CardioBoardExam.com

Library › Devices and mechanical support › Cardiac device follow-up

Cardiac device follow-up: interrogation, remote monitoring, perioperative care, and advisories

Follow-up of pacemakers, defibrillators, resynchronization devices, and loop recorders from implantation to battery depletion, including remote monitoring, perioperative and procedural management, imaging, radiation, and recalls.

Remote monitoring is part of the standard of care for all patients with a pacemaker, defibrillator, resynchronization device, or loop recorder (Class 1, 2023 consensus of the Heart Rhythm Society [HRS], European Heart Rhythm Association [EHRA], Asia Pacific Heart Rhythm Society [APHRS], and Latin American Heart Rhythm Society [LAHRS]). For devices without continuous connectivity, scheduled transmissions run at intervals no longer than 3 to 12 months for pacemakers and 3 to 6 months for defibrillators, and at intervals of 1 to 3 months as the battery nears replacement. Around surgery, four facts decide the plan: the device type, pacing dependence, the magnet response, and where electrosurgical current will flow. A magnet suspends defibrillator therapy but does not make defibrillator pacing asynchronous.

Devices covered by follow-up and what the evaluation protects

Cardiac implantable electronic devices (CIEDs) are pacemakers, cardiac resynchronization therapy pacemakers (CRT-P), cardiac resynchronization therapy defibrillators (CRT-D), transvenous, subcutaneous, and extravascular implantable cardioverter-defibrillators (ICDs), leadless pacemakers, and insertable cardiac monitors (also called implantable loop recorders). Implanted hemodynamic sensors and wearable defibrillators share the remote-monitoring infrastructure but are not pacing or defibrillation systems. Follow-up has five purposes: confirming that the device delivers therapy correctly, detecting hardware failure of the generator, battery, and leads, timing generator replacement, recording the patient's rhythm and heart failure status between visits, and keeping the device programmed for the patient's current condition, goals, and planned procedures.

The questions that follow-up answers depend on the device. A pacing-dependent patient with complete heart block depends on uninterrupted capture. A defibrillator patient depends on correct detection and on not receiving shocks for noise or supraventricular rhythms. A resynchronization patient depends on near-continuous biventricular capture. A loop recorder patient needs an arrhythmia diagnosis, and false detections add workload without benefit.

DeviceTherapiesMain follow-up concernsMagnet behavior (typical)
Single- or dual-chamber pacemakerBradycardia pacing, rate responseCapture, sensing, lead integrity, battery, pacing burden, atrial high-rate episodesAsynchronous pacing at a manufacturer-specific rate
CRT-PBiventricular or conduction system pacingPercentage of effective biventricular pacing, left ventricular threshold, phrenic stimulation, heart failure diagnosticsAsynchronous pacing
Transvenous ICD and CRT-DPacing, antitachycardia pacing, shocksDetection, appropriate and inappropriate therapy, high-voltage impedance, charge time, lead noiseTachyarrhythmia detection and therapy suspended; pacing unchanged
Subcutaneous ICDShocks; brief post-shock pacing onlyT-wave and noise oversensing, battery advisoriesShocks suspended; tone with magnet
Extravascular (substernal) ICDAntitachycardia pacing, shocks, pause-prevention pacingSensing, therapy deliveryTherapy suspended; pause prevention continues
Leadless pacemakerSingle-chamber or atrioventricular synchronous pacingCapture threshold, battery, atrioventricular synchrony algorithmsModel-specific; some models have no magnet response
Insertable cardiac monitorNone (diagnostic)True versus false detections, battery, symptom-rhythm correlationNone clinically relevant

Device types and the concerns that shape their follow-up. Magnet behavior varies by manufacturer, model, and programming and is confirmed from the device report; manufacturer details are given in the magnet section below.

Device design, pacing modes, and troubleshooting of paced rhythms are covered on the pacemaker hub, defibrillator systems and programming on the ICD hub, resynchronization on the cardiac resynchronization hub, and temporary pacing on the temporary pacing hub. Pacing indications are summarized on the bradycardia and atrioventricular block hub. This page covers what happens after implantation: evaluation, remote surveillance, procedures, imaging, advisories, and the end of device therapy.

Identifying an unknown device: registration card, radiograph, and programmer

Patients receive a device identification card listing the manufacturer, generator model, lead models, and implanting center. The card, the patient's device clinic, and the manufacturer's registry are the first sources. Manufacturers maintain technical-support services that can identify a registered device from the patient's name and date of birth. When none of these is available, as in an unresponsive patient in the emergency department, the chest radiograph identifies the class of device and often the manufacturer.

Radiographic features. A pacemaker has thin pace-sense leads without shocking coils. A transvenous ICD lead carries one or two radiopaque coils, in the right ventricle and often the superior vena cava. A subcutaneous ICD has a lateral thoracic generator with a parasternal subcutaneous lead and no intracardiac hardware. A leadless pacemaker is a capsule in the right ventricle, or in the right atrium for atrial leadless systems. A third lead in a lateral coronary vein indicates a resynchronization device. Many generators carry a radiopaque alphanumeric code on the header or within the can that identifies the manufacturer and model family, for example "SJM" for St. Jude Medical (now Abbott) or "BOS" or "BSC" for Boston Scientific. The CaRDIA-X algorithm classifies the manufacturer from the generator outline, battery shape, and circuit-board layout when no code is legible.

Interrogation. Programmers are manufacturer-specific. Once the manufacturer is known, interrogation with the matching programmer or a manufacturer representative retrieves the model, programmed parameters, and stored data. Some leadless systems require a dedicated programmer or programming module.

SourceWhat it providesLimitations
Device identification cardManufacturer, generator and lead models, implant date and centerOften not carried; may predate a generator change
Device clinic record or remote monitoring portalLast interrogation, programmed mode, pacing dependence, magnet response, battery statusAccess limited to the clinic that follows the patient
Manufacturer technical support and registryIdentification of registered devices; programming and advisory informationRequires patient identifiers; registration may be incomplete
Chest radiographDevice class, lead number and position, coils, abandoned leads, radiopaque codeCode may be illegible; does not show programming
Programmer interrogationDefinitive model, settings, stored events, measured parametersRequires the matching programmer and trained personnel

Sources for identifying a cardiac implantable electronic device. The radiograph also documents abandoned, fractured, and epicardial leads, which change MRI eligibility.

The in-person device evaluation: interrogation, testing, and programming

The HRS/EHRA consensus on monitoring of CIEDs distinguishes three activities. Interrogation retrieves programmed parameters and stored data by telemetry. Programming is a noninvasive, reversible change in operating parameters. A complete evaluation combines interrogation with temporary programming to test function, chiefly capture and sensing thresholds, followed by any permanent programming change. An interrogation-only evaluation retrieves data without testing or changing settings, and it is what most remote transmissions provide.

A complete evaluation follows a fixed order. The underlying rhythm and pacing dependence are recorded first. Battery status and, for defibrillators, capacitor charge time follow. The capture threshold, sensed amplitude, and impedance of all leads are measured and compared with the patient's own trend. Stored episodes and therapies are reviewed with their electrograms. Pacing percentages, rate histograms, sensor behavior, and heart failure diagnostics complete the data. The programmed parameters are then reviewed against the current indication: a new atrioventricular block, new atrial fibrillation, a change in antiarrhythmic drugs, or a fall in ejection fraction may require programming changes. The evaluation ends with a printed or electronic summary and confirmation that the final settings were saved.

ElementPacemaker and CRT-PICD and CRT-DInsertable monitor
Underlying rhythm and pacing dependenceYesYesNot applicable
Battery statusVoltage, impedance, or estimated longevityVoltage, estimated longevity, capacitor charge timeRemaining longevity
Pacing leadsCapture threshold, sensed P and R waves, pacing impedanceSame, plus right ventricular sensing integrityR-wave amplitude, noise
High-voltage leadsNot applicableShock impedance trend, noise counters, lead integrity alertsNot applicable
Stored episodesAtrial high-rate episodes, ventricular high-rate episodes, mode switchesVentricular tachyarrhythmia episodes, antitachycardia pacing and shocks, nonsustained runs, atrial arrhythmia burdenDetected pauses, bradycardia, tachycardia, atrial fibrillation; patient-activated recordings
Pacing percentagesAtrial and ventricular pacing; biventricular pacing in CRT-PSame; biventricular pacing in CRT-DNot applicable
Heart failure diagnosticsActivity, heart rate variability, impedance where availableThoracic impedance, multisensor indices where availableNot applicable
Programming reviewMode, rates, atrioventricular delays, sensor, algorithms that minimize right ventricular pacingDetection zones, discriminators, therapies, alerts, pacing parametersDetection settings, sensitivity, alert programming

Elements of a complete device evaluation by device type, drawn from the HRS/EHRA consensus on monitoring of CIEDs and the 2023 remote device clinic consensus.

Lead measurements: capture threshold, sensed amplitude, and impedance

Capture threshold. The threshold is the lowest output that consistently captures the chamber, expressed in volts at a stated pulse width. It is measured by decrementing output while watching for loss of capture on the surface electrocardiogram and electrogram. Thresholds rise in the first weeks after implantation from inflammation at the electrode-tissue interface and then settle at a chronic value. A late rise suggests microdislodgement, fibrosis (exit block), infarction under the electrode, hyperkalemia or acidosis, or drug effect. Automatic capture algorithms measure threshold periodically and adjust output to a small margin above it.

Sensed amplitude. Intrinsic P-wave and R-wave amplitudes determine whether the programmed sensitivity reliably detects native beats. A falling amplitude threatens undersensing; a lead whose amplitude requires very sensitive programming is more prone to oversensing of far-field signals, myopotentials, and noise. In defibrillators, the right ventricular sensed amplitude also determines detection of ventricular fibrillation, whose electrograms are smaller and more variable than those of sinus rhythm.

