Library › Electrocardiography › Ambulatory ECG monitoring
Ambulatory ECG monitoring: Holter, patch, event, implantable loop recorders, and wearables
Ambulatory electrocardiographic monitoring in adults, from recorder types, leads, and analysis through selection by symptom frequency, syncope, palpitations, atrial fibrillation after stroke and in screening, ventricular arrhythmia, Holter risk markers, artifact, and reporting.
The monitor is chosen by how often the symptom occurs and by the question being asked. A 24- to 48-hour Holter suits daily symptoms and burden measurement. A patch, loop recorder, or mobile telemetry worn for 2 to 4 weeks suits weekly symptoms. An implantable loop recorder, which records for about 3 years or longer, is reserved for infrequent events after nondiagnostic external monitoring. A symptom recorded during sinus rhythm is as informative as one recorded during an arrhythmia, because it excludes an arrhythmic cause.
Ambulatory ECG monitoring: purposes and classes of recorder
Ambulatory electrocardiography records the cardiac rhythm during ordinary daily activity and sleep, beyond the 10 seconds of a resting 12-lead electrocardiogram (ECG). It has three purposes. The first is diagnosis: correlating a symptom such as palpitations, presyncope, or syncope with the rhythm at that moment, and documenting an asymptomatic arrhythmia that is itself diagnostic, such as atrial fibrillation after a stroke or intermittent Mobitz II block. The second is quantification: measuring the burden of premature ventricular complexes, atrial fibrillation, or pacing, the ventricular rate in atrial fibrillation, and the response of an arrhythmia to drugs or ablation. The third is risk stratification: detecting nonsustained ventricular tachycardia in hypertrophic or arrhythmogenic cardiomyopathy and, in research and selected practice, measuring autonomic and repolarization markers.
The 2017 expert consensus statement of the International Society for Holter and Noninvasive Electrocardiology and the Heart Rhythm Society (ISHNE-HRS) on ambulatory ECG and external cardiac monitoring is the principal US document. It classifies recorders by duration, recording mode (continuous, event-triggered, or automatically triggered), number of leads, transmission (stored for later analysis or sent in real time), and the analyses available. Guidelines of the American College of Cardiology (ACC), the American Heart Association (AHA), the American College of Chest Physicians (ACCP), and the Heart Rhythm Society add indication-specific recommendations: the 2017 syncope guideline, the 2018 bradycardia guideline, the 2017 ventricular arrhythmia guideline, and the 2023 atrial fibrillation guideline. The 1999 ACC/AHA guideline for ambulatory electrocardiography remains the source of several definitions, including the criteria for ambulatory ST-segment ischemia.
Ambulatory recording began in the late 1940s with a radio-telemetered ECG carried on the back. It developed through magnetic-tape recorders played back at high speed into the present solid-state, wireless, and implantable devices. Modern recorders weigh a few grams to a few hundred grams and many transmit data automatically to a monitoring center.
| Recorder | Usual duration | Recording mode | Leads | Transmission | Main strengths and limits |
|---|---|---|---|---|---|
| Holter monitor | 24–48 h; extended recorders 3–14 days or longer | Continuous, full disclosure | 2 or 3 channels; 12-lead available | Analyzed after return | Morphology, burden, heart rate profile, ST and QT analysis; short window; wires and electrode detachment |
| Adhesive patch monitor | Up to 14 days for most; some longer | Continuous, full disclosure; symptom button | 1 (some 2) closely spaced bipolar leads | Usually after return; some transmit in real time | High adherence, water resistant; limited P-wave and QRS morphology; no localization |
| External loop recorder | 2–6 weeks | Looping memory saved when the patient activates it or an algorithm triggers | Usually 1 | Transmitted after an event | Captures onset of the event; electrodes worn continuously; only stored segments are reviewed |
| Post-event recorder, handheld or smartphone ECG | Weeks to indefinite | Recorded only when applied during symptoms | 1 (some 6-lead) | Immediate | No continuous wear; misses events with loss of consciousness and short episodes |
| Mobile cardiac outpatient telemetry | Up to 30 days | Continuous with automatic and patient-triggered transmission | 1 to 3 | Near real time to an attended center | Detects asymptomatic and dangerous arrhythmias promptly; costly; notification burden |
| Implantable loop recorder | About 3 years; up to about 4.5 years for some current models | Looping memory; automatic and patient-activated storage | 1 subcutaneous bipolar lead | Daily automatic remote transmission | Longest window; minor procedure; single lead with small P waves; undersensing and oversensing |
| Consumer wearable | Indefinite | Photoplethysmography continuously; ECG only when the user records it | Single lead (lead I equivalent) on demand | User's phone | Detects irregular pulse in very large populations; not diagnostic without ECG confirmation |
Characteristics of ambulatory cardiac monitors, adapted from the 2017 ISHNE-HRS consensus statement. Devices and durations vary by manufacturer.
Holter monitoring: recorders, lead systems, and extended recording
The Holter recorder is a small battery-powered device worn on a belt or strap, connected by lead wires to adhesive silver/silver chloride gel electrodes on the chest. It records all beats for the whole period (full disclosure), and the recording is analyzed after the patient returns the device. Conventional recordings last 24 or 48 hours; current recorders store 7 to 14 days or more, and some record continuously for up to 30 days. Patients mark symptoms by pressing an event button and keep a written or electronic diary.
Leads and channels. A two-channel recorder records two independent bipolar chest leads from five electrodes; a three-channel recorder records three leads from five or seven electrodes. The modified chest leads are chosen for the clinical question. CM5 (negative electrode on the manubrium, positive at the V5 position) gives the largest QRS and is the most sensitive single lead for ST depression. A modified V1 lead displays P waves and right bundle branch morphology. An inferior lead displays P-wave polarity and inferior ST changes. A 12-lead Holter uses torso placement of the limb electrodes (the Mason–Likar arrangement, with the arm electrodes below the clavicles and the left leg electrode on the lower left torso) and standard precordial positions. The EASI system derives a 12-lead ECG from five electrodes placed at the sternum and both mid-axillary lines.
The ISHNE-HRS statement recommends 12-lead Holter recording when QRS morphology, ST-segment pattern, or QT dynamics must be analyzed (Class I). Examples are localizing the origin of premature ventricular complexes before ablation, assessing biventricular capture with cardiac resynchronization therapy, detecting a transient type 1 Brugada pattern, and quantifying ischemia. Torso placement of the limb electrodes shifts the frontal axis and changes inferior-lead amplitudes, so a 12-lead Holter tracing is not interchangeable with a standard ECG for serial comparison. The standard resting ECG remains the reference for QT measurement in long QT syndrome.
Extended Holter recording. Lengthening a continuous recording beyond 48 hours raises diagnostic yield for symptoms that do not occur daily while keeping multichannel morphology and full disclosure. Electrode irritation and detachment increase with time, and electrodes are replaced and sites rotated during long recordings.
| Lead system | Electrodes | Information provided | Typical use |
|---|---|---|---|
| Single bipolar lead (patch, loop recorder, implantable recorder) | 2 closely spaced | Rate, rhythm, R-R intervals; small P waves | Arrhythmia detection over long periods |
| Two-channel Holter (for example CM5 and modified V1) | 5 | Two views; better separation of artifact and of aberrant from ventricular beats | Standard rhythm Holter |
| Three-channel Holter | 5 to 7 | Three near-orthogonal views; ST trend in lateral, anterior, and inferior leads | Rhythm and ST analysis |
| EASI derived 12-lead | 5 | Synthesized 12-lead ECG with no limb electrodes | Continuous 12-lead monitoring in active patients |
| Mason–Likar 12-lead Holter | 10 | Full 12-lead morphology and ST segments | PVC origin, Brugada pattern, ischemia, CRT capture, QT dynamics |
Holter and monitor lead configurations described in the 2017 ISHNE-HRS statement. CM5, chest-modified V5; CRT, cardiac resynchronization therapy; PVC, premature ventricular complex.
Skin preparation, electrode placement, and the symptom diary
Signal quality is determined at application. The ISHNE-HRS statement specifies shaving the skin when needed, removing dead surface skin cells by rubbing with a rough paper or cloth, and using electrodes from airtight packages within their expiration dates. Electrodes are placed over bone (the sternum and ribs) to limit muscle and motion artifact, and lead wires are looped and taped so that tension does not pull on the electrode. Patch monitors are applied to the left upper chest over the sternal or pectoral region, as specified by the manufacturer, after the skin is clipped of hair, cleaned, and dried.
The patient is instructed to continue usual activities, including the activity that provokes symptoms, because a recording made during an atypically quiet period has a lower yield. Symptoms are marked with the event button and entered in the diary with the time, the symptom, and the activity. Symptom–rhythm correlation is the main output of a diagnostic recording, so patient instruction determines its yield. In a randomized comparison of loop recorders with Holter monitors, 23% of patients with recurrent symptoms failed to activate the loop recorder correctly despite education.
Adhesive patch monitors: continuous single-lead recording and diagnostic yield
A patch monitor combines electrodes, amplifier, memory, and battery in a single adhesive housing worn on the chest without wires. Most record one bipolar lead from electrodes a few centimeters apart for up to 14 days, and patients can shower and exercise while wearing it. Some patches transmit selected events in near real time through a phone or cellular link, which makes them a form of mobile telemetry. Most store the full recording for analysis after the device is mailed back.
Two studies define the diagnostic advantage of prolonged continuous recording. In 146 patients who wore a 24-hour Holter and a 14-day patch simultaneously, the patch detected 96 arrhythmia events over its wear time compared with 61 for the Holter (P<0.001). In 26,751 consecutive patch recordings, mean wear time was 7.6 days and analyzable signal was a median 99% of wear time. An arrhythmia was found in 60.3% of patients, and in 29.9% of those the first arrhythmia occurred after the first 48 hours; 51.1% of first symptom-triggered arrhythmias occurred after 48 hours. The yield over the full wear period exceeded the yield of the first 48 hours for any arrhythmia (62.2% versus 43.9%) and for any symptomatic arrhythmia (9.7% versus 4.4%).
