CardioBoardExam.com

Library › Imaging › Perfusion imaging artifacts

Artifacts and quality control in myocardial perfusion imaging

The technical causes of false perfusion defects and falsely normal studies on SPECT and PET, their recognition and correction, and the camera, laboratory, and reporting quality controls that protect diagnostic accuracy.

A perfusion defect is a region of reduced detected counts. Reduced coronary flow is one cause; the others are physical and procedural, and on short-axis slices they can look identical to coronary disease. Soft-tissue attenuation, patient motion, extracardiac activity, conduction disease, attenuation-map errors, camera faults, processing errors, and inadequate stress produce false defects or falsely normal studies, and all leave evidence elsewhere in the dataset. This page covers those mechanisms for single-photon emission computed tomography (SPECT) and positron emission tomography (PET), the acquisitions that resolve them, the quality control that prevents them, and the report language that conveys residual uncertainty. Stress protocols, scoring, and prognosis are covered in the myocardial perfusion imaging hub, and stressor pharmacology in the pharmacologic stress testing hub. Tracer properties are covered in the myocardial perfusion radiotracers hub, and the physics of attenuation, detection, and reconstruction in the nuclear cardiology physics hub.

The imaging chain: where a false defect can arise

Between the injected activity and the displayed slice lie seven sequential processes. Three are biological: delivery of tracer by coronary flow, extraction across the capillary and cell membranes, and retention by viable myocytes. Four are physical: radioactive emission, transmission of the photon through the body, detection by the camera, and reconstruction of projections into slices. Regional brightness on the final image is the product of all seven terms after display normalization, and the slice alone cannot show which term is reduced. A normal inferior wall in a large man illustrates the problem. Flow, extraction, retention, and emission are all normal, but photons from that wall cross the diaphragm and abdominal contents, where a substantial fraction is absorbed or scattered. Fewer counts reach the detector, and reconstruction renders the loss as a defect.

ProcessEventFailures that reduce counts without coronary diseaseDataset that exposes the failure
DeliveryCoronary flow carries tracer to the capillary bedSubmaximal exercise; methylxanthine exposure before vasodilator stress; injection after the stress state has ended; infiltrated injectionStress record; projection cine (injection site)
ExtractionTracer crosses endothelium and cell membraneExtraction roll-off at high flow compresses flow differences (a sensitivity limit that creates no false defect)Tracer choice; absolute flow on PET
RetentionViable myocytes hold tracer through acquisitionLabeling failure (free pertechnetate) lowers myocardial uptake globallyThyroid, salivary, and gastric activity on projections; radiochemical purity record
EmissionNucleus decays and emits a photonDecay during long acquisitions or late imaging lowers countsCount density
TransmissionPhoton leaves the thoraxBreast, diaphragm, chest wall, and abdominal attenuation and scatterProjection cine; gated thickening; prone or upright images; attenuation correction
DetectionCamera accepts the event in its energy and position windowsEnergy-window drift, nonuniformity, center-of-rotation error, collimator damage, detector-module failure, dead timeDaily quality control record; repetition across patients
ReconstructionProjections become reoriented slicesMotion, CT-emission misregistration, filter or iteration choice, incorrect axis, normal-database mismatchProjection cine; fused CT display; processing parameters

The seven processes that determine regional counts on a perfusion image. Assigning a defect to one of them identifies the data that can confirm or exclude it.

Reduced flow and scar are the only causes that represent coronary disease, and both have signatures of their own. Ischemia is reversible between stress and rest and follows a coronary distribution. Scar is fixed, follows a coronary distribution, and lacks systolic thickening. All other causes are excluded before a defect is attributed to the coronary circulation. The opposite error is a falsely normal study. It arises when stress was inadequate, when counts are too low to resolve a mild defect, when hot extracardiac activity or a misregistered map adds spurious counts, or when all territories are reduced by a similar proportion so that the relative image is uniform.

The systematic review sequence and the raw projection cine

The order of review is part of quality control. Interpreting slices first tends to bias the later technical review, so the raw data are examined before any perfusion conclusion is reached, and the technical mechanisms are excluded in order.

  1. Raw rotating projections. The cine of the planar projections shows motion, upward creep, breast and diaphragm shadows, arm position, extracardiac activity, the injection site, thyroid and gastric activity from free pertechnetate, body habitus, and count density. Several of these features are visible on no other display. The sinogram and linogram display the same data as a function of angle and expose small shifts that the cine misses.
  2. Tomographic slices, corrected and uncorrected. Short-axis, vertical long-axis, and horizontal long-axis slices are reviewed with stress above rest. Where attenuation correction was applied, the uncorrected slices are displayed alongside the corrected slices. A defect present on one dataset and absent on the other identifies an attenuation or correction mechanism.
  3. Positional acquisitions. A defect that moves or resolves between supine and prone or upright images is attributed to attenuation.
  4. Gated cine, wall motion, and thickening. Preserved thickening in a fixed defect argues against a large transmural scar and supports attenuation.
  5. The stress record and clinical data. Inadequate stress reduces the flow difference on which a defect depends.
  6. Polar map and scores. Quantitative output is compared with the conclusion the raw data already support; a score is not reported when the images do not support it.

American Society of Nuclear Cardiology (ASNC) practice guidance recommends that the initial read be performed with knowledge of only the patient's sex, height, and weight, with clinical data reviewed after a preliminary impression is formed. It also recommends that images be read from the monitor with series normalization, in which a set of slices is scaled to its own brightest pixel. The same guidance states that most artifacts produce mild defects, which is why a severe defect is less likely to be technical and why mild defects receive the most scrutiny.

Soft-tissue attenuation: breast, diaphragm, chest wall, and body habitus

Attenuation is photon loss by photoelectric absorption and Compton scatter in tissue between the heart and the detector. Transmission falls exponentially with the attenuation coefficient and path length (I = I₀e−μx), so a modest increase in tissue depth removes a disproportionate fraction of photons. The 140-keV photons of technetium-99m are attenuated less than the 69- to 83-keV x-rays of thallium-201, which is one reason thallium attenuation artifacts are more pronounced. Attenuation is the most frequent cause of false-positive SPECT studies, and its pattern depends on anatomy.

PatternUsual settingClue on raw projectionsFunctional clueResolving step
Anterior, anterolateral, or anteroseptal reductionBreast tissue, large or dense breasts, breast implants; more frequent in women but set by breast size and positionBreast shadow crossing the heart on the cine; implants show a sharp-edged photopenic bandPreserved thickeningAttenuation correction; prone imaging; identical breast positioning at stress and rest
Inferior and inferoseptal reductionLeft hemidiaphragm, large abdomen, male habitusDiaphragmatic outline overlying the inferior wallPreserved thickeningProne or upright imaging; attenuation correction
Lateral reductionLateral chest-wall fat; left arm at the side and in the fieldArm or chest-wall tissue within the fieldPreserved thickeningArm positioning above the head; attenuation correction
Global count loss with high noiseSevere obesityLow total counts, high scatter, greater detector distanceFunction may be uninterpretable at very low countsTwo-day protocol with full activity; longer acquisition; PET

Attenuation patterns and the data that identify them. Thickening refers to systolic count increase on gated images.

Breast attenuation. The location of the breast relative to the heart determines the defect: a pendulous breast projects over the inferior and anterior walls, a high breast over the anterior wall and base, and a lateral breast over the anterolateral wall. Because breast position can differ between the stress and rest acquisitions, a breast artifact can appear more severe on one and mimic a reversible defect. Imaging the patient in the same position and clothing for both acquisitions, with breast position documented, prevents this. Breast binding or flattening is used in some laboratories.

Diaphragmatic attenuation. In supine imaging the left hemidiaphragm and abdominal contents lie directly beneath the inferior wall, so photons emitted from that wall cross the greatest tissue depth. The defect is inferior and inferoseptal, more prominent in men, and often worse with a large abdomen or elevated hemidiaphragm. It typically spares the apex and does not extend into the inferolateral wall in the way a right coronary or dominant circumflex lesion does.