Impedance. Pacing impedance reflects the conductor, insulation, electrode, and tissue interface. Normal ranges are specific to the lead model; high-impedance leads read substantially higher than standard leads. An abrupt rise from the patient's baseline indicates conductor fracture or a connection problem, and an abrupt fall indicates insulation failure. High-voltage (shock) impedance is measured with painless low-energy pulses and trended in the same way. Single values within the manufacturer's range do not exclude intermittent fracture, which can first appear as bursts of nonphysiologic noise on stored electrograms.

ParameterChange regarded as significant after MRI (2024 Society for Cardiovascular Magnetic Resonance consensus)Common causes of a change during follow-up
Sensed P-wave amplitudeDecrease ≥50%Atrial lead microdislodgement, atrial fibrosis, atrial arrhythmia at the time of measurement
Sensed R-wave amplitudeDecrease ≥25%Ventricular lead microdislodgement, infarction or scar at the lead tip, cardiomyopathy progression
Capture thresholdIncrease ≥0.5 VExit block, microdislodgement, metabolic or drug effect, perforation
Pacing lead impedanceChange ≥50 ΩRise: conductor fracture, loose set screw. Fall: insulation breach
High-voltage lead impedanceChange ≥3 ΩCoil or conductor failure; externalized conductors in some lead families
Battery voltageDecrease ≥0.04 VNormal depletion; premature depletion under advisory; reset

Thresholds for a significant change in device measurements after MRI, from the 2024 Society for Cardiovascular Magnetic Resonance (SCMR) consensus. The same comparisons with the patient's own baseline apply at routine follow-up, where trend matters more than any fixed cutoff.

Battery status: beginning of life, elective replacement, and end of service

Lithium batteries hold a stable voltage over most of their life and fall predictably near depletion, which permits staged indicators. At beginning of life the device functions with all features. The elective replacement indicator (ERI), also called recommended replacement time, signals that the battery can support normal function for a defined period, usually several months, and generator replacement is scheduled. End of service (end of life) means normal function can no longer be guaranteed. At ERI many devices conserve energy by disabling rate response and diagnostics, reverting dual-chamber pacing to a single-chamber mode, and lowering the pacing or magnet rate to a manufacturer-specific value. A defibrillator's capacitor charge time lengthens as the battery depletes, which delays shock delivery.

Longevity depends on pacing burden, output amplitude and pulse width, lead impedance, rate response, diagnostics, wireless telemetry, and, for defibrillators, the number of capacitor charges. An unexpected fall in estimated longevity is itself a finding and points to high-output pacing, a lead with low impedance, an increase in pacing burden, or premature depletion under a manufacturer advisory. In ECOST, fewer capacitor charges with remote monitoring (499 versus 2,081) were associated with longer battery longevity.

In the prospective REPLACE registry, major complications within 6 months occurred in 4.0% of 1,031 generator replacements without lead addition and in 15.3% of 713 procedures that added a transvenous lead for replacement or upgrade; upgrades to or revisions of resynchronization systems had the highest rate (18.7%). The 6-month infection rate was 1.4% and 1.1% in the two cohorts. These figures also frame decisions about prophylactic replacement of advisory devices. Generator replacement is the moment to reassess the indication, the need for defibrillation, and the patient's goals.

Stored electrograms, arrhythmia episodes, and therapy review

The device's classification of an episode reflects its sensed signals and programmed algorithms, and it is confirmed against the stored electrogram before it guides treatment. Defibrillator therapies are classified as appropriate (ventricular tachycardia or fibrillation), inappropriate (supraventricular rhythm, oversensing, or noise), or unnecessary (appropriate detection of a rhythm that would have ended spontaneously). Appropriate therapy prompts a search for ischemia, heart failure decompensation, electrolyte disturbance, and drug change, and review of antiarrhythmic and ablation options. Inappropriate therapy prompts correction of the rhythm, the sensing problem, the lead, or the programming.

Prognosis after a shock depends on the rhythm treated. In the ALTITUDE survival-by-rhythm analysis of 3,809 patients who survived a first shock, mortality was higher than in matched patients without shocks after first shocks for monomorphic ventricular tachycardia (hazard ratio 1.65), ventricular fibrillation or polymorphic ventricular tachycardia (hazard ratio 2.10), and atrial fibrillation or flutter (hazard ratio 1.61). Mortality after first shocks for sinus tachycardia, other supraventricular tachycardia, or noise and oversensing did not differ from that of patients without shocks. The adverse prognosis therefore reflects the underlying arrhythmia more than the shock itself.

Electrogram patternInterpretationAction
Abrupt onset, regular cycle length, ventricular morphology different from sinus, atrioventricular dissociationMonomorphic ventricular tachycardiaAppropriate therapy; trigger search; review antitachycardia pacing efficacy and antiarrhythmic or ablation options
Irregular ventricular intervals, near-template morphology, rapid irregular atrial electrogramAtrial fibrillation with rapid conductionRate or rhythm control; discriminator and detection-rate review
Gradual onset, 1:1 atrioventricular relationship, rate in the lower zoneSinus tachycardiaBeta-blockade; onset and rate criteria review
Very short, irregular intervals incompatible with physiology, signals on the near-field channel only, often with normal far-field channelLead noise from conductor fracture or loose connectionTherapy suspended if repetitive; urgent lead evaluation; impedance and radiograph
Alternating large and small sensed events with QRS and T wave both countedT-wave oversensingSensitivity, vector, or decay adjustment; electrolyte review
High-frequency signals with arm or trunk effortMyopotential oversensingSensing vector or sensitivity change; provocative maneuvers to confirm
Simultaneous noise on all channels with a timed external exposureElectromagnetic interferenceIdentify and remove the source; interrogation for reset

Stored electrogram patterns encountered in device follow-up. Noise episodes on a ventricular channel without therapy, reported by remote monitoring, are often the earliest sign of lead failure.

The management of repeated shocks and electrical storm, including temporary suspension of therapy and reprogramming, is covered in the defibrillator interrogation section of the electrical storm hub.

Device-detected atrial high-rate episodes

Pacemakers and defibrillators with an atrial lead, and insertable monitors, record atrial high-rate episodes. In ASSERT, among 2,580 patients aged 65 years or older with hypertension and no history of atrial fibrillation, subclinical atrial tachyarrhythmias (atrial rate above 190/min for more than 6 minutes) were detected within 3 months in 10.1%. They were associated with clinical atrial fibrillation (hazard ratio 5.56) and with ischemic stroke or systemic embolism (hazard ratio 2.49). The risk is lower than that of clinical atrial fibrillation and rises with episode duration and with the CHA₂DS₂-VASc score (heart failure, hypertension, age, diabetes, stroke, vascular disease, sex category).

The stored electrogram is reviewed before an episode is accepted as atrial fibrillation, because far-field R-wave oversensing, noise, and other atrial tachycardias produce false episodes. The 2023 atrial fibrillation guideline of the American College of Cardiology (ACC), American Heart Association (AHA), American College of Chest Physicians (ACCP), and HRS links anticoagulation to the longest episode and stroke risk. Episodes shorter than 5 minutes do not justify anticoagulation without another indication. Anticoagulation may be reasonable for episodes of 5 minutes to 24 hours with a CHA₂DS₂-VASc score of 3 or more, and it is reasonable for episodes of 24 hours or longer with a score of 2 or more, within shared decision-making. The trials of anticoagulation for device-detected atrial fibrillation and the full recommendations are on the device-detected atrial fibrillation section of the atrial fibrillation hub. The 2023 remote device clinic consensus considers alerts for a first atrial arrhythmia episode, a prolonged episode, or a high burden reasonable (Class 2a). Remote detection identifies atrial fibrillation months earlier than scheduled visits.

Pacing percentages, resynchronization delivery, and right ventricular pacing burden

Biventricular pacing. The benefit of resynchronization rises with the proportion of effectively paced beats, and the target is above 95%. Device counters overestimate effective capture when fusion or pseudofusion beats are counted as paced. Atrial fibrillation with rapid conduction, frequent premature ventricular complexes, an atrioventricular delay longer than intrinsic conduction, and loss of left ventricular capture all reduce effective pacing. The 2023 remote device clinic consensus considers an alert for a low percentage of biventricular pacing reasonable (Class 2a). When heart failure worsens despite resynchronization, the surface electrocardiogram is examined for a narrow resynchronized paced complex, and the tracing takes precedence over the counter.

Right ventricular pacing. A high burden of right ventricular pacing causes dyssynchrony and, in some patients, pacing-induced cardiomyopathy. The 2023 consensus summarizes burden thresholds of about 40% and 33% associated with risk and cites cardiomyopathy in 9% to 19.5% of patients with right ventricular pacing. Alerts for excessive right ventricular pacing are reasonable (Class 2a). Algorithms that minimize unnecessary ventricular pacing and pacing-induced cardiomyopathy are covered on the pacemaker hub, and effective biventricular capture and upgrade to resynchronization on the cardiac resynchronization hub.

Pacing dependence: definition and assessment

Pacing dependence is the absence of an adequate intrinsic rhythm when pacing stops. The EHRA consensus on interference from medical procedures defines it as the absence of an intrinsic escape rhythm, or a slow rate (usually below 50/min) that causes symptoms. It is assessed at interrogation by lowering the pacing rate in steps (a perioperative review uses 30/min) or temporarily inhibiting output while the surface rhythm and symptoms are watched, with backup pacing ready. A high pacing percentage does not establish dependence: a patient paced 100% of the time at a lower rate of 60/min may have a stable escape rhythm at 45/min.

Dependence is likely when the original indication was complete or high-grade atrioventricular block, after atrioventricular node ablation, and after transcatheter or surgical valve procedures with new conduction disease. It is less likely with sinus node dysfunction alone or with atrial fibrillation and a slow ventricular response. Dependence changes with time and with drugs, so it is reassessed when the last evaluation is old (a perioperative review uses more than 6 months). When dependence is unknown before a procedure with expected interference, the patient is managed as pacing-dependent.

