Library › Devices and mechanical support › Intra-aortic balloon pump
Intra-aortic balloon pump: counterpulsation, insertion, timing, waveforms, complications, and weaning
Counterpulsation physiology, balloon selection, and insertion; triggering, timing, and waveform interpretation; anticoagulation, complications, weaning, removal, and transport.
The intra-aortic balloon pump (IABP) is a catheter-mounted balloon in the descending thoracic aorta that inflates in diastole and deflates just before systole, in counterpulsation with the heart. It is the oldest, smallest, and most widely available form of temporary mechanical circulatory support. Device choice among the balloon pump, microaxial pumps, and venoarterial extracorporeal membrane oxygenation (VA-ECMO), shock strategy, and escalation are covered on the temporary mechanical circulatory support and cardiogenic shock pages.
Counterpulsation: design and development of the balloon pump
Counterpulsation was developed in the laboratory in the 1950s, when diastolic pressure augmentation was shown to raise coronary flow in animal models. A balloon catheter in the aorta, inflated in diastole and deflated in systole and driven at first with carbon dioxide, was described in 1962 (Moulopoulos 1962). The first clinical series reported in 1968 used surgically placed balloons in patients with cardiogenic shock after myocardial infarction and showed a rise in arterial pressure and urine output (Kantrowitz 1968). By the mid-1970s it was also used widely to wean patients from cardiopulmonary bypass (Parissis 2016). Percutaneous femoral insertion, introduced around 1980, replaced surgical cutdown and made the device widely available (Bregman 1980).
A current system has three parts. The catheter is a dual-lumen polyurethane shaft of 7 to 8 Fr. The outer (gas) lumen carries helium to and from the balloon; the inner (central) lumen accepts the guidewire during insertion and then transmits central aortic pressure from the tip, either through a fluid column to an external transducer or through a fiber-optic sensor at the tip. The balloon is a cylindrical polyurethane membrane of 25 to 50 mL, about 18 to 27.5 cm long and about 13 to 18 mm in diameter when inflated (Parissis 2016). The console contains the pneumatic drive and helium reservoir, the electrocardiogram (ECG) and pressure inputs, the timing algorithm, the battery, and the alarms.
Physiology of counterpulsation
The balloon displaces a volume of blood equal to its own volume twice per assisted cycle. Inflation at aortic valve closure, marked by the dicrotic notch, displaces blood toward the aortic root and toward the periphery and raises aortic diastolic pressure. Because coronary flow to the left ventricle occurs mainly in diastole, the rise in diastolic pressure raises coronary perfusion pressure, the difference between aortic diastolic pressure and left ventricular diastolic pressure. Deflation immediately before the aortic valve opens removes that volume from the aorta. Aortic end-diastolic pressure falls, isovolumic contraction ends at a lower pressure, and the ventricle ejects against a lower impedance. The balloon is passive during systole: it adds no contractility and generates no forward flow of its own, and its effects depend on native ejection and a stable, detectable rhythm.
Measured effects in clinical series are modest. Cardiac output rises by about 0.5 to 1.0 L/min, systolic pressure falls by up to about 10%, aortic end-diastolic pressure falls by up to about 30%, and the peak rate of rise of left ventricular pressure falls (Parissis 2016; Gillespie 2024). Left ventricular end-diastolic pressure and pulmonary capillary wedge pressure fall slightly. In 218 patients with chronic systolic heart failure and cardiogenic shock, the mean early rise in cardiac output was 0.57 L/min and mean pulmonary artery pressure fell by 5.1 mm Hg. One quarter of patients were strong responders, with a mean rise of 1.21 L/min; systemic vascular resistance above 1,300 dyn·s·cm⁻⁵ and moderate or severe mitral regurgitation predicted that response (Huang 2021). The cardiac output response is larger in heart failure–related shock than in infarct-related shock, in which it is often negligible (Gillespie 2024).
The effect on coronary flow depends on the coronary lesion. Intracoronary Doppler measurements showed that balloon pumping augmented aortic diastolic pressure by a mean of 83% and raised diastolic flow velocity in normal arteries and in arteries opened by angioplasty, but did not raise flow beyond a critical stenosis (Kern 1993). The gain in perfusion is therefore greatest after revascularization and in hypotension, when autoregulation is exhausted.
On the pressure-volume loop the balloon narrows the loop slightly and leaves the end-systolic pressure-volume relationship unchanged; the comparison with other devices is on the temporary mechanical circulatory support page.
| Determinant | Effect on counterpulsation |
|---|---|
| Balloon volume relative to stroke volume | Augmentation is greatest when balloon volume approximates stroke volume; it falls with very low stroke volume, when little blood is available to displace, and when the balloon is small for the aorta (Parissis 2016) |
| Balloon diameter relative to the aorta | An inflated diameter of about 90% to 95% of the descending aortic diameter displaces blood efficiently without occluding the aorta; an oversized balloon traumatizes the wall and an undersized balloon displaces less blood |
| Position | A balloon placed low in the aorta displaces blood toward the visceral and iliac arteries and away from the aortic root, and compromises visceral branches |
| Heart rate | Tachycardia shortens diastole, and with it the duration of augmentation; consoles track high rates, but the effect per beat falls |
| Rhythm | Irregular rhythm makes the dicrotic notch and the onset of systole less predictable and degrades timing |
| Systemic vascular resistance and aortic compliance | Low resistance dissipates the displaced volume and lowers augmentation; higher resistance is associated with a larger cardiac output response (Huang 2021) |
Factors that determine the size of the hemodynamic effect of counterpulsation.
Balloon size, catheter, and sheath
Balloon volume is chosen by patient height, a surrogate for the length of the descending aorta between the left subclavian artery and the celiac trunk. A 40-mL balloon is used in about four of five adults and suits most patients 162 to 182 cm tall; a 50-mL balloon is used above about 182 cm, and 30- to 34-mL balloons below about 162 cm (Parissis 2011; Parissis 2016; Gillespie 2024). Height bands differ between manufacturers and catheter families, and a 25-mL balloon is available for very short adults. The product's instructions for use govern the choice.
| Balloon volume | Typical patient height | Membrane length | Inflated diameter |
|---|---|---|---|
| 25 mL | Very short adults (below about 152 cm in some manufacturers' charts) | About 180 mm | About 13 mm |
| 30 or 34 mL | Below about 162 cm | About 220–230 mm | About 14–15 mm |
| 40 mL | About 162–182 cm | About 260 mm | About 15 mm |
| 50 mL | Above about 182 cm | About 260–270 mm | About 16–18 mm |
Approximate balloon dimensions and height-based sizing compiled from published series and reviews (Parissis 2011; Parissis 2016; Gillespie 2024).