Obesity. Large habitus degrades the study for three reasons: longer photon paths increase attenuation, scatter displaces counts out of the photopeak window, and greater chest depth increases source-to-detector distance, which degrades collimator resolution independent of attenuation. Relative noise varies as 1 ÷ √N, where N is the number of detected counts, so mild defects become indistinguishable from statistical fluctuation. A two-day technetium protocol allows full activity at both phases, a longer acquisition raises counts, and attenuation correction and prone imaging improve specificity. In a large series that combined supine and prone acquisitions with gating, sensitivity and normalcy rates were similar in normal-weight, overweight, and obese patients (normalcy 99%, 98%, and 99%). PET is preferred in severe obesity because attenuation correction is integral, count sensitivity is higher, and image quality is better maintained. Table weight limit, bore clearance, and an adequate CT field of view are confirmed before a patient with severe obesity is scheduled, because truncation of the attenuation map creates its own artifact.

Separating attenuation from scar with gated wall motion and thickening

Attenuation and scar both produce defects that are fixed between stress and rest, and count level does not separate them. Gating does. The myocardial wall is thinner than the system resolution, so recorded counts rise as the wall thickens in systole and more activity falls within a given resolution element; systolic count increase is therefore a measure of thickening. In an attenuation artifact the underlying myocardium contracts normally: absolute counts are depressed, but wall motion and thickening are preserved. Scar contains no contracting tissue, so thickening is absent and motion is reduced.

Among 551 consecutive patients with technetium sestamibi SPECT, 180 had isolated fixed defects. Defect function was abnormal in 96% of those with clinical infarction, and 91% of fixed defects with normal function occurred in women with anterior defects or men with inferior defects, the typical attenuation locations. Reclassifying fixed defects with normal function as normal reduced unexplained fixed defects from 14% to 3%. In 115 women, adding gating to technetium sestamibi SPECT identified fixed defects with preserved function as attenuation and raised specificity for stenosis of 70% or more from 84.4% to 92.2%. Specificity was measured in a group that combined patients with normal angiograms and low-likelihood volunteers.

The discriminator has limits. A small or subendocardial infarct can retain thickening, so preserved thickening argues against a large transmural scar but does not exclude a small one. On SPECT, gated images are acquired after stress, so post-ischemic stunning can reduce function in an ischemic segment. Gated function is also unreliable when gating has failed. Thickening also cannot classify a reversible defect: an attenuation artifact that varies between stress and rest because breast or diaphragm position changed shows preserved thickening at both. Positional imaging or attenuation correction settles the question.

Prone, upright, and combined positional imaging

Positional imaging changes attenuation geometry without changing tracer distribution, because sestamibi and tetrofosmin show minimal clinically useful redistribution. A second acquisition in a different position therefore separates a fixed property of the myocardium from a property of the photon path.

Prone imaging. Turning the patient prone shifts the diaphragm and abdominal contents away from the inferior wall and moves the heart closer to the anterior chest and detector. An inferior defect that resolves prone strongly supports diaphragmatic attenuation. Prone imaging also reduces patient motion, because the prone patient breathes less deeply and moves less. ASNC practice guidance states that prone imaging changes the location of attenuation without eliminating it, so that an artifact moves between supine and prone acquisitions while a true perfusion defect stays fixed. The guidance recommends prone imaging as an addition to supine imaging. Prone imaging lengthens the anterior path, which can create a new anterior or anteroseptal reduction, and it changes the heart's position relative to breast tissue.

Three rows of supine and prone SPECT polar maps and short-axis slices with matching cardiac MRI short-axis images. Row A, a woman with an anterior defect supine that resolves prone; row B, a man with an inferior defect supine that resolves prone; row C, a woman with a defect that persists prone. All three MRI images are normal.
Supine and prone technetium SPECT compared with late gadolinium enhancement cardiac magnetic resonance imaging (MRI) in patients without known coronary disease. A: an anterior defect in a woman that disappears prone (breast attenuation). B: an inferior defect in a man that disappears prone (diaphragmatic attenuation). C: a defect that persists prone although MRI shows no infarction, an attenuation artifact that prone imaging failed to correct. Columns: supine and prone polar maps (prone maps not normalized to a database), supine and prone short-axis slices, and MRI. Credit: Hedén B, Persson E, Carlsson M, Pahlm O, Arheden H, BMC Med Imaging 2009;9:16, Figure 2; CC BY 2.0.

In the study illustrated above, 52 patients without prior infarction or stress-induced ischemia had supine and prone SPECT and late gadolinium enhancement MRI. Of 63 supine defects, 37 disappeared prone, and none of the 37 corresponded to infarction on MRI. Of the 26 that persisted, MRI confirmed infarction in only 2, so a persistent defect does not prove scar. SPECT also missed 3 small scars that MRI detected.

Combined supine-prone quantification. Positional imaging became quantitative when separate supine and prone normal limits were derived. In 369 patients without known coronary disease who had angiography, combined supine-prone total perfusion deficit improved specificity for stenosis of 70% or more from 65% (supine alone) to 86%, and in obese patients with a low likelihood of disease the normalcy rate rose from 78% to 95%. Prone imaging alone had lower sensitivity than supine imaging, which is why the two are read together.

Upright and semi-upright imaging. Several dedicated cardiac cameras image the patient upright or semi-reclining. In the upright position the diaphragm descends, which reduces inferior attenuation, but breast tissue falls over the anterior and inferior walls, and the pattern of breast attenuation changes. Combining upright and supine acquisitions on a high-speed camera raised per-patient specificity to 86%, compared with 59% for upright and 73% for supine imaging alone, with sensitivity of 94% in 56 patients with angiography. Combined acquisitions require position-specific normal limits, because a prone or upright study scored against a supine database attributes the positional difference to perfusion.

CT attenuation correction on SPECT/CT

Attenuation correction estimates the attenuation coefficient for all voxels from a transmission map, now usually a low-dose CT, and reverses the photon loss during iterative reconstruction. CT measures attenuation at an effective x-ray energy far below 140 keV, so CT numbers are converted to coefficients at the emission energy by a scaling relationship. Correction is combined with scatter correction, because attenuation correction applied without scatter correction amplifies scattered counts already present and over-corrects regions next to hot structures.

The clinical effect is mainly on specificity. In a multicenter trial with technetium sestamibi, attenuation and scatter correction with resolution compensation did not change sensitivity for stenosis of 50% or more (76% uncorrected, 75% corrected, 78% corrected with quantitative analysis). The normalcy rate rose from 86% to 96%, and to 97% with quantitative analysis. The same trial showed two drawbacks: detection of multivessel disease fell, and sensitivity for right coronary lesions decreased while specificity improved, because correction raises inferior counts in all patients, including those with true inferior ischemia. The 2002 joint ASNC and Society of Nuclear Medicine position statement recommended that laboratories consider adding attenuation correction hardware and software. It described attenuation correction as a rapidly evolving standard for SPECT perfusion imaging, with corrected and uncorrected images reviewed together. A shorter joint statement on attenuation correction followed in 2004. The 2013 Society of Nuclear Medicine and Molecular Imaging (SNMMI), ASNC, and Society of Cardiovascular Computed Tomography (SCCT) hybrid imaging guideline states that, with SPECT, both uncorrected and attenuation-corrected images are reviewed when both are acquired.

Correction produces predictable changes in the normal appearance that are not perfusion abnormalities:

A defect on uncorrected images that resolves on corrected images supports attenuation; a defect present on both, after technical review, supports a true abnormality; and a defect that appears only after correction is a correction artifact until registration, truncation, and contrast in the map have been excluded.

Attenuation-map errors: misregistration, truncation, and contrast

Attenuation correction is valid only when the transmission map and the emission data describe the same anatomy in the same position. An emission acquisition lasts minutes and averages the cardiac position over many respiratory cycles, whereas a helical CT lasts seconds and captures one respiratory phase. The steepest gradient in attenuation coefficient in the thorax lies at the boundaries between lung and myocardium and between lung and diaphragm, so a small offset there assigns lung values to myocardium or myocardial values to lung. Under-correction of myocardium lying over lung on the map creates a defect, most often anterolateral or lateral when the shift is transaxial and inferior when it is craniocaudal.