The single closely spaced lead records low-amplitude P waves without spatial orientation. Atrial flutter with regular conduction may be difficult to separate from sinus tachycardia. The origin of a premature ventricular complex cannot be localized, aberrant supraventricular beats are difficult to separate from ventricular beats, and ST-segment and QT analysis are limited. Signal quality varies with body habitus. These limits favor a multichannel Holter when morphology is the question and a patch when duration and adherence are the question.
External event and loop recorders and automatic arrhythmia detection
Post-event recorders have built-in electrodes and are not worn continuously. The patient applies the device to the chest, or holds it with both hands or fingers, during symptoms and records a short single-lead tracing, which is then transmitted. Smartphone-based handheld recorders work the same way. They cannot record events that cause loss of consciousness or that end before the device is applied, and they record nothing before the symptom.
External loop recorders are worn continuously with chest electrodes and maintain a looping memory of the most recent ECG. When the patient presses the button, the device saves a programmable period before the activation and after it, so the onset of the arrhythmia is recorded. Automatic triggers save episodes of bradycardia, pauses, tachycardia, and irregular rhythm without patient action, which adds asymptomatic arrhythmias and events with abrupt loss of consciousness.
Randomized comparisons showed a higher diagnostic yield with event recording than with Holter monitoring for intermittent symptoms. In patients with palpitations, event recorders produced a diagnostic symptomatic rhythm strip in 67% compared with 35% with a 48-hour Holter. Clinically important arrhythmias (supraventricular tachycardia, atrial fibrillation, or flutter) were found in 19% compared with none. In a cost-effectiveness analysis of continuous-loop event monitoring for palpitations, new diagnoses per patient were 1.04 in week 1, 0.15 in week 2, and 0.01 in week 3 and beyond. Two weeks was proposed as the standard duration. In patients with syncope or presyncope randomized to a 1-month loop recorder or a 48-hour Holter, an arrhythmia was identified or excluded in 63% versus 24%. In that trial, the overall probability of symptom–rhythm correlation was 56% versus 22%.
Automatic algorithms record the onset of atrial fibrillation but not reliably its offset, so burden cannot be measured accurately from triggered recordings. Only the saved segments are reviewed, and an arrhythmia between triggers is invisible.
Mobile cardiac outpatient telemetry and remote notification
Mobile cardiac outpatient telemetry combines continuous recording, automatic arrhythmia detection, and transmission to an attended monitoring center through a cellular or Wi-Fi link. Technicians review detected events and notify the ordering physician or emergency services according to prespecified criteria, so a dangerous arrhythmia can be acted on during the monitoring period. Sensors are patches, chest belts, or wired electrodes with one to three channels.
A 17-center trial enrolled patients with syncope, presyncope, or severe palpitations and a nondiagnostic 24-hour Holter. They were randomized to up to 30 days of mobile telemetry or a loop recorder. A diagnosis was made in 88% versus 75% (P=0.008), and clinically significant arrhythmias were detected in 41% versus 15% (P<0.001). In patients presenting with syncope or presyncope, diagnosis was reached in 89% versus 69%.
Continuous transmission generates large volumes of data, and vendor notification criteria can produce frequent calls for findings that do not need urgent action. The ISHNE-HRS statement recommends that practices develop their own notification criteria, because the defaults often overstate clinical significance, and that they maintain a mechanism to handle incoming data at any hour.
| Urgency | Example findings | Action |
|---|---|---|
| Emergency | Sustained regular wide-complex tachycardia at 160 beats/min or faster; prolonged asystole | Monitoring center contacts the patient and the physician; emergency medical services are activated if the physician cannot be reached |
| Within 24 hours | Sustained wide-complex tachycardia under 160 beats/min; Mobitz II or third-degree atrioventricular (AV) block; symptomatic pause of 4 s or longer; symptomatic sustained supraventricular tachycardia at 150 beats/min or faster; symptomatic heart rate of 30 beats/min or less; any reported syncope; observed pacemaker malfunction or ICD discharge | Physician or on-call physician notified |
| Office hours | New atrial fibrillation or flutter lasting 30 s or longer; ventricular rate 150 beats/min or faster for 60 s or longer in atrial fibrillation; wide-complex runs at 120 beats/min or faster lasting 3 beats to 30 s; Mobitz I block | Notification the same day when recorded between 9 a.m. and 5 p.m., otherwise the next day; office hours run 7 days a week |
Example notification criteria for ambulatory cardiac telemetry, adapted from the 2017 ISHNE-HRS statement (Table 6). Asymptomatic pauses, supraventricular tachycardia, and bradycardia meeting the 24-hour thresholds move to office-hour notification. ICD, implantable cardioverter-defibrillator.
Implantable loop recorders: device, insertion, programming, and longevity
The implantable loop recorder, also called an insertable cardiac monitor, is a leadless subcutaneous device with two electrodes on its housing that record a single bipolar ECG. Current devices are a few centimeters long and weigh a few grams. The device stores episodes detected automatically (pauses, bradycardia, tachycardia, and atrial fibrillation or tachycardia) and episodes the patient marks with a handheld activator or smartphone application. It transmits stored data automatically, usually daily, through a bedside transmitter or the patient's phone.
Insertion. The device is inserted in the left parasternal or left pectoral region through an incision under local anesthesia, using a dedicated tool that creates a tight subcutaneous tunnel. The orientation is chosen to maximize the R wave and P wave; it is often parallel to the sternum or at about 45 degrees to it over the fourth intercostal space. R-wave amplitude is measured at insertion, because a low R wave causes undersensing that the algorithm interprets as pauses and an irregular rhythm. The procedure takes minutes, is performed in an office, procedure room, or electrophysiology laboratory, and needs no fluoroscopy.
Programming. Detection thresholds are set for the indication. Pause detection is commonly programmed at 3 seconds or more and bradycardia at about 30 to 40 beats/min in syncope. Atrial fibrillation detection sensitivity and a minimum episode duration are set when atrial fibrillation is the question. A tachycardia threshold is set for suspected supraventricular or ventricular tachycardia. Programmable features that reject noise and redetect the R wave reduce false episodes.
Longevity. Battery longevity for current devices is about 3 years, and some newer devices are labeled for about 4.5 years. In CRYSTAL-AF, cumulative detection of atrial fibrillation after cryptogenic stroke continued to rise through 36 months, and median time to first detection was 8.4 months. Yield therefore depends on monitoring for the full battery life when events are infrequent.
Indications. The principal indications are recurrent unexplained syncope with infrequent episodes after nondiagnostic initial evaluation; infrequent palpitations or presyncope suspected to be arrhythmic; detection of atrial fibrillation after stroke or transient ischemic attack of undetermined cause; and surveillance of arrhythmia burden after atrial fibrillation ablation in selected patients.
Complications of loop recorder implantation, magnetic resonance imaging, and removal
Implantation is low risk. In two prospective studies of a miniaturized device, infection occurred in 1.3% and 1.6% of patients and procedure-related serious adverse events in 0.7% and 1.6%. In STROKE-AF, 1.8% of patients who received a monitor had a procedure-related adverse event (one site infection, two incision-site hemorrhages, and one case of implant-site pain). Other complications are pocket hematoma, erosion of the device through the skin, device migration or rotation, which degrades sensing, and persistent pain. Infection or erosion is treated by removal; local wound care alone seldom resolves an infected subcutaneous device.
Current implantable recorders are generally labeled magnetic resonance (MR) conditional at 1.5 T and 3 T according to the manufacturer's conditions. The magnetic field and radiofrequency energy can create artifact that the device stores as arrhythmia episodes, so stored episodes during the scan time are interpreted as noise. A recorder is removed through a small incision when the diagnosis has been made, when the battery is depleted, or when it is no longer useful. Leaving a depleted device in place is an alternative when removal offers no benefit.
Consumer wearables: photoplethysmography, smartwatch ECG, and handheld recorders
Photoplethysmography detects the pulse optically at the wrist or fingertip from changes in light absorption with the arterial pulse. Algorithms examine series of pulse intervals (tachograms) and notify the user when sustained irregularity suggests atrial fibrillation. Photoplethysmography has no P waves and no QRS morphology; frequent premature beats, atrial flutter or tachycardia with variable conduction, second-degree AV block, and motion all produce irregular pulse intervals. The 2023 ACC/AHA/ACCP/HRS guideline states that photoplethysmography can alert a person to obtain an ECG but is not sufficiently reliable to establish a diagnosis of atrial fibrillation.
Single-lead ECG on a watch or handheld device. A user touches an electrode with a finger of the opposite hand (watch) or places fingers of both hands on two electrodes (handheld recorders such as KardiaMobile) for about 30 seconds, recording a lead I equivalent. Some devices record a six-lead tracing using a leg contact. Automated algorithms classify the tracing as sinus rhythm, atrial fibrillation, or unclassifiable. The 2023 guideline recommends that an initial diagnosis of atrial fibrillation be made by a clinician's visual review of the ECG tracing, regardless of the device (Class 1). The guideline also finds a consumer ECG device with a high-quality tracing reasonable for detecting recurrence in patients with known atrial fibrillation in whom monitoring is advised (Class 2a).
| Study | Device and population | Design | Main finding |
|---|---|---|---|
| Apple Heart Study (2019) | Smartwatch photoplethysmography; 419,297 US adults without known AF | Single arm; notified participants mailed a 7-day ECG patch | 0.52% notified over a median of 117 days; AF on the later patch in 34%; positive predictive value of a notification for AF on a simultaneous ECG 84% |
| Fitbit Heart Study (2022) | Wrist photoplethysmography; 455,699 adults, median age 47 years | Single arm; 1-week ECG patch after notification | Irregular rhythm detection in 1% (4% of those aged 65 or older); AF on the patch in 32.2%; positive predictive value of a notification during patch wear 98.2% |
| Huawei Heart Study (2019) | Wristband and watch photoplethysmography; 187,912 adults in China, mean age 35 years | Single arm; clinical confirmation | 0.23% received a suspected AF notification; positive predictive value 91.6% |
| REHEARSE-AF (2017) | Handheld single-lead ECG twice weekly; 1,001 adults 65 or older with CHA₂DS₂-VASc 2 or more | Randomized versus routine care for 12 months | AF diagnosed in 19 versus 5 patients (hazard ratio 3.9); no significant difference in stroke or embolic events |
| mSToPS (2018) | Self-applied continuous ECG patch worn for up to 4 weeks; 2,659 high-risk US adults | Randomized immediate versus delayed monitoring | New AF at 4 months in 3.9% versus 0.9%; more anticoagulation and more cardiology visits at 1 year |
Large studies of consumer and remote rhythm recording. Positive predictive value depends on the prevalence of atrial fibrillation in the population: in younger users a notification is less often confirmed. AF, atrial fibrillation.
