guided Complex PCI. N Engl J Med. 2023;388:1668–1679. doi:10.1056/NEJMoa2216607
- Li X, Ge Z, Kan J, et al. Intravascular ultrasound-guided versus angiography-guided percutaneous coronary intervention in acute coronary syndromes (IVUS-ACS): a two-stage, multicentre, randomised trial. Lancet. 2024;403:1855–1865. doi:10.1016/S0140-6736(24)00282-4
- Kim BK, Shin DH, Hong MK, et al. Clinical Impact of Intravascular Ultrasound-Guided Chronic Total Occlusion Intervention With Zotarolimus-Eluting Versus Biolimus-Eluting Stent Implantation: Randomized Study. Circ Cardiovasc Interv. 2015;8:e002592. doi:10.1161/CIRCINTERVENTIONS.115.002592
- Meneveau N, Souteyrand G, Motreff P, et al. Optical Coherence Tomography to Optimize Results of Percutaneous Coronary Intervention in Patients with Non-ST-Elevation Acute Coronary Syndrome: Results of the Multicenter, Randomized DOCTORS Study (Does Optical Coherence Tomography Optimize Results of Stenting). Circulation. 2016;134:906–917. doi:10.1161/CIRCULATIONAHA.116.024393
- Ali ZA, Maehara A, Généreux P, et al. Optical coherence tomography compared with intravascular ultrasound and with angiography to guide coronary stent implantation (ILUMIEN III: OPTIMIZE PCI): a randomised controlled trial. Lancet. 2016;388:2618–2628. doi:10.1016/S0140-6736(16)31922-5
- Ali ZA, Landmesser U, Maehara A, et al. Optical Coherence Tomography-Guided versus Angiography-Guided PCI. N Engl J Med. 2023;389:1466–1476. doi:10.1056/NEJMoa2305861
- Holm NR, Andreasen LD, Neghabat O, et al. OCT or Angiography Guidance for PCI in Complex Bifurcation Lesions. N Engl J Med. 2023;389:1477–1487. doi:10.1056/NEJMoa2307770
- Hong SJ, Lee SJ, Lee SH, et al. Optical coherence tomography-guided versus angiography-guided percutaneous coronary intervention for patients with complex lesions (OCCUPI): an investigator-initiated, multicentre, randomised, open-label, superiority trial in South Korea. Lancet. 2024;404:1029–1039. doi:10.1016/S0140-6736(24)01454-5
- Kang DY, Ahn JM, Yun SC, et al. Optical Coherence Tomography-Guided or Intravascular Ultrasound-Guided Percutaneous Coronary Intervention: The OCTIVUS Randomized Clinical Trial. Circulation. 2023;148:1195–1206. doi:10.1161/CIRCULATIONAHA.123.066429
- Stone GW, Christiansen EH, Ali ZA, et al. Intravascular imaging-guided coronary drug-eluting stent implantation: an updated network meta-analysis. Lancet. 2024;403:824–837. doi:10.1016/S0140-6736(23)02454-6
- Giacoppo D, Laudani C, Occhipinti G, et al. Coronary Angiography, Intravascular Ultrasound, and Optical Coherence Tomography for Guiding of Percutaneous Coronary Intervention: A Systematic Review and Network Meta-Analysis. Circulation. 2024;149:1065–1086. doi:10.1161/CIRCULATIONAHA.123.067583
- Koo BK, Hu X, Kang J, et al. Fractional Flow Reserve or Intravascular Ultrasonography to Guide PCI. N Engl J Med. 2022;387:779–789. doi:10.1056/NEJMoa2201546
- Yang S, Hu X, Zhang J, et al. Long-Term Outcomes After Fractional Flow Reserve vs Intravascular Ultrasound to Guide PCI: The FLAVOUR Trial Extended Follow-Up. J Am Coll Cardiol. 2025;86:593–606. doi:10.1016/j.jacc.2025.06.042
- Stone GW, Maehara A, Lansky AJ, et al. A prospective natural-history study of coronary atherosclerosis. N Engl J Med. 2011;364:226–235. doi:10.1056/NEJMoa1002358
- Waksman R, Di Mario C, Torguson R, et al. Identification of patients and plaques vulnerable to future coronary events with near-infrared spectroscopy intravascular ultrasound imaging: a prospective, cohort study. Lancet. 2019;394:1629–1637. doi:10.1016/S0140-6736(19)31794-5
- Erlinge D, Maehara A, Ben-Yehuda O, et al. Identification of vulnerable plaques and patients by intracoronary near-infrared spectroscopy and ultrasound (PROSPECT II): a prospective natural history study. Lancet. 2021;397:985–995. doi:10.1016/S0140-6736(21)00249-X
- Prati F, Romagnoli E, Gatto L, et al. Relationship between coronary plaque morphology of the left anterior descending artery and 12 months clinical outcome: the CLIMA study. Eur Heart J. 2020;41:383–391. Corrigendum: Eur Heart J 2020;41:393. doi:10.1093/eurheartj/ehz520