Map errorMechanismAppearancePrevention or correction
MisregistrationRespiration, patient movement between CT and emission scans, upward shift of the heart after vasodilator stress, or arm repositioningDefect present only on corrected images, commonly anterolateral, lateral, or inferior; fused display shows myocardium over lung on the mapFused transaxial, coronal, and sagittal review of all emission-map pairs; manual or automated realignment and re-reconstruction; repeat or averaged (cine) CT
TruncationCT field of view smaller than the emission field; tissue outside it is treated as nonattenuatingUnder-correction at the periphery; body outline ends abruptly on the map; most common with large patients and arms at the sideArms above the head; extended-field reconstruction; check of the map outline
Contrast or metal in the mapIodine or metal raises CT numbers above the value soft tissue would give at the emission energy, inflating the coefficientOver-correction with spurious activity next to enhanced structures or devicesNoncontrast attenuation CT; system-validated scaling where contrast cannot be avoided
Separate rest and stress positions with one mapPatient moved between acquisitionsMisregistration on one of the two acquisitions only, producing an apparent reversible defectSeparate map per acquisition, or registration verified per pair

Errors in the attenuation map. All of these distort the corrected image without any change in perfusion.

The size of the problem was established on PET, where all images depend on correction. In 1,177 consecutive rest-dipyridamole PET perfusion studies with transmission-source attenuation scans, 21.4% had artifactual defects from attenuation-emission misregistration, predicted by horizontal-plane misregistration, body mass index, and heart size. Diaphragm displacement after dipyridamole was larger in obese patients. In 259 consecutive rubidium-82 (Rb-82) PET/CT studies with helical CT attenuation correction, misregistration caused artifactual defects in 40% of patients, moderate to severe in 23%. The degree of misregistration, particularly transaxial misregistration greater than 6 mm, predicted artifact size and severity, and 76% of the artifacts were anterior or lateral, 16% inferior, and 8% apical. The defects normalized when attenuation data averaged over normal breathing (cine CT) or shifted CT data were used.

A defect that appears only on the stress study of a rest-stress pair, when one CT was acquired at rest, deserves particular suspicion. Vasodilator stress and deep breathing commonly shift the heart, so the stress emission no longer matches the rest map, and the resulting artifact mimics reversible ischemia. The fused images are inspected for all emission-map pairs before the corrected study is accepted, and correction is repeated after realignment when the two are misaligned.

PET-specific artifacts and quality control

PET avoids many SPECT problems. Coincidence detection makes attenuation correction integral and accurate when the map is right, count sensitivity is higher, the 511-keV photons are attenuated less per centimeter of tissue, and Rb-82 and N-13 ammonia allow gated images close to peak stress. In meta-analyses PET has higher specificity than SPECT, in part because attenuation artifacts are corrected and spatial resolution is higher (pooled specificity 81% versus 61% for stenosis of 50% or more in one analysis; Rb-82 PET accuracy exceeded SPECT in another). The 2016 ASNC and SNMMI indications statement named PET as preferred after poor-quality or equivocal prior studies and when body habitus degrades image quality. The 2026 ASNC position statement recommends PET with flow quantification, where available, for patients with suspected coronary disease who are candidates for perfusion imaging. The clinical indications are summarized in the myocardial perfusion imaging hub.

PET has its own technical failure modes:

A flow value is accepted only after review of motion, bolus shape, time-activity curves, region placement, model fit, count-rate limits, and registration. These requirements are set out in the 2018 SNMMI/ASNC flow quantification paper and in the myocardial blood flow section of the perfusion hub.

Patient motion and upward creep

Tomographic reconstruction assumes that all projections describe the same object in the same position. When the patient moves partway through acquisition, projections from before and after the shift back-project along lines that fail to converge, and counts are spread across a region set by the direction and size of the shift. The resulting defect follows the geometry of the movement and does not conform to a coronary territory. Typical signatures are paired defects in opposite walls, such as anterior and inferior defects with vertical motion or septal and lateral defects with lateral motion, and a smeared or "hurricane" appearance on short-axis slices, with a tail of activity extending from the ventricle.

Motion typeAppearance on the cine and sinogramEffect on slicesResponse
Abrupt vertical or lateral shiftStep in the rotating sequence; break in the sinogramPaired defects in opposite walls crossing territoriesSoftware realignment for a small single shift; reacquisition when the shift is large or late
Bounce (shift and return)Brief displacement with return to the original positionUsually little effectOften none required
Upward creepGradual cranial drift of the heart through the acquisitionReversible inferior and basal inferoseptal defectsAcquisition started 15 minutes after exercise; motion correction or repeat acquisition
RespiratoryCyclical blurring of the heart and diaphragmInferior blurring; contributes to CT-emission mismatchBreathing instruction; CT matched to average breathing
Rotation or deformationComplex change in cardiac shape between projectionsBizarre defectsReacquisition; translational correction cannot represent it

Patient motion during SPECT. The cine of raw projections, the sinogram, and the linogram are reviewed before any correction is applied.

The effect of motion depends on timing and camera geometry. In simulations with technetium sestamibi, bounce motion produced no significant defects, a shift early in acquisition produced the largest defect with a single-head camera, and a shift in the middle of acquisition produced the largest defect with a dual-head camera. Automated motion correction significantly improved most simulated motion defects. It corrects only translation of the whole frame. Rotation, chest-wall deformation, and cyclical respiration are not represented by the shift model, and correction can leave residual error or create new distortion. ASNC practice guidance recommends motion correction, automatic or manual, when motion exceeds 1 pixel, cross-checking with the sinogram and linogram, and reacquisition, sometimes prone, when motion is complex or large. Persistent retention of technetium tracers allows a repeat acquisition without a new injection.

Upward creep. After exhaustive exercise, mean lung volume is increased and the diaphragm and heart sit low; as breathing normalizes during acquisition, the heart drifts cranially. In the original thallium description, 29% of 102 consecutive patients showed upward movement of 2 pixels or more, and in low-likelihood patients with creep, 9 of 17 had reversible inferior and basal inferoseptal defects, while none of 28 without creep had defects. Starting acquisition 15 minutes after exercise reduced creep among low-likelihood patients to 14%. With technetium agents, the routine delay of 15 to 60 minutes after exercise largely prevents it, but early imaging on high-efficiency cameras reintroduces the risk.

Subdiaphragmatic and other extracardiac activity

Technetium sestamibi and tetrofosmin are cleared through the hepatobiliary system, so the liver, gallbladder, and bowel can contain more activity than the myocardium, especially at rest and after vasodilator stress without exercise, which leaves splanchnic flow high. That activity affects the inferior wall in three ways. Scatter from the hot source adds counts to the adjacent inferior wall and can mask a true defect. Filtered back-projection with a ramp filter creates a negative lobe beside an intense source, which subtracts counts from the adjacent wall and creates a false inferior defect. Display normalization scales the image to its brightest region, so when extracardiac activity is brightest or when hot inferior counts become the reference, the whole myocardium is rendered darker or the remaining walls appear relatively reduced.

The ramp-filter effect is quantifiable. In a chest phantom with a liver insert, a hepatic-to-cardiac activity ratio of 1:1 lowered maximal inferior and inferoseptal activity by 17.8%, and a ratio of 2:1 lowered it by 46.2%, compared with no hepatic activity; with good count statistics, a higher cutoff frequency for the prereconstruction filter lessened the artifact. Iterative reconstruction reduces the negative-lobe effect but does not remove scatter or normalization errors.

Rest SPECT short-axis, horizontal long-axis, and vertical long-axis slices with and without CT attenuation correction; an intense focus of subdiaphragmatic activity lies directly against the inferior wall, marked by yellow arrows.
Interfering subdiaphragmatic activity (yellow arrows) beneath the inferior wall on rest short-axis and vertical long-axis slices, with CT attenuation correction (rows labeled RstCTAC) and without it (rows labeled Rst). Activity this close to the inferior wall can mask a defect through scatter or create one through filtering and normalization. Credit: Strok A, Salobir BG, Stalc M, Zaletel K, Radiol Oncol 2024;58:313–319, Figure 1; CC BY 4.0.
SourceBehaviorEffect on the inferior wallMeasures
LiverDiffuse, highest early after injection and after vasodilator stressSpurious inferior counts from scatter; negative-lobe defect with ramp filtering; normalization errorExercise or low-level exercise with vasodilators; imaging delay (at least 45 minutes for sestamibi and 30 minutes for tetrofosmin in ASNC practice guidance); repeat imaging
GallbladderFocal, intense, adjacent to the inferior wallMasks or creates a defectDelayed imaging; a fatty meal in some protocols
BowelMobile; may differ between stress and restVariable artifact that mimics reversibilityDelay and repeat; prone imaging moves bowel away
Gastric wallIncreased with proton pump inhibitor use; Rb-82 gastric uptakeInferolateral and inferior spurious countsDrink of water or carbonated beverage before imaging; repeat imaging
Stomach with thyroid and salivary activityFree pertechnetate from a labeling failureLow myocardial contrast and gastric activity near the inferior wallRadiochemical purity testing; repeat with a new preparation

Extracardiac sources of activity near the heart and the measures used to limit them.