A notification is followed by review of any ECG tracing the device recorded and, when no diagnostic tracing exists, by a prescribed ambulatory monitor or a 12-lead ECG during symptoms. Episodes of a few seconds labeled as atrial fibrillation are reviewed visually, because ectopy and artifact are common causes. Consumer devices are not a substitute for a prescribed monitor when syncope or a potentially dangerous arrhythmia is being evaluated, because they are not worn continuously as ECG recorders and transmit nothing to a clinician.
Choosing a monitor by symptom frequency and clinical question
The 2017 syncope guideline states that the choice of a specific monitor should be determined by the frequency and nature of the events (Class 1). The 2018 bradycardia guideline gives the same recommendation for suspected bradycardia (Class 1). The ISHNE-HRS statement recommends a 24- to 48-hour Holter when frequent symptoms are expected within the recording window (Class I). It recommends extended monitoring for 15 to 30 days when symptoms occur less than daily or at an uncertain frequency (Class I).
Other determinants are the consequence of the suspected arrhythmia, the patient's ability to activate a device, loss of consciousness during the event, the need to measure morphology or burden, and the need for results during the monitoring period. A patient with structural heart disease and unexplained syncope who has not been admitted needs real-time transmission. A patient with daily palpitations needs a Holter with a diary. A patient with monthly palpitations who can activate a device needs a loop recorder or patch for 2 to 4 weeks. A patient with two syncopal episodes a year needs an implantable recorder.
| Recording duration | Recorder | Palpitations | Syncope | Silent AF after cryptogenic stroke |
|---|---|---|---|---|
| Under 60 s | Event recorder | 50–60% | Not applicable | Not applicable |
| 24–48 h | Holter | 10–15% | 1–5% | 1–5% |
| 3–7 days | Patch, external loop recorder, mobile telemetry | 50–70% | 5–10% | 5–10% |
| 1–4 weeks | External loop recorder, patch, mobile telemetry | 70–85% | 15–25% | 10–15% |
| Up to 36 months | Implantable loop recorder | 80–90% | 30–50% | 15–20% |
Estimated diagnostic yield by recorder and duration, from the 2017 ISHNE-HRS statement (Table 2). Figures for stroke were marked uncertain in the source, and later trials reported 30% detection at 3 years with an implantable recorder after cryptogenic stroke.
| Situation | Recommendation | Source and class |
|---|---|---|
| Frequent symptoms expected within the recording window | 24- to 48-hour Holter | ISHNE-HRS, Class I |
| Symptoms less than daily or of uncertain frequency | Extended monitoring for 15 to 30 days | ISHNE-HRS, Class I |
| Analysis of QRS morphology, ST pattern, or QT dynamics | 12-lead Holter | ISHNE-HRS, Class I |
| Suspected arrhythmic syncope, ambulatory patient | Monitor chosen by frequency and nature of events | ACC/AHA/HRS syncope, Class 1 |
| Suspected arrhythmic syncope, selected ambulatory patients | Holter, transtelephonic monitor, external loop recorder, patch, or mobile telemetry | ACC/AHA/HRS syncope, Class 2a |
| Suspected arrhythmic syncope, selected ambulatory patients | Implantable cardiac monitor | ACC/AHA/HRS syncope, Class 2a (randomized evidence) |
| Hospitalized for syncope with suspected cardiac cause | Continuous ECG monitoring | ACC/AHA/HRS syncope, Class 1 |
| Suspected bradycardia, infrequent symptoms more than 30 days apart | Implantable cardiac monitor after nondiagnostic noninvasive evaluation | ACC/AHA/HRS bradycardia, Class 2a |
| Symptoms possibly caused by ventricular arrhythmia | Ambulatory ECG monitoring | AHA/ACC/HRS ventricular arrhythmia, Class 1 |
| Sporadic symptoms possibly caused by ventricular arrhythmia | Implantable cardiac monitor | AHA/ACC/HRS ventricular arrhythmia, Class 2a |
| Cryptogenic stroke | Extended ambulatory monitoring to detect undiagnosed AF | ISHNE-HRS, Class I |
| Stroke or TIA of undetermined cause | Initial monitoring and, if needed, an implantable loop recorder | ACC/AHA/ACCP/HRS AF, Class 2a |
| Systemic embolism without known AF, maximum sensitivity sought | Implantable cardiac monitor | ACC/AHA/ACCP/HRS AF, Class 2a |
| Cryptogenic stroke without contraindication to anticoagulation | Long-term rhythm monitoring (mobile telemetry, implantable loop recorder, or other) | AHA/ASA secondary stroke prevention, Class 2a |
Guideline and consensus recommendations for the choice of ambulatory monitor. AF, atrial fibrillation; TIA, transient ischemic attack.
Probability of recording an episode: monitor duration and burden calculator
When episodes occur at random at a constant average rate, the probability that a monitor worn for a given time records at least one episode is 1 − e−t/T, where t is the effective recording time and T is the average interval between episodes. A monitor worn for one average interval records an episode with a probability of 63%, and one worn for three intervals with a probability of 95%. Clustered episodes, poor adherence, and episodes that the patient does not recognize lower the real yield. The second part of the calculator gives premature ventricular complex burden and atrial fibrillation burden from report counts.
Monitor yield and arrhythmia burden calculator
Poisson capture model; burden from report countsCapture probability = 1 − exp(−duration × proportion worn ÷ interval). PVC burden = PVC count ÷ total beats. AF burden = minutes in AF ÷ analyzable minutes. PVC thresholds are from a cohort referred for PVC ablation; anticoagulation thresholds for device-detected AF are from the 2023 ACC/AHA/ACCP/HRS guideline and are discussed in the atrial fibrillation hub.
With one episode per 7 days and 95% wear, the probability of capture is 13% with a 24-hour Holter, 24% with a 48-hour Holter, 61% at 7 days, 85% at 14 days, and more than 98% at 30 days. With one episode per 90 days and the same 95% wear, a 30-day monitor captures an episode in 27% of patients, and a 3-year implantable recorder in nearly all who have a recurrence during its life.
Syncope: ambulatory monitoring and the implantable loop recorder
The initial evaluation of syncope (history, examination with orthostatic measurements, and a 12-lead ECG) establishes a cause or a likely cause in many patients and identifies those at high risk who need admission or urgent evaluation. Ambulatory monitoring is indicated when an arrhythmic cause is suspected or must be excluded and the initial evaluation is not diagnostic. Features that raise the probability of arrhythmic syncope are syncope during exertion or while supine, abrupt loss of consciousness without prodrome, palpitations immediately before the event, injury, structural heart disease, an abnormal ECG, and a family history of sudden death. The abnormal ECG findings are bifascicular block, Mobitz II block, sinus bradycardia below 40 beats/min or sinus pauses longer than 3 seconds while awake, pre-excitation, a Brugada pattern, a long or short QT interval, and Q waves.
What monitoring can establish. A recording during a typical event either shows an arrhythmia that explains it, which leads to treatment, or shows sinus rhythm, which excludes an arrhythmic cause of that event. Some asymptomatic findings are diagnostic on their own because they indicate the mechanism and warrant treatment. Examples are Mobitz II or third-degree AV block, ventricular pauses longer than 3 seconds while awake, rapid paroxysmal supraventricular tachycardia, and sustained or rapid ventricular tachycardia. Sinus bradycardia, Mobitz I block, and short pauses during sleep are physiologic in many adults and do not explain syncope without symptom correlation.
Yield of external monitoring. Holter monitoring rarely captures the event that caused syncope; the ISHNE-HRS statement estimates a 1% to 5% yield. In one series of short-term Holter monitoring, symptoms correlated with an arrhythmia in 4% and occurred without an arrhythmia in 17%. Extended external recording for 2 to 4 weeks raises the yield when events occur at least monthly.
Implantable loop recorder. In the Randomized Assessment of Syncope Trial, 60 patients with unexplained syncope were assigned to conventional testing (external loop recorder, tilt testing, and electrophysiology study) or to an implantable recorder with 1 year of monitoring. A diagnosis was obtained in 52% with prolonged monitoring versus 20% with conventional testing, and bradycardia was detected in 40% of monitored patients versus 8%. In the PICTURE registry of 570 patients with recurrent unexplained syncope who had undergone a median of 13 tests before implantation, syncope recurred in 36% within 12 months. Of 218 recorded events, the recorder guided a diagnosis in 78%, and 75% of these were cardiac. The 2017 ACC/AHA/HRS guideline finds an implantable monitor useful in selected ambulatory patients with suspected arrhythmic syncope (Class 2a). The 2018 bradycardia guideline finds it reasonable when symptoms occur more than 30 days apart and noninvasive evaluation is not diagnostic (Class 2a).
Reflex syncope with documented asystole. In ISSUE-3, 511 patients 40 years or older with three or more syncopal episodes in 2 years received an implantable recorder. Eighty-nine had syncope with asystole of 3 seconds or more, or asymptomatic asystole of 6 seconds or more. Seventy-seven were randomized to dual-chamber pacing with rate-drop response switched on or off. The 2-year syncope recurrence rate was 25% with pacing on versus 57% with pacing off, a 57% relative reduction. Pacing treats the cardioinhibitory component and does not prevent vasodepressor syncope. Pacing indications in reflex syncope and carotid sinus syndrome are given in the bradycardia and AV block hub.