Dependence determines the consequence of oversensing. In a dependent patient, electrosurgical noise sensed as cardiac activity inhibits pacing and produces asystole for as long as the interference lasts. In a nondependent patient, the same inhibition produces a brief pause filled by the native rhythm.

Follow-up schedule from implantation through battery depletion

Follow-up begins at implantation and continues for the life of the system, combining in-person evaluation with remote monitoring. The 2008 HRS/EHRA consensus set minimum frequencies. Evaluation is in person within 72 hours of implantation and again at 2 to 12 weeks. It then occurs at 3- to 12-month intervals for pacemakers and CRT-P and at 3- to 6-month intervals for ICDs and CRT-D (in person or remote), with at least one in-person evaluation per year. At signs of battery depletion, evaluation occurs at 1- to 3-month intervals. The 2015 HRS consensus recommended remote monitoring with at least one in-person evaluation per year in preference to a calendar-based schedule of in-person visits alone. The 2023 HRS/EHRA/APHRS/LAHRS consensus on the remote device clinic updated these intervals for patients with continuous connectivity.

Time pointEvaluationSource and class
Before implantationEducation about remote monitoring, its capabilities and limits, and clinic policies, as part of shared decision-making2023 consensus, Class 1
Before discharge or within 2 weeksEnrollment in remote monitoring and confirmation of connectivity2023 consensus, Class 2a (insertable monitors before discharge, Class 1)
Within 72 hoursIn-person evaluation, wound and device function2008 consensus
2 to 12 weeksIn-person evaluation after lead maturation; output programmed to chronic thresholds2008 consensus
Ongoing, pacemaker and CRT-PRemote transmission at intervals no longer than 3 to 12 months without continuous connectivity2023 consensus, Class 1
Ongoing, ICD and CRT-DRemote transmission at intervals no longer than 3 to 6 months without continuous connectivity2023 consensus, Class 1
Ongoing, all devices on remote monitoringRoutine remote surveillance of lead function and battery status2023 consensus, Class 1
In-person interval with continuous connectivityA 24-month interval is reasonable for pacemaker and ICD patients with consistent connectivity, no recent alerts, and no cardiac comorbidity requiring review2023 consensus, Class 2a
Approaching elective replacementTransmission at 1- to 3-month intervals2023 consensus, Class 1
Component under safety advisoryContinuous connectivity added to scheduled interrogation2023 consensus, Class 1
Insertable monitor with consistent connectivityRoutine in-office visits not indicated2023 consensus, Class 3: No Benefit

Follow-up schedule for cardiac implantable electronic devices, combining the 2008 HRS/EHRA monitoring consensus with the 2023 HRS/EHRA/APHRS/LAHRS remote device clinic consensus. Additional in-person or remote evaluations follow symptoms, alerts, shocks, procedures with interference, and changes in medication or clinical status.

Transtelephonic monitoring, in which the patient transmitted a rhythm strip over a telephone line, provided only a brief snapshot of pacemaker function. The 2008 consensus held that it should not be the sole means of pacemaker follow-up. It has been replaced by wireless systems that retrieve the full device memory.

Remote monitoring systems, connectivity, and transmissions

Remote monitoring systems consist of the implanted device's telemetry, a home transmitter or a smartphone application paired with the device, a cellular or internet connection, the manufacturer's secure server, and a clinic portal. Older systems required the patient to place a wand over the device (inductive telemetry); current systems use radiofrequency or low-energy Bluetooth telemetry and transmit automatically, usually at night. Transmissions are of three kinds: scheduled full interrogations, alert-triggered transmissions when a programmed condition occurs, and patient-initiated transmissions after symptoms or a shock.

Remote monitoring is distinct from remote programming. Transmissions retrieve data; programming changes still require a programmer in the clinic, hospital, or a site with a representative. Site-based remote monitoring is reasonable where onsite interrogation is unavailable and where it speeds care (2023 consensus, Class 2a). A transmitter in an emergency department or a hospital without a device service sends an interrogation to the clinic that follows the patient.

Two opened bedside pacemaker monitoring units: on the left, a gray home transmitter with its wand stored in the base tray; on the right, an older red transtelephonic kit containing a telephone cradle, wrist electrodes, and a white ring magnet.
Two generations of home pacemaker monitoring equipment. The gray unit (left) is a home transmitter with a wand that interrogates the device and sends the data to the clinic. The red kit (right) is an older transtelephonic system: wrist electrodes recorded a rhythm strip sent over a telephone line. The ring magnet converted the pacemaker to its magnet mode so that output, capture, and magnet rate could be judged. Current systems communicate wirelessly without a wand or magnet. Credit: Wikimedia Commons, public domain.

Connectivity. Monitoring protects only patients whose transmitters are connected. The 2023 consensus notes that up to half of patients fail to activate their receiver, and it recommends maintaining connectivity for devices capable of continuous connectivity (Class 1) and an established process with dedicated staff to reconnect patients who lose it (Class 1). Manufacturers notify clinics of disconnection (Class 1). Smartphone-based connectivity depends on the phone's battery, application permissions, and operating system updates, and patients receive explicit instruction on all three.

Limits of remote monitoring. Remote monitoring is not an emergency system. Alerts are reviewed during clinic hours, and patients are told that symptoms, shocks, or syncope require direct contact or emergency care (2023 consensus, Class 1). Remote data do not replace in-person evaluation for wound problems, pocket concerns, symptoms needing examination, lead problems requiring testing, or reprogramming.

Randomized trials and registries of remote monitoring

Randomized trials established that remote monitoring safely replaces most routine in-person visits and shortens the time from an event to a clinical decision. Effects on mortality and hospitalization have been consistent only with daily automatic transmission and a defined clinical response.

StudyPopulation and designMain results
TRUST (2010)1,339 ICD recipients randomized 2:1 to daily automatic home monitoring or office follow-up for 1 yearIn-hospital device evaluations reduced by 45% without increased morbidity; 85.8% of 6-, 9-, and 12-month checks done remotely only; median time to evaluation of arrhythmic events <2 days versus 36 days
CONNECT (2011)1,997 ICD and CRT-D recipients, wireless monitoring with automatic clinician alerts versus office care, 15 monthsMedian time from clinical event to clinical decision 4.6 versus 22 days; length of stay per cardiovascular hospitalization 3.3 versus 4.0 days
EVOLVO (2012)200 heart failure patients with ICD or CRT-D, remote monitoring versus standard care with audible alerts, 16 monthsEmergency department or urgent in-office visits 35% lower; time from alert condition to data review 1.4 versus 24.8 days
ECOST (2013)433 ICD recipients, daily home monitoring with annual visits versus visits at 6-month intervals, 24 monthsNoninferior for major adverse events (38.5% versus 41.5%); fewer shocks (193 versus 657); 52% fewer patients with inappropriate shocks; fewer capacitor charges
IN-TIME (2014)664 patients with New York Heart Association (NYHA) class II–III heart failure, left ventricular ejection fraction (LVEF) ≤35%, dual-chamber ICD or CRT-D; daily multiparameter telemonitoring versus standard care, 1 yearWorsened composite clinical score 18.9% versus 27.2% (odds ratio 0.63); 10 versus 27 deaths
MORE-CARE (2017)865 CRT-D recipients, remote checks alternating with office visits versus office visits alone, median 24 monthsNo difference in death or cardiovascular or device-related hospitalization (hazard ratio 1.02); 38% fewer health care resource uses, mainly office visits
REM-HF (2017)1,650 heart failure patients with a CIED, weekly downloads with a formal remote protocol versus usual care, median 2.8 yearsDeath or unplanned cardiovascular hospitalization 42.4% versus 40.8% (hazard ratio 1.01); no secondary end point improved
Pooled IN-TIME, ECOST, TRUST (2017)Patient-level analysis, 2,405 patients with daily home monitoring1-year mortality reduced by 1.9% absolute (risk ratio 0.62); death or worsening heart failure hospitalization reduced by 5.6% (risk ratio 0.64)
ALTITUDE (2010), registry185,778 ICD and CRT-D recipients; 69,556 followed on a remote networkHigher 1- and 5-year survival with remote follow-up (50% reduction in mortality); observational
Adherence analysis (2015), registry269,471 US pacemaker, ICD, CRT-P, and CRT-D recipients with wireless monitoringGraded association between the percentage of time on remote monitoring and survival; patients who never used monitoring had the worst survival

Randomized trials and large registries of remote monitoring of cardiac implantable electronic devices. Registry associations are subject to selection: patients who use remote monitoring differ from those who do not.

The neutral trials differed in design from IN-TIME: MORE-CARE alternated scheduled remote checks with office visits, supported by automatic wireless alerts, and REM-HF reviewed weekly downloads under a formal protocol. IN-TIME used daily transmission with a defined physician response, including telephone contact, medication change, and visits, which supports the interpretation that benefit depends on the alert-to-action pathway. The 2023 consensus recommends remote monitoring as part of the standard of care (Class 1, Level A) on the strength of these data.

Remote monitoring of heart failure: impedance and multisensor algorithms

Defibrillators and resynchronization devices record variables that change before heart failure decompensation: intrathoracic impedance (falling as lung fluid rises), night heart rate, heart rate variability, activity, atrial fibrillation burden and ventricular rate, biventricular pacing percentage, respiration, and heart sounds. The 2023 remote device clinic consensus considers remote monitoring of heart failure diagnostics reasonable to detect new heart failure or its progression (Class 2a).

Single-parameter impedance. Intrathoracic impedance monitoring with an audible patient alert did not improve outcome in DOT-HF. Among 335 patients, the composite of death and heart failure hospitalization occurred in 29% with access to impedance alerts and 20% without (hazard ratio 1.52, not significant), with more heart failure hospitalizations (hazard ratio 1.79) and more than three times as many outpatient visits in the access arm. Impedance alone has limited specificity, and alerts sent to patients generated visits without clear benefit.