Several drugs and preparations influence gastric activity. In a matched case-control study of 5,432 perfusion studies, the odds of rescanning were higher with proton pump inhibitor use (adjusted odds ratio 1.6), and the association was concentrated in rescans for gastric wall uptake (adjusted odds ratio 6.3). The culprit sites among rescanned studies were the left hepatic lobe (48%), gastric wall (31%), gastric lumen (12%), spleen (7%), and bowel (less than 1%). Among 600 consecutive Rb-82 PET studies, gastric uptake was 23% higher in patients taking a proton pump inhibitor. No published trial has shown that routinely withholding a proton pump inhibitor before imaging improves diagnostic accuracy or reduces repeat imaging. Interventions to reduce interfering activity, including water, milk, carbonated water, fatty meals, walking, and delayed imaging, have produced inconsistent results. A 2024 review concluded that exercise stress when possible, later imaging, fluid intake, and carbonated water immediately before imaging are reasonable choices. ASNC practice guidance advises that when liver or bowel activity precludes interpretation, imaging is repeated after a delay that allows the activity to clear or move away from the heart.

Other extracardiac findings. The projection cine and the low-dose CT can show breast and axillary lesions, lung tumors, lymphoma, thyroid nodules, and other tracer-avid tissue, because sestamibi and thallium concentrate in some tumors. A pericardial effusion appears as a photopenic halo around the ventricle, and a hiatal hernia can produce activity or a photopenic structure behind the heart. These are described and handled under the laboratory policy for incidental findings; they are not coronary findings.

Injection infiltration, low count density, and radiopharmaceutical failure

Infiltrated injection. When part of the dose is deposited subcutaneously, less activity reaches the myocardium, total counts fall, and statistical noise rises throughout the study. The depot also emits photons and appears as an intense focus at the injection site, which can enter the field of view, alter normalization, and add scatter. The site is visible on the raw projections, and no processing recovers activity that never reached the circulation. Residual activity at the injection site can also enter selected projections and create an apparent defect. One study included 50 men imaged on a cadmium zinc telluride (CZT) camera with residual injection-site activity in the field, most often in the prone position. Repeat acquisition with the injection site covered by a lead shield changed the final interpretation in almost one in five, usually by removing an apparent stress-induced defect. A stress study with a substantial infiltration is repeated, because the stress state cannot be recreated by reconstruction.

Low count density. A study with too few counts is uniformly noisy. Mild defects disappear into the noise, random fluctuation creates mottled pseudo-defects, gated function becomes unreliable, and contour placement fails. The causes are large habitus, low administered activity, short acquisition, infiltration, a camera with low sensitivity, and heavy beat rejection during gating. Activity and acquisition time are chosen to reach adequate counts on the installed camera, and protocols transferred from another camera or reduced-dose protocols require local validation of count density, gating, artifact behavior, and the normal database.

Labeling failure. Free pertechnetate from an incomplete kit reaction has no myocardial affinity and distributes to the thyroid, salivary glands, gastric mucosa, and soft tissue. Myocardial contrast falls, background rises, and gastric activity can lie against the inferior wall. Thyroid and gastric activity on a perfusion study indicate a radiochemical purity failure, which the preparation record confirms. Tracer properties, kit preparation, and purity testing are covered in the technetium section of the radiotracer hub.

Dose calibrator error. A calibrator that reads high causes under-administration and low counts, and one that reads low causes over-administration and unnecessary dose. Daily constancy and scheduled accuracy, linearity, and geometry testing protect against both.

Left bundle branch block and ventricular pacing

Left bundle branch block and right ventricular pacing produce septal or anteroseptal perfusion defects without obstructive left anterior descending disease. The mechanism is multifactorial. Activation of the septum is early and asynchronous, so the septum contracts against a ventricle that is not yet pressurized and then remains under raised intramyocardial pressure during the period when coronary inflow would otherwise occur. Diastolic perfusion time shortens as heart rate rises. Reduced septal thickening lowers counts through the partial-volume effect. The defect is rate dependent and is commonly reversible, so with exercise it reproduces the pattern of left anterior descending ischemia; dobutamine also raises heart rate and carries the same theoretical risk. The diagnostic criteria and clinical significance of the block itself are covered in the bundle branch block hub.

Vasodilator stress raises heart rate far less than exercise and is the preferred stress in left bundle branch block and ventricular pacing. In 173 consecutive patients with left bundle branch block, 73 of whom had catheterization, adenosine thallium imaging had an overall predictive accuracy of 93% compared with 68% for exercise thallium imaging. Specificity in the territories that supply the septum was 82% compared with 42%. In a series of 383 patients, 154 of whom had angiography, the false-positive rate for septal defects was 46% with exercise and 10% with pharmacologic stress. Specificity for left anterior descending stenosis was 36% with exercise, 81% with adenosine, and 80% with dobutamine. The dobutamine estimate rests on only 27 patients with angiography. US guidelines nonetheless favor vasodilator stress, because dobutamine raises heart rate and the dobutamine experience in this setting is small. Regadenoson raises heart rate modestly. Low-level exercise combined with a vasodilator is avoided in left bundle branch block and pacing because it reintroduces the rate effect. Agent selection is covered in the pharmacologic stress hub.

Several features favor true left anterior descending ischemia over a conduction artifact: a reversible defect that involves the apex or the anterior wall beyond the septum, a severe defect, a defect that is more extensive at end diastole, and systolic dysfunction that matches the perfusion defect. An isolated septal or anteroseptal defect sparing the apex, with septal dyssynchrony on the gated cine and normal anterior thickening, favors the conduction pattern. Abnormal septal motion alone does not establish septal infarction, because conduction disease alters septal mechanics. Rate-related left bundle branch block that appears during exercise produces the same artifact and is recorded in the stress report with the heart rate at onset.

Normal variants that mimic perfusion defects

A normal heart contains regions with lower counts for anatomical reasons, and segment scoring software has no anatomical knowledge. Four patterns account for most variant defects.

Apical thinning. The apical myocardium is thinner than the adjacent walls, and a structure at or below system resolution is measured with reduced apparent activity through the partial-volume effect. The reduction is small, symmetric about the apex, identical on stress and rest, and accompanied by normal apical thickening. It is more conspicuous after attenuation correction and with high-resolution cameras. Apical infarction is usually asymmetric, extends into the supplying territory, and lacks thickening, although a small apical infarct can retain thickening and the apex is itself vulnerable to partial-volume error on gated analysis.

Basal septal reduction. The membranous septum contains no myocardium, and partial volume and the geometry of the outflow tract lower counts at the most basal septal slices. The finding is confined to those slices, identical on stress and rest, and does not extend into the mid septum. It is absent from some normal studies. When the normal database or segmentation does not match the study, it can be scored as a small fixed defect that inflates both summed scores.

Papillary muscle hot spots. Papillary muscles take up tracer and can appear as focal areas of increased activity projecting into the cavity at the anterolateral and inferolateral walls. Because display normalization uses the brightest region, an intense papillary muscle can make adjacent myocardium appear relatively reduced, and with ramp filtering it can create a small adjacent negative-lobe defect. Its position and motion on the gated cine identify it.

Right ventricular uptake. The normal right ventricular free wall is thin and faint; ASNC practice guidance gives a normal right ventricular intensity of approximately 50% of peak left ventricular intensity. Increased right ventricular activity at rest occurs with right ventricular hypertrophy, most often from pulmonary hypertension. Right ventricular prominence that appears or increases at stress usually reflects reduced left ventricular counts, a marker of balanced left ventricular ischemia, so left ventricular count level is assessed before a right ventricular cause is assigned. The interpretation of these variants within the full perfusion study is summarized in the variants section of the perfusion hub.