Height predicts aortic length imperfectly. In 63 cardiac surgical patients who underwent computed tomography (CT) during balloon support, the radiographic tip position appeared correct in 96.8% but was appropriate on CT in only 38.1%. Balloon length exceeded the measured subclavian-to-celiac distance in 68.2%. At least one visceral artery was covered in 61 of 63 patients: the celiac trunk in 96.8%, the superior mesenteric artery in 87.3%, and the renal arteries in 66.7% (Rastan 2010). In a cadaver study, height-based selection agreed only modestly with the size predicted from measured aortic length (kappa 0.38) (Parissis 2011). Where CT is available, the measured subclavian-to-celiac distance can guide selection of a shorter balloon.
The catheter is placed through an 8 Fr sheath or without a sheath. Sheathless insertion places the catheter over the wire and reduces the size of the arterial entry by the thickness of the sheath wall; it is used in small or diseased femoral arteries and when limb ischemia is a concern. A sheath allows easier exchange of a malfunctioning catheter. Some kits include a sidearm sheath that can transduce pressure if the central lumen fails (Gillespie 2024).
Femoral insertion and positioning
The balloon is placed most often through the common femoral artery, at the bedside in an emergency or, preferably, under fluoroscopy, with ultrasound used to choose the less diseased side and to puncture over the femoral head. Principles of arterial access and limb protection that apply to all devices are covered on the temporary mechanical circulatory support page.
- Preparation: available aortoiliac imaging is reviewed for disease, aneurysm, and tortuosity, and baseline pulses, ankle Doppler signals, platelet count, and coagulation tests are recorded.
- Access: the common femoral artery is punctured with ultrasound guidance, a J-tipped guidewire is advanced into the descending thoracic aorta under fluoroscopy, and the sheath is placed or the tract is dilated for sheathless insertion.
- Balloon preparation: the balloon is evacuated with a syringe and one-way valve so that it stays tightly wrapped, and the central lumen is flushed.
- Advancement: the catheter is advanced over the wire until the radiopaque tip lies about 1 to 2 cm below the origin of the left subclavian artery, with the whole membrane beyond the sheath, because a balloon partly inside the sheath cannot unwrap.
- Connection: the wire is removed, the central lumen is aspirated, flushed, and connected to the pressure transducer (or the fiber-optic sensor is connected and zeroed), and the gas lumen is connected to the console.
- Start of pumping: the console fills the balloon with helium, pumping starts at 1:1 or briefly at 1:2 for a timing check, and augmentation and timing are confirmed on the arterial tracing.
- Securing: the catheter is fixed, and the position is confirmed on a chest radiograph.
The target position places the tip just distal to the left subclavian artery and the lower end of the membrane above the renal and visceral arteries. On a frontal chest radiograph the radiopaque tip marker projects about 2 cm above the carina, or 1 to 2 cm below the aortic knob (Gillespie 2024). A tip below the carina suggests a low balloon whose lower end may lie over the celiac trunk, superior mesenteric artery, or renal arteries. A tip above the aortic knob suggests that the balloon may occlude the left subclavian artery or abrade the arch. Radiographic landmarks are only approximate, and the lower marker, when visible on an abdominal film, adds information about the visceral segment (Chaturvedi 2023).
Transesophageal echocardiography can confirm the tip just below the left subclavian artery and exclude dissection. The radiographic appearance of support devices is described on the chest radiography page.
Axillary and subclavian insertion for prolonged support
Femoral access confines the patient to bed with the hip straight and the head of the bed at or below about 30 degrees to avoid kinking the catheter (Gillespie 2024). When support is expected to last days to weeks, as in heart failure–related shock awaiting transplantation or a durable left ventricular assist device (LVAD), the balloon can be placed through the left axillary or subclavian artery so that the patient can sit, stand, and walk. Percutaneous access is obtained under ultrasound and fluoroscopy, and the catheter passes around the arch into the descending aorta; surgical exposure with a graft is an alternative.
In the first series of 50 patients bridged to transplantation, 42 (84%) underwent transplantation and survival on support was 92%. Median support lasted 18 days, and 4 patients (8%) had clinically significant thromboembolic or bleeding events. Malposition was the most common minor event (44%), and pulmonary artery pressure, creatinine, and bilirubin improved on support (Estep 2013). The expanded series of 195 patients reported planned heart replacement in 68% (120 transplants and 13 LVADs), death on support in 16, and escalation in 18. Catheter replacement for malfunction was needed in 37% of patients (Bhimaraj 2020). Arm movement displaces the catheter, so position is rechecked after mobilization. A balloon pump placed through either route qualifies a candidate for status 2 on the US heart allocation system when hemodynamic criteria are met, as described on the temporary mechanical circulatory support page.
Helium, the console, and gas-handling safety
Helium is the shuttle gas because its low density and viscosity allow fast transfer between console and balloon through a long, narrow gas lumen, which permits complete inflation and deflation within a short diastole at high heart rates (Parissis 2016; Gillespie 2024). Carbon dioxide, used in early systems, dissolves in blood far more readily than helium, so an embolus of carbon dioxide is absorbed faster, but it shuttles more slowly. The console fills the closed circuit from a helium cylinder and refills it at intervals. Alarms for gas loss, high or low fill pressure, catheter restriction (a kink), and suspected blood in the circuit indicate a leak, a kink, a balloon that is not unwrapped, or a balloon perforation (Getinge Cardiosave guide).
The console also displays the balloon pressure waveform, a square wave with a plateau during inflation; a rounded or low plateau suggests a balloon partly in the sheath or not unwrapped, a kink, or a leak.
Trigger sources and operating modes
The trigger is the signal the console uses to identify successive cardiac cycles. Deflation is linked to the trigger event, and inflation is placed at a set point within the cycle.
| Trigger | Event detected | Use | Limitations |
|---|---|---|---|
| ECG | R wave; deflation at the R wave, inflation near the middle of the T wave | Trigger of choice when a clear R wave is present | Electrocautery, motion artifact, low-voltage QRS, and chest compressions interfere; pacing spikes are rejected automatically on current consoles |
| Arterial pressure | Systolic upstroke | Regular rhythm without a usable ECG; chest compressions | Requires a pulse pressure the console can detect; a damped line or a narrow pulse pressure causes trigger loss |
| Ventricular or atrioventricular pacing | Ventricular pacing spike | Patient fully paced, when ECG and pressure triggers fail | Safe only with 100% ventricular pacing; intrinsic beats are not tracked |
| Atrial pacing | R wave, with atrial spikes ignored | Atrially paced rhythm, in which pacing artifacts interfere with R-wave detection | Requires an R wave |
| Internal (asynchronous) | Fixed rate, 80 per minute by default, adjustable from 40 to 120 on one console | No cardiac ejection: cardiopulmonary bypass or asystole, to keep the balloon moving | Not synchronized with any native activity; never used while the heart ejects |
Trigger sources for counterpulsation (Gillespie 2024; Getinge Cardiosave guide). Internal rates and available modes vary by console.