Bundle branch block and syncope. Unexplained syncope with bifascicular block raises suspicion of paroxysmal infranodal AV block. An electrophysiology study with a His–ventricular (HV) interval of 70 ms or more, or infra-His block provoked by atrial pacing or procainamide, supports pacing. A normal study does not exclude intermittent block, and an implantable recorder is the next step when syncope remains unexplained. Empiric pacing is an option when suspicion is high.
Structural heart disease. In patients with ischemic or nonischemic cardiomyopathy and syncope, an implantable recorder does not replace an electrophysiology study or a defibrillator when an indication for a defibrillator already exists. It does not protect the patient while monitoring continues. Monitoring in this group is used when the arrhythmic risk is judged low enough to wait for a recorded event.
Palpitations and presyncope: indications for monitoring and interpretation of a normal recording
Palpitations are the most common indication for ambulatory monitoring. The history defines the likely mechanism. Isolated skips or thumps suggest premature beats. Sudden onset and offset of a rapid regular rhythm, sometimes ended by a vagal maneuver, suggests paroxysmal supraventricular tachycardia. An irregular rapid rhythm suggests atrial fibrillation. Gradual onset with exercise or anxiety suggests sinus tachycardia. Palpitations with syncope, exertion, structural heart disease, or a family history of sudden death raise concern for ventricular arrhythmia. A detailed history, examination, and 12-lead ECG establish a diagnosis in up to one-third of patients; in the rest, ambulatory monitoring is the main diagnostic test.
The ISHNE-HRS statement indicates monitoring for symptom–ECG correlation (Class I for unexplained palpitations) in four situations. These are a history, examination, or ECG that suggests an arrhythmia; structural heart disease, a family history of sudden death, or an inherited channelopathy; a patient who needs a specific explanation and reassurance; and treatment, such as ablation or drug therapy, that depends on a documented arrhythmia. A Holter suits daily palpitations or symptoms that can be reproduced. Event or loop recorders worn for about 2 weeks suit less frequent episodes. An implantable recorder suits infrequent sustained episodes that external devices have missed. The 2015 ACC/AHA/HRS supraventricular tachycardia guideline lists ambulatory monitoring and patient-activated event recorders as the means of documenting paroxysmal tachycardia that the resting ECG has not captured.
A normal recording during the patient's usual symptom is a diagnostic result. Sinus rhythm or isolated premature beats during typical palpitations exclude a sustained arrhythmia as their cause. When monitoring repeatedly misses a history strongly suggestive of paroxysmal supraventricular tachycardia, an electrophysiology study is a reasonable alternative to further monitoring, because it both diagnoses and treats the mechanism. The electrocardiographic diagnosis of the recorded rhythm is covered in the supraventricular rhythms hub.
Presyncope and dizziness are evaluated like syncope when they are abrupt or accompanied by palpitations, and like palpitations when they are recurrent and the patient can activate a recorder. Lightheadedness that is gradual, positional, or continuous is seldom arrhythmic, and orthostatic measurements, medication review, and tilt testing are often more informative than prolonged monitoring.
Detection of atrial fibrillation by prolonged monitoring after stroke and transient ischemic attack
About one-quarter of ischemic strokes have no cause identified after standard evaluation. Paroxysmal atrial fibrillation is a plausible mechanism in many, and its detection changes secondary prevention from antiplatelet therapy to anticoagulation. Detection increases with the duration of monitoring, and most first episodes are asymptomatic.
| Trial | Population | Comparison | AF definition | AF detection |
|---|---|---|---|---|
| CRYSTAL-AF (2014) | 441 patients 40 years or older, cryptogenic stroke or TIA within 90 days, at least 24 h of monitoring without AF | Implantable monitor versus conventional follow-up | Over 30 s | 8.9% vs 1.4% at 6 months; 12.4% vs 2.0% at 12 months; 30.0% vs 3.0% at 36 months |
| EMBRACE (2014) | 572 patients 55 years or older, cryptogenic stroke or TIA within 6 months | 30-day event-triggered recorder versus 24-h Holter | 30 s or longer | 16.1% vs 3.2% at 90 days (number needed to screen 8); 9.9% vs 2.5% for 2.5 min or longer; anticoagulation 18.6% vs 11.1% |
| Find-AFRANDOMISED (2017) | 398 patients 60 years or older with acute ischemic stroke, any presumed cause except severe large-artery stenosis | 10-day Holter at baseline, 3 months, and 6 months versus at least 24 h of monitoring | 30 s or longer | 14% vs 5% at 6 months (number needed to screen 11) |
| PER DIEM (2021) | 300 patients with ischemic stroke without known AF | Implantable recorder for 12 months versus 30-day external loop recorder | 2 min or longer | 15.3% vs 4.7% at 12 months |
| STROKE-AF (2021) | 492 patients with stroke attributed to large- or small-vessel disease | Implantable monitor versus usual care | Over 30 s | 12.1% vs 1.8% at 12 months |
Randomized trials of prolonged monitoring for atrial fibrillation after stroke or transient ischemic attack (TIA). None was powered for recurrent stroke. In CRYSTAL-AF, 81% of first episodes were asymptomatic.
The trials established that longer monitoring detects more atrial fibrillation; none showed that the strategy reduces recurrent stroke. In PER DIEM, recurrent ischemic stroke occurred in 3.3% versus 5.3%, a nonsignificant difference. The 2023 ACC/AHA/ACCP/HRS guideline finds initial cardiac monitoring and, if needed, extended monitoring with an implantable loop recorder reasonable after stroke or transient ischemic attack of undetermined cause (Class 2a). It also finds an implantable monitor reasonable after systemic embolism when maximum sensitivity is sought (Class 2a). The 2021 guideline of the AHA and the American Stroke Association on secondary stroke prevention finds long-term rhythm monitoring with mobile telemetry, an implantable loop recorder, or another approach reasonable after cryptogenic stroke in patients without a contraindication to anticoagulation (Class 2a).
Monitoring is most productive in older patients, those with embolic-appearing infarcts, left atrial enlargement, or frequent atrial ectopy, and those who would accept anticoagulation. The evaluation of patent foramen ovale in cryptogenic stroke, including the requirement to exclude atrial fibrillation before closure, is covered in the atrial septal defect and PFO hub.
Screening for atrial fibrillation in asymptomatic adults
Screening increases detection of atrial fibrillation. Whether it reduces stroke has been tested in randomized trials with mixed results.
| Trial | Population | Screening method | AF detection | Clinical outcome |
|---|---|---|---|---|
| STROKESTOP (2021) | 28,768 residents aged 75–76 years in two Swedish regions; 51% of those invited participated | Intermittent handheld ECG twice daily for 14 days | Higher in the screened group | Composite of stroke, systemic embolism, bleeding leading to hospitalization, and death: hazard ratio 0.96 (P=0.045) over 6.9 years |
| LOOP (2021) | 6,004 adults aged 70–90 years with hypertension, diabetes, prior stroke, or heart failure | Implantable loop recorder; anticoagulation for episodes of 6 min or longer | 31.8% vs 12.2% over a median of 64.5 months | Stroke or systemic embolism 4.5% vs 5.6% (hazard ratio 0.80, P=0.11); major bleeding 4.3% vs 3.5% (not significant) |
| GUARD-AF (2024) | 11,905 US primary care patients 70 years or older; enrollment stopped early | 14-day single-lead patch | 5% vs 3.3% | Stroke hospitalization 0.7% vs 0.6% (hazard ratio 1.10) over a median of 15.3 months |
| mSToPS (2018) | 2,659 high-risk US adults | Continuous ECG patch for up to 4 weeks, immediate versus delayed | 3.9% vs 0.9% at 4 months | Not designed for outcomes |
Randomized trials of screening for atrial fibrillation (AF) with ambulatory or intermittent ECG recording.
In 2022, the US Preventive Services Task Force concluded that the evidence is insufficient to assess the balance of benefits and harms of screening for atrial fibrillation in asymptomatic adults 50 years or older (an I statement). The 2023 ACC/AHA/ACCP/HRS guideline reviews risk stratification and population screening but makes no graded recommendation for or against screening. It states that a benefit of screening on ischemic stroke, systemic embolism, and survival has not yet been established, including in adults at high predicted risk of atrial fibrillation.
Subclinical and device-detected atrial fibrillation. Atrial high-rate episodes from pacemakers and defibrillators with an atrial lead, and subclinical atrial fibrillation from implantable or wearable monitors, carry a lower stroke risk than clinical atrial fibrillation. The risk rises with episode duration and stroke-risk score. Stored electrograms are reviewed to confirm the rhythm before an episode count is acted on (Class 1). The 2023 guideline's duration-based anticoagulation recommendations and the results of NOAH-AFNET 6 and ARTESIA are summarized in the atrial fibrillation hub.
Monitoring established atrial fibrillation: burden, rate control, drugs, and ablation
Burden. Atrial fibrillation burden is the proportion of monitored time spent in atrial fibrillation. Continuous recorders (Holter, patch, mobile telemetry, implantable recorders, and devices with an atrial lead) measure it; triggered recorders do not, because they record onset but not offset. The 2023 guideline finds it reasonable to infer frequency, duration, and burden from the automated algorithms of continuous monitors and devices, with periodic review to exclude other arrhythmias (Class 2a). The minimum duration of 30 seconds used to define an episode is a convention for trials. It is not a threshold linked to outcomes.