A general principle separates variants from disease: a variant occupies a predictable location, appears similarly on stress and rest, and preserves thickening, whereas a pathological defect follows a coronary distribution, may change between stress and rest, and is accompanied by dysfunction when the myocardium is scarred or severely ischemic. These patterns are clues and have exceptions.

Ventricular size, hypertrophy, and cardiac position

Small left ventricle. When the end-systolic cavity is smaller than the system resolution, counts from opposing walls blur across it, and the software finds no count minimum. End-systolic volume is underestimated and ejection fraction is overestimated. For example, a true end-diastolic volume of 55 mL and end-systolic volume of 20 mL give an ejection fraction of 64%; if end-systolic volume is measured as 7 mL, the reported value is 87%. Phantom studies of mathematically defined ventricles showed marked overestimation at low end-diastolic volumes and high ejection fractions, worse with smoothing, and the effect is most common in small women. An ejection fraction of 85% or more in a small ventricle is reviewed for this artifact. By the same mechanism, a mildly reduced ejection fraction can appear normal. A small cavity also blunts transient ischemic dilation and can make small defects harder to resolve.

Left ventricular hypertrophy. In hypertrophic cardiomyopathy and asymmetric septal hypertrophy, the thick septum contains more tracer than the normal lateral wall, and normalization to the bright septum makes the lateral wall appear relatively reduced. Hypertrophy also blunts partial-volume effects in a way that differs from the normal database. Hypertrophic cardiomyopathy can cause true perfusion abnormalities and transient ischemic dilation from microvascular disease, so a defect in this setting is interpreted with wall thickness from echocardiography or MRI.

Dilated left ventricle. A large, thin-walled ventricle spreads counts over a wide cavity, lowers wall thickness relative to resolution, and often contains large defects that leave the contour algorithm nothing to fit. Contour placement becomes unreliable in either direction, and volumes and ejection fraction are accepted only after the contours are inspected on the gated cine.

Cardiac position. Dextrocardia, mesocardia, and marked rotation from pectus excavatum or prior thoracic surgery change the orientation that reorientation software expects and can change the attenuation pattern. Acquisition orbits and reorientation are adapted to the actual position, and wall labels are verified before interpretation.

Electrocardiographic gating errors and unreliable function measurements

Electrocardiographic gating divides the R-R interval into 8 or 16 frames and sums counts from thousands of accepted beats into one composite cycle. The method assumes that the accepted beats have comparable length and filling. ASNC practice guidance describes rejection of beats that fall more than 20% outside a stable R-R interval as usual practice. When long and short cycles contribute to the same frame, counts from different phases of contraction are summed together, and volumes, ejection fraction, and regional wall motion become unreliable. Intermittent flashing of the gated cine, in which the image brightens and dims from frame to frame, is the visual sign of a gating error from arrhythmia or dropped beats.

Beat histogram appearanceInterpretationConsequence for the report
Narrow dominant peakConsistent cycle lengthGated function reliable
Broad distributionVariable cycle length, as in atrial fibrillationFunction uncertain; perfusion often interpretable
Two or more peaksBigeminy or frequent ectopyFunction unreliable
Large rejected fractionAcceptance window too narrow or frequent irregular beatsCounts reduced in all frames and in the summed image

Interpretation of the gating beat histogram, which is reviewed before ejection fraction is read.

The perfusion image is reconstructed from summed counts and does not depend on which frame a count occupied, so perfusion often remains interpretable when gating has failed. Heavy beat rejection, however, lowers total counts and can degrade the perfusion images too. When rhythm is irregular, the report states the perfusion findings and qualifies or omits the functional measurements with the reason. Poor electrode contact, low-voltage QRS complexes, and tall T waves cause missed or double triggers and are corrected by repositioning the leads. Eight-frame gating systematically gives slightly lower ejection fractions than 16-frame gating, so serial comparisons use the same framing and software. Automatic ejection fraction algorithms fit endocardial surfaces to the count profile, and contour errors from small cavities, the valve-plane limit, large defects, or adjacent extracardiac activity produce precise but incorrect numbers. The end-diastolic and end-systolic contours are inspected before volumes and ejection fraction are accepted.

Camera quality control and instrument artifacts

Patient factors produce defects that vary between individuals; instrument failures produce defects that repeat. An off-peak energy window, a nonuniform detector, a center-of-rotation error, a damaged collimator, or a failed detector module reproduces the same abnormality in successive patients, and the same defect in several patients on one camera in one session is an equipment problem until shown otherwise. Workstation processing cannot correct an acquisition fault. The physics of these tests is covered in the quality control section of the nuclear cardiology physics hub.

TestPriority and frequency (ASNC practice guidance)Failure it detectsArtifact if missed
Energy peakingMandatory, daily, all detectorsPhotopeak window driftLow counts, high scatter, and poor contrast in all patients on that detector
Uniformity (intrinsic or extrinsic flood)Mandatory, dailyPhotomultiplier gain drift, crystal damage, collimator defectsRing or focal artifacts on tomographic slices
Center of rotation and multidetector registrationMandatory, per manufacturerMechanical axis offset from the electronic matrix"Doughnut" artifact with 360-degree orbits, "tuning fork" with 180-degree orbits, blurring; most evident above 2 pixels on a 64 × 64 matrix
Uniformity calibrationMandatory, per manufacturerCorrection map out of dateRegional nonuniformity
Resolution and linearityRecommended, per manufacturerLoss of resolution; spatial distortionBlurring; distorted wall contours
Total performance (phantom)Recommended, quarterlySystem-level degradationGeneral loss of contrast and resolution
SensitivityOptional, per manufacturerDrift in counts per unit activityLow counts at standard activity
CT tube warm-up, air calibration, water phantomRecommended, daily (SPECT/CT); monthly slice and positioning checksCT number driftIncorrect attenuation coefficients and correction errors
Detector map (CZT systems)Per manufacturerModule drift or failureStraight-edged, strip-like defects crossing coronary territories
Dose calibrator constancy, accuracy, linearityDaily constancy; others on scheduleActivity measurement errorLow count density or excess dose

Camera and laboratory quality control for cardiac SPECT and SPECT/CT, after the ASNC SPECT instrumentation and quality assurance practice points and the 2018 ASNC SPECT guideline. Manufacturer, licensing, and accreditation requirements govern the exact schedule.

Energy window. A symmetric photopeak window of 15% to 20% is standard; for technetium-99m at 140 keV a 20% window spans approximately 126 to 154 keV. If the window drifts off the photopeak, true photopeak events fall outside it while scattered photons fall inside it, total counts fall, and contrast deteriorates. Peaking for the wrong isotope, such as a thallium window on a technetium study, produces a severely degraded, low-count study. A sudden fall in count rate across consecutive patients is investigated as a window or detector fault.

Nonuniformity. A regional nonuniformity barely visible on a planar flood is projected into all angular views and reinforced at the same location by reconstruction, so a defect of a few percent on a flood becomes a conspicuous ring or focal artifact on tomographic slices. This is why daily high-count floods are required. Collimator dents and septal damage produce focal extrinsic nonuniformity at a fixed detector location.

Center of rotation. Reconstruction assumes all projections were acquired around a known common axis. An offset back-projects the views along displaced lines, so counts from opposite angles fail to converge, resolution falls, and a point source becomes a ring. Because the error is geometric and fixed, it affects all studies on that system. ASNC practice guidance notes that the effect is most evident when the error exceeds 2 pixels on a 64 × 64 matrix and that smaller errors reduce resolution.

Collimator and septal penetration. Collimator geometry trades resolution against sensitivity: longer, narrower holes improve resolution and reduce counts. Resolution also degrades approximately linearly with source-to-detector distance, which is why body-contouring orbits keep detectors close to the chest. A low-energy collimator used with higher-energy photons allows septal penetration, which produces star-shaped streaks around intense sources and loss of contrast.