Modern consoles operate in an automatic mode that selects the ECG lead and the trigger, switches between ECG and pressure triggering when one signal becomes unreliable, places inflation and deflation, and adjusts timing beat by beat with changes in rate and rhythm. In a semiautomatic mode the operator selects the lead and trigger and sets inflation and deflation, after which the console adjusts the settings with heart rate; pacer and internal triggers are available only in that mode on some consoles (Getinge Cardiosave guide). The automatic mode is preferred in most patients (Gillespie 2024).
Timing of inflation and deflation
Inflation is set to coincide with aortic valve closure, the dicrotic notch on the central aortic pressure waveform. Deflation is set so that the balloon is empty at the moment the aortic valve opens, which on the pressure waveform is just before the systolic upstroke and on the ECG corresponds to the R wave. Peripheral arterial waveforms are delayed relative to central aortic events by the transit time of the pulse wave, and the dicrotic notch in a peripheral tracing is less distinct. Timing is therefore assessed on the central lumen or fiber-optic signal, or on a radial tracing with allowance for the delay.
Conventional timing places deflation at a point predicted from the preceding cardiac cycle length, so that the balloon is empty before the next R wave; it works well in regular rhythm and fails when a beat comes early. Real-time (R-wave) deflation deflates the balloon on detection of the R wave, so that the balloon cannot remain inflated into an early systole; inflation is still placed at the predicted dicrotic notch. Consoles switch to it automatically in irregular rhythm.
Timing is assessed with the console set to 1:2 (or 1:3) so that an assisted beat can be compared with an unassisted beat on the same screen. Correct timing meets five conditions (Getinge Cardiosave guide; Gillespie 2024):
- Inflation occurs at the dicrotic notch of the unassisted beat.
- Inflation produces a sharp V at the notch, with a steep rise of pressure from the notch to the augmented peak.
- Peak diastolic augmentation is higher than the unassisted systolic pressure.
- Assisted aortic end-diastolic pressure is lower than unassisted end-diastolic pressure, typically by 15 to 20 mm Hg.
- Assisted systolic pressure, the peak after an assisted diastole, is lower than unassisted systolic pressure, typically by about 5 mm Hg.
Augmentation below unassisted systolic pressure with correct timing is not a timing error; it reflects low stroke volume, hypovolemia, tachycardia, low vascular resistance, or a balloon that is small, low, underfilled, or not fully unwrapped.
The assisted arterial waveform
Five pressures are read from a 1:2 tracing: unassisted systolic pressure, peak diastolic augmentation, unassisted aortic end-diastolic pressure, assisted aortic end-diastolic pressure, and assisted systolic pressure. Their relations define the quality of timing; their absolute values define the effect.
The bedside monitor often misreads the augmented wave. Most patient monitors report the highest pressure in a sampling window as systolic, so diastolic augmentation is displayed as "systolic" pressure and the true systolic pressure is not shown. The balloon pump console identifies systolic pressure, augmentation, and assisted and unassisted end-diastolic pressures separately, and its values are used for decisions; noninvasive cuff pressures are unreliable during counterpulsation (Getinge Cardiosave guide). Mean arterial pressure rises with augmentation because both pressure and time under the diastolic curve increase, and it is the value used for perfusion targets.
Timing errors: recognition, consequences, and correction
The timing errors have recognizable waveforms and specific physiologic consequences. Errors that leave the balloon inflated while the ventricle ejects, early inflation and late deflation, raise afterload and oxygen demand and are the most harmful. Errors that shorten the time the balloon is inflated, late inflation and early deflation, waste benefit.
| Error | Waveform | Physiologic consequence | Correction |
|---|---|---|---|
| Early inflation | Inflation before the dicrotic notch; augmentation encroaches on systole and the notch is lost; the augmented peak may merge with the systolic peak | Balloon inflates while the aortic valve is still open: premature valve closure, reduced stroke volume, aortic regurgitation, higher end-diastolic volume and pressure, higher wall stress and oxygen demand | Inflation moved later, to the notch |
| Late inflation | Inflation after the dicrotic notch; the notch is visible, pressure falls before the balloon inflates, the sharp V is absent, augmentation is lower and may not exceed unassisted systolic pressure | Shorter and lower diastolic augmentation; less coronary perfusion pressure gained | Inflation moved earlier, to the notch |
| Early deflation | Sharp fall soon after the augmented peak, followed by a U-shaped plateau before systole; assisted end-diastolic pressure may be equal to or lower than unassisted; assisted systolic pressure equal to or higher than unassisted | Augmentation is cut short, afterload reduction has dissipated before ejection, and retrograde coronary and carotid flow can occur and can cause angina | Deflation moved later, toward the onset of systole |
| Late deflation | Assisted end-diastolic pressure equal to or higher than unassisted; augmented wave widened; slow rise of the assisted systolic upstroke | Balloon still inflated as the valve opens: afterload rises, isovolumic contraction is prolonged, oxygen demand rises, and ejection is impeded; afterload reduction is absent | Deflation moved earlier |
Timing errors on a 1:2 tracing (Getinge Cardiosave guide; Gillespie 2024).
Late deflation converts the device from an afterload-reducing to an afterload-raising one while the waveform still shows marked augmentation. Early inflation appears when a delayed peripheral tracing is used as the timing reference, after a rise in rate in semiautomatic mode, and in irregular rhythm. Corrections are made in small steps and checked on the next 1:2 tracing.
Timing-error recognizer for a 1:2 arterial tracing
The recognizer applies the criteria in the timing-error table to pressures read from one unassisted and one assisted beat of a 1:2 tracing. End-diastolic reduction = unassisted minus assisted end-diastolic pressure; systolic reduction = unassisted minus assisted systolic pressure. With correct deflation, end-diastolic reduction is typically 15 to 20 mm Hg and systolic reduction about 5 mm Hg (Gillespie 2024). More than one error can coexist, and the result applies only to a tracing in regular rhythm.
Fiber-optic pressure sensing and automatic timing
Conventional catheters transmit aortic pressure through a fluid column in the central lumen to an external transducer. The long, narrow column is prone to damping, whip, and air bubbles; the transducer must be leveled and zeroed; and the signal reaches the console with a delay. Fiber-optic catheters place a pressure sensor at the catheter tip. The signal arrives about 50 milliseconds faster than a fluid-filled signal, the sensor moves with the patient so that leveling is unnecessary, and the console recalibrates (zeroes) the sensor against a reference inside the patient at regular intervals (Yarham 2013; Getinge Cardiosave guide). Timing algorithms use the high-fidelity signal to locate the dicrotic notch and the onset of systole beat by beat.