Rate control. Resting heart rate on a 12-lead ECG underestimates the ventricular rate during activity. Holter or patch monitoring measures the mean, maximum, and minimum ventricular rates, the rate during ordinary activity, and pauses after termination or during sleep that may limit rate-slowing drugs. The 2023 guideline sets a resting heart rate target of below 100 to 110 beats/min for most patients without heart failure. A stricter target of below 80 beats/min at rest and below 110 beats/min during moderate exercise applies when cardiomyopathy induced by atrial fibrillation is suspected. Ambulatory monitoring documents that the stricter target is met. Before cardiac resynchronization therapy and in patients with a device, monitoring also shows whether the ventricular rate in atrial fibrillation limits biventricular pacing. Rate-control targets and drugs are covered in the atrial fibrillation hub.
Antiarrhythmic drugs. Monitoring during outpatient initiation detects proarrhythmia: conversion of atrial fibrillation to atrial flutter with 1:1 conduction or QRS widening with class IC drugs, QT prolongation (particularly after pauses) and torsades de pointes with class III drugs, and sinus pauses or AV block. The ISHNE-HRS statement recommends ambulatory monitoring for the outpatient initiation of some antiarrhythmic drugs (Class I). Dofetilide is started in hospital with continuous monitoring, as described in the antiarrhythmic drugs hub. Monitoring also documents the efficacy and safety of the pill-in-the-pocket strategy, including post-conversion pauses.
After ablation. Symptoms are an unreliable guide to recurrence after ablation, because ablation increases the proportion of asymptomatic episodes. The 2017 international expert consensus statement on catheter and surgical ablation of atrial fibrillation defines recurrence as atrial fibrillation, atrial flutter, or atrial tachycardia lasting 30 seconds or longer after a 3-month blanking period. It recommends ECG documentation at follow-up visits, ambulatory monitoring at intervals, and event recording during symptoms; implantable recorders give the most complete assessment. Early recurrence within the blanking period predicts later recurrence but does not by itself define failure. The 2023 AF guideline notes that the 30-second definition may be too simplistic and that burden may be a more meaningful end point.
Young patients with atrial fibrillation. A triggering supraventricular tachycardia (AV nodal or AV re-entrant tachycardia) is found in about one-quarter of patients younger than 30 years. Documenting the onset of episodes on ambulatory recordings can identify a re-entrant tachycardia whose ablation prevents further atrial fibrillation.
Bradyarrhythmias, pauses, and conduction disease
Ambulatory monitoring is the main test for intermittent bradycardia. Its tasks are to correlate symptoms with the rhythm, to identify findings that indicate disease of the conduction system whether or not symptoms occur, and to separate physiologic bradycardia from pathologic bradycardia.
Sinus node dysfunction appears as sinus bradycardia, sinus pauses or arrest, sinoatrial exit block, chronotropic incompetence, or tachy-brady syndrome with pauses on termination of atrial fibrillation. The ISHNE-HRS statement notes that formal thresholds are arbitrary and that a sinus rate below 40 beats/min is considered potentially causative in a symptomatic patient. Pacing is indicated for symptoms attributable to the bradycardia. A rate or pause length alone is not an indication. Asymptomatic sinus bradycardia and nocturnal pauses do not warrant pacing.
AV block. Paroxysmal AV block on a monitor appears as sudden failure of conduction with a constant or accelerating sinus rate, which indicates intrinsic conduction system disease. Gradual sinus slowing with PR prolongation before block, often at night or with nausea or pain, indicates a vagal mechanism and does not by itself warrant pacing. Mobitz II, high-grade, and third-degree block are significant whether symptomatic or not.
Chronotropic assessment. The heart rate histogram and the maximum rate during daily activity supplement exercise testing in suspected chronotropic incompetence and in patients taking rate-slowing drugs.
Pacing indications, the localization of AV block, and the full diagnostic evaluation are covered in the bradycardia and AV block hub, and temporary pacing for unstable bradycardia in the temporary cardiac pacing hub.
Premature ventricular complexes, nonsustained ventricular tachycardia, and ventricular arrhythmia
PVC burden. Ambulatory monitoring quantifies premature ventricular complexes as a count and as a percentage of all beats, identifies their morphologies, coupling intervals, and relation to heart rate, and documents couplets and nonsustained runs. In 174 patients referred for ablation of frequent idiopathic PVCs, burden on 24-hour Holter was 33% in those with reduced ejection fraction versus 13% in those with normal function. A burden above 24% best separated the groups (sensitivity 79%, specificity 78%), and the lowest burden associated with reversible cardiomyopathy was 10%. Burden varies from day to day, so a single 24-hour recording can misclassify a patient near a threshold, and longer recordings give a more stable estimate. The ISHNE-HRS statement recommends monitoring in newly diagnosed nonischemic cardiomyopathy when arrhythmia-induced dysfunction is suspected (Class I). A 12-lead Holter localizes the origin before ablation. Evaluation and treatment of PVC-induced cardiomyopathy are covered in the ventricular arrhythmia hub.
Nonsustained ventricular tachycardia (NSVT) is three or more consecutive ventricular beats at a rate above 100 beats/min lasting less than 30 seconds. Its significance depends on the substrate. In structurally normal hearts, monomorphic outflow-tract runs are usually benign. In hypertrophic cardiomyopathy, NSVT on 24- to 48-hour monitoring is part of sudden death risk assessment, recommended at the initial evaluation and at 1- to 2-year intervals (Class 1, 2024 AHA/ACC guideline), as described in the hypertrophic cardiomyopathy hub. In arrhythmogenic right ventricular cardiomyopathy, more than 500 ventricular extrasystoles in 24 hours is a minor diagnostic criterion of the 2010 Task Force criteria, and NSVT of left bundle branch morphology with superior axis is a major criterion. After myocardial infarction, NSVT has limited independent prognostic value in reperfused patients receiving beta-blockers once ejection fraction is considered. The ISHNE-HRS statement finds a 24- to 48-hour Holter useful in patients with healed infarction and an ejection fraction of 35% to 40% to detect NSVT and select patients for electrophysiology study (Class IIa), following the approach of the Multicenter Unsustained Tachycardia Trial (MUSTT).
Symptoms suspected to be ventricular arrhythmia. The 2017 AHA/ACC/HRS guideline recommends ambulatory ECG monitoring to evaluate whether palpitations, presyncope, or syncope are caused by ventricular arrhythmia (Class 1). It finds an implantable cardiac monitor useful for sporadic symptoms (Class 2a). A recorded wide-complex tachycardia on a single-lead monitor is often difficult to classify, and aberrant supraventricular tachycardia, pre-excited tachycardia, and artifact are the main alternatives. The criteria for wide-complex tachycardia are covered in the ventricular arrhythmia hub.
Monitoring after ablation or drug therapy. Older criteria for drug efficacy required a 75% reduction in PVCs and a 90% reduction in runs on repeated 24-hour Holters, but spontaneous variability limits the reproducibility of 24-hour recordings. Recurrence after ventricular tachycardia ablation in patients without a defibrillator is documented by 12-lead ECG or ambulatory monitoring.
Inherited arrhythmia syndromes, pre-excitation, and QT monitoring
Long QT syndrome. The resting 12-lead ECG is the diagnostic standard. Twelve-lead Holter recording adds the QT–RR relation across the full range of heart rates, nocturnal and post-pause QT prolongation (characteristic of long QT syndrome types 2 and 3), T-wave alternans, and arrhythmias during treatment, including the effect of beta-blockers on heart rate. QT measurement methods and QT monitoring with high-risk drugs are covered in the long QT hub.
Brugada syndrome. The type 1 pattern fluctuates. In a registry of 251 patients, a spontaneous intermittent type 1 pattern appeared on 12-lead 24-hour recordings in 20% of patients whose diagnosis had been made by drug challenge. Placing the V1 and V2 electrodes in the third intercostal space on a 12-lead Holter detects the pattern more often than conventional lead placement. Extended monitoring is reasonable in unexplained syncope when Brugada syndrome is suspected.
Catecholaminergic polymorphic ventricular tachycardia. Exercise testing is the main test; ambulatory monitoring records ectopy and runs during daily exertion and emotion and assesses treatment. Ectopy typically appears at heart rates of about 110 to 130 beats/min and increases in complexity with rate.
Pre-excitation. Intermittent pre-excitation on an ambulatory recording, or abrupt loss of the delta wave with sinus acceleration, indicates an accessory pathway with a long antegrade refractory period and a lower risk of rapid conduction during atrial fibrillation. The ISHNE-HRS statement recommends monitoring to evaluate pathway conduction and detect runs of atrial fibrillation (Class I). Risk stratification of the asymptomatic pattern is covered in the supraventricular rhythms hub.
Families of sudden death victims. Ambulatory monitoring is part of the phenotype-guided evaluation of first-degree relatives, with the ECG, exercise testing, echocardiography, and cardiac magnetic resonance.
Drug development. Continuous 12-lead Holter recordings in early-phase clinical studies allow QT measurement over many beats and concentration–QT modeling, which has reduced the need for separate thorough QT studies.
Ambulatory ST-segment monitoring and silent ischemia
Holter monitoring can record transient ST-segment depression during daily life. The 1999 ACC/AHA guideline defines an ambulatory ischemic episode as horizontal or downsloping ST-segment depression of 1 mm (0.1 mV) or more lasting at least 1 minute and separated from other episodes by at least 1 minute. Valid ST analysis requires a recorder with a low-frequency response to 0.05 Hz, leads chosen to match the territory (CM5 and an inferior lead), a supine and upright baseline, and exclusion of positional and rate-related changes.
In stable coronary disease, about half of patients have transient ST depression on ambulatory monitoring, most of it asymptomatic, and ambulatory ischemia predicted outcome in observational studies. In acute coronary syndromes, ischemia on continuous monitoring predicted short- and long-term risk. Ambulatory ST monitoring has been replaced in practice by stress imaging and coronary CT angiography for diagnosis. The method cannot separate ischemia from the ST changes produced by left ventricular hypertrophy, bundle branch block, digoxin, hyperventilation, body position, and electrolyte shifts.
The remaining clinical use is in suspected vasospastic angina, in which ambulatory recording during spontaneous rest pain can document transient ST elevation.