Combined diagnostics. In PARTNERS HF, a positive combined device diagnostic in 694 CRT-D recipients was followed by a 5.5-fold higher risk of heart failure hospitalization with pulmonary congestion in the next month. The diagnostic was positive with two abnormal criteria in a month among atrial fibrillation duration, ventricular rate during atrial fibrillation, fluid index, activity, heart rate variability or night heart rate, and therapy delivery, or with a very high fluid index alone. Monthly review identified a high-risk window.

Multisensor index. The HeartLogic index combines heart sounds, respiration, thoracic impedance, heart rate, and activity. In the MultiSENSE validation cohort, it detected heart failure events with a sensitivity of 70% and an unexplained alert rate of 1.47 per patient-year, with a median warning of 34 days. In the first phase of MANAGE-HF, 200 patients with reduced ejection fraction had 1.76 alert cases per patient-year. Heart failure medications were increased in 74% of alert cases. Serious adverse events related to alert-prompted medication changes were rare (0.015 per patient-year), and the median N-terminal pro-B-type natriuretic peptide fell from 1,316 to 743 pg/mL over 12 months. Outcome benefit has not been shown in a completed randomized trial of the algorithm.

Implantable pulmonary artery pressure sensors, a separate technology, and their guideline recommendation are covered in the remote hemodynamic monitoring section of the heart failure with reduced ejection fraction hub. Device-based diagnostics are most useful when the heart failure team receives the alerts and has a protocol for diuretic and medication adjustment, atrial fibrillation management, and reprogramming.

Alert programming, triage, and the remote device clinic workflow

Alerts convert continuous data into clinical events. Manufacturers offer a menu of alerts, some fixed and some programmable, and their volume determines clinic workload. Unscheduled transmissions account for 27% to 40% of remote clinic work, and false atrial fibrillation detections and pause detections by insertable monitors are the largest source of nonactionable alerts. The 2023 consensus recommends customizing alerts to the clinical indication (Class 1) and reprogramming them to avoid nonactionable notifications (Class 1).

AlertClinical significance2023 consensus
Critical device or lead integrity alerts (impedance out of range, lead noise, capture loss, high-voltage circuit)Risk of failed pacing or inappropriate or failed shocksPrompt clinic notification, Class 1
Battery at elective replacement or rapid depletionLimited remaining service; replacement planningRoutine surveillance of battery status, Class 1
Any ventricular shockAppropriate or inappropriate therapy; more than 2 shocks is a high-priority alertAlert for all shocks, Class 1
Antitachycardia pacing deliveredVentricular tachycardia treated painlessly, or inappropriate pacing for supraventricular rhythmAlert for all antitachycardia pacing, Class 2a
Atrial arrhythmia: first episode, prolonged episode, or high burdenAnticoagulation and rhythm decisions; loss of resynchronizationClass 2a
Low biventricular pacing percentageReduced resynchronization benefitClass 2a
Excessive right ventricular pacingRisk of pacing-induced cardiomyopathyClass 2a
Heart failure diagnosticsImpending decompensationClass 2a
Loss of connectivityNo surveillanceManufacturer notification and clinic reconnection process, Class 1

Remote monitoring alerts and the corresponding 2023 HRS/EHRA/APHRS/LAHRS recommendations. High-priority alerts are reviewed and acted on within 1 business day (Class 2a).

Triage of remote transmissions. All transmissions are reviewed by trained staff, who confirm actionable findings from the electrograms (Class 1 for insertable monitors), classify urgency, and escalate to the responsible physician. High-priority (red) alerts are reviewed first. Results are communicated to the patient and the referring clinicians, and sharing all transmission results with patients according to their preference is reasonable (Class 2a). A complete pathway names who reviews alerts, the response time, how patients are reached, and how urgent findings are escalated outside clinic hours.

Insertable cardiac monitors. The indication determines alert programming. For unexplained syncope, patients activate a recording or transmit manually immediately after an event (Class 1). For cryptogenic stroke, adjustment of detection sensitivity to improve atrial fibrillation detection is reasonable (Class 2a). False-positive atrial fibrillation detections in reported series range from 46% to 86%, and frequent undersensing or oversensing requires reprogramming (Class 1).

Staffing, documentation, reimbursement, and data privacy

Staffing of the remote device clinic. The 2023 consensus recommends a team-based model with formal policies and defined roles (Class 1), dedicated time for all remote monitoring tasks (Class 1), and staff-to-patient ratios that reflect the workload from unscheduled transmissions (Class 1). A minimum of 3.0 full-time equivalents (clinical and administrative combined) per 1,000 patients on remote monitoring is reasonable (Class 2a). The consensus workload model estimated 1.61 full-time equivalents in the United States and 1.34 in Europe per 1,000 patients for combined remote and in-clinic follow-up before other tasks, rising to about 2.3 when 30% of patients have insertable monitors. US staff time per patient per year ranged from 2.1 hours for pacemakers to 9.3 hours for insertable monitors. Clinicians who prescribe, interpret, or document remote monitoring hold appropriate education or certification (Class 1), and clinics conduct regular quality review (Class 1). Third-party monitoring services are reasonable to relieve workload (Class 2a), and patients are told when such services are used (Class 2a).

Documentation and reporting. Clinics maintain manufacturer portals so that device changes, deaths, and transfers of care are recorded promptly (Class 1). Incorporation of reports into the electronic health record maintains patient privacy (Class 1). Universally accepted data definitions and formats are beneficial (Class 2a), and vendor-neutral software that aggregates transmissions from several manufacturers may be beneficial (Class 2b).

Reimbursement. The 2023 consensus recommends that payers adopt adequate reimbursement for remote monitoring tailored to regional care patterns (Class 1). In the United States, remote interrogation of pacemakers and defibrillators is reported per 90-day period with separate professional and technical components, and remote evaluation of insertable cardiac monitors and implantable physiologic monitors is reported per 30-day period. Per-visit payment structures do not reflect the continuous work of alert review.

Data privacy and cybersecurity. Transmissions contain protected health information and travel through manufacturer servers and clinic portals. Wireless telemetry also creates a cybersecurity exposure. In 2018, a US government advisory described two vulnerabilities in radiofrequency-enabled defibrillators and resynchronization defibrillators of one manufacturer: an authentication weakness that could allow a nearby attacker to send commands, and repeated wake-up commands that could deplete the battery. No known public exploit of these vulnerabilities was reported. The remedy was a firmware update applied with the clinic programmer in a monitored setting with external defibrillation available, because firmware updates carry a small risk of backup-mode pacing, incomplete update, or loss of programmed settings. Clinics weigh such updates against the device's remaining life and the patient's dependence on it. Disabling wireless communication ends remote monitoring and is generally not advisable.

Lead failure recognized during follow-up

Lead failure is the most common hardware failure of transvenous systems, and remote monitoring has changed how it presents. Conductor fracture produces intermittent high impedance and nonphysiologic noise; in a defibrillator, noise on the rate-sensing channel is counted as ventricular fibrillation and can trigger repeated inappropriate shocks in a conscious patient in sinus rhythm. Insulation failure produces low impedance, oversensing, loss of capture, and rapid battery drain. A loose set screw or incompletely inserted pin produces erratic high impedance soon after implantation or generator change.

Recalled lead families. The Sprint Fidelis defibrillator lead (Medtronic) was recalled in 2007 after 665 reported failures and 5 deaths, with about 150,000 US patients affected. In a two-center registry, it failed at 3.75% per year compared with 0.58% per year for other defibrillator leads, its 3-year survival was 87.9% versus 98.5%, and the hazard of failure increased with time. Pace-sense conductor fracture caused 87.5% of failures and inappropriate shocks in half of the patients with failure. The Riata and Riata ST leads (St. Jude Medical, now Abbott) developed externalized conductors, in which the cables migrate outside the insulation. In a seven-center study, 8.0% of Riata leads failed during follow-up, externalized conductors were found on fluoroscopy in 32% of Riata leads examined, and 27% of Riata leads with externalized conductors were malfunctioning. These findings supported systematic fluoroscopic screening of Riata leads.

Lead noise algorithms. Manufacturer algorithms combine impedance trends with counts of very short intervals and nonsustained high-rate episodes to identify likely fracture and to extend detection or alert the patient before a shock. In 12,793 device-lead combinations followed on a remote network, the Lead Integrity Alert identified more than 66% more lead system events and lead failures than conventional impedance monitoring, across lead families. Remote monitoring with systematic review of all transmissions and dedicated noise alerts has reduced inappropriate shocks from lead failure to low rates in large monitored cohorts.

Management of a failed lead (abandonment with a new lead, or extraction) depends on lead age and number, venous access, infection, patient age, and extraction risk. Device infection, an indication for complete system removal, is covered in the CIED infection section of the infective endocarditis hub.

Device recalls, advisories, and safety alerts

A device advisory is a manufacturer or regulatory notice that a device or lead may not perform as intended. In the United States, the Food and Drug Administration (FDA) classifies recalls by risk. Advisories are common: a review of FDA enforcement reports from 2000 to 2008 cited in the 2023 consensus found that 26.4% of generators or leads were subject to a recall or advisory.

FDA categoryDefinitionTypical device examples
Class I recallReasonable probability that use of or exposure to the product will cause serious adverse health consequences or deathLead fracture or premature battery depletion with failure to deliver therapy
Class II recallUse may cause temporary or medically reversible adverse consequences, or serious harm is remoteSoftware anomalies with reversible effects
Class III recallUse is unlikely to cause adverse health consequencesLabeling issues
Medical device safety alertA device may present an unreasonable risk of substantial harm; sometimes also considered a recallCommunication of a hazard with recommended clinical actions

FDA recall categories. For implanted devices, a "recall" rarely means removal from patients; it usually means notification, enhanced surveillance, reprogramming, or a software update.