In a two-center audit comparing fluid-filled with fiber-optic pressure monitoring in cardiac surgical patients, suboptimal timing fell from 55 of 98 to 2 of 94 assessments (Yarham 2013). The central lumen of a fiber-optic catheter still requires a pressurized flush to remain patent, and an alternative arterial source such as a radial line is used only when the fiber-optic signal is unavailable (Getinge Cardiosave guide). Fiber-optic sensing improves the pressure signal and timing; it does not change the indications for the device or, by itself, outcomes.
Assist ratio and augmentation volume
The assist ratio (frequency) is the proportion of cardiac cycles in which the balloon inflates: all beats at 1:1, one beat in two at 1:2, and one beat in three at 1:3. Full support is 1:1. Ratios of 1:2 and 1:3 are used to assess timing, to wean, and occasionally at very fast heart rates. Augmentation volume is a separate setting: the balloon can be filled to less than its full volume, which reduces both the displaced volume and the effect.
At reduced ratios the balloon is motionless for longer periods. Thrombus forms on a motionless balloon, which is the basis for anticoagulation at low ratios and for the rule that the balloon is not left still in the aorta. When pumping must stop, as for a procedure or a console change, the balloon is inflated and deflated manually at intervals with the console fill function or, during transport failure, with a syringe of air through a stopcock, which is intended to prevent thrombus and not to provide support (Gillespie 2024).
Arrhythmia, tachycardia, and pacing
In atrial fibrillation the dicrotic notch moves with the cycle length, and an early beat can arrive while the balloon is still inflated. Consoles respond with R-wave deflation on all beats and an inflation point predicted from the preceding cycle length. The interval from the R wave to the dicrotic notch correlates with the preceding R–R interval, so the notch can be predicted beat by beat (Sakamoto 1995). In atrial fibrillation the ECG trigger is preferred, because the pressure trigger responds after ejection has begun.
Frequent ectopy causes erratic triggering, and an ECG lead with a tall R wave and a distinct T wave is selected while the cause is treated. In sustained ventricular tachycardia or fibrillation the ECG trigger is lost; defibrillation and cardioversion can be delivered with the pump running (Gillespie 2024).
Tachycardia shortens diastole. Some current systems follow 1:1 counterpulsation at rates up to about 220 per minute, but the duration of augmentation and the effect on coronary perfusion fall as rate rises, and reduced performance has been reported above about 110 per minute (Gillespie 2024). At very high rates the cause of the tachycardia is treated, and a 1:2 ratio may give a more complete inflation and deflation per assisted beat.
Balloon pump during cardiac arrest and cardiopulmonary resuscitation
During cardiac arrest the ECG trigger fails: ventricular fibrillation has no R wave, and in pulseless electrical activity the console detects QRS complexes that produce no ejection, while chest compressions add artifact. The pressure trigger is preferred during compressions, so that the balloon inflates in the relaxation phase of the compression and deflates before the next; when neither ECG nor pressure provides a trigger, the internal asynchronous mode keeps the balloon moving (Gillespie 2024; Getinge Cardiosave guide). Benefit of inflation during compressions has not been established in patients. After return of spontaneous circulation, the trigger is returned to ECG, timing is reassessed, and the catheter position is checked, because compressions can displace it. The resuscitation sequence itself is covered on the adult cardiac arrest page.
Anticoagulation during counterpulsation
Anticoagulation on the balloon pump aims to prevent thrombus on the balloon, sheath, and catheter and distal embolism. Its risk is bleeding. Unfractionated heparin is the usual agent, given as an infusion without a bolus and monitored by activated partial thromboplastin time (aPTT) or anti-factor Xa activity to a low therapeutic range. No guideline sets a target, and practice varies.
Randomized evidence is limited. In 153 patients randomized to heparin (aPTT 50 to 70 seconds) or no heparin during counterpulsation, minor limb ischemia occurred in 3 and 2 patients. No major limb ischemia or balloon-surface clot occurred in either group, and bleeding was more frequent with heparin, including 2 major bleeds against none (Jiang 2003). A structured review of that trial and two cohort studies concluded that omitting heparin, or using it selectively, reduced bleeding without increasing ischemic events (Pucher 2012). Many centers therefore omit systemic heparin during 1:1 support in patients at high bleeding risk. They give it when the ratio is reduced to 1:2 or 1:3, when the balloon will be motionless, or when another indication for anticoagulation exists, such as atrial fibrillation, mechanical valves, or acute coronary syndrome. Strategies for all devices are compared on the temporary mechanical circulatory support page.
Antiplatelet therapy is governed by the underlying disease, usually acute coronary syndrome treated with percutaneous coronary intervention (PCI), and not by the balloon. When heparin-induced thrombocytopenia is suspected, heparin is stopped and a direct thrombin inhibitor such as bivalirudin or argatroban is used, as described on the heparin-induced thrombocytopenia page.
Monitoring during counterpulsation
| Variable | Frequency | Abnormal finding and its meaning |
|---|---|---|
| Arterial waveform on the console in 1:2 or 1:3 | Once per shift and after any change in rate, rhythm, trigger, or position | Timing error; falling augmentation from gas loss, malposition, or falling stroke volume |
| Balloon pressure waveform and alarms | Continuously | Gas loss, kink, balloon not unwrapped, blood in the circuit |
| Distal pulses, ankle Doppler signals, skin temperature, and color of the cannulated limb | Hourly at first | Limb ischemia from sheath obstruction, thrombus, embolism, or dissection |
| Left radial pulse and left arm perfusion | Regularly | Proximal migration occluding the left subclavian artery (Gillespie 2024) |
| Urine output and creatinine | Hourly urine output; daily creatinine | Sudden oliguria after insertion or repositioning suggests renal artery compromise by a low balloon |
| Abdominal examination, lactate, and liver enzymes | Daily and with clinical change | Mesenteric ischemia from a low balloon, embolism, or low output |
| Chest radiograph | After insertion, daily, and after any change in augmentation | Migration of the tip above the knob or below the carina |
| Platelet count and hemoglobin | Daily | Thrombocytopenia, bleeding, hemolysis |
Bedside monitoring during counterpulsation (Gillespie 2024; Getinge Cardiosave guide). Hemodynamic and perfusion targets for shock are set out on the cardiogenic shock page.