Holter risk markers: heart rate variability, heart rate turbulence, signal-averaged ECG, and T-wave alternans
These markers quantify autonomic tone, repolarization instability, and slow conduction in scar. All of them predict mortality in some populations, and none has been shown to select patients who benefit from a defibrillator. The ISHNE-HRS statement and US guidelines do not recommend them for routine sudden death risk stratification.
Heart rate variability measures beat-to-beat variation of normal-to-normal (NN) intervals. Time-domain measures include SDNN (standard deviation of all NN intervals over 24 hours, reflecting total variability), SDANN (standard deviation of the 5-minute mean NN intervals), RMSSD (root mean square of successive differences, reflecting vagal modulation), and pNN50 (percentage of successive NN intervals that differ by more than 50 ms). Frequency-domain analysis divides spectral power into high-frequency (0.15 to 0.40 Hz, vagal and respiratory), low-frequency (0.04 to 0.15 Hz, sympathetic and vagal with baroreflex), very-low-frequency, and ultra-low-frequency bands. Ectopic beats and artifact must be removed before analysis. After myocardial infarction, an SDNN below 50 ms was associated with a 5.3-fold higher mortality than an SDNN above 100 ms in a cohort of 808 patients. In ATRAMI, an SDNN below 70 ms carried a relative risk of cardiac death of 3.2. Low variability predicts death from heart failure more than sudden death.
Heart rate turbulence describes the response of the sinus rate to a single premature ventricular complex: brief acceleration followed by deceleration, mediated by the baroreflex. Turbulence onset is the relative change of the RR intervals immediately after the PVC compared with those before it; turbulence slope is the steepest slope of the RR intervals over the following beats. Turbulence onset of 0% or more and turbulence slope of 2.5 ms per RR interval or less are abnormal, and at least five suitable PVCs are needed. Abnormal turbulence predicted mortality after myocardial infarction in the MPIP and EMIAT populations and in later reperfusion-era cohorts.
Signal-averaged ECG averages several hundred QRS complexes to reduce noise below about 1 µV and detects late potentials, low-amplitude high-frequency signals at the end of the QRS that reflect slow conduction in scar. With a 40-Hz high-pass filter, late potentials are present when the filtered QRS duration exceeds 114 ms, the root-mean-square voltage of the terminal 40 ms is below 20 µV, or the duration of low-amplitude signals below 40 µV exceeds 38 ms. Two of three abnormal values are usually required. Late potentials are a minor criterion in arrhythmogenic right ventricular cardiomyopathy. Their negative predictive value after infarction is high and their positive predictive value low.
Microvolt T-wave alternans is beat-to-beat alternation of T-wave amplitude at the microvolt level, measured during exercise or pacing by spectral analysis or from ambulatory recordings by the modified moving average method. In the SCD-HeFT substudy of 490 patients with heart failure, alternans testing did not predict arrhythmic events (hazard ratio 1.24, not significant).
| Marker | What it measures | Abnormal value | Evidence and current role |
|---|---|---|---|
| SDNN (heart rate variability) | Overall autonomic modulation of sinus rate over 24 h | Below 50 ms (below 70 ms in ATRAMI) | Predicts total and heart failure mortality after MI and in heart failure; does not select ICD candidates |
| Heart rate turbulence | Baroreflex response to single PVCs | Turbulence onset 0% or more; slope 2.5 ms/RR or less | Predicts mortality after MI and in heart failure; needs at least 5 PVCs |
| Signal-averaged ECG | Late potentials from slow conduction in scar | Filtered QRS over 114 ms; RMS40 under 20 µV; LAS40 over 38 ms (40-Hz filter) | High negative predictive value after MI; minor ARVC criterion |
| Microvolt T-wave alternans | Repolarization instability | Positive or indeterminate spectral test | Did not predict arrhythmic events in the SCD-HeFT substudy |
| Nonsustained VT on Holter | Substrate and trigger | 3 or more beats at a rate above 100 beats/min | Part of HCM risk assessment; EF 35%–40% after MI leads to EP study |
Holter-based risk markers. ARVC, arrhythmogenic right ventricular cardiomyopathy; EF, ejection fraction; EP, electrophysiology; HCM, hypertrophic cardiomyopathy; ICD, implantable cardioverter-defibrillator; LAS40, low-amplitude signal duration; MI, myocardial infarction; PVC, premature ventricular complex; RMS40, terminal 40-ms root-mean-square voltage.
Ambulatory monitoring in patients with pacemakers, defibrillators, and resynchronization devices
Implanted devices record their own diagnostics, and routine ambulatory monitoring after implantation is not needed. External monitoring remains useful in specific situations. The ISHNE-HRS statement gives a Class IIa recommendation for symptoms that device interrogation does not explain. Other situations are suspected intermittent loss of capture, undersensing, or oversensing; upper-rate behavior during exercise; slow ventricular tachycardia below the detection zone of a defibrillator; and verification that the biventricular pacing percentage reported by a resynchronization device is not inflated by fusion and pseudofusion beats. In a 12-lead Holter study of 19 resynchronization patients with permanent atrial fibrillation and more than 90% biventricular pacing on the device counter, 53% had ineffective pacing, with fewer than 90% fully paced beats because of fusion and pseudofusion. Effective capture and the biventricular pacing percentage are covered in the cardiac resynchronization therapy hub. Monitoring is also used before device reimplantation after extraction, to establish whether pacing is still needed.
A recorded pacemaker rhythm is interpreted with knowledge of the programmed mode and timing; reading the paced tracing is covered in the pacemaker hub.
Wearable cardioverter-defibrillator. A wearable defibrillator is a vest with electrodes that detects and shocks ventricular tachycardia and fibrillation and stores ECG recordings that can be reviewed remotely. It is a temporary treatment for selected patients at high short-term risk and not a diagnostic monitor. Its indications, evidence, and wear time are covered in the implantable cardioverter-defibrillator hub.
Ambulatory monitoring in athletes, older adults, pregnancy, kidney disease, sleep apnea, and congenital heart disease
Athletes. Monitoring is most useful when symptoms occur during sport and cannot be reproduced on an exercise test. The monitor is worn during the athlete's actual training or competition when possible, because the symptom may depend on the sport's intensity and conditions. Resting sinus bradycardia, sinus pauses during sleep, first-degree AV block, and nocturnal Mobitz I block are training-related findings. Frequent or complex ventricular ectopy found on screening warrants quantification and evaluation for underlying heart disease.
Older adults. Atrial and ventricular ectopy and short runs of atrial tachycardia increase with age and are common in asymptomatic older adults. Monitoring identifies tachy-brady syndrome and drug-related bradycardia, and the evaluation of falls of uncertain cause may include monitoring when syncope cannot be excluded. Skin fragility under adhesive electrodes and difficulty operating patient-activated devices favor continuous recorders with automatic detection.
Pregnancy. Palpitations are common in pregnancy and are usually due to sinus tachycardia or premature beats. Ambulatory monitoring involves no radiation and is the standard test for documenting suspected arrhythmia during pregnancy.
Kidney failure treated with hemodialysis. Patients receiving hemodialysis have high rates of sudden death. Arrhythmias cluster around dialysis sessions; atrial fibrillation begins more often on dialysis days and during dialysis, and monitoring around dialysis can detect asymptomatic atrial fibrillation that changes anticoagulation decisions.
Sleep apnea. Nocturnal bradycardia, pauses, AV block, and paroxysmal atrial fibrillation are common in sleep apnea, and nocturnal bradyarrhythmia on a Holter in an otherwise asymptomatic patient raises the question of untreated sleep-disordered breathing.
Adult congenital heart disease. Patients with repaired tetralogy of Fallot, atrial switch or Fontan operations, and other repairs are monitored for intra-atrial re-entrant tachycardia, sinus node dysfunction, AV block, and ventricular arrhythmia. Atrial flutter with 2:1 conduction can be difficult to recognize on a single lead.
Normal findings in ambulatory recordings of healthy adults
In 24-hour recordings from 50 healthy young men, waking maximal sinus rates ranged from 107 to 180 beats/min, marked sinus arrhythmia was present in 50%, sinus pauses longer than 1.75 seconds in 28%, and transient nocturnal Mobitz I AV block in 6%. Premature atrial complexes occurred in 56% and premature ventricular complexes in 50%, but only one subject had more than 100 atrial complexes and one more than 50 ventricular complexes in 24 hours.
These observations underlie several conventions. Nocturnal sinus rates in the 30s beats/min and Mobitz I block during sleep are physiologic in young and athletic adults, and isolated atrial and ventricular ectopy is common. Frequent ectopy, complex ventricular ectopy, Mobitz II or higher-grade block, and pauses longer than 3 seconds while awake are abnormal and warrant evaluation. Premature atrial complexes increase with age; excessive supraventricular ectopy in older adults and its relation to atrial fibrillation are discussed in the supraventricular rhythms hub.
The Holter and patch report: elements and interpretation
Automated software detects and classifies all beats by comparing them with templates of the dominant morphology, then groups events and generates trends. A technician reviews the classification, corrects misclassified beats, verifies events, and correlates diary entries with the rhythm. The physician interprets the reviewed recording. Proprietary algorithms are rarely validated in published studies, and the physician's review of the strips is the basis of the report. The summary counts are not.
| Report element | Content | Interpretive points |
|---|---|---|
| Recording quality | Total and analyzable time; leads; artifact | Low analyzable time limits a negative result |
| Underlying rhythm and heart rate | Minimum, maximum, and mean rates with times; hourly trend; histogram | Minimum rate during sleep is often physiologic; maximum rate shows chronotropic response |
| Pauses | Number, longest pause with time, and mechanism (sinus arrest, AV block, post-conversion) | Pauses longer than 3 s while awake are abnormal; nocturnal sinus pauses in young adults are common |
| AV conduction | PR interval; episodes of second- or third-degree block and their type | Vagal Mobitz I is separated from Mobitz II or paroxysmal block |
| Supraventricular ectopy | Count, percentage, couplets, runs; longest and fastest supraventricular tachycardia | Onset and termination identify mechanism; frequent ectopy predicts atrial fibrillation |
| Atrial fibrillation or flutter | Episodes, burden, longest episode, ventricular rate during episodes | Visual confirmation of new diagnoses |
| Ventricular ectopy | Count, burden, morphologies, couplets, bigeminy, runs; longest and fastest run | Burden and morphology guide evaluation for PVC cardiomyopathy and ablation |
| Paced rhythm | Percentage paced; capture and sensing | Fusion and pseudofusion overstate effective pacing |
| ST segment | Episodes, maximal deviation, lead, timing, heart rate at onset | Valid only with suitable leads and frequency response |
| QT interval | QTc trend, maximum QTc with time | Automated values require manual verification |
| Symptom correlation | Symptoms with the rhythm at that time | The central finding of a diagnostic study, including symptoms during sinus rhythm |
Elements of an ambulatory ECG report. AV, atrioventricular; PVC, premature ventricular complex.