Dedicated cardiac cameras. CZT systems convert photons directly to charge, have better energy resolution, and use focused or multi-pinhole geometry that raises sensitivity. A focused field samples a limited volume; when the heart lies outside it because of positioning, unusual orientation, or ventricular dilation, part of the myocardium is truncated and appears absent without any attenuating structure to explain it. Inspection of the heart's position within the sampled volume identifies the problem, and repositioning and reacquisition solve it. A failed or drifting CZT module produces a defect with straight borders following the module boundary, which repeats in all patients until the detector map is corrected. Attenuation patterns differ between CZT and Anger cameras, and normal databases are camera specific.

ASNC practice guidance states that when energy peaking, uniformity, sensitivity and resolution, or center of rotation are outside acceptable limits, clinical imaging does not begin, the service engineer is notified, and imaging resumes only after the system is examined and all quality control is repeated within limits. A normal report depends on that record: a uniform image acquired on a failing camera may be uniform because of the camera.

Reconstruction, filtering, and reorientation errors

Filtering. Filtered back-projection requires a ramp filter to remove the blur of simple back-projection, and a smoothing window, commonly a Butterworth filter with a cutoff frequency and order, controls noise. A cutoff set too low over-smooths and conceals small or apical defects; one set too high leaves statistical noise that creates mottled pseudo-defects. Neither error is visible as a filter problem on the slices, and both repeat across all studies processed with the same settings. Filter parameters are fixed in protocol and matched to the normal database.

Iterative reconstruction. Ordered-subset expectation maximization models attenuation, scatter, and distance-dependent collimator blur. Too few iterations understate contrast, so a true defect appears milder; too many amplify noise. Resolution recovery sharpens images and allows lower counts, but it can overshoot at sharp boundaries and produce a bright rim beside an apparently darker region. Iteration and subset numbers, resolution recovery, and corrections are fixed protocol parameters, and a study reconstructed differently from the normal database cannot be scored against it. Machine-learning denoising, synthetic attenuation maps, and automated registration are available in clinical software; their output depends on training data and requires validation on the installed hardware, protocol, and population. The raw projections, maps, and contours are still reviewed.

Reorientation and axis errors. Reconstruction produces transaxial slices aligned to the scanner, and reorientation rotates the volume onto the long and short axes of the left ventricle. If the long axis is set incorrectly, short-axis slices cut obliquely through the ventricle, apparent wall thickness varies around the circumference, the apex is distorted, and wall labels no longer match anatomy; a defect can appear where none exists or be assigned to the wrong territory. Stress and rest axes that differ produce apparent reversibility. ASNC guidance requires long-axis lines parallel to the walls and consistent between stress and rest. Incorrect selection of the apical and basal limits, or of the valve plane, distorts volumes and polar maps in the same way.

Display. The color scale determines how a normal study appears. A high-contrast map exaggerates small differences and turns variants into defects, a low-contrast map conceals mild reductions, and when the brightest region is itself abnormal, normalization shifts the whole display. The scale is kept constant across studies and in serial comparison, and the slices are reviewed beside the polar map.

Quantitative analysis and normal-database mismatch

Quantitative perfusion values are comparisons between the patient's count distribution and a reference population acquired and processed under defined conditions. A sample more than a set number of standard deviations below the reference mean is classified as abnormal, and total perfusion deficit integrates the extent and depth of abnormal samples. Software can generate scores from studies affected by motion, misregistration, or an inappropriate database, and the scores themselves carry no warning.

Four properties of the database must match the study:

Thresholding converts a continuous distribution into a binary classification, so a small change in count density moves a region across the cutoff and changes the score discontinuously. Values near a decision threshold are interpreted with more uncertainty and checked against the images. Serial quantitative comparison is valid only when position, camera, reconstruction, correction, software, and database are matched. No universal smallest significant change in summed scores, total perfusion deficit, or volumes applies across systems. The 17-segment model, semiquantitative scores, and total perfusion deficit are described in the scoring section of the perfusion hub. Automated and machine-learning interpretation improves reproducibility, but acquisition, attenuation, motion, and contour errors propagate into its output, and a learned model can perform poorly in a population that differs from its training set.

Balanced reduction and the ancillary markers as quality signals

Relative perfusion images are normalized to the patient's own brightest region. When left main or three-vessel disease limits hyperemic flow in all territories by a similar proportion, the reference falls with the rest and the display can appear uniform. This is the most important false-negative mechanism of relative imaging, although most patients with severe multivessel disease still show a regional defect that underestimates the burden. The clues lie outside the perfusion pattern: ischemic ST depression, exercise hypotension, limiting angina at low workload, transient ischemic dilation, a fall in ejection fraction after stress, increased lung uptake, transient right ventricular visualization, and, on PET, globally reduced stress flow and flow reserve. The interpretation of balanced ischemia is covered in the perfusion hub, and the thresholds for high-risk markers in its high-risk section.

These markers have technical failure modes of their own:

Stress adequacy: a technically good image of the wrong physiology

Image quality and stress adequacy are independent. A patient who stops exercising at a low workload for leg fatigue, or who has consumed caffeine before a vasodilator, can produce images free of motion, attenuation, and count limitation in which the hyperemic stress was never delivered. The flow difference that creates a defect is reduced, sensitivity falls, and a uniform image has a weaker negative predictive value. Common failures are early termination of exercise, injection after exercise has stopped, an interrupted adenosine infusion, tracer injected through shared tubing, premature reversal with aminophylline or caffeine, and antianginal drugs taken on the day of a diagnostic study. Methylxanthine exposure is another, from coffee, decaffeinated coffee, tea, chocolate, energy drinks, or caffeine-containing analgesics. On PET, absent augmentation of stress flow may reflect inadequate vasodilation. It also arises from injection infiltration, count-rate saturation, frame motion, an abnormal input function, and model failure, so a low stress flow requires explanation. Caffeine holds and drug withholding are covered in the pharmacologic stress hub, and exercise end points in the exercise stress testing hub.

The report states the achieved stress conditions before the perfusion conclusion: method and agent, dose, achieved workload and heart rate or vasodilator conditions, symptoms, electrocardiographic and hemodynamic response, the reason for termination, and any reversal agent with its timing relative to tracer injection.

Hybrid CT: coronary calcium and incidental findings

Hybrid SPECT/CT and PET/CT systems acquire a low-dose, usually nongated CT for attenuation correction and localization. That CT also shows coronary calcium. Visible calcium indicates atherosclerosis even when perfusion is normal, because nonobstructive plaque does not limit flow. In 492 patients from three centers who had both a CT attenuation scan and a standard gated calcium scan, visual estimation of calcium on the attenuation CT placed 63% of patients in the same Agatston category and 93% within one category, with a weighted kappa of 0.89. The 2016 SCCT and Society of Thoracic Radiology guideline recommends that coronary calcium be reported on all noncontrast chest CT, using a visual ordinal estimate (none, mild, moderate, severe) when the scan is not gated. The 2022 ASNC, American Association of Physicists in Medicine (AAPM), SCCT, and SNMMI hybrid CT guideline recommends reporting calcium visible on the attenuation CT. A formal Agatston score requires a dedicated electrocardiographically gated acquisition with standard slice thickness, so the attenuation CT is described visually without a numeric score. The 2013 SNMMI/ASNC/SCCT hybrid guideline notes that visual review misses small amounts of calcium in about 8% of patients. Heavy calcium with a uniform relative perfusion study raises the question of balanced ischemia when the clinical and ancillary findings are high risk. The calcium score method and reporting system are covered in the coronary CT hub.

The CT field of view also contains lung nodules and masses, pleural and pericardial effusions, aortic dilation, hiatal hernia, and breast or mediastinal lesions. The attenuation CT is not a diagnostic chest CT, and its limitations are stated in the report; a clinically important finding is still documented and communicated. The 2022 hybrid CT guideline describes laboratory models for this review, including who reads the CT field, the training required, and when a diagnostic CT or specialist review is recommended.

Diagnostic accuracy, women, and the impact of artifacts

Artifacts lower specificity more than sensitivity. In a meta-analysis of studies against angiographic stenosis of 50% or more, pooled SPECT sensitivity was 88% and specificity 61%, compared with 84% and 81% for PET; part of the specificity gap reflects attenuation artifacts and part reflects referral bias, since patients with normal studies are rarely sent to angiography. Attenuation correction, gating, positional imaging, and PET all raise specificity or normalcy rates in the studies described above. The full comparison with stress echocardiography, stress MRI, and coronary CT angiography is in the accuracy section of the perfusion hub.