- Kedhi E, Berta B, Roleder T, et al. Thin-cap fibroatheroma predicts clinical events in diabetic patients with normal fractional flow reserve: the COMBINE OCT-FFR trial. Eur Heart J. 2021;42:4671–4679. doi:10.1093/eurheartj/ehab433
- Park SJ, Ahn JM, Kang DY, et al. Preventive percutaneous coronary intervention versus optimal medical therapy alone for the treatment of vulnerable atherosclerotic coronary plaques (PREVENT): a multicentre, open-label, randomised controlled trial. Lancet. 2024;403:1753–1765. doi:10.1016/S0140-6736(24)00413-6
- Nicholls SJ, Puri R, Anderson T, et al. Effect of Evolocumab on Progression of Coronary Disease in Statin-Treated Patients: The GLAGOV Randomized Clinical Trial. JAMA. 2016;316:2373–2384. doi:10.1001/jama.2016.16951
- de la Torre Hernandez JM, Hernández Hernandez F, Alfonso F, et al. Prospective application of pre-defined intravascular ultrasound criteria for assessment of intermediate left main coronary artery lesions results from the multicenter LITRO study. J Am Coll Cardiol. 2011;58:351–358. doi:10.1016/j.jacc.2011.02.064
- Park SJ, Ahn JM, Kang SJ, et al. Intravascular ultrasound-derived minimal lumen area criteria for functionally significant left main coronary artery stenosis. JACC Cardiovasc Interv. 2014;7:868–874. doi:10.1016/j.jcin.2014.02.015
- Kang SJ, Lee JY, Ahn JM, et al. Validation of intravascular ultrasound-derived parameters with fractional flow reserve for assessment of coronary stenosis severity. Circ Cardiovasc Interv. 2011;4:65–71. doi:10.1161/CIRCINTERVENTIONS.110.959148
- Kang SJ, Lee JY, Ahn JM, et al. Intravascular ultrasound-derived predictors for fractional flow reserve in intermediate left main disease. JACC Cardiovasc Interv. 2011;4:1168–1174. doi:10.1016/j.jcin.2011.08.009
- Waksman R, Legutko J, Singh J, et al. FIRST: Fractional Flow Reserve and Intravascular Ultrasound Relationship Study. J Am Coll Cardiol. 2013;61:917–923. doi:10.1016/j.jacc.2012.12.012
- Kang SJ, Ahn JM, Song H, et al. Comprehensive intravascular ultrasound assessment of stent area and its impact on restenosis and adverse cardiac events in 403 patients with unprotected left main disease. Circ Cardiovasc Interv. 2011;4:562–569. doi:10.1161/CIRCINTERVENTIONS.111.964643
- Fujino A, Mintz GS, Matsumura M, et al. A new optical coherence tomography-based calcium scoring system to predict stent underexpansion. EuroIntervention. 2018;13:e2182–e2189. doi:10.4244/EIJ-D-17-00962
- Hill JM, Kereiakes DJ, Shlofmitz RA, et al. Intravascular Lithotripsy for Treatment of Severely Calcified Coronary Artery Disease. J Am Coll Cardiol. 2020;76:2635–2646. doi:10.1016/j.jacc.2020.09.603
- Jia H, Abtahian F, Aguirre AD, et al. In vivo diagnosis of plaque erosion and calcified nodule in patients with acute coronary syndrome by intravascular optical coherence tomography. J Am Coll Cardiol. 2013;62:1748–1758. doi:10.1016/j.jacc.2013.05.071
- Jia H, Dai J, Hou J, et al. Effective anti-thrombotic therapy without stenting: intravascular optical coherence tomography-based management in plaque erosion (the EROSION study). Eur Heart J. 2017;38:792–800. doi:10.1093/eurheartj/ehw381
- Soeda T, Uemura S, Park SJ, et al. Incidence and Clinical Significance of Poststent Optical Coherence Tomography Findings: One-Year Follow-Up Study From a Multicenter Registry. Circulation. 2015;132:1020–1029. doi:10.1161/CIRCULATIONAHA.114.014704
- Souteyrand G, Amabile N, Mangin L, et al. Mechanisms of stent thrombosis analysed by optical coherence tomography: insights from the national PESTO French registry. Eur Heart J. 2016;37:1208–1216. doi:10.1093/eurheartj/ehv711
- Kuramitsu S, Iwabuchi M, Haraguchi T, et al. Incidence and clinical impact of stent fracture after everolimus-eluting stent implantation. Circ Cardiovasc Interv. 2012;5:663–671. doi:10.1161/CIRCINTERVENTIONS.112.969238
- Jia R, Nie X, Li H, et al. Impact of attenuated plaques on TIMI grade flow and clinical outcomes of coronary artery disease patients: a systematic review and meta analysis. J Thorac Dis. 2016;8:527–536. doi:10.21037/jtd.2016.02.51