Artifact and false detections in external and implantable recorders
Artifact is the most frequent source of error. Motion and muscle artifact, loose electrodes, and broken lead wires create deflections that simulate atrial flutter or fibrillation, ventricular tachycardia, or ventricular fibrillation; intermittent loss of signal simulates sinus arrest, AV block, or pacemaker failure. On a multichannel recording, artifact usually appears in one channel only, and native QRS complexes can be identified marching through it at the underlying rate. The absence of a change in heart rate before and after an apparent arrhythmia, normal QRS complexes within it, and an accelerometer signal showing vigorous activity support artifact. On single-lead devices, these clues are fewer.
Implantable recorders have device-specific false detections. Undersensing of low-amplitude R waves produces false pauses and false atrial fibrillation, and oversensing of T waves or myopotentials produces false tachycardia. False automatic detections are common, and all stored episodes are reviewed before a diagnosis is made. A device-recorded ventricular fibrillation episode in an ambulatory patient without symptoms is reviewed for artifact before any treatment is escalated.
Errors of interpretation carry consequences in both directions. Errors of commission include unnecessary drugs, catheterization, electrophysiology study, and device implantation for artifact; errors of omission include missing a true arrhythmia that the software labeled as noise. Artifact on the 12-lead ECG and in-hospital monitors is described in the ECG interpretation hub.
Skin reactions, adherence, and data quality
Adhesive electrodes cause erythema after removal in most patients and contact dermatitis in some, from the acrylate adhesive or the gel. Pruritus and blistering lead to early removal. Rotating electrode sites, hypoallergenic electrodes, and an adhesive-free interval between devices reduce irritation. Patch monitors have higher adherence than wired Holters. Holters are limited by electrode detachment and wire tension, external loop recorders by the need to wear electrodes for weeks, and patient-activated devices by failure to activate during symptoms.
A negative study is interpreted in light of the analyzable time, whether the usual symptom occurred, and whether the device was worn during the activities that provoke it.
Inpatient telemetry compared with ambulatory monitoring
In-hospital monitoring detects arrhythmias in patients at short-term risk, with immediate response by staff. The 2017 ACC/AHA/HRS syncope guideline recommends continuous ECG monitoring for patients hospitalized for syncope with a suspected cardiac cause (Class 1). Hospital telemetry has low yield in low-risk patients and generates alarm fatigue: 72% to 99% of hospital alarms have no clinical relevance. In one study of low-risk patients with chest pain and normal or nonspecific ECGs, arrhythmia yield was 1.5%. A short period of inpatient telemetry does not exclude intermittent arrhythmia, and patients discharged without a diagnosis are evaluated with ambulatory monitoring matched to their symptom frequency. The AHA practice standards for in-hospital monitoring are summarized in the ECG interpretation hub.
Ambulatory monitoring after hospital discharge is also used for new conduction abnormalities after transcatheter aortic valve replacement (TAVR). Some US pathways have used about 14 days of monitoring in selected higher-risk patients to detect delayed high-grade AV block, as described with conduction disturbances after TAVR.
Coding and coverage of ambulatory monitoring in the United States
Ambulatory monitoring services are billed under separate Current Procedural Terminology (CPT) code families according to duration and type of service, and the code family determines what the physician interprets and what the service includes. Medicare coverage for electrocardiographic services is defined in National Coverage Determination 20.15, and individual payers define indications and the frequency of repeat studies.
| Service | CPT codes | Duration |
|---|---|---|
| Continuous recording (Holter) | 93224–93227 | Up to 48 hours |
| Extended continuous recording | 93241–93244 | More than 48 hours to 7 days |
| Extended continuous recording | 93245–93248 | More than 7 days to 15 days |
| External event recording with memory loop | 93268–93272 | Up to 30 days |
| Mobile cardiovascular telemetry | 93228–93229 | Up to 30 days |
| Implantable loop recorder insertion and removal | 33285, 33286 | Procedure |
CPT code families for ambulatory ECG services. The extended continuous recording codes replaced temporary category III codes in 2021. All of these code families contain global, technical, and professional components; payer rules govern which codes can be reported for the same monitoring period.
Differences between US and European guidance on implantable loop recorders and atrial fibrillation screening
The 2018 European Society of Cardiology syncope guideline gives the implantable loop recorder a Class I recommendation early in the evaluation of recurrent syncope of uncertain origin in patients without high-risk criteria who are likely to have a recurrence within the battery life. It gives the same class to high-risk patients in whom a comprehensive evaluation has not found a cause and who have no conventional indication for a defibrillator or pacemaker. The 2017 ACC/AHA/HRS guideline gives the implantable monitor a Class 2a recommendation for selected patients. The European guideline also discusses implantable recorders in unexplained falls and suspected epilepsy. For atrial fibrillation screening, European guidelines endorse opportunistic screening in older adults. The US Preventive Services Task Force found the evidence insufficient to assess benefits and harms (an I statement), and the 2023 ACC/AHA/ACCP/HRS guideline makes no graded recommendation for or against screening.
Sources of error in ambulatory ECG monitoring
- Using a monitor too short for the symptom frequency: a 24-hour Holter in a patient with monthly symptoms has a low probability of recording an event, and a negative result does not exclude an arrhythmia.
- Treating a negative study as excluding arrhythmia when the usual symptom did not occur during recording. Only a recording during the typical symptom excludes an arrhythmic cause.
- Treating artifact as arrhythmia: motion and electrode artifact simulate ventricular tachycardia, atrial fibrillation, and pauses, and single-lead recordings offer few clues. Native QRS complexes marching through the artifact and an unchanged heart rate before and after identify it.
- Acting on an implantable recorder's automated episode list without reviewing the stored electrograms. Undersensing produces false pauses and false atrial fibrillation.
- Attributing symptoms to physiologic findings: nocturnal sinus bradycardia, sinus pauses during sleep, and nocturnal Mobitz I block are common in healthy adults and do not indicate pacing without symptom correlation.
- Mistaking vagally mediated AV block, preceded by sinus slowing and PR prolongation, for paroxysmal infranodal block, or the reverse. Block with a constant or accelerating sinus rate indicates conduction system disease.
- Diagnosing atrial fibrillation from a photoplethysmography notification or an automated label without visual review of an ECG tracing.
- Measuring atrial fibrillation burden from a triggered recorder, which captures onset but not offset of episodes.
- Classifying a wide-complex rhythm on a single-lead recording as ventricular tachycardia without considering aberrancy, pre-excitation, and artifact; a multichannel or 12-lead recording resolves many cases.
- Relying on a patient-activated recorder when the event causes abrupt loss of consciousness; an automatically triggered or continuous recorder is required.
- Using a 12-lead Holter recorded with torso limb electrodes for serial comparison of axis or QRS amplitude with standard ECGs.
- Interpreting ST depression on a Holter as ischemia without suitable leads, frequency response, and exclusion of positional, rate-related, and drug effects.
- Implanting a loop recorder in a patient who already meets an indication for a defibrillator or pacing; monitoring then delays treatment.
- Concluding that a single 24-hour PVC count near a threshold classifies the patient, when day-to-day variability is large.