Framework for management. The decision for an individual patient balances the probability and consequence of device failure against the risk of replacement. A decision analysis found that the advisory's failure rate and the likely effect of failure on survival drive the choice, with procedural mortality a secondary factor. For pacing-dependent patients, advisory failure rates above 0.3% per year favored replacement in most situations. For defibrillator patients, a failure rate of about 3% was needed to favor replacement in most cases, falling toward 1% when procedural mortality was low or the risk of fatal arrhythmia high. For pacemakers implanted for conditions that are not life-threatening, most advisories did not warrant replacement. Replacement carries its own risk: among 533 Canadian patients whose advisory ICD generators were replaced, complications occurred in 8.1%, major complications requiring reoperation in 5.8%, and 2 patients died after extraction for pocket infection. Only 3 advisory-related malfunctions occurred among 2,915 patients with advisory devices.

Practice after an advisory. Affected patients are identified from the clinic registry with the manufacturer's help, contacted, and enrolled in remote monitoring with continuous connectivity (2023 consensus, Class 1). Manufacturers notify clinics, help identify affected patients, and advise on alert programming (Class 1). Recommended reprogramming or software updates are applied, and the decision about prophylactic replacement is individualized and documented with the patient. The HRS task force on device performance policies set expectations for manufacturer reporting and physician communication with patients.

Battery advisories. Premature battery depletion is a generator advisory that remote monitoring is well suited to detect. In 2016, a manufacturer advisory concerned rapid depletion from lithium clusters (deposits that create an internal short circuit) in certain ICDs and CRT-Ds; battery depletion could progress from the elective replacement indicator to loss of function within a day to a few weeks. Clinicians were advised to act promptly on elective replacement alerts and keep patients on remote monitoring, prophylactic replacement was not routinely recommended, and a later firmware update added a battery performance alert. In subcutaneous ICDs, a capacitor-related advisory affected nearly 40,000 patients. In a nine-center registry of 1,112 subcutaneous ICDs, 547 had the affected component; clinically suspected premature depletion occurred in 29.1% of those, at a median of 65 months to removal or replacement, and none occurred in devices outside the advisory.

Magnet response of pacemakers and defibrillators by manufacturer

Clinical magnets, usually stronger than 80 gauss, close the generator's magnet sensor; the EHRA consensus notes that fields of 10 gauss or more can activate it. Over most pacemakers, the response is asynchronous pacing at a manufacturer-specific rate that falls as the battery nears replacement, so the magnet rate also reports battery status. Over most ICDs, the response is suspension of tachyarrhythmia detection and therapy while the magnet remains, with bradycardia pacing unchanged. The response can be programmed differently or off in some models, and the device report is the definitive source.

A white ring-shaped clinical magnet for implanted pacemakers and defibrillators lying on green surgical drape beside a ballpoint pen for scale; the magnet is slightly smaller in diameter than the pen is long.
A ring magnet used over pacemakers and defibrillators. Placed over the generator, it converts most pacemakers to asynchronous pacing and suspends tachyarrhythmia therapy in most ICDs for as long as it remains in place; removal restores the programmed function. The pen shows scale. Credit: Wikimedia Commons, CC0 1.0.

Pacemaker magnet rates at beginning of life are 100/min for Abbott and Boston Scientific (85/min at ERI), 90/min for Biotronik (80/min), 85/min for Medtronic (65/min), and 96/min for MicroPort (80/min). The Boston Scientific magnet rate also codes battery status: 100/min with more than 1 year remaining, 90/min with 1 year or less, and 85/min when replacement is indicated. The nominal Biotronik Auto setting gives only 10 asynchronous cycles, and its magnet response lasts 8 hours unless the magnet is removed and reapplied. Some older Medtronic models suppress the magnet response for 1 hour after programmer communication. Some Abbott models respond best with the magnet placed off-center over the edge of the generator. The full manufacturer table is on the pacemaker magnet section.

DeviceEffect of a magnetNotes
Transvenous ICD and CRT-D, all major manufacturersAntitachycardia pacing and shocks suspendedNo transvenous ICD paces asynchronously with a magnet; in MicroPort (Sorin) ICDs the magnet raises the pacing rate to the magnet rate (96/min, falling to 80/min near replacement) without changing the pacing mode
Medtronic ICDTherapies suspendedSteady tone for about 30 seconds confirms the response
Boston Scientific ICDTherapies suspended at nominal settingsTone synchronized to sensed R waves; magnet response programmable, including off
Biotronik ICDTherapies suspendedNo tone; therapies resume automatically after 8 hours of continuous application
Abbott ICDTherapies suspended (programmable)No tone in older models; newer models sound a brief tone on application and a higher tone on removal
Subcutaneous ICD (Boston Scientific)Shocks suspendedMagnet placed over the upper or lower edge of the generator in current models; a tone for about 60 seconds confirms the response; no bradycardia pacing beyond 30 seconds after a shock
Extravascular ICD (Medtronic)Antitachycardia pacing and shocks suspendedPause-prevention pacing continues if programmed on
Micra leadless pacemakers (Medtronic)No magnet responseAsynchronous pacing requires reprogramming
AVEIR leadless pacemakers (Abbott)Asynchronous pacing100/min, falling to as low as 85/min with battery depletion

Magnet effects on defibrillators and leadless systems, from the 2022 review of magnet application and the 2026 state-of-the-art perioperative review. When the expected tone or rate is absent, the magnet position is adjusted or the device is reprogrammed.

A magnet is the fastest way to stop inappropriate shocks in a patient in sinus rhythm or atrial fibrillation, provided monitoring and external defibrillation are available; detection resumes when it is removed. Over a pacemaker, magnet-induced asynchronous pacing terminates pacemaker-mediated tachycardia and overrides inhibition by oversensing. Loss of capture during magnet pacing indicates a capture failure, because asynchronous pacing removes oversensing as a cause.

Electromagnetic interference: mechanisms and device responses

Electromagnetic interference is any external electrical or magnetic signal that alters device function. Devices defend against it with bipolar sensing, bandpass filters, feedthrough capacitors that block radiofrequency energy, and noise-detection algorithms. Interference that passes these defenses produces several effects:

Reset is recognized by unexpected pacing at an unfamiliar rate in a single-chamber mode after a procedure. Backup settings differ by manufacturer.

ManufacturerReset modeReset ratePacing polarity (pacemaker / ICD)
AbbottVVI67/minUnipolar / bipolar
BiotronikVVI70/minUnipolar / bipolar
Boston ScientificVVI72.5/minUnipolar / unipolar
MedtronicVVI65/minUnipolar / bipolar
MicroPortVVI70/min (pacemaker), 60/min (ICD)Unipolar / bipolar

Electrical reset (backup) modes by manufacturer, from the 2022 review of perioperative management. A reset device is reprogrammed to the patient's settings; in a defibrillator, reset typically re-enables tachyarrhythmia therapies.

Preoperative evaluation of the patient with a device

The HRS and American Society of Anesthesiologists (ASA) consensus statement bases perioperative care on a device-team recommendation communicated to the procedural team before elective surgery. The plan answers a defined set of questions: the device type and manufacturer; the indication and pacing dependence; the date and findings of the last evaluation, including battery status and lead function; the programmed magnet response; the procedure, its location relative to the generator and leads, the energy sources, and the patient's position; and who will restore any changed settings afterward. Unresolved device problems, such as a battery at ERI or a failing lead, are addressed before elective surgery.

DocumentMost recent device evaluation before elective surgeryInterrogation after surgery
2024 ACC/AHA perioperative guideline (reaffirmed 2026)Within 12 months for pacemakers and 6 months for ICDsNo specific recommendation
2020 ASA practice advisoryWithin 3 to 6 monthsAfter emergency surgery without preoperative interrogation, suspected permanent change from magnet use, interference near the device, or delivered antitachycardia therapy
2022 EHRA consensus (with HRS, APHRS, LAHRS)Not specified; reprogramming planned before the procedure startsComplete interrogation immediately after procedures with a risk of interference or malfunction
2022 British Heart Rhythm Society guidanceWithin 12 monthsNot specified
2024 AHA scientific statement on periprocedural managementWithin 3 monthsRoutine follow-up unless concerning events occur

Timing of preoperative and postoperative device evaluation across society documents, as compiled in a 2026 state-of-the-art review. Any suspicion of malfunction prompts interrogation regardless of timing.

Antithrombotic interruption around surgery and device procedures is covered on the periprocedural anticoagulation and reversal hub.

Electrosurgery: monopolar and bipolar energy, distance, and the umbilicus rule

Monopolar electrosurgery is the main intraoperative source of interference. Current passes from the active electrode through the patient to a dispersive (return) pad, and the device senses it when that path crosses or approaches the generator and leads. Bipolar instruments confine current between the two tips and carry negligible risk of interference; ultrasonic scalpels produce none. Cutting current interferes less than coagulation current, and short bursts limit the duration of any inhibition.

Distance and the current path decide the risk. A 2022 review summarized a multicenter randomized trial of 80 patients with pectoral ICDs and its registry of procedures within 6 inches (15 cm) of the device. Interference occurred in 47% of operations within 15 cm of the generator, 7% of abdominal or pelvic operations, and none of lower abdominal or lower extremity operations. The EHRA consensus reports interference in 18 of 40 thoracic procedures (45%) and in none of 143 procedures below the iliac crest. US documents use the umbilicus as the boundary: procedures with monopolar electrosurgery above it carry meaningful risk with a pectoral device, and those below it carry little. The EHRA consensus uses a 15-cm distance from the generator and leads. Abdominal generators, abandoned leads, and unusual system anatomy require individual assessment.

Technique to reduce interference. Bipolar instruments are used where feasible. Monopolar energy is applied in short bursts at the lowest effective setting. The dispersive pad is placed so that the line from the active electrode to the pad does not cross the generator or leads. For procedures above the umbilicus, the pad is also placed at least 15 cm from the generator. Full-body return electrodes beneath the patient are avoided unless the device has been programmed appropriately.