- Mariani J Jr, Guedes C, Soares P, et al. Intravascular ultrasound guidance to minimize the use of iodine contrast in percutaneous coronary intervention: the MOZART (Minimizing cOntrast utiliZation With IVUS Guidance in coRonary angioplasTy) randomized controlled trial. JACC Cardiovasc Interv. 2014;7:1287–1293. doi:10.1016/j.jcin.2014.05.024
- Gagne PJ, Tahara RW, Fastabend CP, et al. Venography versus intravascular ultrasound for diagnosing and treating iliofemoral vein obstruction. J Vasc Surg Venous Lymphat Disord. 2017;5:678–687. doi:10.1016/j.jvsv.2017.04.007
- Ge J, Jeremias A, Rupp A, et al. New signs characteristic of myocardial bridging demonstrated by intracoronary ultrasound and Doppler. Eur Heart J. 1999;20:1707–1716. doi:10.1053/euhj.1999.1661
- van der Sijde JN, Karanasos A, van Ditzhuijzen NS, et al. Safety of optical coherence tomography in daily practice: a comparison with intravascular ultrasound. Eur Heart J Cardiovasc Imaging. 2017;18:467–474. doi:10.1093/ehjci/jew037
- Hu F, Ding D, Westra J, et al. Diagnostic accuracy of optical flow ratio: an individual patient-data meta-analysis. EuroIntervention. 2023;19:e145–e154. doi:10.4244/EIJ-D-22-01098
- Fazel R, Yeh RW, Cohen DJ, et al. Intravascular imaging during percutaneous coronary intervention: temporal trends and clinical outcomes in the USA. Eur Heart J. 2023;44:3845–3855. doi:10.1093/eurheartj/ehad430
- Blessing E, Hausmann D, Sturm M, et al. Intravascular ultrasound and stent implantation: intraobserver and interobserver variability. Am Heart J. 1999;137:368–371. doi:10.1053/hj.1999.v137.93032
- Jeong GJ, Lee G, Lee JG, et al. Deep Learning-Based Lumen and Vessel Segmentation of Intravascular Ultrasound Images in Coronary Artery Disease. Korean Circ J. 2024;54:30–39. doi:10.4070/kcj.2023.0166
- Velagapudi P, Abbott JD. Intracoronary imaging-I: step-by-step technique for intravascular ultrasound (IVUS). In: Coronary Interventions Handbook, Chapter 12. American College of Cardiology; 2021. acc.org
- Ben-Assa E, Young MN, Jang IK. Step-by-step approach to intracoronary imaging-II: optical coherence tomography (OCT). In: Coronary Interventions Handbook, Chapter 13. American College of Cardiology; 2021. acc.org
- Paul TK, Kerrigan JL, Bagai J. Practical aspects of IVUS-guided percutaneous coronary intervention. Society for Cardiovascular Angiography & Interventions, Quality Improvement Tip. 2022. scai.org
- Ono M, Kawashima H, Hara H, et al. Advances in IVUS/OCT and future clinical perspective of novel hybrid catheter system in coronary imaging. Front Cardiovasc Med. 2020;7:119. doi:10.3389/fcvm.2020.00119
- Alasnag M, Ahmed W, Al-Bawardy R, et al. Optimising PCI by intracoronary image-guidance. Front Cardiovasc Med. 2022;9:878801. doi:10.3389/fcvm.2022.878801
- Gupta A, Shrivastava A, Vijayvergiya R, et al. Optical coherence tomography: an eye into the coronary artery. Front Cardiovasc Med. 2022;9:854554. doi:10.3389/fcvm.2022.854554
- Tsigkas G, Spyropoulou P, Bousoula E, et al. Intracoronary imaging: current practice and future perspectives. Rev Cardiovasc Med. 2023;24(2):39. doi:10.31083/j.rcm2402039
- Kobashigawa JA, Tobis JM, Starling RC, et al. Multicenter intravascular ultrasound validation study among heart transplant recipients: outcomes after five years. J Am Coll Cardiol. 2005;45:1532–1537. doi:10.1016/j.jacc.2005.02.035
- Kang SJ, Ahn JM, Han S, et al. Sex differences in the visual-functional mismatch between coronary angiography or intravascular ultrasound versus fractional flow reserve. JACC Cardiovasc Interv. 2013;6:562–568. doi:10.1016/j.jcin.2013.02.016
- Johnson TW, Gonzalo N, de la Torre Hernandez JM, et al. Intracoronary imaging for left main percutaneous coronary intervention: a clinical consensus statement of the European Association of Percutaneous Cardiovascular Interventions (EAPCI) of the ESC and the European Bifurcation Club (EBC). Eur Heart J. 2026;47:5247–5265. doi:10.1093/eurheartj/ehag353