Imaging in women combines several technical limits: breast attenuation, smaller ventricles with partial-volume effects and overestimated ejection fraction, and a higher prevalence of microvascular dysfunction. With microvascular dysfunction, a normal relative perfusion study does not exclude abnormal coronary physiology. In women, technetium sestamibi SPECT had higher specificity than thallium-201 (82.4% compared with 58.8% for stenosis of 70% or more), and gating raised specificity further. Sex-specific normal databases, consistent breast positioning, attenuation correction, prone or upright imaging, and PET with flow quantification address these limits. Obese patients face attenuation, scatter, low counts, and misregistration together, and the measures in the attenuation section apply.

False-negative studies have technical causes too: inadequate stress, balanced reduction, low counts that bury a mild defect, over-smoothing, hot extracardiac activity that fills in an inferior defect, attenuation correction that raises counts in a truly ischemic inferior wall, and rest shine-through in one-day protocols, in which residual rest activity adds counts to a stress defect and reduces its apparent severity.

Reporting image quality and confidence

The ASNC reporting guideline treats image quality as a required report element, alongside the stress data, perfusion findings, function, ancillary findings, and comparison with prior studies. A report describes count density, motion, attenuation, extracardiac activity, and gating reliability, and it names the methods used to resolve them (attenuation correction, prone imaging, motion correction). The impression then uses a confidence category that matches what the study can support.

CategoryMeaningReport content
NormalNo inducible flow limitation under adequate stress with diagnostic image qualityStress conditions stated with the finding
AbnormalA defect is presentLocation by segment, extent, severity, reversibility, likely territory, and function
EquivocalA specific uncertainty exists in the imagesThe specific issue named, with the data that would resolve it
Technically limitedInterpretation possible within a stated constraintThe limitation and the wall or question it affects
NondiagnosticThe study cannot answer the clinical questionThe reason, and the recommended alternative; no negative-study language

Confidence categories for perfusion reports. ASNC standardized reporting also allows probably normal and probably abnormal as graded impressions. A technically limited study has a qualified reading; a nondiagnostic study has none.

Unexplained caveats such as "cannot exclude ischemia" leave the referring clinician without information. A useful statement names the finding, its suspected mechanism, and the data that would resolve it: for example, a mild fixed inferior reduction with preserved thickening, consistent with diaphragmatic attenuation, resolved on prone imaging. Common errors of overstatement are naming a percentage stenosis in a vessel (perfusion imaging does not measure luminal narrowing), declaring the absence of coronary disease (the study shows only that no inducible flow limitation was demonstrated), and reporting a flow value without its technical validity. Common errors of understatement are omitting a fall in ejection fraction or transient dilation, failing to address an ischemic stress electrocardiogram with normal perfusion, and omitting the reason a study was technically limited. Urgent or unexpected findings that can alter immediate care are communicated directly to the responsible clinician and documented with the recipient, date, and time.

Laboratory quality programs, metrics, and accreditation

Quality control verifies equipment and acquisition for individual studies; quality improvement evaluates laboratory processes and performance over time. The 2023 ASNC information statement on quality metrics for SPECT organizes laboratory performance into eight quality axes, broken into measurable metrics with worksheets that let a laboratory find weak areas, set targets, and reassess. The axes include appropriate patient selection and test choice, patient preparation and stress, protocol selection with patient-centered and stress-only imaging, radiation dose, acquisition and processing quality, interpretation and reporting, and correlation with outcomes or other tests.

A laboratory quality program built on these principles contains:

Accreditation by the Intersocietal Accreditation Commission requires documented protocols, staff qualifications and continuing education, equipment and radiopharmaceutical quality control, radiation safety, peer and quality review, and complete reports. Algorithmic tools do not displace physician responsibility and require validation and oversight within the laboratory.