Reprogramming or magnet: the intraoperative strategy

Two aims govern the intraoperative plan. In pacing-dependent patients, pacing must continue despite interference, which requires asynchronous pacing. In defibrillator patients, noise must not trigger shocks, which requires suspension of tachyarrhythmia therapy with continuous monitoring and external defibrillation available. A magnet achieves the first aim in most pacemakers and the second in most ICDs, but not the first in an ICD. A pacing-dependent patient with an ICD who is exposed to interference therefore requires reprogramming to an asynchronous pacing mode with therapies off.

SituationNot pacing-dependentPacing-dependent
Pacemaker, monopolar electrosurgery below the umbilicusNo change; magnet availableNo reprogramming in US guidance; magnet available (EHRA favors routine magnet application)
Pacemaker, monopolar electrosurgery above the umbilicusNo change to pacing mode; rate response disabled; magnet availableAsynchronous pacing by reprogramming, or by magnet when the response is confirmed asynchronous and the generator is accessible
Transvenous ICD or CRT-D, below the umbilicusNo reprogramming in US guidance; magnet availableNo reprogramming in US guidance; magnet available
Transvenous ICD or CRT-D, above the umbilicusTherapies suspended by magnet or programming; pads appliedReprogramming: therapies off and asynchronous pacing
Field within 15 cm of the generator, or generator inaccessible (prone or lateral position, head and neck surgery)Reprogramming for ICDsReprogramming
Micra leadless pacemakerNo changeReprogramming to asynchronous pacing (no magnet response)
Subcutaneous ICDShocks suspended by magnet or programming when interference is above the groinSeparate plan for pacing; the device provides no bradycardia pacing

Intraoperative device strategy by device type, pacing dependence, and site of monopolar electrosurgery, synthesized from the HRS/ASA consensus, the 2020 ASA advisory, the 2024 ACC/AHA guideline, and the 2022 EHRA consensus as compiled in the 2026 review. Minute-ventilation rate response is disabled for procedures with electrosurgery or impedance-based monitoring.

Magnet or programming. A magnet protocol is simple and reversible, and it cannot leave a patient discharged with therapies off. In a randomized trial of 80 ICD patients undergoing surgery with monopolar electrosurgery more than 6 inches from the device, magnet application compared with programming therapies off reduced excess time without therapy (28 versus 115 minutes) and caregiver handoffs, with no arrhythmic events or resets in either group. A magnet is unsuitable in pacing-dependent ICD patients. It is also unsuitable when the generator is inaccessible, the magnet cannot be secured or cannot reliably close the sensor (as in obesity), the patient's position changes, or the magnet response is programmed off or unknown. Reprogramming requires personnel and a programmer at both ends of the procedure and carries the risk that therapies are not restored.

Asynchronous pacing. The asynchronous rate is set above the intrinsic rate to avoid competition. Asynchronous pacing during an intrinsic rhythm can, rarely, deliver a stimulus in the vulnerable period; the risk is low with modern outputs, and pacing dependence outweighs it.

Intraoperative monitoring. Electrosurgery obscures the electrocardiogram, so pulse oximetry plethysmography or an arterial waveform confirms perfusion during cautery. When tachyarrhythmia therapies are suspended, external defibrillation pads are applied before the procedure and the patient remains continuously monitored until therapies are restored (EHRA consensus). Transcutaneous pacing is available for pacing-dependent patients.

Perioperative device plan by device type, pacing dependence, and electrosurgery site

The plan follows from the device type, pacing dependence, site of monopolar electrosurgery, magnet accessibility, and time since the last device evaluation, applying the US interrogation intervals (12 months for pacemakers, 6 months for defibrillators) and the strategy table above.

—
Before surgery. —
During surgery. —
Monitoring. —
After surgery. —

The plan follows the HRS/ASA consensus, the 2020 ASA advisory, the 2024 ACC/AHA guideline, and the 2022 EHRA consensus. Generators in the abdomen, abandoned or epicardial leads, cardiac surgery, and emergency procedures without device information require individual device-team planning.

External defibrillation and cardioversion in patients with devices

External shocks can reset the generator, raise capture thresholds through current conducted along the lead to the myocardium, and, rarely, damage the device. The EHRA consensus recommends anteroposterior pad placement, with the anterior pad at least 15 cm from the generator, biphasic shocks at the lowest effective energy, and interrogation after cardioversion and before discharge in higher-risk circumstances: legacy devices, a short pad-to-device distance, or pads not in the anteroposterior position. In a survey it cites, complications occurred in 0.6% of 1,809 patients. Device considerations never delay defibrillation of ventricular fibrillation; pads are placed away from the generator when this is possible without delay. Pad position and energy for resuscitation are covered in the defibrillation section of the adult cardiac arrest hub.

Elective cardioversion of atrial arrhythmias in a defibrillator patient can use the device's own shock through the programmer, which avoids external current; the device is interrogated afterward in either case.

Postoperative interrogation and restoration of therapies

The plan names who will restore the original programming and when. A patient whose defibrillator therapies are suspended, by magnet or by programming, remains on continuous monitoring with external defibrillation immediately available until therapies are restored. A patient left in an asynchronous pacing mode remains monitored until the original mode is restored. When a magnet was used, its removal restores programmed function in most devices, and immediate interrogation is unnecessary unless an event occurred.

CircumstancePostoperative device action
Device reprogrammed before the procedureOriginal settings restored and confirmed by interrogation before the patient leaves the monitored setting
Defibrillator therapy delivered, or suspectedInterrogation with electrogram review
External cardioversion or defibrillationInterrogation for reset, threshold rise, and damage
Emergency surgery without preoperative device informationInterrogation after the procedure
Substantial interference near the device, or cardiac surgeryInterrogation as soon as feasible
Unexpected pacing behavior (single-chamber pacing at an unfamiliar rate)Interrogation for electrical reset
Magnet used, no events, procedure away from the deviceRoutine follow-up

Indications for postoperative device interrogation, drawn from the 2020 ASA advisory, the 2022 EHRA consensus, and the 2022 review of magnet application.

Procedures other than surgery: ablation, lithotripsy, nerve stimulation, electroconvulsive therapy, endoscopy, and dental care

Many procedures outside the operating room expose devices to electrical or mechanical energy. The 2022 EHRA consensus graded the incidence of interference as high for electrocautery, cardiac radiofrequency ablation, and transcutaneous electrical nerve stimulation and related stimulators; moderate for noncardiac radiofrequency ablation, external cardioversion, and electroconvulsive therapy; and low for lithotripsy, therapeutic diathermy, radiofrequency identification, electromyography, and tissue expanders with magnets.

ProcedureRiskDevice management
Cardiac radiofrequency catheter ablationOversensing, inhibition, reset; lead dislodgement with immature leads; tissue injury at electrodesDevice programmed before ablation; ICD therapies disabled; asynchronous pacing if dependent; no ablation in contact with leads, and great caution within 1 cm; interrogation and reprogramming before the patient leaves the room
Noncardiac radiofrequency ablation (tumor, nerve)Interference along the current pathManaged as electrosurgery by site; ICD therapies suspended when the path is near the system
Extracorporeal shock-wave lithotripsyMechanical damage if focused on the generator; triggered oversensingBeam focused away from the generator; electrocardiographic (ECG) monitoring; reprogramming or a magnet not routinely required with modern devices
Transcutaneous electrical nerve stimulation, muscle stimulation, spinal cord stimulationOversensing, inhibition, inappropriate detectionElectrodes far from the device, unilateral current; first session monitored with electrograms in dependent patients and with a defibrillator nearby in ICD patients
Electroconvulsive therapyBrief current (about 1 to 2 seconds); seizure-related sinus tachycardia and myopotentialsECG monitoring; ICD therapies suspended by magnet; pacemakers usually unchanged
Gastrointestinal endoscopy and capsule endoscopyNone for diagnostic endoscopy or capsule; electrocautery during therapeutic endoscopyNo change for diagnostic procedures; ECG monitoring and magnet available for radiofrequency endoscopy
Dental equipmentMinimal at 20 cm except electrocauteryNo routine change; magnet if symptoms occur
Left ventricular assist deviceTelemetry interference and sensing of pump noiseRegular interrogation; programming adjustments; remote monitoring encouraged

Management of nonsurgical procedures in device patients, from the 2022 EHRA consensus on interference from medical procedures and the 2026 perioperative review.

Radiation therapy in patients with devices

Ionizing radiation damages the device's semiconductor circuits, and the effect is cumulative and unpredictable. Malfunction occurs in about 2% to 7% of treated patients, most often through damage to generator memory. Neutrons, generated by photon beams above 10 MV, electron beams above 20 MeV, and proton therapy, cause most failures, often as resets, independent of the dose to the generator. The device is never placed in the direct beam.

Cumulative dose to the generatorNot pacing-dependent pacemakerPacing-dependent pacemakerICD
<2 GyLow risk: interrogation before the first and after the last fractionIntermediate riskHigh risk
2–5 GyIntermediate risk: interrogation before, during (mid-course), and after treatmentIntermediate riskHigh risk
>5 Gy, or neutron-producing treatmentHigh risk: remote monitoring or at least weekly interrogationHigh riskHigh risk

Risk categories for radiation therapy in device patients, from the 2022 EHRA consensus, which follows the American Association of Physicists in Medicine (AAPM) Task Group 203 framework. High-risk patients are followed by remote monitoring or at least weekly interrogation, and a magnet of more than 10 gauss is kept in the treatment room.