Complications of counterpulsation
Complications are mostly vascular and hematologic, and they become more frequent as support is prolonged. In the Benchmark Registry of 16,909 patients from 203 hospitals, major complications, defined as major limb ischemia, severe bleeding, balloon leak, or death directly due to insertion or failure, occurred in 2.6%. Female sex, older age, and peripheral vascular disease independently predicted them (Ferguson 2001). Among its 5,495 patients with acute myocardial infarction, the rate was 2.7% over a median of three days (Stone 2003). In IABP-SHOCK II, the balloon pump and control groups did not differ in major bleeding (3.3% and 4.4%), peripheral ischemic complications (4.3% and 3.4%), sepsis (15.7% and 20.5%), or stroke (0.7% and 1.7%) (Thiele 2012). Vascular ischemic complications of any severity are reported in 8% to 18% of patients in surgical series, while major limb ischemia occurs in fewer than 1% (Parissis 2016). Autopsy findings of unsuspected aortoiliac dissection indicate that clinical series understate complications (Parissis 2016).
| Complication | Mechanism | Recognition | Management |
|---|---|---|---|
| Limb ischemia | Sheath or catheter obstruction of a small or diseased femoral artery; thrombus at the entry site; distal embolism; dissection | Cold, pale, or mottled limb; loss of pulse or Doppler signal; pain; rising lactate or creatine kinase | Catheter moved to the other femoral artery or to the axillary route, or device removed; embolectomy, thrombectomy, or bypass if perfusion does not return; fasciotomy for compartment syndrome |
| Bleeding and access-site injury | Arterial puncture, anticoagulation, thrombocytopenia; high puncture with retroperitoneal bleeding; pseudoaneurysm | Hematoma, falling hemoglobin, hypotension, flank or back pain | Compression, reduction or interruption of heparin, transfusion, CT for retroperitoneal bleeding, endovascular or surgical repair |
| Malposition: too low | Balloon over the celiac, superior mesenteric, or renal arteries; balloon too long for the aorta | Oliguria, rising creatinine, abdominal pain, rising lactate or liver enzymes, tip below the carina | Repositioning under fluoroscopy; shorter balloon; CT if visceral ischemia is suspected |
| Malposition: too high | Tip in the arch or the left subclavian artery; migration with arm movement after axillary insertion | Weak left radial pulse, left arm ischemia, stroke; tip above the aortic knob | Withdrawal to the correct position under fluoroscopy |
| Thrombocytopenia | Mechanical platelet consumption on the balloon surface | Steady fall in platelet count over the first days | Monitoring; heparin-induced thrombocytopenia excluded when the fall is large or accompanied by thrombosis; recovery after removal |
| Balloon perforation, rupture, and entrapment | Abrasion of the membrane against a calcified aorta; manufacturing defect | Blood or brown flecks in the gas tubing; gas-loss alarms; sudden loss of augmentation | Pumping stopped, catheter clamped and removed promptly in the supine position; surgical extraction if entrapped |
| Stroke and systemic embolism | Thrombus or atheroma dislodged from the aorta; helium embolism after rupture; high position | Focal neurologic deficit, blue toes, visceral infarction | Neuroimaging and stroke management; review of position and anticoagulation |
| Infection | Femoral catheter in place for days | Fever, site erythema or discharge, bacteremia | Cultures, antibiotics, removal or exchange to another site |
| Aortic dissection or perforation | Guidewire or catheter in a false lumen or through diseased aorta during insertion | Back or abdominal pain, absent augmentation, loss of distal pulses; often unrecognized | Device removal, CT angiography, aortic management |
Complications of the balloon pump, their recognition, and management (Ferguson 2001; Parissis 2016; Gillespie 2024; Chaturvedi 2023; Getinge Cardiosave guide).
Thrombocytopenia. A fall in platelet count is predictable. In a prospective study in which all patients received heparin, platelet counts in patients with a balloon pump fell to a nadir of 63% of baseline on day 4 and then stabilized. A count below 150,000/mm³ developed in 47% against 12% of patients without a balloon, a fall of at least half occurred in 26% against 4%, and counts rose quickly after removal (Vonderheide 1998). In 252 consecutive coronary care unit patients, thrombocytopenia (nadir below 150,000/mm³ or a fall of more than 50%) occurred in 43%, with a similar incidence with and without heparin (45% and 40%). Lower body weight, cardiogenic shock, and longer support predicted it, and after adjustment it did not independently predict major bleeding or death (Roy 2010). A fall of more than 50% beginning 5 to 10 days after heparin exposure, or thrombosis with a falling count, raises the possibility of heparin-induced thrombocytopenia.
Limb ischemia. Limb ischemia is the most frequent serious complication and is more common in women, older patients, and patients with peripheral vascular disease (Ferguson 2001); acute limb ischemia is covered on the peripheral artery disease page.
Balloon rupture, blood in the tubing, and gas embolism
Perforation of the balloon membrane, usually from abrasion against calcified plaque, lets blood enter the balloon and the gas lumen, and lets helium escape into the aorta. It presents as bright red blood, dried brown flecks, or serosanguineous fluid in the catheter or helium extender tubing, as a blood-detection or gas-loss alarm, as autofill failure, or as a sudden change in augmentation (Getinge Cardiosave guide; Gillespie 2024).
When perforation is suspected, pumping is stopped, the catheter tubing is clamped and disconnected from the console, the physician is notified, and the catheter is removed promptly with the patient supine. If continued support is needed, a new catheter is placed, and the console is replaced if blood may have entered it (Getinge Cardiosave guide). The balloon is not refilled, because any further inflation can embolize more helium. The catheter is not left in place, because blood inside the collapsed membrane clots and can bind the folds into a rigid mass that cannot be withdrawn through the arteriotomy. This complication is called balloon entrapment. An entrapped balloon is removed surgically or by endovascular techniques.
A helium embolus is usually small, but helium dissolves slowly, and embolism to the brain or viscera can cause deficit or organ ischemia. Management is supportive, and hyperbaric oxygen is considered for cerebral arterial gas embolism in centers that provide it.
Indications, contraindications, and guideline recommendations for the balloon pump
Device selection among all forms of temporary support and escalation pathways are covered on the temporary mechanical circulatory support page, which also summarizes the randomized evidence in infarct-related shock.
Selective uses. The balloon pump is used as a bridge to surgery in papillary muscle rupture and ventricular septal rupture, where afterload reduction lowers the regurgitant or shunt fraction. Other uses are refractory ischemia and ischemia-driven ventricular arrhythmias awaiting revascularization; heart failure–related shock as a bridge to transplantation or LVAD; high-risk cardiac surgery and failure to wean from bypass; selected high-risk PCI; and use as a left ventricular vent on VA-ECMO (left ventricular unloading). In the Benchmark Registry, the leading indications were support during or after catheterization (20.6%), cardiogenic shock (18.8%), weaning from bypass (16.1%), preoperative support in high-risk patients (13.0%), and refractory unstable angina (12.3%) (Ferguson 2001). Mechanical complications of infarction are covered on the STEMI and mitral regurgitation pages.