References
- Steinberg JS, Varma N, Cygankiewicz I, et al. 2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Heart Rhythm. 2017;14:e55–e96. doi:10.1016/j.hrthm.2017.03.038
- Shen WK, Sheldon RS, Benditt DG, et al. 2017 ACC/AHA/HRS guideline for the evaluation and management of patients with syncope: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2017;136:e60–e122. doi:10.1161/CIR.0000000000000499
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation. 2019;140:e382–e482. doi:10.1161/CIR.0000000000000628
- Joglar JA, Chung MK, Armbruster AL, et al. 2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2024;149:e1–e156. doi:10.1161/CIR.0000000000001193
- Al-Khatib SM, Stevenson WG, Ackerman MJ, et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Circulation. 2018;138:e272–e391. doi:10.1161/CIR.0000000000000549
- Page RL, Joglar JA, Caldwell MA, et al. 2015 ACC/AHA/HRS guideline for the management of adult patients with supraventricular tachycardia. Circulation. 2016;133:e506–e574. doi:10.1161/CIR.0000000000000311
- Kleindorfer DO, Towfighi A, Chaturvedi S, et al. 2021 guideline for the prevention of stroke in patients with stroke and transient ischemic attack: a guideline from the American Heart Association/American Stroke Association. Stroke. 2021;52:e364–e467. doi:10.1161/STR.0000000000000375
- Ommen SR, Ho CY, Asif IM, et al. 2024 AHA/ACC/AMSSM/HRS/PACES/SCMR guideline for the management of hypertrophic cardiomyopathy. Circulation. 2024;149:e1239–e1311. doi:10.1161/CIR.0000000000001250
- Crawford MH, Bernstein SJ, Deedwania PC, et al. ACC/AHA guidelines for ambulatory electrocardiography. J Am Coll Cardiol. 1999;34:912–948. doi:10.1016/s0735-1097(99)00354-x
- Brignole M, Moya A, de Lange FJ, et al. 2018 ESC guidelines for the diagnosis and management of syncope. Eur Heart J. 2018;39:1883–1948. doi:10.1093/eurheartj/ehy037
- Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation: executive summary. Europace. 2018;20:157–208. doi:10.1093/europace/eux275
- US Preventive Services Task Force, Davidson KW, Barry MJ, et al. Screening for atrial fibrillation: US Preventive Services Task Force recommendation statement. JAMA. 2022;327:360–367. doi:10.1001/jama.2021.23732
- Sanna T, Diener HC, Passman RS, et al. Cryptogenic stroke and underlying atrial fibrillation. N Engl J Med. 2014;370:2478–2486. doi:10.1056/NEJMoa1313600
- Brachmann J, Morillo CA, Sanna T, et al. Uncovering atrial fibrillation beyond short-term monitoring in cryptogenic stroke patients: three-year results from the Cryptogenic Stroke and Underlying Atrial Fibrillation trial. Circ Arrhythm Electrophysiol. 2016;9:e003333. doi:10.1161/CIRCEP.115.003333
- Gladstone DJ, Spring M, Dorian P, et al. Atrial fibrillation in patients with cryptogenic stroke. N Engl J Med. 2014;370:2467–2477. doi:10.1056/NEJMoa1311376
- Wachter R, Gröschel K, Gelbrich G, et al. Holter-electrocardiogram-monitoring in patients with acute ischaemic stroke (Find-AFRANDOMISED): an open-label randomised controlled trial. Lancet Neurol. 2017;16:282–290. doi:10.1016/S1474-4422(17)30002-9
- Buck BH, Hill MD, Quinn FR, et al. Effect of implantable vs prolonged external electrocardiographic monitoring on atrial fibrillation detection in patients with ischemic stroke: the PER DIEM randomized clinical trial. JAMA. 2021;325:2160–2168. doi:10.1001/jama.2021.6128
- Bernstein RA, Kamel H, Granger CB, et al. Effect of long-term continuous cardiac monitoring vs usual care on detection of atrial fibrillation in patients with stroke attributed to large- or small-vessel disease: the STROKE-AF randomized clinical trial. JAMA. 2021;325:2169–2177. doi:10.1001/jama.2021.6470
- Svendsen JH, Diederichsen SZ, Højberg S, et al. Implantable loop recorder detection of atrial fibrillation to prevent stroke (the LOOP Study): a randomised controlled trial. Lancet. 2021;398:1507–1516. doi:10.1016/S0140-6736(21)01698-6
- Svennberg E, Friberg L, Frykman V, et al. Clinical outcomes in systematic screening for atrial fibrillation (STROKESTOP): a multicentre, parallel group, unmasked, randomised controlled trial. Lancet. 2021;398:1498–1506. doi:10.1016/S0140-6736(21)01637-8
- Lopes RD, Atlas SJ, Go AS, et al. Effect of screening for undiagnosed atrial fibrillation on stroke prevention. J Am Coll Cardiol. 2024;84:2073–2084. doi:10.1016/j.jacc.2024.08.019
- Steinhubl SR, Waalen J, Edwards AM, et al. Effect of a home-based wearable continuous ECG monitoring patch on detection of undiagnosed atrial fibrillation: the mSToPS randomized clinical trial. JAMA. 2018;320:146–155. doi:10.1001/jama.2018.8102
- Halcox JPJ, Wareham K, Cardew A, et al. Assessment of remote heart rhythm sampling using the AliveCor heart monitor to screen for atrial fibrillation: the REHEARSE-AF study. Circulation. 2017;136:1784–1794. doi:10.1161/CIRCULATIONAHA.117.030583
- Perez MV, Mahaffey KW, Hedlin H, et al. Large-scale assessment of a smartwatch to identify atrial fibrillation. N Engl J Med. 2019;381:1909–1917. doi:10.1056/NEJMoa1901183
- Lubitz SA, Faranesh AZ, Selvaggi C, et al. Detection of atrial fibrillation in a large population using wearable devices: the Fitbit Heart Study. Circulation. 2022;146:1415–1424. doi:10.1161/CIRCULATIONAHA.122.060291
- Guo Y, Wang H, Zhang H, et al. Mobile photoplethysmographic technology to detect atrial fibrillation. J Am Coll Cardiol. 2019;74:2365–2375. doi:10.1016/j.jacc.2019.08.019
- Barrett PM, Komatireddy R, Haaser S, et al. Comparison of 24-hour Holter monitoring with 14-day novel adhesive patch electrocardiographic monitoring. Am J Med. 2014;127:95.e11–95.e17. doi:10.1016/j.amjmed.2013.10.003
- Turakhia MP, Hoang DD, Zimetbaum P, et al. Diagnostic utility of a novel leadless arrhythmia monitoring device. Am J Cardiol. 2013;112:520–524. doi:10.1016/j.amjcard.2013.04.017
- Kinlay S, Leitch JW, Neil A, et al. Cardiac event recorders yield more diagnoses and are more cost-effective than 48-hour Holter monitoring in patients with palpitations: a controlled clinical trial. Ann Intern Med. 1996;124:16–20. doi:10.7326/0003-4819-124-1_part_1-199601010-00003
- Zimetbaum PJ, Kim KY, Josephson ME, et al. Diagnostic yield and optimal duration of continuous-loop event monitoring for the diagnosis of palpitations: a cost-effectiveness analysis. Ann Intern Med. 1998;128:890–895. doi:10.7326/0003-4819-128-11-199806010-00002
- Sivakumaran S, Krahn AD, Klein GJ, et al. A prospective randomized comparison of loop recorders versus Holter monitors in patients with syncope or presyncope. Am J Med. 2003;115:1–5. doi:10.1016/s0002-9343(03)00233-x
- Rothman SA, Laughlin JC, Seltzer J, et al. The diagnosis of cardiac arrhythmias: a prospective multi-center randomized study comparing mobile cardiac outpatient telemetry versus standard loop event monitoring. J Cardiovasc Electrophysiol. 2007;18:241–247. doi:10.1111/j.1540-8167.2006.00729.x
- Krahn AD, Klein GJ, Yee R, et al. Randomized assessment of syncope trial: conventional diagnostic testing versus a prolonged monitoring strategy. Circulation. 2001;104:46–51. doi:10.1161/01.cir.104.1.46
- Edvardsson N, Frykman V, van Mechelen R, et al. Use of an implantable loop recorder to increase the diagnostic yield in unexplained syncope: results from the PICTURE registry. Europace. 2011;13:262–269. doi:10.1093/europace/euq418
- Brignole M, Menozzi C, Moya A, et al. Pacemaker therapy in patients with neurally mediated syncope and documented asystole: Third International Study on Syncope of Uncertain Etiology (ISSUE-3): a randomized trial. Circulation. 2012;125:2566–2571. doi:10.1161/CIRCULATIONAHA.111.082313
- Mittal S, Sanders P, Pokushalov E, et al. Safety profile of a miniaturized insertable cardiac monitor: results from two prospective trials. Pacing Clin Electrophysiol. 2015;38:1464–1469. doi:10.1111/pace.12752
- Baman TS, Lange DC, Ilg KJ, et al. Relationship between burden of premature ventricular complexes and left ventricular function. Heart Rhythm. 2010;7:865–869. doi:10.1016/j.hrthm.2010.03.036
- Brodsky M, Wu D, Denes P, et al. Arrhythmias documented by 24 hour continuous electrocardiographic monitoring in 50 male medical students without apparent heart disease. Am J Cardiol. 1977;39:390–395. doi:10.1016/s0002-9149(77)80094-5
- Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Heart rate variability: standards of measurement, physiological interpretation and clinical use. Circulation. 1996;93:1043–1065. doi:10.1161/01.cir.93.5.1043
- Kleiger RE, Miller JP, Bigger JT Jr, et al. Decreased heart rate variability and its association with increased mortality after acute myocardial infarction. Am J Cardiol. 1987;59:256–262. doi:10.1016/0002-9149(87)90795-8
- La Rovere MT, Bigger JT Jr, Marcus FI, et al. Baroreflex sensitivity and heart-rate variability in prediction of total cardiac mortality after myocardial infarction (ATRAMI). Lancet. 1998;351:478–484. doi:10.1016/s0140-6736(97)11144-8
- Schmidt G, Malik M, Barthel P, et al. Heart-rate turbulence after ventricular premature beats as a predictor of mortality after acute myocardial infarction. Lancet. 1999;353:1390–1396. doi:10.1016/S0140-6736(98)08428-1
- Bauer A, Malik M, Schmidt G, et al. Heart rate turbulence: standards of measurement, physiological interpretation, and clinical use: International Society for Holter and Noninvasive Electrophysiology consensus. J Am Coll Cardiol. 2008;52:1353–1365. doi:10.1016/j.jacc.2008.07.041
- Breithardt G, Cain ME, el-Sherif N, et al. Standards for analysis of ventricular late potentials using high-resolution or signal-averaged electrocardiography: a statement by a task force committee of the European Society of Cardiology, the American Heart Association, and the American College of Cardiology. Eur Heart J. 1991;12:473–480. doi:10.1093/oxfordjournals.eurheartj.a059926
- Gold MR, Ip JH, Costantini O, et al. Role of microvolt T-wave alternans in assessment of arrhythmia vulnerability among patients with heart failure and systolic dysfunction: primary results from the T-wave alternans sudden cardiac death in heart failure trial substudy. Circulation. 2008;118:2022–2028. doi:10.1161/CIRCULATIONAHA.107.748962
- Kamath GS, Cotiga D, Koneru JN, et al. The utility of 12-lead Holter monitoring in patients with permanent atrial fibrillation for the identification of nonresponders after cardiac resynchronization therapy. J Am Coll Cardiol. 2009;53:1050–1055. doi:10.1016/j.jacc.2008.12.022
- Marcus FI, McKenna WJ, Sherrill D, et al. Diagnosis of arrhythmogenic right ventricular cardiomyopathy/dysplasia: proposed modification of the Task Force criteria. Circulation. 2010;121:1533–1541. doi:10.1161/CIRCULATIONAHA.108.840827