The 2017 HRS consensus statement on MRI and radiation recommends a complete device evaluation before and after the course of treatment and weekly evaluation for patients receiving neutron-producing treatment (Class 1). It recommends non-neutron-producing treatment in preference to neutron-producing treatment to reduce the risk of reset (Class 1), and weekly evaluation may be reasonable for pacing-dependent patients receiving non-neutron-producing treatment (Class 2b). It recommends relocating the device when it would otherwise interfere with adequate tumor treatment (Class 1) and recommends against relocation for a maximum cumulative incident dose below 5 Gy (Class 3: Harm). Planning brings the radiation oncologist, the medical physicist, and the device clinic together before treatment; the planning dose to the generator is calculated, and pacing-dependent and defibrillator patients are monitored during fractions with programmers available.

MRI in patients with MR-conditional and non-MR-conditional systems

MRI exposes a device to a static magnetic field (force, torque, and magnet-sensor activation), radiofrequency fields (lead-tip heating and threshold rise), and gradient fields (induced currents that can capture the heart or be sensed and inhibit output). The EHRA consensus estimates that 50% to 75% of device patients will need MRI during their device's life.

MR-conditional systems. A system is MR-conditional only when the generator and all leads are labeled together for the conditions of the scan: field strength, region, specific absorption rate or B1+rms limits, and time since implantation (typically at least 6 weeks). Scanning under these conditions is on-label care. The device is programmed to an MRI mode before the scan and restored afterward.

Non-MR-conditional (legacy) systems. Two prospective registries at 1.5 tesla, MagnaSafe (1,500 nonthoracic examinations) and a study of 2,103 thoracic and nonthoracic examinations in 1,509 patients, found no long-term clinically significant adverse events under a standardized protocol. Resets were uncommon (0.4% of examinations in the second study), and one pacemaker with less than 1 month of battery remaining could not be reprogrammed. The registry details are on the pacemaker MRI section. The 2017 HRS consensus considers MRI reasonable in non-MR-conditional systems when there are no fractured, epicardial, or abandoned leads, MRI is the best test for the clinical question, and an institutional protocol with a designated MR physician and device physician is in place (Class 2a).

SCMR categoryDefinitionApproach
Category 1MR-conditional generator and leads scanned within labeled conditionsStandard of care; labeled field strength, region, and power limits
Category 2Non-MR-conditional system, or conditional generator with nonconditional leads, without abandoned, fractured, or epicardial leads1.5 tesla; institutional protocol; informed consent; pre- and post-scan interrogation
Category 3Any system with abandoned or fractured intracardiac leads or surgically placed epicardial leadsExperienced centers; individualized risk-benefit decision

Device categories for cardiovascular MRI from the 2024 SCMR consensus. A chest radiograph identifies abandoned and fractured leads when no recent film is available.

Scan protocol for patients with devices. The device is interrogated before the scan. Patients who are not pacing-dependent are programmed to a nonpacing (OVO or ODO) or inhibited (VVI or DDI) mode, and pacing-dependent patients to asynchronous pacing (VOO or DOO) at a rate above the intrinsic rate. Tachyarrhythmia therapies, rate response, magnet response, and other advanced features are disabled. Continuous ECG and pulse oximetry monitoring, personnel trained in advanced life support, and a defibrillator outside the scanner room are required. Whole-body specific absorption rate is kept below 2.0 W/kg, preferably well below. The device is interrogated immediately after the scan and restored to its settings; a post-scan change exceeding the SCMR thresholds in the lead measurements table prompts further evaluation.

Higher-risk circumstances. Abandoned leads, epicardial leads, lead fragments, very recent implantation, and a battery at or near ERI increase risk. In a series of 139 patients with 243 abandoned leads who underwent 200 scans, no serious safety events occurred, although one scan was stopped for heating with an abandoned subcutaneous array. Devices near ERI are more likely to reset and may fail to accept reprogramming.

Image quality. Generators produce susceptibility artifact in the anterior left ventricle. Spoiled gradient-echo readouts in place of balanced steady-state free precession, wideband inversion-recovery preparation for late gadolinium enhancement and T1 mapping, and elevation of the ipsilateral arm reduce artifact.

Computed tomography and nuclear imaging. Paced rhythms, pacing stimuli, and alternation between paced and intrinsic beats can disrupt ECG triggering of cardiac CT, and lead streak artifact commonly obscures the right coronary artery near the tricuspid annulus. Paced activation alters septal motion and can produce septal perfusion abnormalities on myocardial perfusion imaging.

Electromagnetic interference in daily life and at work

Household appliances, mobile phones in normal use, and airport screening are safe at normal distances for modern bipolar systems. Strong static magnets are the most common modern exposure. Smartphones, smartwatches, wireless earbuds, and their cases incorporate magnets for wireless charging and accessory attachment, and these can activate a device's magnet mode when held directly over it. The FDA recommends keeping such consumer electronics at least 6 inches (15 cm) from implanted devices, particularly defibrillators, and not carrying them in a breast pocket over the generator. At this distance, clinically significant events are unlikely.

Walk-through and handheld security detectors and retail anti-theft gates passed at a normal pace carry very low risk. Arc welding, induction furnaces, large motors, and strong industrial magnets require a workplace assessment of field strength, sometimes with monitoring at the workplace, and implants with magnets, such as tissue expanders, are reviewed by the device team before placement. Sources and precautions are tabulated on the pacemaker electromagnetic interference section, and occupational exposure with a defibrillator on the ICD magnet and interference section.

Driving and occupational restrictions

Driving rules are set by licensing authorities and differ for private and commercial licenses. Private driving is restricted for 1 week after a primary-prevention ICD and for 6 months after a secondary-prevention ICD or after appropriate therapy for ventricular tachycardia or fibrillation. For commercial drivers, the 2007 AHA/HRS addendum advises against driving with any ICD. The 2024 Federal Motor Carrier Safety Administration advisory criteria, which are guidance and not regulation, describe ICDs as likely to cause syncope or collapse, and medical examiners generally do not certify drivers with an ICD. After pacemaker implantation, the restriction is short when pacing corrects the cause of symptoms. The details are on the ICD driving section and the pacemaker driving section.

Occupational assessment covers exposure to industrial electromagnetic fields, the consequences of sudden incapacity at work (heights, heavy machinery, public safety roles), and contact sports, in which blows to the generator or leads near the clavicle can damage hardware. Arm elevation and heavy lifting on the device side are limited for several weeks after transvenous implantation while the leads mature.

Device therapy at the end of life

A patient with decision-making capacity, or the legal surrogate of a patient without capacity, may request withdrawal of any device therapy, including pacing in a pacing-dependent patient, and withdrawal is ethically and legally equivalent to withdrawal of any other life-sustaining treatment (HRS consensus). The ethical framework and guideline recommendations are on the ICD deactivation section and the pacemaker end-of-life section.

Deactivating tachyarrhythmia therapy stops shocks without stopping the heart, and bradycardia pacing can remain active; deactivating pacing in a pacing-dependent patient can be followed by death within a short time, so symptom control and family preparation precede reprogramming. The functions turned off and the responsible clinicians are documented. In urgent situations, a magnet taped over most defibrillators suspends shocks until programming is available, and hospice teams are told how to reach the device service. Goals of care during electrical storm are covered in the electrical storm hub.

Device emergencies presenting outside the device clinic

Repeated shocks. The patient and the surface rhythm are assessed first. Ventricular tachycardia or fibrillation is treated as an arrhythmic emergency. When shocks occur in sinus rhythm or atrial fibrillation, they are presumed inappropriate (noise, oversensing, or a supraventricular rhythm), and a magnet over the generator suspends further therapy while the patient remains monitored with external defibrillation available until interrogation. The stored electrograms then identify the cause.

Pacing failure. Absent or ineffective pacing in a dependent patient is treated with transcutaneous or temporary transvenous pacing and chronotropic drugs while the cause is established. Over a pacemaker, a magnet distinguishes oversensing (magnet pacing restores output and capture) from capture or output failure (stimuli without capture, or no stimuli). The chest radiograph shows dislodgement, fracture, twiddler syndrome, or a disconnected pin. Hyperkalemia and acidosis raise capture thresholds and are corrected.

Apparent device tachycardia. Rapid ventricular pacing usually represents tracking of an atrial tachyarrhythmia up to the upper rate or pacemaker-mediated tachycardia. A magnet converts the pacemaker to asynchronous pacing, which terminates pacemaker-mediated tachycardia immediately and stops atrial tracking while it remains.

A device emergency is followed by interrogation and review in the device clinic, with reprogramming, lead evaluation, and medication changes as indicated, and an assessment of driving restrictions and psychological effects of shocks.