Contraindications. Moderate or severe aortic regurgitation is a contraindication, because diastolic augmentation increases regurgitant volume. Aortic dissection and aortic aneurysm at risk of rupture are contraindications, as is severe aortoiliac disease for the femoral route, in which the axillary route is an alternative. Uncontrolled bleeding, which precludes the anticoagulation needed at low assist ratios, and a rhythm that defeats timing are relative contraindications (Rihal 2015). Dynamic left ventricular outflow obstruction, as in hypertrophic cardiomyopathy or takotsubo syndrome with systolic anterior motion of the mitral valve, is worsened by afterload reduction, and the balloon is avoided in that phenotype.
| Source | Recommendation relevant to the balloon pump | Class |
|---|---|---|
| 2025 ACC/AHA/ACEP/NAEMSP/SCAI acute coronary syndrome guideline | Routine IABP in patients with acute myocardial infarction and cardiogenic shock | 3: No Benefit |
| 2022 AHA/ACC/HFSA heart failure guideline | Temporary mechanical circulatory support, of any type, in cardiogenic shock when end-organ function cannot be maintained by pharmacologic means | 2a |
| 2021 ACC/AHA/SCAI revascularization guideline | Elective insertion of an appropriate hemodynamic support device during PCI in selected high-risk patients | 2b |
| 2025 ACC concise clinical guidance on cardiogenic shock | Routine temporary mechanical circulatory support in all patients with shock strongly discouraged; device chosen by phenotype and goals | Consensus (not graded) |
US recommendations that bear on balloon pump use (Rao 2025; Heidenreich 2022; Lawton 2021; Sinha 2025). The full table for all devices is on the temporary mechanical circulatory support page.
The 2023 European Society of Cardiology acute coronary syndrome guideline likewise recommends against routine balloon pump use in cardiogenic shock without mechanical complications (Class III) (Byrne 2023).
Randomized trials of the balloon pump in shock, high-risk PCI, and cardiac surgery
IABP-SHOCK II randomized 600 patients with infarct-related shock. Mortality at 30 days was 39.7% with the balloon and 41.3% without, which defines the device's place in that setting. Mortality was 52% and 51% at 12 months and 66.3% and 67.0% at 6 years (Thiele 2012; Thiele 2013; Thiele 2019). The trial is discussed on the cardiogenic shock page, and comparison with microaxial pumps, including IMPRESS in Severe Shock (Ouweneel 2017), on the temporary mechanical circulatory support page.
Altshock-2: heart failure–related cardiogenic shock. Altshock-2 randomized an intra-aortic balloon pump specifically in shock arising from chronic heart failure, whereas the major balloon and microaxial pump shock trials enrolled patients with acute myocardial infarction. Eligible patients were in Society for Cardiovascular Angiography and Interventions (SCAI) shock stages B, C, or D and were candidates for heart replacement therapy (transplantation or durable LVAD). They were randomized to early balloon pump plus standard care (53 patients) or standard care alone (48 patients). The trial was stopped for futility at a prespecified interim analysis. The primary end point, survival or successful bridge to heart replacement therapy at 60 days, occurred in 81% with the balloon pump and 75% with standard care. The difference was not statistically significant (reported hazard ratio 0.72; 95% confidence interval [CI] 0.31 to 1.68; P = 0.45). Six patients (13%) assigned to standard care crossed over to a balloon pump, and complications were similar (Morici 2025). The trial was small and open label, and crossover reduced its power. It does not support routine early balloon use in heart failure–related shock. It does not exclude benefit in selected patients with high systemic vascular resistance or significant mitral regurgitation, the profile associated with the largest hemodynamic response (Huang 2021).
CRISP AMI: anterior infarction without shock. CRISP AMI randomized 337 patients with anterior ST-elevation myocardial infarction without shock to balloon counterpulsation started before primary PCI and continued for at least 12 hours, or to PCI alone (Patel 2011). Infarct size on cardiac magnetic resonance imaging 3 to 5 days later did not differ (42.1% and 37.5% of left ventricular mass; P = 0.06). Major vascular complications (4.3% and 1.1%; P = 0.09) and major bleeding or transfusion (3.1% and 1.7%) were numerically more frequent with the balloon, and 6-month mortality was 1.9% with the balloon and 5.2% without (P = 0.12).
BCIS-1: elective support during high-risk PCI. BCIS-1 randomized 301 patients with ejection fraction of 30% or less and extensive coronary disease (jeopardy score 8 or more of 12) to elective balloon insertion before PCI or to no planned balloon. Major adverse cardiac and cerebrovascular events at discharge did not differ (15.2% and 16.0%; P = 0.85), and 6-month mortality was 4.6% and 7.4% (P = 0.32). Major procedural complications were less frequent with elective balloon support (1.3% and 10.7%; P < 0.001). Major or minor bleeding (19.2% and 11.3%; P = 0.06) and access-site complications (3.3% and 0%) were numerically more frequent with the balloon (Perera 2010). At a median of 51 months, all-cause mortality was lower in the elective balloon group (42 against 58 deaths; hazard ratio 0.66; 95% CI 0.44 to 0.98; P = 0.039) (Perera 2013). This long-term result is a secondary analysis. The broader evidence on protected PCI, including PROTECT II, is on the temporary mechanical circulatory support page.
Preoperative balloon support in coronary bypass surgery. A Cochrane review of six small randomized trials found fewer hospital deaths with preoperative balloon support in high-risk patients undergoing on-pump coronary bypass (4 of 132 against 23 of 123 in the pooled on-pump trials; odds ratio 0.18) but judged the trials of limited quality and generalizability (Theologou 2011). A later single-center trial randomized 110 hemodynamically stable patients with ejection fraction below 35% to preincision balloon support or none and found no difference in major morbidity (40% and 31%) (Ranucci 2013). Preoperative insertion in stable patients is therefore not routine. Postcardiotomy shock is covered on the temporary mechanical circulatory support page.