Sources of error in perfusion image interpretation

References

  1. Dorbala S, Ananthasubramaniam K, Armstrong IS, et al. Single photon emission computed tomography (SPECT) myocardial perfusion imaging guidelines: instrumentation, acquisition, processing, and interpretation. J Nucl Cardiol. 2018;25:1784–1846. doi:10.1007/s12350-018-1283-y
  2. Henzlova MJ, Duvall WL, Einstein AJ, Travin MI, Verberne HJ. ASNC imaging guidelines for SPECT nuclear cardiology procedures: stress, protocols, and tracers. J Nucl Cardiol. 2016;23:606–639. doi:10.1007/s12350-015-0387-x
  3. Duvall WL, Skali H, Aggarwal NR, et al. 2026 American Society of Nuclear Cardiology guideline for stress testing in single-photon emission computed tomography and positron emission tomography myocardial perfusion imaging. J Nucl Cardiol. 2026;64:102970. doi:10.1016/j.nuclcard.2026.102970
  4. Dilsizian V, Bacharach SL, Beanlands RS, et al. ASNC imaging guidelines/SNMMI procedure standard for positron emission tomography (PET) nuclear cardiology procedures. J Nucl Cardiol. 2016;23:1187–1226. doi:10.1007/s12350-016-0522-3
  5. Al-Mallah MH, Bateman TM, Branch KR, et al. 2022 ASNC/AAPM/SCCT/SNMMI guideline for the use of CT in hybrid nuclear/CT cardiac imaging. J Nucl Cardiol. 2022;29:3491–3535. doi:10.1007/s12350-022-03089-z
  6. Dorbala S, Di Carli MF, Delbeke D, et al. SNMMI/ASNC/SCCT guideline for cardiac SPECT/CT and PET/CT 1.0. J Nucl Med. 2013;54:1485–1507. doi:10.2967/jnumed.112.105155
  7. Abbott BG, Case JA, Dorbala S, et al. Contemporary cardiac SPECT imaging—innovations and best practices: an information statement from the American Society of Nuclear Cardiology. J Nucl Cardiol. 2018;25:1847–1860. doi:10.1007/s12350-018-1348-y
  8. Hage FG, Einstein AJ, Ananthasubramaniam K, et al. Quality metrics for single-photon emission computed tomography myocardial perfusion imaging: an ASNC information statement. J Nucl Cardiol. 2023;30:864–907. doi:10.1007/s12350-022-03162-7
  9. Tilkemeier PL, Bourque J, Doukky R, Sanghani R, Weinberg RL. ASNC imaging guidelines for nuclear cardiology procedures: standardized reporting of nuclear cardiology procedures. J Nucl Cardiol. 2017;24:2064–2128. doi:10.1007/s12350-017-1057-y
  10. American Society of Nuclear Cardiology. Practice Points: SPECT instrumentation and quality assurance; SPECT acquisition; SPECT processing and interpretation. Bethesda, MD: ASNC; updated August 2011. asnc.org
  11. Hendel RC, Corbett JR, Cullom SJ, DePuey EG, Garcia EV, Bateman TM. The value and practice of attenuation correction for myocardial perfusion SPECT imaging: a joint position statement from the American Society of Nuclear Cardiology and the Society of Nuclear Medicine. J Nucl Cardiol. 2002;9:135–143. doi:10.1067/mnc.2002.120680
  12. Heller GV, Links J, Bateman TM, et al. American Society of Nuclear Cardiology and Society of Nuclear Medicine joint position statement: attenuation correction of myocardial perfusion SPECT scintigraphy. J Nucl Cardiol. 2004;11:229–230. doi:10.1016/j.nuclcard.2003.12.001
  13. Hendel RC, Berman DS, Cullom SJ, et al. Multicenter clinical trial to evaluate the efficacy of correction for photon attenuation and scatter in SPECT myocardial perfusion imaging. Circulation. 1999;99:2742–2749. doi:10.1161/01.cir.99.21.2742
  14. Burrell S, MacDonald A. Artifacts and pitfalls in myocardial perfusion imaging. J Nucl Med Technol. 2006;34:193–211. tech.snmjournals.org
  15. DePuey EG, Rozanski A. Using gated technetium-99m-sestamibi SPECT to characterize fixed myocardial defects as infarct or artifact. J Nucl Med. 1995;36:952–955. PubMed 7769451
  16. Taillefer R, DePuey EG, Udelson JE, Beller GA, Latour Y, Reeves F. Comparative diagnostic accuracy of Tl-201 and Tc-99m sestamibi SPECT imaging (perfusion and ECG-gated SPECT) in detecting coronary artery disease in women. J Am Coll Cardiol. 1997;29:69–77. doi:10.1016/s0735-1097(96)00435-4
  17. Nishina H, Slomka PJ, Abidov A, et al. Combined supine and prone quantitative myocardial perfusion SPECT: method development and clinical validation in patients with no known coronary artery disease. J Nucl Med. 2006;47:51–58. PubMed 16391187
  18. Berman DS, Kang X, Nishina H, et al. Diagnostic accuracy of gated Tc-99m sestamibi stress myocardial perfusion SPECT with combined supine and prone acquisitions to detect coronary artery disease in obese and nonobese patients. J Nucl Cardiol. 2006;13:191–201. doi:10.1007/BF02971243
  19. Nakazato R, Tamarappoo BK, Kang X, et al. Quantitative upright-supine high-speed SPECT myocardial perfusion imaging for detection of coronary artery disease: correlation with invasive coronary angiography. J Nucl Med. 2010;51:1724–1731. doi:10.2967/jnumed.110.078782
  20. Hedén B, Persson E, Carlsson M, Pahlm O, Arheden H. Disappearance of myocardial perfusion defects on prone SPECT imaging: comparison with cardiac magnetic resonance imaging in patients without established coronary artery disease. BMC Med Imaging. 2009;9:16. doi:10.1186/1471-2342-9-16
  21. Friedman J, Van Train K, Maddahi J, et al. "Upward creep" of the heart: a frequent source of false-positive reversible defects during thallium-201 stress-redistribution SPECT. J Nucl Med. 1989;30:1718–1722. PubMed 2795212
  22. Matsumoto N, Berman DS, Kavanagh PB, et al. Quantitative assessment of motion artifacts and validation of a new motion-correction program for myocardial perfusion SPECT. J Nucl Med. 2001;42:687–694. PubMed 11337561
  23. Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med. 2007;48:1112–1121. doi:10.2967/jnumed.107.039792
  24. Loghin C, Sdringola S, Gould KL. Common artifacts in PET myocardial perfusion images due to attenuation-emission misregistration: clinical significance, causes, and solutions. J Nucl Med. 2004;45:1029–1039. PubMed 15181138
  25. Germano G, Chua T, Kiat H, Areeda JS, Berman DS. A quantitative phantom analysis of artifacts due to hepatic activity in technetium-99m myocardial perfusion SPECT studies. J Nucl Med. 1994;35:356–359. PubMed 8295010
  26. Strok A, Salobir BG, Stalc M, Zaletel K. Subdiaphragmatic activity-related artifacts in myocardial perfusion scintigraphy. Radiol Oncol. 2024;58:313–319. doi:10.2478/raon-2024-0053
  27. Rose DS, Rowe JM, Love AE. Effect of proton pump inhibitors and other commonly prescribed drugs on rescanning of patients undergoing myocardial perfusion imaging: a case-control study. Nucl Med Commun. 2022;43:494–501. doi:10.1097/MNM.0000000000001538
  28. Alzahrani AH, Arasaratnam P, Massalha S, et al. Effect of proton pump inhibitors on rubidium-82 gastric uptake using positron emission tomography myocardial perfusion imaging. J Nucl Cardiol. 2020;27:1443–1451. doi:10.1007/s12350-019-01954-y
  29. O'Keefe JH Jr, Bateman TM, Barnhart CS. Adenosine thallium-201 is superior to exercise thallium-201 for detecting coronary artery disease in patients with left bundle branch block. J Am Coll Cardiol. 1993;21:1332–1338. doi:10.1016/0735-1097(93)90305-k
  30. Vaduganathan P, He ZX, Raghavan C, Mahmarian JJ, Verani MS. Detection of left anterior descending coronary artery stenosis in patients with left bundle branch block: exercise, adenosine or dobutamine imaging? J Am Coll Cardiol. 1996;28:543–550. doi:10.1016/0735-1097(96)00224-0
  31. Higgins JP, Williams G, Nagel JS, Higgins JA. Left bundle-branch block artifact on single photon emission computed tomography with technetium Tc 99m (Tc-99m) agents: mechanisms and a method to decrease false-positive interpretations. Am Heart J. 2006;152:619–626. doi:10.1016/j.ahj.2006.06.009
  32. Ford PV, Chatziioannou SN, Moore WH, Dhekne RD. Overestimation of the LVEF by quantitative gated SPECT in simulated left ventricles. J Nucl Med. 2001;42:454–459. PubMed 11337523
  33. Germano G, Kiat H, Kavanagh PB, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med. 1995;36:2138–2147. PubMed 7472611
  34. Cerqueira MD, Weissman NJ, Dilsizian V, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–542. doi:10.1161/hc0402.102975
  35. Abidov A, Bax JJ, Hayes SW, et al. Transient ischemic dilation ratio of the left ventricle is a significant predictor of future cardiac events in patients with otherwise normal myocardial perfusion SPECT. J Am Coll Cardiol. 2003;42:1818–1825. doi:10.1016/j.jacc.2003.07.010
  36. Murthy VL, Bateman TM, Beanlands RS, et al. Clinical quantification of myocardial blood flow using PET: joint position paper of the SNMMI Cardiovascular Council and the ASNC. J Nucl Med. 2018;59:273–293. doi:10.2967/jnumed.117.201368
  37. Bateman TM, Al-Mallah MH, Alnabelsi TS, et al. Clinical indications for positron emission tomography myocardial perfusion imaging and myocardial blood flow quantification: an American Society of Nuclear Cardiology position statement. J Nucl Cardiol. 2026;57:102619. doi:10.1016/j.nuclcard.2025.102619
  38. Bateman TM, Dilsizian V, Beanlands RS, et al. American Society of Nuclear Cardiology and Society of Nuclear Medicine and Molecular Imaging joint position statement on the clinical indications for myocardial perfusion PET. J Nucl Cardiol. 2016;23:1227–1231. doi:10.1007/s12350-016-0626-9
  39. Owczarek A, Adamczewski Z, Plachcinska A, Cichocki P. Elimination of artifacts caused by residual radiopharmaceutical activity in injection site in myocardial perfusion imaging on Discovery NM 530c semiconductor gamma camera. EJNMMI Res. 2025;15:27. doi:10.1186/s13550-025-01222-w
  40. Einstein AJ, Johnson LL, Bokhari S, et al. Agreement of visual estimation of coronary artery calcium from low-dose CT attenuation correction scans in hybrid PET/CT and SPECT/CT with standard Agatston score. J Am Coll Cardiol. 2010;56:1914–1921. doi:10.1016/j.jacc.2010.05.057
  41. Hecht HS, Cronin P, Blaha MJ, et al. 2016 SCCT/STR guidelines for coronary artery calcium scoring of noncontrast noncardiac chest CT scans. J Cardiovasc Comput Tomogr. 2017;11:74–84. doi:10.1016/j.jcct.2016.11.003
  42. Jaarsma C, Leiner T, Bekkers SC, et al. Diagnostic performance of noninvasive myocardial perfusion imaging using single-photon emission computed tomography, cardiac magnetic resonance, and positron emission tomography imaging for the detection of obstructive coronary artery disease: a meta-analysis. J Am Coll Cardiol. 2012;59:1719–1728. doi:10.1016/j.jacc.2011.12.040
  43. Mc Ardle BA, Dowsley TF, deKemp RA, Wells GA, Beanlands RS. Does rubidium-82 PET have superior accuracy to SPECT perfusion imaging for the diagnosis of obstructive coronary disease? A systematic review and meta-analysis. J Am Coll Cardiol. 2012;60:1828–1837. doi:10.1016/j.jacc.2012.07.038
  44. Miller RJH, Slomka PJ. Artificial intelligence in nuclear cardiology: an update and future trends. Semin Nucl Med. 2024;54:648–657. doi:10.1053/j.semnuclmed.2024.02.005
  45. Intersocietal Accreditation Commission. IAC standards and guidelines for nuclear/PET accreditation. intersocietal.org