Sources of error in device follow-up and perioperative care

References

  1. Ferrick AM, Raj SR, Deneke T, et al. 2023 HRS/EHRA/APHRS/LAHRS expert consensus statement on practical management of the remote device clinic. Heart Rhythm. 2023;20:e92–e144. doi:10.1016/j.hrthm.2023.03.1525 (also J Arrhythm. 2023;39:250–302)
  2. Slotwiner D, Varma N, Akar JG, et al. HRS expert consensus statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. Heart Rhythm. 2015;12:e69–e100. doi:10.1016/j.hrthm.2015.05.008
  3. Wilkoff BL, Auricchio A, Brugada J, et al. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm. 2008;5:907–925. doi:10.1016/j.hrthm.2008.04.013
  4. Crossley GH, Poole JE, Rozner MA, et al. The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) expert consensus statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management. Heart Rhythm. 2011;8:1114–1154. doi:10.1016/j.hrthm.2010.12.023
  5. American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Implantable Electronic Devices. Practice advisory for the perioperative management of patients with cardiac implantable electronic devices: pacemakers and implantable cardioverter-defibrillators 2020. Anesthesiology. 2020;132:225–252 (erratum 2020;132:938). doi:10.1097/ALN.0000000000002821
  6. Thompson A, Fleischmann KE, Smilowitz NR, et al. 2024 AHA/ACC/ACS/ASNC/HRS/SCA/SCCT/SCMR/SVM guideline for perioperative cardiovascular management for noncardiac surgery. Circulation. 2024;150:e351–e442 (correction 2024;150:e466). doi:10.1161/CIR.0000000000001285
  7. Stühlinger M, Burri H, Vernooy K, et al. EHRA consensus on prevention and management of interference due to medical procedures in patients with cardiac implantable electronic devices. Europace. 2022;24:1512–1537. doi:10.1093/europace/euac040
  8. Fernando RJ, Crimi E, Puttur Rajkumar K, et al. Perioperative management of cardiac implantable electronic devices: a state of the art narrative review. J Thorac Dis. 2026;18:536. doi:10.21037/jtd-2026-1-0434
  9. Özkartal T, Demarchi A, Caputo ML, et al. Perioperative management of patients with cardiac implantable electronic devices and utility of magnet application. J Clin Med. 2022;11:691. doi:10.3390/jcm11030691
  10. Gifford J, Larimer K, Thomas C, et al. Randomized controlled trial of perioperative ICD management: magnet application versus reprogramming. Pacing Clin Electrophysiol. 2014;37:1219–1224. doi:10.1111/pace.12417
  11. Indik JH, Gimbel JR, Abe H, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017;14:e97–e153. doi:10.1016/j.hrthm.2017.04.025
  12. Kim D, Collins JD, White JA, et al. SCMR expert consensus statement for cardiovascular magnetic resonance of patients with a cardiac implantable electronic device. J Cardiovasc Magn Reson. 2024;26:100995. doi:10.1016/j.jocmr.2024.100995
  13. Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376:755–764. doi:10.1056/NEJMoa1603265
  14. Nazarian S, Hansford R, Rahsepar AA, et al. Safety of magnetic resonance imaging in patients with cardiac devices. N Engl J Med. 2017;377:2555–2564. doi:10.1056/NEJMoa1604267
  15. Miften M, Mihailidis D, Kry SF, et al. Management of radiotherapy patients with implanted cardiac pacemakers and defibrillators: a report of the AAPM TG-203. Med Phys. 2019;46:e757–e788. doi:10.1002/mp.13838
  16. Varma N, Epstein AE, Irimpen A, et al. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation. 2010;122:325–332. doi:10.1161/CIRCULATIONAHA.110.937409
  17. Crossley GH, Boyle A, Vitense H, et al. The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial: the value of wireless remote monitoring with automatic clinician alerts. J Am Coll Cardiol. 2011;57:1181–1189. doi:10.1016/j.jacc.2010.12.012
  18. Guédon-Moreau L, Lacroix D, Sadoul N, et al. A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trial. Eur Heart J. 2013;34:605–614. doi:10.1093/eurheartj/ehs425
  19. Landolina M, Perego GB, Lunati M, et al. Remote monitoring reduces healthcare use and improves quality of care in heart failure patients with implantable defibrillators: the EVOLVO study. Circulation. 2012;125:2985–2992. doi:10.1161/CIRCULATIONAHA.111.088971
  20. Hindricks G, Taborsky M, Glikson M, et al. Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet. 2014;384:583–590. doi:10.1016/S0140-6736(14)61176-4
  21. Hindricks G, Varma N, Kacet S, et al. Daily remote monitoring of implantable cardioverter-defibrillators: insights from the pooled patient-level data from three randomized controlled trials (IN-TIME, ECOST, TRUST). Eur Heart J. 2017;38:1749–1755. doi:10.1093/eurheartj/ehx015
  22. Boriani G, Da Costa A, Quesada A, et al. Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial. Eur J Heart Fail. 2017;19:416–425. doi:10.1002/ejhf.626
  23. Morgan JM, Kitt S, Gill J, et al. Remote management of heart failure using implantable electronic devices. Eur Heart J. 2017;38:2352–2360. doi:10.1093/eurheartj/ehx227
  24. Saxon LA, Hayes DL, Gilliam FR, et al. Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation. 2010;122:2359–2367. doi:10.1161/CIRCULATIONAHA.110.960633
  25. Varma N, Piccini JP, Snell J, et al. The relationship between level of adherence to automatic wireless remote monitoring and survival in pacemaker and defibrillator patients. J Am Coll Cardiol. 2015;65:2601–2610. doi:10.1016/j.jacc.2015.04.033
  26. Powell BD, Saxon LA, Boehmer JP, et al. Survival after shock therapy in implantable cardioverter-defibrillator and cardiac resynchronization therapy-defibrillator recipients according to rhythm shocked: the ALTITUDE survival by rhythm study. J Am Coll Cardiol. 2013;62:1674–1679. doi:10.1016/j.jacc.2013.04.083
  27. Boehmer JP, Hariharan R, Devecchi FG, et al. A multisensor algorithm predicts heart failure events in patients with implanted devices: results from the MultiSENSE study. JACC Heart Fail. 2017;5:216–225. doi:10.1016/j.jchf.2016.12.011
  28. Hernandez AF, Albert NM, Allen LA, et al. Multiple cArdiac seNsors for mAnaGEment of Heart Failure (MANAGE-HF) – phase I evaluation of the integration and safety of the HeartLogic multisensor algorithm in patients with heart failure. J Card Fail. 2022;28:1245–1254. doi:10.1016/j.cardfail.2022.03.349
  29. van Veldhuisen DJ, Braunschweig F, Conraads V, et al. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation. 2011;124:1719–1726. doi:10.1161/CIRCULATIONAHA.111.043042
  30. Whellan DJ, Ousdigian KT, Al-Khatib SM, et al. Combined heart failure device diagnostics identify patients at higher risk of subsequent heart failure hospitalizations: results from PARTNERS HF. J Am Coll Cardiol. 2010;55:1803–1810. doi:10.1016/j.jacc.2009.11.089
  31. Healey JS, Connolly SJ, Gold MR, et al. Subclinical atrial fibrillation and the risk of stroke. N Engl J Med. 2012;366:120–129. doi:10.1056/NEJMoa1105575
  32. Joglar JA, Chung MK, Armbruster AL, et al. 2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation. Circulation. 2024;149:e1–e156 (correction 2024;149:e1413). doi:10.1161/CIR.0000000000001193
  33. Carlson MD, Wilkoff BL, Maisel WH, et al. Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines. Heart Rhythm. 2006;3:1250–1273. doi:10.1016/j.hrthm.2006.08.029
  34. Amin MS, Matchar DB, Wood MA, Ellenbogen KA. Management of recalled pacemakers and implantable cardioverter-defibrillators: a decision analysis model. JAMA. 2006;296:412–420. doi:10.1001/jama.296.4.412
  35. Gould PA, Krahn AD; Canadian Heart Rhythm Society Working Group on Device Advisories. Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories. JAMA. 2006;295:1907–1911. doi:10.1001/jama.295.16.1907
  36. Poole JE, Gleva MJ, Mela T, et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation. 2010;122:1553–1561. doi:10.1161/CIRCULATIONAHA.110.976076
  37. Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Heart Rhythm. 2009;6:605–610. doi:10.1016/j.hrthm.2009.02.024
  38. Abdelhadi RH, Saba SF, Ellis CR, et al. Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm. 2013;10:361–365. doi:10.1016/j.hrthm.2012.10.045
  39. Ellenbogen KA, Gunderson BD, Stromberg KD, Swerdlow CD. Performance of Lead Integrity Alert to assist in the clinical diagnosis of implantable cardioverter defibrillator lead failures: analysis of different implantable cardioverter defibrillator leads. Circ Arrhythm Electrophysiol. 2013;6:1169–1177. doi:10.1161/CIRCEP.113.000744
  40. Wörmann J, Strik M, Jurisic S, et al. Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients. Europace. 2024;26:euae170. doi:10.1093/europace/euae170
  41. Jacob S, Shahzad MA, Maheshwari R, et al. Cardiac rhythm device identification algorithm using X-rays: CaRDIA-X. Heart Rhythm. 2011;8:915–922. doi:10.1016/j.hrthm.2011.01.012
  42. Epstein AE, Baessler CA, Curtis AB, et al. Addendum to “Personal and public safety issues related to arrhythmias that may affect consciousness”: public safety issues in patients with implantable defibrillators. A scientific statement from the American Heart Association and the Heart Rhythm Society. Circulation. 2007;115:1170–1176. doi:10.1161/CIRCULATIONAHA.106.180203
  43. Lampert R, Hayes DL, Annas GJ, et al. HRS expert consensus statement on the management of cardiovascular implantable electronic devices (CIEDs) in patients nearing end of life or requesting withdrawal of therapy. Heart Rhythm. 2010;7:1008–1026. doi:10.1016/j.hrthm.2010.04.033
  44. US Food and Drug Administration. Magnets in cell phones and smart watches may affect pacemakers and other implanted medical devices. fda.gov
  45. Cybersecurity and Infrastructure Security Agency. ICS medical advisory ICSMA-18-107-01: Abbott Laboratories defibrillator (radiofrequency authentication and battery-depletion vulnerabilities; firmware update). April 2018. cisa.gov
  46. Centers for Medicare & Medicaid Services. Billing and coding: cardiac rhythm device evaluation (A56602). Medicare Coverage Database. cms.gov
  47. Thompson A, Fleischmann KE, Smilowitz NR, et al. 2026 AHA/ACC/ACS/ASNC/HRS/SCA/SCCT/SCMR/SVM guideline for perioperative cardiovascular management for noncardiac surgery (reaffirmation of the 2024 guideline). Circulation. 2026;154:e413–e504. doi:10.1161/CIR.0000000000001472
  48. Wan EY, Rogers AJ, Lavelle M, et al. Periprocedural management and multidisciplinary care pathways for patients with cardiac implantable electronic devices: a scientific statement from the American Heart Association. Circulation. 2024;150:e183–e196. doi:10.1161/CIR.0000000000001264
  49. Federal Motor Carrier Safety Administration. Medical advisory criteria for evaluation under 49 CFR part 391, subpart E. 49 CFR part 391, appendix A. Fed Regist. 2024;89:3577. govinfo.gov/content/pkg/FR-2024-01-19/html/2024-00980.htm