Weaning from counterpulsation
Weaning begins when the cause has been treated and pressure, cardiac index, filling pressures, urine output, and lactate meet predefined targets on stable or falling vasoactive doses. Ratio weaning reduces the assist ratio from 1:1 to 1:2 and then 1:3, with both steps held for one or more hours while hemodynamics and perfusion are reassessed. Volume weaning keeps 1:1 assistance and reduces the balloon fill in steps. The two were compared in a randomized trial of 30 patients with low output after cardiac surgery. Weaning began when cardiac index exceeded 2.5 L/min/m², central venous pressure was 12 mm Hg or less, lactate was below 2.5 mmol/L, mean arterial pressure exceeded 65 mm Hg, and urine output was at least 1 mL/kg/h for 5 hours. Ratio weaning used 4 hours at 1:2 and 1 hour at 1:3; volume weaning removed 10% of the fill per hour for 5 hours. All but one patient were weaned, and volume weaning preserved cardiac index, vascular resistance, central venous pressure, and lactate better and shortened the interval from removal to discharge from intensive care (Onorati 2013).
Weaning is completed over hours, and anticoagulation is given during low-ratio support when bleeding risk allows. Failure of weaning, shown by falling pressure or cardiac output, rising filling pressures, or rising lactate at a lower level of support, leads to a return to full support, a search for unresolved causes, and, when indicated, escalation as described on the temporary mechanical circulatory support page.
Removal and hemostasis
Heparin is stopped before removal, and the aPTT or activated clotting time is allowed to fall to near normal. Pumping is stopped, the catheter is disconnected from the console, and the balloon is deflated completely by aspiration with a syringe. The catheter is withdrawn until the balloon meets the sheath, and the catheter and sheath are then removed together: a balloon that has been used no longer refolds into its original wrap and cannot be pulled back through the sheath. In sheathless insertion the catheter is withdrawn through the arteriotomy.
As the catheter leaves the artery, brief back-bleeding flushes thrombus from the entry site, and firm manual compression is applied above the puncture, for longer than after a diagnostic sheath because the arteriotomy is larger. Distal pulses are rechecked after compression and over the following hours. Surgical repair is used for a damaged arteriotomy and after insertion through a graft. Hemostasis after axillary insertion follows center protocols and uses compression, a closure device, or surgical repair.
Interhospital and air transport
The balloon pump allows transfer of a patient with shock from a referring hospital to a center that can provide surgery, durable LVAD, transplantation, or a higher level of support. Transport consoles operate on batteries; current consoles run for about 2 to 2.25 hours on battery. A spare battery is carried, AC power or an inverter is used whenever available, and the helium supply is filled before departure (Gillespie 2024). The catheter is secured, and the tip position is confirmed by radiograph before departure, because movement during transfer can displace it. Automatic mode is preferred, because ECG and pressure signals degrade with motion and vibration (Gillespie 2024).
Altitude changes the pressure of the gas in the closed circuit. During ascent the gas expands and during descent it contracts, and a console with automatic fill compensates by purging and refilling the circuit; one console does so when atmospheric pressure falls by 25 mm Hg, corresponding to about 1,000 feet of ascent, or rises by 50 mm Hg, corresponding to about 2,000 feet of descent. Rapid ascent can trigger high-baseline or high-fill alarms that stop pumping, and rapid descent can trigger gas-loss alarms (Gillespie 2024).
Sources of error in balloon pump management
- Marked augmentation taken as proof of correct timing, while assisted end-diastolic pressure at or above the unassisted value shows late deflation and increased afterload.
- A radiographic tip position above the carina accepted as proof of correct position, when a balloon too long for the aorta still covers the celiac, mesenteric, or renal arteries.
- A weak left radial pulse after mobilization or transport not linked to proximal migration over the left subclavian artery.
- Brown flecks in the gas tubing dismissed, or the balloon refilled after a blood alarm, which increases the risk of helium embolism and entrapment.
References
- Rao SV, O'Donoghue ML, Ruel M, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of patients with acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2025;85:2135–2237. doi:10.1016/j.jacc.2024.11.009
- Sinha SS, Morrow DA, Kapur NK, Kataria R, Roswell RO. 2025 Concise clinical guidance: an ACC expert consensus statement on the evaluation and management of cardiogenic shock. J Am Coll Cardiol. 2025;85:1618–1641. doi:10.1016/j.jacc.2025.02.018
- Rihal CS, Naidu SS, Givertz MM, et al. 2015 SCAI/ACC/HFSA/STS clinical expert consensus statement on the use of percutaneous mechanical circulatory support devices in cardiovascular care. J Am Coll Cardiol. 2015;65:e7–e26. doi:10.1016/j.jacc.2015.03.036
- Heidenreich PA, Bozkurt B, Aguilar D, et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2022;145:e895–e1032. doi:10.1161/CIR.0000000000001063
- Lawton JS, Tamis-Holland JE, Bangalore S, et al. 2021 ACC/AHA/SCAI guideline for coronary artery revascularization: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2022;79:e21–e129. doi:10.1016/j.jacc.2021.09.006
- Byrne RA, Rossello X, Coughlan JJ, et al. 2023 ESC guidelines for the management of acute coronary syndromes. Eur Heart J. 2023;44:3720–3826. doi:10.1093/eurheartj/ehad191
- Moulopoulos SD, Topaz S, Kolff WJ. Diastolic balloon pumping (with carbon dioxide) in the aorta: a mechanical assistance to the failing circulation. Am Heart J. 1962;63:669–675. doi:10.1016/0002-8703(62)90012-1
- Kantrowitz A, Tjonneland S, Freed PS, Phillips SJ, Butner AN, Sherman JL Jr. Initial clinical experience with intraaortic balloon pumping in cardiogenic shock. JAMA. 1968;203:113–118. PMID 5694059.
- Bregman D. Clinical experience with percutaneous intraaortic balloon pumping. Cardiovasc Dis. 1980;7:318–324. PMCID PMC287871.
- Parissis H, Graham V, Lampridis S, Lau M, Hooks G, Mhandu PC. IABP: history-evolution-pathophysiology-indications: what we need to know. J Cardiothorac Surg. 2016;11:122. doi:10.1186/s13019-016-0513-0
- Gillespie LE, Lane BH, Shaw CR, et al. The intra-aortic balloon pump: a focused review of physiology, transport logistics, mechanics, and complications. J Soc Cardiovasc Angiogr Interv. 2024;3:101337. doi:10.1016/j.jscai.2024.101337
- Kern MJ, Aguirre F, Bach R, Donohue T, Siegel R, Segal J. Augmentation of coronary blood flow by intra-aortic balloon pumping in patients after coronary angioplasty. Circulation. 1993;87:500–511. doi:10.1161/01.cir.87.2.500
- Huang AL, Fried JA, Malick W, et al. Predictors of hemodynamic response to intra-aortic balloon pump therapy in patients with acute decompensated heart failure and cardiogenic shock. J Invasive Cardiol. 2021;33:E275–E280. doi:10.25270/jic/20.00473
- Parissis H, Soo A, Leotsinidis M, Dougenis D. A statistical model that predicts the length from the left subclavian artery to the celiac axis; towards accurate intra aortic balloon sizing. J Cardiothorac Surg. 2011;6:95. doi:10.1186/1749-8090-6-95
- Rastan AJ, Tillmann E, Subramanian S, et al. Visceral arterial compromise during intra-aortic balloon counterpulsation therapy. Circulation. 2010;122(11 Suppl):S92–S99. doi:10.1161/CIRCULATIONAHA.109.929810
- Chaturvedi A, Rotman Y, Hoang T, Jew G, Mandalapu A, Narins C. CT and chest radiography in evaluation of mechanical circulatory support devices for acute heart failure. Insights Imaging. 2023;14:122. doi:10.1186/s13244-023-01469-8
- Estep JD, Cordero-Reyes AM, Bhimaraj A, et al. Percutaneous placement of an intra-aortic balloon pump in the left axillary/subclavian position provides safe, ambulatory long-term support as bridge to heart transplantation. JACC Heart Fail. 2013;1:382–388. doi:10.1016/j.jchf.2013.06.002
- Bhimaraj A, Agrawal T, Duran A, et al. Percutaneous left axillary artery placement of intra-aortic balloon pump in advanced heart failure patients. JACC Heart Fail. 2020;8:313–323. doi:10.1016/j.jchf.2020.01.011
- Yarham G, Clements A, Morris C, et al. Fiber-optic intra-aortic balloon therapy and its role within cardiac surgery. Perfusion. 2013;28:97–102. doi:10.1177/0267659112454156
- Sakamoto T, Arai H, Maruyama T, Suzuki A. New algorithm of intra aortic balloon pumping in patients with atrial fibrillation. ASAIO J. 1995;41:79–83. PMID 7727826.
- Getinge. Cardiosave intra-aortic balloon pump operation: quick reference guide (software version D.03). Document MCA00005161 Rev A. Getinge AB.
- Jiang CY, Zhao LL, Wang JA, Mohammod B. Anticoagulation therapy in intra-aortic balloon counterpulsation: does IABP really need anti-coagulation? J Zhejiang Univ Sci. 2003;4:607–611. doi:10.1631/jzus.2003.0607
- Pucher PH, Cummings IG, Shipolini AR, McCormack DJ. Is heparin needed for patients with an intra-aortic balloon pump? Interact Cardiovasc Thorac Surg. 2012;15:136–139. doi:10.1093/icvts/ivs017
- Vonderheide RH, Thadhani R, Kuter DJ. Association of thrombocytopenia with the use of intra-aortic balloon pumps. Am J Med. 1998;105:27–32. doi:10.1016/s0002-9343(98)00128-4
- Roy SK, Howard EW, Panza JA, Cooper HA. Clinical implications of thrombocytopenia among patients undergoing intra-aortic balloon pump counterpulsation in the coronary care unit. Clin Cardiol. 2010;33:30–35. doi:10.1002/clc.20694
- Ferguson JJ 3rd, Cohen M, Freedman RJ Jr, et al. The current practice of intra-aortic balloon counterpulsation: results from the Benchmark Registry. J Am Coll Cardiol. 2001;38:1456–1462. doi:10.1016/s0735-1097(01)01553-4
- Stone GW, Ohman EM, Miller MF, et al. Contemporary utilization and outcomes of intra-aortic balloon counterpulsation in acute myocardial infarction: the Benchmark Registry. J Am Coll Cardiol. 2003;41:1940–1945. doi:10.1016/s0735-1097(03)00400-5
- Thiele H, Zeymer U, Neumann FJ, et al. Intraaortic balloon support for myocardial infarction with cardiogenic shock (IABP-SHOCK II). N Engl J Med. 2012;367:1287–1296. doi:10.1056/NEJMoa1208410
- Thiele H, Zeymer U, Neumann FJ, et al. Intra-aortic balloon counterpulsation in acute myocardial infarction complicated by cardiogenic shock (IABP-SHOCK II): final 12 month results of a randomised, open-label trial. Lancet. 2013;382:1638–1645. doi:10.1016/S0140-6736(13)61783-3
- Thiele H, Zeymer U, Thelemann N, et al. Intraaortic balloon pump in cardiogenic shock complicating acute myocardial infarction: long-term 6-year outcome of the randomized IABP-SHOCK II trial. Circulation. 2019;139:395–403. doi:10.1161/CIRCULATIONAHA.118.038201
- Morici N, Sacco A, Frea S, et al. Early intra-aortic balloon support for heart failure-related cardiogenic shock: a randomized clinical trial (Altshock-2). J Am Coll Cardiol. 2025;85:1587–1597. doi:10.1016/j.jacc.2025.03.003
- Patel MR, Smalling RW, Thiele H, et al. Intra-aortic balloon counterpulsation and infarct size in patients with acute anterior myocardial infarction without shock: the CRISP AMI randomized trial. JAMA. 2011;306:1329–1337. doi:10.1001/jama.2011.1280
- Perera D, Stables R, Thomas M, et al. Elective intra-aortic balloon counterpulsation during high-risk percutaneous coronary intervention: a randomized controlled trial (BCIS-1). JAMA. 2010;304:867–874. doi:10.1001/jama.2010.1190
- Perera D, Stables R, Clayton T, et al. Long-term mortality data from the Balloon Pump-Assisted Coronary Intervention Study (BCIS-1): a randomized, controlled trial of elective balloon counterpulsation during high-risk percutaneous coronary intervention. Circulation. 2013;127:207–212. doi:10.1161/CIRCULATIONAHA.112.132209
- Theologou T, Bashir M, Rengarajan A, et al. Preoperative intra aortic balloon pumps in patients undergoing coronary artery bypass grafting. Cochrane Database Syst Rev. 2011;(1):CD004472. doi:10.1002/14651858.CD004472.pub3
- Ranucci M, Castelvecchio S, Biondi A, et al. A randomized controlled trial of preoperative intra-aortic balloon pump in coronary patients with poor left ventricular function undergoing coronary artery bypass surgery. Crit Care Med. 2013;41:2476–2483. doi:10.1097/CCM.0b013e3182978dfc
- Onorati F, Santini F, Amoncelli E, et al. How should I wean my next intra-aortic balloon pump? Differences between progressive volume weaning and rate weaning. J Thorac Cardiovasc Surg. 2013;145:1214–1221. doi:10.1016/j.jtcvs.2012.03.063
- Ouweneel DM, Eriksen E, Sjauw KD, et al. Percutaneous mechanical circulatory support versus intra-aortic balloon pump in cardiogenic shock after acute myocardial infarction (IMPRESS in Severe Shock). J Am Coll Cardiol. 2017;69:278–287. doi:10.1016/j.jacc.2016.10.022